Home > GE > Medical Equipment > GE Logiq P5 Service Manual

GE Logiq P5 Service Manual

Here you can view all the pages of manual GE Logiq P5 Service Manual. The GE manuals for Medical Equipment are available online for free. You can easily download all the documents as PDF.

Page 101

GE HEALTHCARE
DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL
Section 4-2 - Required Equipment 4 - 15
4-3-10 Using the DVD-R
4-3-10-1 Using the DVD-R
1.) Go to the Utility->Connectivity->Removable Media.
2.) Insert CD/DVD media.
3.) Press Verify 
4.) The properties will be like below. a.) Capacity, Free space depends on media type.
b.) Formatted / Finalized : No
5.) Go to Patient.
6.) Press Data Transfer (located at the upper-left side of Patient screen)
7.) Select Export task.
NOTICENever move...

Page 102

GE HEALTHCARE
DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL 
4 - 16 Section 4-2 - Required Equipment
4-3-10-1              Using the DVD-R (cont’d)
8.) Select the type of Removable CD/DVD on the [To] combo box.
9.) It will do auto -formatting.
10.) After auto-formatting, select some patients on  the upper list area and then press the [Transfer] 
button.
11.) it will do trans ferring to CD/DVD.Figure 4-8   Formatting
Figure 4-9   Transferring 

Page 103

GE HEALTHCARE
DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL
Section 4-2 - Required Equipment 4 - 17
4-3-10-1              Using the DVD-R (cont’d)
12.) The selected data is transferred correctly.
13.) Press F3 and select the [Yes and Verify files].
14.) It will do finalizing. Figure 4-10   Transfer
Figure 4-11   Finalizing 

Page 104

GE HEALTHCARE
DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL 
4 - 18 Section 4-2 - Required Equipment
4-3-10-1              Using the DVD-R (cont’d)
15.) It will do Verifying.
16.) It will eject the CD/DVD and Go  to Utility and reinsert the CD/DVD.
17.) Go to Utility- >Connectivity->Removable Media.
18.) Press Verify 
19.) The properties will be like below.  a.) Capacity is the size of transferred data.
b.) Free space is 0.
c.) Formatted / Database  Present / Finalized : Yes
(DICOMDIR Present :...

Page 105

GE HEALTHCARE
DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL
Section 4-2 - Required Equipment 4 - 19
4-3-11 Backup and Restore Database, Preset Configurations and Images
NOTE: When upgrading from R1.0.x or R2.0.x to R3.0.x or R4.0.x ,preset cannot be transferred. Refer 
to the table below.
4-3-11-1 Formatting Media 1.) To format the backup media, CD- R or DVD-R, select the UTILITY
 key on the Front Panel.
2.) Select CONNECTIVITY
, then RERNOVABLE MEDIA. Properly label and Insert the backup...

Page 106

GE HEALTHCARE
DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL 
4 - 20 Section 4-2 - Required Equipment
spaces or punctuation marks.
5.) The Ultrasound system displays a pop-up menu, as shown in Figure 4-15. When the formatting  has been completed, press OK
 to continue.
6.) If desired, verify that the form at was successful by returning to  Utility > Connectivity > 
Removable Media  and selecting VERIFY
 as shown in Figure 4-12. 
Figure 4-14   Format and Verify Media
Figure 4-15   Format...

Page 107

GE HEALTHCARE
DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL
Section 4-2 - Required Equipment 4 - 21
4-3-11-2 Backup System Presets and Configurations
NOTE: Always backup any preset configurations before a so ftware reload. This ensures that if the presets need 
to be reloaded, afte r the software update, they will be the same ones the cu stomer was using prior to 
service.
1.) Insert a formatted CD-R or DVD-R into the drive.
2.) Press UTILITY
 key 
3.) Select SYSTEM
. on the monitor screen
4.)...

Page 108

GE HEALTHCARE
DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL 
4 - 22 Section 4-2 - Required Equipment
4-3-11-3 Restore System Presets and Configurations
1.) Insert the Backup/Restore CD-R or DVD-R into the drive. 
2.) Press the UTILITY
. key
3.) Select SYSTEM
.menu on the monitor screen
4.) On the monitor display,  select BACKUP/RESTORE
.
5.) In the Restore list, select Pa tient Archive, Report Archive, User Defined Configuration.
6.) In the Media field, select the Backup/Restore DVD-R.
7.)...

Page 109

GE HEALTHCARE
DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL
Section 4-2 - Required Equipment 4 - 23
4-3-11-4 Archiving Images
To export an exam(s) to a compatible Ultrasound system:
1.) Format the removable media (CD-ROM/DVD-R).  Label the removable media. Answer Yes/OK 
to the messages.Press Patient. Deselect any select ed patient(s) in the search portion of the 
Patient screen. Press Exam Data Transfer (located at the upper, left-hand corner of the Patient 
menu).
2.) Select Export task...

Page 110

GE HEALTHCARE
DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL 
4 - 24 Section 4-2 - Required Equipment
4-3-11-4              Archiving Images (cont’d)
4.) In the patient list at the top of the Patient me nu, select the patient(s) you want to export. You 
can use Windows commands to select more than  one patient. To select a consecutive list of 
patients, click the cursor on the first name, move  the cursor to the last name, then press and 
hold down the Shift+right Set ke y to select all the...
Start reading GE Logiq P5 Service Manual

Related Manuals for GE Logiq P5 Service Manual

All GE manuals