GE Logiq P5 Service Manual

Here you can view all the pages of manual GE Logiq P5 Service Manual. The GE manuals for Medical Equipment are available online for free. You can easily download all the documents as PDF.

Page 51

GE HEALTHCARE
DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL
Section 2-2 - General Console Requirements 2 - 3
2-2-2 Electrical Requirements
NOTE: GE Medical Systems requires a dedicated power and ground for the proper operation of its 
Ultrasound equipment. This dedicated power shall originate at the last distribution panel before 
the system.
Sites with a mains power system with defined Neutral and Live:
The dedicated line shall consist of  one phase, a neutral (not shared with any other...

Page 52

GE HEALTHCARE
DIRECTION 5394141, REVISION  1LOGIQ™ P5 SERVICE MANUAL 
2 - 4 Section 2-2 - General Console Requirements
2-2-2-4 Site Power Outlets
A desiccated AC power outlet must be within reach of the unit without extens ion cords. Other outlets 
adequate for the external peripherals, medical and test equipment needed to support this unit must also 
be present within 1 m (3.2 ft.) of th e unit. Electrical installation must meet all current local, state, and 
national electrical codes.
2-2-2-5 Unit...

Page 53

GE HEALTHCARE
DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL
Section 2-2 - General Console Requirements 2 - 5
2-2-3 EMI Limitations
Ultrasound machines are susceptible to Electromagnetic Interference (EMI) from radio frequencies, 
magnetic fields, and transient in the air wiring.  They also generate EMI. The LOGIQ™ P5/A5/A5Pro 
complies with limits as stated on the EMC label. Howe ver there is no guarantee that interface will not 
occur in a particular installation.
Possible EMI sources should be...

Page 54

GE HEALTHCARE
DIRECTION 5394141, REVISION  1LOGIQ™ P5 SERVICE MANUAL 
2 - 6 Section 2-2 - General Console Requirements
2-2-4 Scan Probe Environmental Requirements
Operation:10° to 30° C
Storage:--10° to 60° C
NOTE: Temperature in degrees C. Conversion to Degrees F = (Degrees C * 9/5) + 32).
NOTICESYSTEMS AND ELECTRONIC PROBES ARE  DESIGNED FOR STORAGE TEMPERATURES 
OF -10 TO + 60 degrees C. WHEN EXPOSED TO LARGE TEMPERATURE VARIATIONS, THE 
PRODUCT SHOULD BE KEPT IN ROOM TE MPERATURE FOR 10 HOURS BEFORE...

Page 55

GE HEALTHCARE
DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL
Section 2-3 - Facility Needs 2 - 7
Section 2-3
Facility Needs
2-3-1 Purchaser Responsibilities
The work and materials needed to prepare the site is the responsibility of the purchaser. Delay, 
confusion, and waste of manpower can be avoided by  completing pre installation work before delivery. 
User the Pre Installation checklist to veri fy that all needed steps have been taken,
Purchaser reasonability includes:
• Procuring the...

Page 56

GE HEALTHCARE
DIRECTION 5394141, REVISION  1LOGIQ™ P5 SERVICE MANUAL 
2 - 8 Section 2-3 - Facility Needs
2-3-2 Required Features
NOTE: GE Medical Systems requires a dedicated power and ground for the proper operation of its 
Ultrasound equipment. This dedicated power shall originate at the last distribution panel before 
the system.
Sites with a mains power system with defined Neutral and Live:
The dedicated line shall consist of one phase, a neutral (not shared with any other circuit), and a full 
size...

Page 57

GE HEALTHCARE
DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL
Section 2-3 - Facility Needs 2 - 9
2-3-4 Recommended and Alternate Ultrasound Room Layout
Recommended standard floor plan and a mini mal floor plan for ultrasound equipment:
Figure 2-1   RECOMMENDED ULTRASOUND ROOM LAYOUT
LINEN SUPPLY
EXAMINATION
TABLE
FILM VIEWER
COUNTER TOP    FILE
CABINET
SECRETARYS OR
DOCTOR’S DESK
FILM 
PROCESSING 
ROOM,
   OVERHEAD
LIGHTS DIMMER
DEDICATED POWER
RECEPTACLE
EXTERNAL
PERIPHERALS
SINK
COUNTER
TOP
FILM...

Page 58

GE HEALTHCARE
DIRECTION 5394141, REVISION  1LOGIQ™ P5 SERVICE MANUAL 
2 - 10 Section 2-3 - Facility Needs
2-3-5 Networking Pre-installation Requirements
2-3-5-1 Purpose of DICOM Network Function
DICOM services provide the operator  with clinically useful features for moving images and patient 
information over a hospital network. Examples of DI COM services include the transfer of images to 
workstations for viewing or transferring images to  remote printers. As an added benefit, transferring 
images in...

Page 59

GE HEALTHCARE
DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL
Section 2-3 - Facility Needs 2 - 11
2-3-5-2              DICOM Option Pre-installation Requirements  (cont’d)
.
Figure 2-2   Worksheet for DICOM Network Information
LOGIQ™ 
Host Name
AE TitleLocal PortIP Address
Net Mask. .
..
.
..
ROUTING INFORMATION
ROUTER2
ROUTER3 ROUTER1
.
..
.
..
.
..
.
..
.
..
.
..
.
.. Destination 
IP Addresses
GATEWAY IP Addresses
Default
DICOM APPLICATION INFORMATION NAME
Store 2 Store 3 Store 1
MAKE/REVISION...

Page 60

GE HEALTHCARE
DIRECTION 5394141, REVISION  1LOGIQ™ P5 SERVICE MANUAL 
2 - 12 Section 2-3 - Facility Needs
This page was intentionally left blank. 