GE S5 Service Manual
Have a look at the manual GE S5 Service Manual online for free. It’s possible to download the document as PDF or print. UserManuals.tech offer 45 GE manuals and user’s guides for free. Share the user manual or guide on Facebook, Twitter or Google+.
Service Menu 51 Document no. M1031519-01 Table 3 Module hardware status Module keyModule keyboard status. See Table 4 Module keyboard status for a detailed description of the message. Table 4 Module keyboard status Error Statuses (HEX): General errorModule general error status. See Table 5 Module general error status for a detailed description of the message. Table 5 Module general error status Pneuma errorModule pneumatics error status. See Table 6 Module pneumatics error status for a detailed description of the message. bit 0 Lamp ON bit 1 Stepper motor ON bit 2 Zeroing valve ON bit 3 Pump valve ON bit 4-15 Not used bit 0 Button 1 (start/stop) pressed bit 1 Button 2 (Lab.Data) pressed bit 2-15 Not used Bit Description What to do bit 0-1 Not used. - bit 2 Zeroing error. Zero gas signal out of limits or pressure or temperature signal varying during measurement.CO 2 sensor faulty. Send the module to the factory for repair. bit 3 CO 2 measurement error. CO2 signal out of limits or pressure or temperature signal varying during measurement.CO2 sensor faulty. Send the module to the factory for repair. bit 4 Measurement error. Sensor temperature, measurement pressure or absorbance is out of limits.CO 2 sensor faulty. Send the module to the factory for repair. bit 5 Calibration error. Failure in calibration. Recalibrate. bit 6 Pneumatic error. See pneumatic error status bits for details. bit 7 Lamp pulsing error. CO 2 sensor faulty. Send the module to the factory for repair. bit 8-15 Not used. -
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors 52 Document no. M1031519-01 Table 6 Module pneumatics error status HW errorModule hardware error status. See Table 7 Module hardware error status for a detailed description of the message. Bit Error label Description What to do 0 pumping error Stepper motor is not able to generate negative pressure at the pump pressure sensor during zeroing.Check the function of the stepper motor. 1 eq. to amb. press failed Before filling the catheter, the pressure in the module tubing is equilibrated to the ambient pressure by switching the pump valve on and off. If ambient pressure is not reached, this error is shown.Calibrate the pressure sensors. Check the pump valve function. 2 tubing/zero block occl Pressure at the pump pressure sensor gets too low, when air is aspirated though the zeroing valve.Check tubing/valves for occlusion. 8 press sens values differ Pump pressure and catheter pressure readings differ more than 50 mbars.Calibrate the pressure sensors. 9 pneuma fatal error This error is shown, if the pneumatic error occurred is considered fatal.Check for other pneumatics errors. 10 cath press too low Catheter pressure reading is below reasonable pressure values.Make sure that you are not applying too low pressure to the catheter connector. Calibrate the pressure sensors. 11 cath press too high Catheter pressure reading is above reasonable pressure values.Make sure that you are not applying too high pressure to the catheter connector. Calibrate the pressure sensors. 12 pump press too low Pump pressure reading is below reasonable pressure values.Make sure that you are not applying too low pressure to the catheter connector. Calibrate the pressure sensors. 13 pump press too high Pump pressure reading is above reasonable values.Make sure that you are not applying too high pressure to the catheter connector. Calibrate the pressure sensors.
