Page 231
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL Section 8-2 - Software Loading Procedures for BT09 (R8.x.x or later) 8-25 8-2-8-5 TCP/IP Filter Setup (cont’d) 7.) If your system has no TCP port numbers as follows, perform these procedure to add the TCP port numbers. If the TCP port already have been set, skip these procedures. (This depends on the network card installed in the PC box.) * At first check mark Enable TCP/IP Filtering (All adapters) . *Check mark Permit Only in TCP...
Page 232
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL 8-26Section 8-2 - Software Loading Procedures for BT09 (R8.x.x or later) 8-2-8-5 TCP/IP Filter Setup (cont’d) 8.) Verify that the proper TCP port numbers are entered in TCP Ports box, then click OK buttons for all opened window to complete the TCP/IP Filter Setup . 9.) The following dialog box appears. Click No to proceed the other settings. (After completion of all system checks, the system is shut down.) Figure 8-82 TCP/IP Filter...
Page 233
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL Section 8-2 - Software Loading Procedures for BT09 (R8.x.x or later) 8-27 8-2-8-6 Date and Time Setting 1.) Select Start > Settings > Control Panel > Date and Time. I 2.) Click on Time Zone tab. 3.) Select the appropriate time zone. Then click on Apply. 4.) Then click on Date & Time tab. Figure 8-84 Date and Time Figure 8-85 Time Zone
Page 234
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL 8-28Section 8-2 - Software Loading Procedures for BT09 (R8.x.x or later) 8-2-8-6 Date and Time Setting (cont’d) 5.) Set the current date and time. 6.) Click on Apply. 7.) Click on OK to close the Date and Time window. 8.) Select Start > Shutdown > OK to power OFF the scanner. Figure 8-86 Date & Time Month and Year Day Time
Page 235
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL Section 8-2 - Software Loading Procedures for BT09 (R8.x.x or later) 8-29 8-2-9 Installing R8.x.x Application Software 8-2-9-1 Installing R8.x.x Application Software 1.) Power ON the scanner. The windows desktop appears. 2.) Insert the Application Software CD-ROM into the DVD drive. NOTICEYou must install BT09 Base image before Application Software installation. If not, Application shows the error message, “Service platform is out of...
Page 236
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL 8-30Section 8-2 - Software Loading Procedures for BT09 (R8.x.x or later) 8-2-9-1 Installing R8.x.x Application Software (cont’d) 3.) Double-click on LoadSoftware.bat. 4.) The Installation program will start, and the command prompt screen will be shown on the monitor as shown. 5.) Select TYPE A - BT09 with BEP4 does not have PRO version available. Figure 8-88 LoadSoftware.bat NOTICE1.If wrong application software installed, the...
Page 237
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL Section 8-2 - Software Loading Procedures for BT09 (R8.x.x or later) 8-31 8-2-9-1 Installing R8.x.x Application Software (cont’d) 6.) Type Y key. NOTE: Press the N key to cancel this procedures and to power OFF the system. NOTE: When the Information Window appears asking you to select OK, do nothing. This window will disappear automatically. 7.) The installation procedures will start automatically. It will take 10 minutes. 8.) After...
Page 238
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL 8-32Section 8-2 - Software Loading Procedures for BT09 (R8.x.x or later) 8-2-10 System Settings After application is successfully installed, various LOGIQ™ 7 system parameters must be properly set. Table 8-7 System Settings ItemMethodProper Value Software Version Utility System About s/w version : R8.x.x Image Part No. : 5305206-3 Regional Preset Utility System About Additional About...
Page 239
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL Section 8-2 - Software Loading Procedures for BT09 (R8.x.x or later) 8-33 Language (For Japanese Only)Utility System General Language Regional Option Regional Option (tab) Standard and formats Location Language (tab) Details Settings (tab) Installed Service Default Input Language Advanced (tab) Language JPN Japanese Japan Japanese Keyboard Japanese Japanese Language (For Russian/ Greek Only)Utility...
Page 240
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL 8-34Section 8-2 - Software Loading Procedures for BT09 (R8.x.x or later) 8-2-10-1 Peripheral Device Connection/Setup 1.) Connect all of the peripheral devices to the scanner.2.) Power ON all of the peripheral devices. 3.) Verify that the system boots up with no error. (It will take a few minutes.) 4.) Set up customer data saved at 8-2-6 Pre-installation Procedures. 8-2-10-2 Confirmation of the software version 1.) Touch the Utility...