GE Logiq P5 Service Manual
Here you can view all the pages of manual GE Logiq P5 Service Manual. The GE manuals for Medical Equipment are available online for free. You can easily download all the documents as PDF.
Page 51
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL Section 2-2 - General Console Requirements 2 - 3 2-2-2 Electrical Requirements NOTE: GE Medical Systems requires a dedicated power and ground for the proper operation of its Ultrasound equipment. This dedicated power shall originate at the last distribution panel before the system. Sites with a mains power system with defined Neutral and Live: The dedicated line shall consist of one phase, a neutral (not shared with any other...
Page 52
GE HEALTHCARE DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL 2 - 4 Section 2-2 - General Console Requirements 2-2-2-4 Site Power Outlets A desiccated AC power outlet must be within reach of the unit without extens ion cords. Other outlets adequate for the external peripherals, medical and test equipment needed to support this unit must also be present within 1 m (3.2 ft.) of th e unit. Electrical installation must meet all current local, state, and national electrical codes. 2-2-2-5 Unit...
Page 53
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL Section 2-2 - General Console Requirements 2 - 5 2-2-3 EMI Limitations Ultrasound machines are susceptible to Electromagnetic Interference (EMI) from radio frequencies, magnetic fields, and transient in the air wiring. They also generate EMI. The LOGIQ™ P5/A5/A5Pro complies with limits as stated on the EMC label. Howe ver there is no guarantee that interface will not occur in a particular installation. Possible EMI sources should be...
Page 54
GE HEALTHCARE DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL 2 - 6 Section 2-2 - General Console Requirements 2-2-4 Scan Probe Environmental Requirements Operation:10° to 30° C Storage:--10° to 60° C NOTE: Temperature in degrees C. Conversion to Degrees F = (Degrees C * 9/5) + 32). NOTICESYSTEMS AND ELECTRONIC PROBES ARE DESIGNED FOR STORAGE TEMPERATURES OF -10 TO + 60 degrees C. WHEN EXPOSED TO LARGE TEMPERATURE VARIATIONS, THE PRODUCT SHOULD BE KEPT IN ROOM TE MPERATURE FOR 10 HOURS BEFORE...
Page 55
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL Section 2-3 - Facility Needs 2 - 7 Section 2-3 Facility Needs 2-3-1 Purchaser Responsibilities The work and materials needed to prepare the site is the responsibility of the purchaser. Delay, confusion, and waste of manpower can be avoided by completing pre installation work before delivery. User the Pre Installation checklist to veri fy that all needed steps have been taken, Purchaser reasonability includes: • Procuring the...
Page 56
GE HEALTHCARE DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL 2 - 8 Section 2-3 - Facility Needs 2-3-2 Required Features NOTE: GE Medical Systems requires a dedicated power and ground for the proper operation of its Ultrasound equipment. This dedicated power shall originate at the last distribution panel before the system. Sites with a mains power system with defined Neutral and Live: The dedicated line shall consist of one phase, a neutral (not shared with any other circuit), and a full size...
Page 57
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL Section 2-3 - Facility Needs 2 - 9 2-3-4 Recommended and Alternate Ultrasound Room Layout Recommended standard floor plan and a mini mal floor plan for ultrasound equipment: Figure 2-1 RECOMMENDED ULTRASOUND ROOM LAYOUT LINEN SUPPLY EXAMINATION TABLE FILM VIEWER COUNTER TOP FILE CABINET SECRETARYS OR DOCTOR’S DESK FILM PROCESSING ROOM, OVERHEAD LIGHTS DIMMER DEDICATED POWER RECEPTACLE EXTERNAL PERIPHERALS SINK COUNTER TOP FILM...
Page 58
GE HEALTHCARE DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL 2 - 10 Section 2-3 - Facility Needs 2-3-5 Networking Pre-installation Requirements 2-3-5-1 Purpose of DICOM Network Function DICOM services provide the operator with clinically useful features for moving images and patient information over a hospital network. Examples of DI COM services include the transfer of images to workstations for viewing or transferring images to remote printers. As an added benefit, transferring images in...
Page 59
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL Section 2-3 - Facility Needs 2 - 11 2-3-5-2 DICOM Option Pre-installation Requirements (cont’d) . Figure 2-2 Worksheet for DICOM Network Information LOGIQ™ Host Name AE TitleLocal PortIP Address Net Mask. . .. . .. ROUTING INFORMATION ROUTER2 ROUTER3 ROUTER1 . .. . .. . .. . .. . .. . .. . .. Destination IP Addresses GATEWAY IP Addresses Default DICOM APPLICATION INFORMATION NAME Store 2 Store 3 Store 1 MAKE/REVISION...
Page 60
GE HEALTHCARE DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL 2 - 12 Section 2-3 - Facility Needs This page was intentionally left blank.