Page 111

GE HEALTHCARE
DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL
Section 4-4 - Software Configuration Checks 4 - 25
4-3-12 ECG Check Out 
Connect the ECG Harness and check:
Section 4-4
Software Configuration Checks
Table 4-5     ECG Harness Check
StepTask to doExpected Result(s)
1.Connect the ECG at the Connector on the Front of the 
scanner. It will display a curve along the bottom edge of the image 
sector
Table 4-6     Software Configuration Checks
StepTask to doExpected Result(s)
1.
Check Date...

Page 112

GE HEALTHCARE
DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL 
4 - 26 Section 4-4 - Software Configuration Checks
Section 4-5
Peripheral Checks
Check that peripherals work as described below:
Table 4-7    Peripheral Checks
StepTask to do Expected Result(s)
1.
Press ( FREEZE) Stop image acquisition.
2. Press ( P1
) or (P2) on the Keyboard  The image displayed on the screen is printed on B&W or Color printer, 
depending on the key assignment configuration
3. Press  RECORD
 on the Keyboard. To start...

Page 113

GE HEALTHCARE
DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL
Section 4-4 - Software Configuration Checks 4 - 27
Section 4-6Monitor Function Checks
4-6-1 15 LCDMON II Assy & LCD Panel with filter Function Validation Procedure
1.) Turn on the system.
2.) Select Utility Key” on the Key board and then choose “Test Pattern” and “R esolution” marked in the 
figure below.
3.) All the Numbers from 0 to 100 on the fi gure above should be easily distinguished. 
Figure 4-23   15” LCDMON Assy Function Check 

Page 114

GE HEALTHCARE
DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL 
4 - 28 Section 4-4 - Software Configuration Checks
4-6-2 15 OSD Key Assy Functi on Validation Procedure
1.) Turn on the system.
2.) Select “OSD Button” and “Lamp Button” marked below in the figure below.
3.) OSD Menu should be displayed within 5sec and the Lamp should be turned on as soon as Lamp 
Button is pushed.
4.) Select “OSD Button” marked below in the figure below and check that contrast value is changed as  soon as the button...

Page 115

GE HEALTHCARE
DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL
Section 4-7 - Keyboard Function Checks 4 - 29
Section 4-7Keyboard Function Checks
4-7-1 Main Keyboard Assy Function Validation Procedure
1.) Boot up the system.
2.) Insert the Service Dongle and go to the maintenance mode.
3.) Click Start and go to Run.
4.) Click the Run and then you can get the Run dialog box.
5.) Write the cmd command into Run dialog box  and press OK button to display command prompt. 
Figure 4-26   Start and run...

Page 116

GE HEALTHCARE
DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL 
4 - 30 Section 4-7 - Keyboard Function Checks
4-7-1              Main Keyboard Assy Fu nction Validation Procedure (cont’d)
6.) Write the scfpapitest.exe on the command prompt and press Enter button to pop up the  scfpapitest program.
7.) Press the keyboard button and rotate rotary butt on. When you press or rotary keys, Log dialog box 
displays the key log. Button name and displaying log name is like the table. Figure 4-29...

Page 117

GE HEALTHCARE
DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL
Section 4-7 - Keyboard Function Checks 4 - 31
8.) Check the Fast radio button into Indicator box on scfpapitest program and press Set All button.
R (Right)Right Button
BodyPattern/Ellipse (Rotary BodyPattern/Ellipse PushEncoder 
Rotary
THI Harmonics Button
Commentcomment Button
AO Auto Right Button
ClearClear Button
B (Button) B PushEncoder Button
TGC SlideSlider : n=8 s1 ~ s8
B (Rotary) B PushEncoder Rotary
NOTICEWhen you can check the...

Page 118

GE HEALTHCARE
DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL 
4 - 32 Section 4-7 - Keyboard Function Checks
4-7-2 Sub Keyboard Assy Func tion Validation Procedure
1.) Boot up the system and run the scfpapitest program.
2.) Press and Rotary sub keyboard, and Paddle (up/down, right/left) button.
3.) Check the log dialog box in scfpapitest program
4-7-3 Trackball Assy validation
1.) Boot up the system.
2.) Press the measurement button.
3.) Move the Trackball.
When you move the trackball in...

Page 119

GE HEALTHCARE
DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL
Section 4-7 - Keyboard Function Checks 4 - 33
4-7-6 TGC key assay / TGC Knob Set validation
1.) Follow up the procedure 4-7-1 Main Keyboard Assy Function Validation Procedure on page 4-29 
to run the scfpapitest program.
2.) Slide TGC pot and check up the value.
4-7-7 Sub Keyboard Encoder Knob Set validation
Follow up  4-7-2 Sub Keyboard Assy Function Validation Procedure on page 4-32.
NOTICEWhile sliding TGC po t, TGC message should be...

Page 120

GE HEALTHCARE
DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL 
4 - 34 Section 4-8 - Mechanical Function Checks
Section 4-8Mechanical Function Checks
4-8-1 Cover Parts Function Validation
• Check if the FRU parts are assembled tightly by naked eye and hands. 
• Check if there are dents, scratches, or cracks on the FRU parts.
• Check if screws are in place. Table 4-10    Cover parts of the FRU
NoItem
1
Right and Left cover
2 Front cover
3 Rear cover
4 Middle cover
5 Top cover
6 Top bottom cover
7...