Page 231

GE HEALTHCARE
DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
Section 8-2 - Software Loading Procedures for BT09 (R8.x.x or later) 8-25
8-2-8-5 TCP/IP Filter Setup (cont’d)
7.) If your system has no TCP port numbers as follows, perform these procedure to add the TCP port numbers. If the TCP port already have been set, skip these procedures.  (This depends on the 
network card installed in the PC box.)
* At first check mark  Enable TCP/IP Filtering (All adapters) .
*Check mark  Permit Only in TCP...

Page 232

GE HEALTHCARE
DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
8-26Section 8-2 - Software Loading Procedures for BT09 (R8.x.x or later)
8-2-8-5 TCP/IP Filter Setup (cont’d)
8.) Verify that the proper TCP port numbers are entered in TCP Ports box, then click  OK buttons for all 
opened window to complete the TCP/IP Filter Setup .
9.) The following dialog box appears. Click  No to proceed the other settings. (After completion of all 
system checks, the system is shut down.) Figure 8-82   TCP/IP Filter...

Page 233

GE HEALTHCARE
DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
Section 8-2 - Software Loading Procedures for BT09 (R8.x.x or later) 8-27
8-2-8-6 Date and Time Setting
1.) Select Start > Settings > Control Panel > Date and Time. I
2.) Click on  Time Zone tab.
3.) Select the appropriate time zone.  Then click on  Apply.
4.) Then click on Date & Time tab. Figure 8-84   Date and Time
Figure 8-85   Time Zone 

Page 234

GE HEALTHCARE
DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
8-28Section 8-2 - Software Loading Procedures for BT09 (R8.x.x or later)
8-2-8-6 Date and Time Setting (cont’d)
5.) Set the current date and time.
6.) Click on  Apply.
7.) Click on  OK to close the Date and Time window.
8.) Select Start > Shutdown > OK  to power OFF the scanner.
Figure 8-86   Date & Time
Month and Year
Day Time 

Page 235

GE HEALTHCARE
DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
Section 8-2 - Software Loading Procedures for BT09 (R8.x.x or later) 8-29
8-2-9 Installing R8.x.x Application Software
8-2-9-1 Installing R8.x.x Application Software
1.) Power ON the scanner. The windows desktop appears. 
2.) Insert the Application Software CD-ROM into the DVD drive. 
NOTICEYou must install BT09 Base image before Application Software installation. If not, Application shows the 
error message, “Service platform is out of...

Page 236

GE HEALTHCARE
DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
8-30Section 8-2 - Software Loading Procedures for BT09 (R8.x.x or later)
8-2-9-1 Installing R8.x.x Application Software  (cont’d)
3.) Double-click on  LoadSoftware.bat.
4.) The Installation program will start, and the command prompt screen will be shown on the monitor  as shown.
5.) Select TYPE A - BT09 with BEP4 does not have PRO version available. Figure 8-88   LoadSoftware.bat
NOTICE1.If wrong application software installed, the...

Page 237

GE HEALTHCARE
DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
Section 8-2 - Software Loading Procedures for BT09 (R8.x.x or later) 8-31
8-2-9-1 Installing R8.x.x Application Software (cont’d)
6.) Type  Y key.
NOTE: Press the N key to cancel this procedures and to power OFF the system.
NOTE: When the Information Window appears asking you to select OK, do nothing. This window will  disappear automatically.
7.) The installation procedures will start automatically. It will take 10 minutes.
8.) After...

Page 238

GE HEALTHCARE
DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
8-32Section 8-2 - Software Loading Procedures for BT09 (R8.x.x or later)
8-2-10 System Settings
After application is successfully installed, various LOGIQ™ 7 system parameters must be properly set.
Table 8-7   System Settings
ItemMethodProper Value
Software Version Utility
    
System
        
About s/w version : R8.x.x
Image Part No. : 5305206-3
Regional Preset
Utility
    
System
        
About
             
Additional About...

Page 239

GE HEALTHCARE
DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
Section 8-2 - Software Loading Procedures for BT09 (R8.x.x or later) 8-33
Language (For Japanese
Only)Utility
 
System 
General
 
Language
  
Regional Option
  
Regional Option (tab)
  
Standard and formats
  
Location
  
Language (tab)
  
Details   
Settings (tab)
  
Installed Service
  
Default Input Language
  
Advanced (tab)
  
Language JPN
Japanese
Japan
Japanese Keyboard
Japanese
Japanese
Language (For Russian/
Greek Only)Utility...

Page 240

GE HEALTHCARE
DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
8-34Section 8-2 - Software Loading Procedures for BT09 (R8.x.x or later)
8-2-10-1 Peripheral Device Connection/Setup
1.) Connect all of the peripheral devices to the scanner.2.) Power ON all of the peripheral devices.
3.) Verify that the system boots up with no error. (It will take a few minutes.) 4.) Set up customer data saved at 8-2-6 Pre-installation Procedures.
8-2-10-2 Confirmation of the software version 1.) Touch the  Utility...