GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL Section 3-4 - Preparing for Installation 3-11 Section 3-5Completing the Installation 3-5-1 Power On/Boot Up NOTE: After turning off the system, wait at least ten seconds before turning it on again. The system may not be able to boot if power is recycled too quickly. 3-5-1-1 Scanner Power On 1.) Connect the Main Power Cable at the rear of the System. 2.) Connect the Main Power cable to an appropriate mains power outlet. 3.) Switch ON...
Page 72
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL 3-12 Section 3-4 - Preparing for Installation 3-5-1-1 Scanner Power On (cont’d) Figure 3-9 Power On /Off Switch Location
Page 73
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL Section 3-4 - Preparing for Installation 3-13 3-5-1-2 Power Up Sequence 1.) The Start Up Screen will be shown on the Mo nitor display when the system is turned ON. 2.) After initialization is complete , all lighted buttons on the Control Panel light and the default B-Mode screen or Patient screen (no probes are conn ected) is displayed on the monitor display. 3-5-1-3 Power Up Sequence NOTE: To enter the Maintenance Mode, select...
Page 74
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL 3-14 Section 3-4 - Preparing for Installation 3-5-2 Power Off/ Shutdown NOTE: After turning off a system, wait at least ten seconds before turning it on again. The system may not be able to boot if power is recycled too quickly. 3-5-2-1 Power Shutdown 1.) Press the ON/OFF key at the front of the System. refer to Figure 3-9 on page 3-12 . 2.) Click the shutdown icon in the shutdown dialog box and wa it until the system shutdown....
Page 75
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL Section 3-4 - Preparing for Installation 3-15 3-5-3 Transducer Connection 1.) Connect a transducer to one of the three rightmost transducer receptacle as follows: A.) Ensure that the transducer twist lo ck lever to the horizontal position. B.) Insert the transducer connector on the receptacle guide pi n until it touches the receptacle mating surface. C.) Twist the transducer twis t lock lever to horizontal position to lock it in...
Page 76
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL 3-16 Section 3-4 - Preparing for Installation Section 3-6 System Configuration 3-6-1 System Specifications 3-6-1-1 System Settings 1.) Press Insert key on the A/N keyboard to enter the utility menu. 2.) Select Utility > System > General. 3.) Set the Hospital name, Department, Date and Time, Language, and Units. 4.) Click on Exit to terminate the utility function. Table 3-1 System Configuration Configuration...
Page 77
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL Section 3-4 - Preparing for Installation 3-17 3-6-1-2 Physical Dimensions The physical dimensions of the LOGIQ™ P5/A5/A5Pro unit are summarized in Table 3-2 on page 3-17 . The Size of LOGIQ™ P5/A5/A5Pro, with monitor and peripherals Table 3-2 Physical Dimensions of LOGIQ™ P5/A5/A5Pro HeightWidthDepthUnit 135/141 43.064.0 cm 53.15 - 55.50 16.97 25.2inches Figure 3-13 Overall Dimensions WEIGHT : 75Kg (165bs) including Monitor NOTE :...
Page 78
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL 3-18 Section 3-4 - Preparing for Installation 3-6-1-3 Weight without Monitor and Peripherals 3-6-1-4 Acoustic Noise Output: Less than 70dB (A) according to DIN 45635 - 19 - 01 - KL2. 3-6-2 Electrical Specifications 3-6-3 On-Board Optional Peripherals See each option installation instructions for installation and connection procedures. NOTE: There are no external peripheral options for the LOGIQ™ P5/A5/A5Pro. Table 3-3 Weight of...
Page 79
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL Section 3-4 - Preparing for Installation 3-19 3-6-4 External I/O Connector Panel Located on the rear panel are video input and output connectors, audio input and output, USB, footswitch connector power connector and control connections for VCR or DVD Recorder, printer, and service tools. This section indicates the pin assignment for each connector. NOTE: Each outer (case) ground line of peripheral/accessory connectors are...
Page 80
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL 3-20 Section 3-4 - Preparing for Installation 3-6-4-1 External I/O Pin Outs Pin No.SignalPin No.Signal 1RED9 N /A 2 GREEN 10SGND 3B LU E1 1 N /A 4N /A1 2N /A 5 GND 13HSYNC 6 RGND 14VSYNC 7 GGND 15N/A 8B GND Table 3-6 Pin Assignments of DB15 connector for External VGA Pin No.SignalPin No.Signal 1+ 5 V DC3D ATA + 2 DATA - 4GND Table 3-7 Pin Assignments of USB Pin No.SignalPin No.Signal 1T X+5 N C 2 TX- 6RX- 3R X+7 N C 4N C8N C...