Page 91

GE HEALTHCARE
DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL
Section 4-2 - Required Equipment 4 - 5
4-3-3 System Features
4-3-3-1 Control Panel for LOGIQ™ P5
1.) Power on/off switch
2.) Alpha Numeric key
3.) Patient Key
4.) TGC
5.) Track Ball
6.) Function Select key
7.) Freeze Key
8.) Gain Knob
9.) User Define Key Figure 4-4   Control Panel Tour LOGIQ™ P5
1
2
3
4
5
6 7
8
9
10 

Page 92

GE HEALTHCARE
DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL 
4 - 6 Section 4-2 - Required Equipment
4-3-3-2 Control Panel for LOGIQ™ A5/LOGIQ™ A5Pro
1.) Power on/off switch
2.) Alpha Numeric key
3.) Patient Key
4.) TGC
5.) Track Ball
6.) Function Select key
7.) Freeze Key
8.) Gain Knob
9.) User Define Key
NOTE: If the LOGIQ™ A5 have been upgrade with Color  Upgrade kit, LOGIQ™ A5 have same keyboard as 
LOGIQ™ P5.
NOTE: LOGIQ™ A5Pro has the same control panel  of LOGIQ™ A5, but PW Mode button is...

Page 93

GE HEALTHCARE
DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL
Section 4-2 - Required Equipment 4 - 7
4-3-3-3 Monitor Display for LOGIQ™ P5
Figure 4-6   P5 Monitor Display Tour
1. Institution/Hospit al Name, Date, Time, 
Operator Identification.
2.  Patient Name, Patient Identification.
3. Power Output Readout
4.  GE Symbol: Probe Orientation Marker.
5. Image Preview.
6. Gray/Color Bar.
7. Cine Gauge.
8.  Measurement Summary Window.
9. Image.
10. Measurement Calipers.
11. Measurement Results...

Page 94

GE HEALTHCARE
DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL 
4 - 8 Section 4-2 - Required Equipment
4-3-4 System B/M-Mode Checks
For a basic functional check of the system’s different modes,  the Quick Guide will familiarize you with 
image optimization for  B-Mode, M-Mode , Color Flow, and Doppler .
Table 4-2    B/M-Mode Functions
Power Output (Acoustic Power) Optimizes image quality and allows user to reduce beam 
intensity. 2% increments between 0-100%.
Dynamic Range Dynamic Range controls...

Page 95

GE HEALTHCARE
DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL
Section 4-2 - Required Equipment 4 - 9
4-3-5 System CFM and PWD Checks 
Table 4-3    CFM and PWD Functions
BaselineAdjusts the baseline to accommodate faster or slower blood 
flows to eliminate aliasing.
PRF/Wall Filter Velocity scale determines pulse
 repetition frequency. If the 
sample volume gate range exceeds single gate PRF capability, 
the system automatically switches to high PRF mode. Multiple 
gates appear, and HPRF is...

Page 96

GE HEALTHCARE
DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL 
4 - 10 Section 4-2 - Required Equipment
4-3-6 Basic Measurements
NOTE: The following instructions assume that you first scan the patient and then press  Freeze.
4-3-6-1 Distance and Tissue Depth Measurements 1.) Press MEASURE
 once an active caliper displays.
2.) To position the active caliper at the start poin t (distance) or the most anterior point (tissue 
depth), move the TRACKBALL
.
3.) To fix the start point, press  Set. The...

Page 97

GE HEALTHCARE
DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL
Section 4-2 - Required Equipment 4 - 11
4-3-6-3 Worksheets
Measurement/Calculation worksheets are available to  display and edit measurements and calculations. 
There are generic worksheets as well as Application specific worksheets.  

Page 98

GE HEALTHCARE
DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL 
4 - 12 Section 4-2 - Required Equipment
4-3-7 Probe/Connectors Usage
4-3-7-1 Connecting a probe
1.) Place the probes carrying case on a stable surface and open the case.
2.) Carefully remove the prob e and unwrap the probe cable.
3.) DO NOT allow the probe head to hang free. Impact to the probe head could result in irreparable  damage.
4.) Turn the connector locking handle counterclockwise.
5.) Align the connector with the probe port...

Page 99

GE HEALTHCARE
DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL
Section 4-2 - Required Equipment 4 - 13
4-3-8 Using Cine
4-3-8-1 Activating CINE
Press FREEZE
, then roll the TRACKBALL to activate CINE. To start CINE Loop playback, press Run/
Stop. To stop CINE Loop playback. press Run/Stop.
4-3-8-2 Quickly Move to Start/End Frame Press FIRST
 to move to the first CINE frame; press LAST to move to the last CINE frame.
4-3-8-3 Start Frame/End Frame Turn the START FRAME
 dial to the left to move to the...

Page 100

GE HEALTHCARE
DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL 
4 - 14 Section 4-2 - Required Equipment
4-3-9 Image Management (QG) 
For Image Management functionality refer to the LOGIQ™ P5/A5/A5Pro Quick Guide. It talks about 
several topics:
• Clipboard
• Printing Images
• Browsing and Managing an Exam’s Stored Image
• Connectivity, and Dataflow Concept and Creation
• Starting an Exam
• Configuring Connectivity
•TCP/IP
• Services (Destinations)
• Buttons
• Views
Verifying and Pinging a Device 
Start reading GE Logiq P5 Service Manual