GE Logiq 7 Service Manual
Here you can view all the pages of manual GE Logiq 7 Service Manual. The GE manuals for Medical Equipment are available online for free. You can easily download all the documents as PDF.
Page 191
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL Section 6-6 - BW Printer Setting / Adjustment 6-29 6-6-2 Parameters for UP-D895 1.) When system started, Go to Utility > System >Peripherals, then hit Printers button in the Setup group. 2.) Click SONY UP-D895 in the Printer List, then right click and select Properties . 3.) Select General tab, then hit Printing Preferences ... 4.) Set Landscape into Orientation, then click on Advanced... 5.) Set 1920x1280 into PaperFigure 6-37...
Page 192
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL 6-30Section 6-6 - BW Printer Setting / Adjustment 6-6-2 Parameters for UP-D895 (cont’d) 6.) Click on Scaling..., then check off into Enlarge to Paper and set Bilinear into Interpolation Method. 7.) Click on Density Adjust... , set TONE1 into GAMMA and set 7 into Sharpness. 6-6-3 Service Tips - Print Quality When printer is set to “Nearest Neighbor” interepolation method, comments and body pattern may become hard to read....
Page 193
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL Section 6-6 - BW Printer Setting / Adjustment 6-31 6-6-4 Service Tips - Print Speed 1.) Make sure that the UP-D897 is switched ON. (The back light of the LCD light is in green.) If not, press the ON/OFF switch. 2.) Confirm that “READY” is displayed on the LCD, then press the jog dial. The UP-D897 enters the menu mode. When you perform a menu operation for the first time after you purchase the unit , the top item of the “BEEP” menu...
Page 194
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL 6-32Section 6-6 - BW Printer Setting / Adjustment 6-6-4 Service Tips - Print Speed (cont’d) 3.) Turn the jog dial up or down until “ COMD” is displayed on the LCD. 4.) Press the jog dial. “ CO:897” is displayed. The UP-D897 enters the mode in which you can select the driver setting. CO:897, which is the driver setting for D897, is displayed on the LCD. You need to change to CO:895 which is the driver setting for the D895. 5.) Turn the...
Page 195
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL Section 6-7 - Cleaning the Trackball 6-33 Section 6-7 Cleaning the Trackball 1.) Power OFF the scanner. 2.) Place your fingers onto the notches of the trackball retainer ring. 3.) Rotate the retainer ring counterclockwise until it can be removed from the keyboard. 4.) Lift off the inner retainer and trackball from the keyboard.Figure 6-41 Rotating the Retainer Ring Figure 6-42 Removing Inner Retainer and Trackball Retainer Ring Inner...
Page 196
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL 6-34Section 6-7 - Cleaning the Trackball Section 6-7 Cleaning the Trackball (cont’d) 5.) Wipe off any oil or dust from the trackball using a cleaner or dry cloth. 6.) Wipe off any oil or dust from the trackball housing, rollers, and small ball, using a cleaner or cotton bud. NOTICEWhen cleaning the trackball housing, make sure not to spill or spray any liquid into the trackball housing (Keyboard or system). Avoid organic solvents...
Page 197
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL Section 6-7 - Cleaning the Trackball 6-35 Section 6-7 Cleaning the Trackball (cont’d) 7.) Wipe off any oil or dust from the two rollers using a cleaner or cotton bud. NOTICEWhen cleaning the roller, make sure not to spill or spray any liquid into the trackball housing (Keyboard or system). Use either ethanol, isopropyl alcohol or VCR head cleaner to clean the trackball assembly. Avoid other solvents that may damage the mechanical...
Page 198
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL 6-36Section 6-7 - Cleaning the Trackball Section 6-7 Cleaning the Trackball (cont’d) 8.) Insert the trackball into the housing. 9.) Place the trackball and inner retainer into the housing with its stopper facing down.Lift off the inner retainer and trackball from the keyboard. 10.) Install the trackball retainer ring onto the inner retainer, then rotate it clockwise until its notches are set in the horizontal position. Figure 6-45...
Page 199
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL Section 6-8 - Jumper and Dip Switch Setting 6-37 Section 6-8 Jumper and Dip Switch Setting 6-8-1 Dip Switch Setting Normally the dip switches shall not be adjusted. In case of special needs, such as scanner upgrade, refer to each instructions/documentations for proper setting. The dip switch must be set properly according to the table below. The value in the table (Dip Switch setting) represents the bit location as shown. Figure 6-47...
Page 200
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL 6-38Section 6-8 - Jumper and Dip Switch Setting 6-8-1 Dip Switch Setting (cont’d) Figure 6-49 Bit Settings (5 to 8) 0 (h) 1 (h) 2 (h) 3 (h) 4 (h) 5 (h) 6 (h) 7 (h) O N 1 2 3 4 5 6 7 8O N 1 2 3 4 5 6 7 8O N 1 2 3 4 5 6 7 8O N 1 2 3 4 5 6 7 8O N 1 2 3 4 5 6 7 8O N 1 2 3 4 5 6 7 8O N 1 2 3 4 5 6 7 8O N 1 2 3 4 5 6 7 8 O N 1 2 3 4 5 6 7 8O N 1 2 3 4 5 6 7 8O N 1 2 3 4 5 6 7 8O N 1 2 3 4 5 6 7 8O N 1 2 3 4 5 6 7 8O N 1 2 3 4 5 6 7 8O N 1 2...