Page 81
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL Section 3-4 - Preparing for Installation 3-21 3-6-5 Video Specification Video specifications may be needed to be able to connect the VCR or DVD Video Recorder to the LOGIQ™ P5/A5/A5Pro. Table 3-9 Video Specifications Timing ParameterNTSCPAL 640x480 60Hz800x60060Hz800x60075Hz1024x768 60Hz Horizontal Rate [kHz] 60.02 31.25 31.47 37.88 46.88 48.36 Horizontal Period [µ s]16.66 32.00 31.78 26.40 21.33 20.68 Pixel Clock [MHz] 78.75 29.50...
Page 82
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL 3-22 Section 3-4 - Preparing for Installation 3-6-6 Setting VCR (Mitsubishi MD3000) USB Interface 1.) Turn off the Power of the VCR. 2.) Open the rear panel of the VCR. 3.) Set dip switches. Off 1~6 pins, On 7 pin, and Off 8 pin. NOTE: 7 Pin ON = USB Interface, OFF = RS232C Interface. Figure 3-15 Setting VCR USB Interface
Page 83
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL Section 3-8 - Software/Option Configuration 3-23 Section 3-7 Available Probes See Specifications in the LOGIQ™ P5/A5/A5Pro User Manual for probes and intended use. See Chapter 9 , for part numbers to be used when ordering new or replacement probes. Section 3-8 Software/Option Configuration refer to the LOGIQ™ P5/A5/A5Pro Basic User Manual, Chapter 16, Customizing Your System for information on conf iguring items like Hospital,...
Page 84
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL 3-24 Section 3-8 - Software/Option Configuration Section 3-9 Connectivity Installation Worksheet Site System Information Services (Destination Devices) TCP/IP Settings Device Type Manufacturer Name IP Address Port AE Title 1 2 3 4 5 6 7 8 9 10 11 12 Name - AE Title: Dept: LOGIQ SN:Type: Floor: Room: REV: Comments: CONTACT INFORMATION Name Title PhoneE-Mail Address Site: IP Settings IP Address: Subnet Mask: Default Gateway: Remote...
Page 85
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL Section 3-11 - Loading Base System Software 3-25 Section 3-10 Insite IP Address Configuration This information needed to facilitate the OnLine Center engineers in connecting to the system is found in the InSite Installation Manual. Section 3-11 Loading Base System Software 3-11-1 Loading Base system Software For more information on loading base system software, refer to Chapter 8. Section 3-12 Warnings 3-12-1 Warnings 3-12-1-1 USB...
Page 86
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL 3-26 Section 3-12 - Warnings Section 3-13 Paperwork NOTE: During and after installation, the documentation (i.e. User Manuals, Installation Manuals...) for the peripheral units must be ke pt as part of the original system do cumentation. This will ensure that all relevant safety and user information is available during the operation and service of the complete system. 3-13-1 Product Locator Installation NOTE: The Product Locator...
Page 87
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL Section 4-1 - Overview 4 - 1 Chapter 4 Functional Checks Section 4-1 Overview 4-1-1 Purpose of Chapter 4 This chapter provides procedures for quickly checking major functions of the LOGIQ™ P5/A5/A5Pro scanner diagnostics by using the built-in service software, and power supply adjustments. Section 4-2 Required Equipment • An empty (blank) DVD-R or CD-R Disk. • At least one transducer. Table 4-1 Contents in Chapter 4...
Page 88
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL 4 - 2 Section 4-2 - Required Equipment Section 4-3 General Procedure 4-3-1 Power On/Boot Up NOTE: After turning off the system, wait at least ten seconds before turning it on again. The system may not be able to boot if power is recycled too quickly. 4-3-1-1 Scanner Power On 1.) Connect the Main Power cable to an appropriate mains power outlet. 2.) Switch ON the Main Circuit Breaker at the rear of the System. When power is applied to...
Page 89
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL Section 4-2 - Required Equipment 4 - 3 4-3-1-1 Scanner Power On (cont’d) 3.) Press the ON/OFF key at the front of the System once. Figure 4-2 Power On/Off Standby Switch Location
Page 90
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL 4 - 4 Section 4-2 - Required Equipment 4-3-2 Power Off / Shutdown NOTE: After turning off a system, wait at least ten seco nds before turning it on again. The system may not be able to boot if power is recycled too quickly. 4-3-2-1 Scanner Shutdown 1.) Press the On/Off Key at the front of the System once to display the SYSTEM - EXIT menu. 2.) Select Shutdown from the SYSTEM - EXIT menu. 4-3-2-2 Switch off the System 1.) Switch OFF...