Page 121

GE HEALTHCARE
DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL
Section 4-8 - Mechanical Function Checks 4 - 35
4-8-2 LCD plastic filter Function Validation Procedure
Check if LCD plastic filter is fastened so that it wont move.
4-8-3 Air filter set Func tion Validation Procedure
Check if the air filter is placed deep inside.Figure 4-31   Check the LCD Plastic Filter
Figure 4-32   Check the air filter 

Page 122

GE HEALTHCARE
DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL 
4 - 36 Section 4-8 - Mechanical Function Checks
4-8-4 TGC key assy/TGC Knob Se t, Sub keyboard encoder Knob Set, Keycap set, Main 
Keyboard Encoder Knob set  Function Validation Procedure
• Check if the each button operates properly.
• Check if encoder knobs are in the center.
• Check if the keycap set can not be removed easily.
Figure 4-33   Check the each button 

Page 123

GE HEALTHCARE
DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL
Section 4-8 - Mechanical Function Checks 4 - 37
4-8-5 Cup holder Function Validation Procedure
Check if the cup holder can not be removed easily.
Figure 4-34   Check the cup holder 

Page 124

GE HEALTHCARE
DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL 
4 - 38 Section 4-8 - Mechanical Function Checks
4-8-6 Front caster/Rear caster  Function Validation Procedure
• In case of Bi break caster, check all caster lo cks and caster swivel locks for proper operation.
• In case of front caster, check caster locks for proper operation.
• In case of rear caster, check casters motion for proper operation.
• Check if screws are in place.
Figure 4-35   Check the caster 

Page 125

GE HEALTHCARE
DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL
Section 4-8 - Mechanical Function Checks 4 - 39
4-8-7 Probe holder Function Validation Procedure
Check if the probe holders are assembled tightly by naked eye and hands.
Check if the probe holder can not be removed easily.
Figure 4-36   Check the probe holder 

Page 126

GE HEALTHCARE
DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL 
4 - 40 Section 4-8 - Mechanical Function Checks
4-8-8 Flexible LCD Arm Functi on Validation Procedure
• Check if the flexible arm operates properl y.(Up-down, rotation and tilting mechanism)
• Check if the flexible arm cover parts can not be  removed easily.Check if the lock operates properly.
Figure 4-37   Check the Flexible LCD Arm
 
  

Page 127

GE HEALTHCARE
DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL
Section 4-9 - Board Function Checks 4 - 41
Section 4-9Board Function Checks
4-9-1 CL1TRX Assy / BL1TRX Assy Function Validation Procedure
1.) Turn on the system and check the completion of Echoloader loading.
2.) Do the Service Login. Refer to the 5-10-2-2 Service Login on page 5-41.
3.) Select Diagnostics => LOGIQ P5 Diagnostics  => FRU Test => CL1TRX Assy / BL1TRX overall.
4.) Execute the diagnostic test.
5.) 5)All diagnostic test...

Page 128

GE HEALTHCARE
DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL 
4 - 42 Section 4-9 - Board Function Checks
4-9-2 ACWD Assy Function Validation Procedure
1.) Insert the Sector probe in probe port.
2.) Turn on the system.
3.) Press Virtual Convex button to enable  Virtual Convex if not already enabled.
Figure 4-39   Virtual Convex Enabled
Figure 4-40   Virtual Convex 

Page 129

GE HEALTHCARE
DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL
Section 4-9 - Board Function Checks 4 - 43
4.) Press “CW” button to start CWD Mode.
5.) Scan in CWD mode (with Virtual Convex enabled).
6.) Press “B” button to scan in “B” mode.
7.) Press Virtual Convex button to disable  Virtual Convex if not already disabled.
8.) Press “CW” button to start CWD Mode.
9.) Scan in CWD mode (with  Virtual Convex disabled)
10.) Compare the CWD (with Virtua l Convex enabled) sensitivity and CWD (with...

Page 130

GE HEALTHCARE
DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL 
4 - 44 Section 4-9 - Board Function Checks
ACWD Diagnostics Menu,” on page 18.Reboot up the system.
16.) Reboot up the system.
17.) Insert Probe in Probe port (Sector or Pencil Probe).
18.) After system boot up, enter the CWD mode.
19.) Check the display Image. Refer to  Section 10-5 System Maintenance on page 10-7  and 4-3-5 
System CFM and PWD Checks on page 4-9 .
Figure 4-43   ACWD Diagnostic Test  