Page 111
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL Section 4-4 - Software Configuration Checks 4 - 25 4-3-12 ECG Check Out Connect the ECG Harness and check: Section 4-4 Software Configuration Checks Table 4-5 ECG Harness Check StepTask to doExpected Result(s) 1.Connect the ECG at the Connector on the Front of the scanner. It will display a curve along the bottom edge of the image sector Table 4-6 Software Configuration Checks StepTask to doExpected Result(s) 1. Check Date...
Page 112
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL 4 - 26 Section 4-4 - Software Configuration Checks Section 4-5 Peripheral Checks Check that peripherals work as described below: Table 4-7 Peripheral Checks StepTask to do Expected Result(s) 1. Press ( FREEZE) Stop image acquisition. 2. Press ( P1 ) or (P2) on the Keyboard The image displayed on the screen is printed on B&W or Color printer, depending on the key assignment configuration 3. Press RECORD on the Keyboard. To start...
Page 113
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL Section 4-4 - Software Configuration Checks 4 - 27 Section 4-6Monitor Function Checks 4-6-1 15 LCDMON II Assy & LCD Panel with filter Function Validation Procedure 1.) Turn on the system. 2.) Select Utility Key” on the Key board and then choose “Test Pattern” and “R esolution” marked in the figure below. 3.) All the Numbers from 0 to 100 on the fi gure above should be easily distinguished. Figure 4-23 15” LCDMON Assy Function Check
Page 114
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL 4 - 28 Section 4-4 - Software Configuration Checks 4-6-2 15 OSD Key Assy Functi on Validation Procedure 1.) Turn on the system. 2.) Select “OSD Button” and “Lamp Button” marked below in the figure below. 3.) OSD Menu should be displayed within 5sec and the Lamp should be turned on as soon as Lamp Button is pushed. 4.) Select “OSD Button” marked below in the figure below and check that contrast value is changed as soon as the button...
Page 115
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL Section 4-7 - Keyboard Function Checks 4 - 29 Section 4-7Keyboard Function Checks 4-7-1 Main Keyboard Assy Function Validation Procedure 1.) Boot up the system. 2.) Insert the Service Dongle and go to the maintenance mode. 3.) Click Start and go to Run. 4.) Click the Run and then you can get the Run dialog box. 5.) Write the cmd command into Run dialog box and press OK button to display command prompt. Figure 4-26 Start and run...
Page 116
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL 4 - 30 Section 4-7 - Keyboard Function Checks 4-7-1 Main Keyboard Assy Fu nction Validation Procedure (cont’d) 6.) Write the scfpapitest.exe on the command prompt and press Enter button to pop up the scfpapitest program. 7.) Press the keyboard button and rotate rotary butt on. When you press or rotary keys, Log dialog box displays the key log. Button name and displaying log name is like the table. Figure 4-29...
Page 117
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL Section 4-7 - Keyboard Function Checks 4 - 31 8.) Check the Fast radio button into Indicator box on scfpapitest program and press Set All button. R (Right)Right Button BodyPattern/Ellipse (Rotary BodyPattern/Ellipse PushEncoder Rotary THI Harmonics Button Commentcomment Button AO Auto Right Button ClearClear Button B (Button) B PushEncoder Button TGC SlideSlider : n=8 s1 ~ s8 B (Rotary) B PushEncoder Rotary NOTICEWhen you can check the...
Page 118
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL 4 - 32 Section 4-7 - Keyboard Function Checks 4-7-2 Sub Keyboard Assy Func tion Validation Procedure 1.) Boot up the system and run the scfpapitest program. 2.) Press and Rotary sub keyboard, and Paddle (up/down, right/left) button. 3.) Check the log dialog box in scfpapitest program 4-7-3 Trackball Assy validation 1.) Boot up the system. 2.) Press the measurement button. 3.) Move the Trackball. When you move the trackball in...
Page 119
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL Section 4-7 - Keyboard Function Checks 4 - 33 4-7-6 TGC key assay / TGC Knob Set validation 1.) Follow up the procedure 4-7-1 Main Keyboard Assy Function Validation Procedure on page 4-29 to run the scfpapitest program. 2.) Slide TGC pot and check up the value. 4-7-7 Sub Keyboard Encoder Knob Set validation Follow up 4-7-2 Sub Keyboard Assy Function Validation Procedure on page 4-32. NOTICEWhile sliding TGC po t, TGC message should be...
Page 120
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL 4 - 34 Section 4-8 - Mechanical Function Checks Section 4-8Mechanical Function Checks 4-8-1 Cover Parts Function Validation • Check if the FRU parts are assembled tightly by naked eye and hands. • Check if there are dents, scratches, or cracks on the FRU parts. • Check if screws are in place. Table 4-10 Cover parts of the FRU NoItem 1 Right and Left cover 2 Front cover 3 Rear cover 4 Middle cover 5 Top cover 6 Top bottom cover 7...