GE Logiq P5 Service Manual
Here you can view all the pages of manual GE Logiq P5 Service Manual. The GE manuals for Medical Equipment are available online for free. You can easily download all the documents as PDF.
Page 91
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL Section 4-2 - Required Equipment 4 - 5 4-3-3 System Features 4-3-3-1 Control Panel for LOGIQ™ P5 1.) Power on/off switch 2.) Alpha Numeric key 3.) Patient Key 4.) TGC 5.) Track Ball 6.) Function Select key 7.) Freeze Key 8.) Gain Knob 9.) User Define Key Figure 4-4 Control Panel Tour LOGIQ™ P5 1 2 3 4 5 6 7 8 9 10
Page 92
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL 4 - 6 Section 4-2 - Required Equipment 4-3-3-2 Control Panel for LOGIQ™ A5/LOGIQ™ A5Pro 1.) Power on/off switch 2.) Alpha Numeric key 3.) Patient Key 4.) TGC 5.) Track Ball 6.) Function Select key 7.) Freeze Key 8.) Gain Knob 9.) User Define Key NOTE: If the LOGIQ™ A5 have been upgrade with Color Upgrade kit, LOGIQ™ A5 have same keyboard as LOGIQ™ P5. NOTE: LOGIQ™ A5Pro has the same control panel of LOGIQ™ A5, but PW Mode button is...
Page 93
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL Section 4-2 - Required Equipment 4 - 7 4-3-3-3 Monitor Display for LOGIQ™ P5 Figure 4-6 P5 Monitor Display Tour 1. Institution/Hospit al Name, Date, Time, Operator Identification. 2. Patient Name, Patient Identification. 3. Power Output Readout 4. GE Symbol: Probe Orientation Marker. 5. Image Preview. 6. Gray/Color Bar. 7. Cine Gauge. 8. Measurement Summary Window. 9. Image. 10. Measurement Calipers. 11. Measurement Results...
Page 94
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL 4 - 8 Section 4-2 - Required Equipment 4-3-4 System B/M-Mode Checks For a basic functional check of the system’s different modes, the Quick Guide will familiarize you with image optimization for B-Mode, M-Mode , Color Flow, and Doppler . Table 4-2 B/M-Mode Functions Power Output (Acoustic Power) Optimizes image quality and allows user to reduce beam intensity. 2% increments between 0-100%. Dynamic Range Dynamic Range controls...
Page 95
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL Section 4-2 - Required Equipment 4 - 9 4-3-5 System CFM and PWD Checks Table 4-3 CFM and PWD Functions BaselineAdjusts the baseline to accommodate faster or slower blood flows to eliminate aliasing. PRF/Wall Filter Velocity scale determines pulse repetition frequency. If the sample volume gate range exceeds single gate PRF capability, the system automatically switches to high PRF mode. Multiple gates appear, and HPRF is...
Page 96
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL 4 - 10 Section 4-2 - Required Equipment 4-3-6 Basic Measurements NOTE: The following instructions assume that you first scan the patient and then press Freeze. 4-3-6-1 Distance and Tissue Depth Measurements 1.) Press MEASURE once an active caliper displays. 2.) To position the active caliper at the start poin t (distance) or the most anterior point (tissue depth), move the TRACKBALL . 3.) To fix the start point, press Set. The...
Page 97
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL Section 4-2 - Required Equipment 4 - 11 4-3-6-3 Worksheets Measurement/Calculation worksheets are available to display and edit measurements and calculations. There are generic worksheets as well as Application specific worksheets.
Page 98
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL 4 - 12 Section 4-2 - Required Equipment 4-3-7 Probe/Connectors Usage 4-3-7-1 Connecting a probe 1.) Place the probes carrying case on a stable surface and open the case. 2.) Carefully remove the prob e and unwrap the probe cable. 3.) DO NOT allow the probe head to hang free. Impact to the probe head could result in irreparable damage. 4.) Turn the connector locking handle counterclockwise. 5.) Align the connector with the probe port...
Page 99
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL Section 4-2 - Required Equipment 4 - 13 4-3-8 Using Cine 4-3-8-1 Activating CINE Press FREEZE , then roll the TRACKBALL to activate CINE. To start CINE Loop playback, press Run/ Stop. To stop CINE Loop playback. press Run/Stop. 4-3-8-2 Quickly Move to Start/End Frame Press FIRST to move to the first CINE frame; press LAST to move to the last CINE frame. 4-3-8-3 Start Frame/End Frame Turn the START FRAME dial to the left to move to the...
Page 100
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL 4 - 14 Section 4-2 - Required Equipment 4-3-9 Image Management (QG) For Image Management functionality refer to the LOGIQ™ P5/A5/A5Pro Quick Guide. It talks about several topics: • Clipboard • Printing Images • Browsing and Managing an Exam’s Stored Image • Connectivity, and Dataflow Concept and Creation • Starting an Exam • Configuring Connectivity •TCP/IP • Services (Destinations) • Buttons • Views Verifying and Pinging a Device