Service Menu 53 Document no. M1031519-01 Table 7 Module hardware error status TestbitTestbit is a module production phase test status and is always 0000 (HEX). If the testbit status is ≠ 0000, then the module should be returned to the factory. Voltages Vdd; 5 V digital units driving voltage received from the module frame. 5vdc; 5 Volts of the module derived internally from 15vdc. 15vdc; 15 Volts of the module received from the module frame. Pneuma voltage; 15 V voltage driving pneumatic elements of the module, derived from 15V dirty. 15V dirty; 15 V voltage received from the module frame. bit 0 Lamp error bit 1, 2, 3 not used bit 4 CPU EEPROM error bit 5 Sensor EEPROM error bit 6 Internal AD converter error bit 7 External AD converter error bit 8 CO 2 termopile error bit 9 Reserved bit 10 Temperature detector error bit 11 Pump pressure sensor error bit 12 Catheter pressure sensor error bit 13 Pneumatic power failure bit 14-15 Not used
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors 54 Document no. M1031519-01 4.12.1 Tonometry This is another route to reach Tonometry setup menu, Others - Tonometry. For a detailed description on how to use this menu see “User’s Reference Manual”. How to calibrate PgCO2 1. Connect the calibration gas sampling line to the regulator and to the module’s catheter connector. 2. Select PgCO2 Calibration. 3. Wait until the text ‘Start feeding gas and press ComWheel. Feed gas until Adjust message is displayed’ appears. Open the regulator and start feeding gas. Press the ComWheel and continue feeding gas until the text ‘Adjust’ appears on the display. 4. Check that the displayed values match the values on the calibration gas container. Adjust with the ComWheel, if necessary. 5. If airway gases are monitored, it is recommended to calibrate the airway gases at the same time. Start / Stop Cycling is for starting or stopping the measurement. Display with PgCO2 is for defining the calculated value displayed with PgCO2. The choices are Pg-Pa, Pg-ET or pHI. Lab Data is for entering the blood gas values nad calculations. PgCO2 Calibration is for calibrating the tonometry. PgCO2/Pg-ET Alarms is for disabling/enabling alarms and adjusting the alarm limits.
Service Menu 55 Document no. M1031519-01 4.12.2 Press Sensor Start Calib.: A selection to start a module pressure sensor calibration sequence. The procedure is guided online. Record Data: A selection to print out the shown service data to the device defined in the ...Service - Record Data menu. TONO service data Amb Press; ambient pressure measured by the pressure sensor. All the pressure values are given in both mmHg and mbar on this page. Pump Press; pressure measured from the pumping unit. Cath Press; pressure measured from the catheter line. User Cal Press; pressure that the user sets during the calibration sequence according to the pressure that is applied to the catheter port. The set value should be close to 100 mmHg. User Amb Press; pressure that the user sets during the calibration sequence according to the current ambient pressure in the room. Store; a selection for storing/discarding the newly gained calibration values. Last Press calibration date; data read from the module that tells when the module has last been calibrated.
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors 56 Document no. M1031519-01 4.12.3 System Test Table 8 Description of system test Test label Description If test fails Pump unit function Stepper motor aspirates/pushes room air through the zeroing valve. If the stepper is working properly, it is able to generate small negative/positive pressure at the pump unit pressure sensor. Check the function of stepper motor. Cath press = pump press Pump pressure sensor and catheter pressure sensor readings are compared at three pressures: at ambient pressure, at approx. +100 mbar and –100 mbar. Calibrate the pressure sensors. Tubing leak test The air tightness of the module inner tubing from the zeroing valve to the pump valve is tested. Check tubing and connections for leaks. Pump unit leak test The air tightness of the pump unit is tested. Check pump unit for leaks. Start Test: A selection to start the automatic system test sequence. The procedure is guided online. Stop Test: A selection to stop the automatic system test sequence. Record Data: A selection to print out the shown service data to the device defined in the ...Service - Record Data menu. TONO Cathpres; pressure measured from the catheter line [mbar]. Ambpres; ambient pressure in the room [mbar]. Pumppres; pressure measured from the pumping unit [mbar]. System test results; detailed information of the test results of different system parts. The meaning of the messages: - “OK”, the part successfully passed the test - “FAIL”, the part failed in the test - “N/A”, the test could not be carried out properly. See Table 8 Description of system test for a detailed description of the tested parts.
