Page 71

GE HEALTHCARE
DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL
Section 3-4 - Preparing for Installation 3-11
Section 3-5Completing the Installation
3-5-1 Power On/Boot Up
NOTE: After turning off the system, wait at least ten seconds before turning it on again. The system 
may not be able to boot if power is recycled too quickly.
3-5-1-1 Scanner Power On 1.) Connect the Main Power Cable at the rear of the System.
2.) Connect the Main Power cable to an appropriate mains power outlet.
3.) Switch ON...

Page 72

GE HEALTHCARE
DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL 
3-12 Section 3-4 - Preparing for Installation
3-5-1-1              Scanner Power On (cont’d)
Figure 3-9   Power On /Off Switch Location 

Page 73

GE HEALTHCARE
DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL
Section 3-4 - Preparing for Installation 3-13
3-5-1-2 Power Up Sequence
1.) The Start Up Screen will be shown on the Mo nitor display when the system is turned ON.
2.) After initialization is complete , all lighted buttons on the Control Panel light and the default B-Mode 
screen or Patient screen (no probes are conn ected) is displayed on the monitor display.
3-5-1-3 Power Up Sequence  NOTE: To enter the Maintenance Mode, select...

Page 74

GE HEALTHCARE
DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL 
3-14 Section 3-4 - Preparing for Installation
3-5-2 Power Off/ Shutdown
NOTE: After turning off a system, wait at least ten  seconds before turning it on again. The system may 
not be able to boot if power is recycled too quickly.
3-5-2-1 Power Shutdown 1.) Press the  ON/OFF
 key at the front of the System. refer to Figure 3-9 on page 3-12 .
2.) Click the shutdown icon in  the shutdown dialog box and wa it until the system shutdown....

Page 75

GE HEALTHCARE
DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL
Section 3-4 - Preparing for Installation 3-15
3-5-3 Transducer Connection
1.) Connect a transducer to one of the three rightmost transducer receptacle as follows:
A.) Ensure that the transducer twist lo ck lever to the horizontal position.
B.) Insert the transducer  connector on the receptacle guide pi n until it touches the receptacle 
mating surface.
C.) Twist the transducer twis t lock lever to horizontal position to lock it in...

Page 76

GE HEALTHCARE
DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL 
3-16 Section 3-4 - Preparing for Installation
Section 3-6
System Configuration
3-6-1 System Specifications
3-6-1-1 System Settings
1.) Press Insert key on the A/N keyboard to enter  the utility menu. 
2.) Select Utility > System > General.
3.) Set the  Hospital name, Department, Date and Time, Language,  and Units.
4.) Click on  Exit to terminate the  utility function.
Table 3-1    System Configuration
Configuration...

Page 77

GE HEALTHCARE
DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL
Section 3-4 - Preparing for Installation 3-17
3-6-1-2 Physical Dimensions
The physical dimensions of the LOGIQ™ P5/A5/A5Pro  unit are summarized in Table 3-2 on page 3-17 .
The Size of LOGIQ™ P5/A5/A5Pro, with monitor and peripherals
Table 3-2    Physical Dimensions of LOGIQ™ P5/A5/A5Pro
HeightWidthDepthUnit
135/141 43.064.0 cm
53.15 - 55.50 16.97 25.2inches
Figure 3-13   Overall Dimensions
WEIGHT : 75Kg (165bs)
including Monitor
NOTE :...

Page 78

GE HEALTHCARE
DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL 
3-18 Section 3-4 - Preparing for Installation
3-6-1-3 Weight without Monitor and Peripherals
3-6-1-4 Acoustic Noise Output:
Less than 70dB (A) according to DIN 45635 - 19 - 01 - KL2.
3-6-2 Electrical  Specifications
3-6-3 On-Board Optional Peripherals
See each option installation instructions  for installation and connection procedures.
NOTE: There are no external peripheral options for the LOGIQ™ P5/A5/A5Pro. Table 3-3    Weight of...

Page 79

GE HEALTHCARE
DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL
Section 3-4 - Preparing for Installation 3-19
3-6-4 External I/O Connector Panel
Located on the rear panel are video input and output connectors, audio input and output, USB, 
footswitch connector power connector and control connections for VCR or DVD Recorder, printer, and 
service tools. 
This section indicates the pin  assignment for each connector.
NOTE: Each outer (case) ground line of peripheral/accessory connectors are...

Page 80

GE HEALTHCARE
DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL 
3-20 Section 3-4 - Preparing for Installation
3-6-4-1 External I/O Pin Outs
Pin No.SignalPin No.Signal
1RED9 N /A
2 GREEN 10SGND
3B LU E1 1 N /A
4N /A1 2N /A
5 GND 13HSYNC
6 RGND 14VSYNC
7 GGND 15N/A
8B GND
Table 3-6    Pin Assignments of DB15 connector for External VGA
Pin No.SignalPin No.Signal
1+ 5 V DC3D ATA  +
2 DATA - 4GND
Table 3-7    Pin Assignments of USB
Pin No.SignalPin No.Signal
1T X+5 N C
2 TX- 6RX-
3R X+7 N C
4N C8N C...