Service Menu 57 Document no. M1031519-01 Cath conn. leak test The air tightness of the zeroing unit and catheter connector is tested.Make sure that the catheter connector is properly closed during the test. Check the zeroing unit and the catheter connector for leaks. Block test Air is aspirated through the zeroing valve. If pressure at the pump pressure sensor drops too much, the tubing/valves from the zeroing valve to the pump pressure sensor is blocked.Check the tubing/valves for blocks. Zero valve function Module pneumatic system pressurized and then the zeroing valve is opened to room air. If the pressure does not drop, the zeroing valve may not be functioning.Check the function of the zeroing valve. Pump valve function Module pneumatic system pressurized and then the pump valve is opened to room air. If the pressure does not drop, the pump valve may not be functioning.Check the function of the pump valve. Test label Description If test fails
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors 58 Document no. M1031519-01 4.12.4 CO2 Sensor Lamp ON/OFF: A selection to toggle the sensor lamp on and off. Meas Signal: A selection to start an automatic signal measurement sequence. This sequence measures sensor lamp signal levels. Meas Zero: A selection to start an automatic zeroing measurement sequence. Meas Gas: A selection to start an automatic gas measurement sequence. This sequence first performs the zeroing measurement. Then it measures the CO 2 concentration of the sample of arbitrary gas aspirated from the catheter port. Record Data: A selection to print out the shown service data to the device defined in the ...Service - Record Data menu. TONO Calib gain; CO2 measurement gain factor that is set in calibration. Lamp on current; sensor lamp current when lit. Lamp off current; sensor lamp current when off. Block temp; sensor temperature [°C] (in degrees centigrade). Detector offset; CO 2 detector offset voltage. Lamp status; displays whether the sensor lamp is on or off. CO2 AD; signal from CO 2 sensor thermopile (AD counts). CO2 signal; thermopile signal maximum – minimum when the lamp is blinking (AD counts). CO2 zero; thermopile signal maximum - minimum during sensor zeroing when the lamp is blinking (AD counts). CO2 absorbance; calculated light absorbance during CO 2 measurement. CO2 concentration; CO 2 concentration (%) in sensor (does not include compensations related to catheter measurement). MEAS TEMP; CO 2 sensor temperature during signal measurement [°C]. MEAS PRESS; CO 2 sensor pressure during signal measurement [mbar]. ZERO TEMP; CO 2 sensor temperature during zeroing measurement [°C]. ZERO PRESS; CO 2 sensor pressure during zeroing measurement [mbar].
Service Menu 59 Document no. M1031519-01 4.12.5 Pneumatics Zero valve: A selection to toggle the zero valve on/off. Press valve: A selection to toggle the pump valve on/off. Record Data: A selection to print out the shown service data to the device defined in the ...Service - Record Data menu. TONO Cathpres; pressure measured from the catheter line [mbar]. Ambpres; ambient pressure in the room [mbar]. Pumppres; pressure measured from the pumping unit [mbar]. Pneuma Power; displays whether electricity has been connected to the pneumatics system or not. Zero valve; displays whether the zero valve is on or off. Pump valve; displays whether the pump valve is on or off. Gas removed count; displays how many times gas is removed from the catheter during a measurement cycle. Pneumatic errors; detailed information of the performance of different pneumatics system parts. The meaning of the messages: ‘NO’, the part performed OK during the measurement = ‘this error did not occur’ ‘YES’, the part failed during the measurement = ‘this error did occur’ See Table 6 Module pneumatics error status for a detailed description of the messages.
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors 60 Document no. M1031519-01 4.12.6 AD Record Data: A selection to print out the shown service data to the device defined in the ...Service - Record Data menu. TONO All the numbers in this section are AD counts, i.e. computer internal data. EXTERNAL ADC CO2; CO 2 sensor CO2 signal. temp; CO 2 sensor temperature signal. cath press; pressure measured from the catheter line. pump press; pressure measured from the pumping unit. INTERNAL ADC VDD; 5 V digital units driving voltage received from the module frame. 5V; 5 Volts of the module derived internally from 15vdc. 15V; 15 Volts of the module received from the module frame. pneuma voltage; 15 V voltage driving pneumatic elements of the module, derived from 15 V dirty. CO2 det offset; CO2 detector offset voltage. lamp current; CO2 sensor lamp current. 15V dirty; 15 V voltage received from the module frame.