Page 221
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL Section 8-2 - Software Loading Procedures for BT09 (R8.x.x or later) 8-15 8-2-7 Install Base System (cont’d) 4.) The following screen appears. Press B then Enter to initialize the C drive 5.) Check if the message appears as reloading C: partition only, then press Enter key. If the message appears as reloading the entire disk, press CTRL-ALT-DEL to reboot the sysytem and retry from procedures described in previous page. Figure 8-65...
Page 222
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL 8-16Section 8-2 - Software Loading Procedures for BT09 (R8.x.x or later) 8-2-7 Install Base System (cont’d) 6.) If the following screen appears, verify that OK buttton is selected, then press Enter key. (If it is not selected, select OK button using the arrow keys then press Enter key.) 7.) Approx. fifteen minutes later, the following screen appears. Eject the Base System Software Load Image DVD from the drive. 8.) To restart the...
Page 223
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL Section 8-2 - Software Loading Procedures for BT09 (R8.x.x or later) 8-17 8-2-8 Modify System/Windows Settings After Base System is successfully installed, various parameters must be properly set before proceeding to Application load. The below outlines necessary system/windows setting. 8-2-8-1 UPS Setting UPS automatically setup by Ghost. You do not need to setup it manually. You can see the message below on the desktop. Please...
Page 224
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL 8-18Section 8-2 - Software Loading Procedures for BT09 (R8.x.x or later) 8-2-8-2 Drive Letter Setting If you CANNOT see this dialog at the first boot up, this dialog appears at 2nd boot up. Please choose OK. You should check the Drive Letter manually by Windows Explorer for double check. - DVD Drive should be assigned to G: If Drive letter is wrong, please double-click Remap Drive Letter icon on the desktop. 8-2-8-3 Hardware Device...
Page 225
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL Section 8-2 - Software Loading Procedures for BT09 (R8.x.x or later) 8-19 2.) The System Properties window appears. Select Hardware > Divice Manager... 3.) Verify that X or ! is not displayed. Figure 8-72 Divice Manager NOTICEThe Eusc (Mlultifunction adapters > Eusc) might contain “ !” mark, but it is no problem. Skip any of device driver installation procedures described below. “!” mark on Eusc is no problem.
Page 226
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL 8-20Section 8-2 - Software Loading Procedures for BT09 (R8.x.x or later) 8-2-8-3 Hardware Device Check: (cont’d) 4.) If X or ! is displayed: - Left-click on the device with x mark, then rigtht-click on Enable. - Left-click on the device with ! mark, then rigtht-click on Properties to perform the device driver reconfiguration. When some files are required, designate the file in the directories of C:\WINDOWS\SYSTEM32 or...
Page 227
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL Section 8-2 - Software Loading Procedures for BT09 (R8.x.x or later) 8-21 8-2-8-4 LAN Parameter Setting 1.) On the desktop, left-click on My Network Places, then right-click on Properties... 2.) Verify that two network icons appears. (If “Local Area Connection” icon does not appear, the hardware failure might occur.) 3.) Using the right trackball button, double-click on Local Area Connection. 4.) When the following window appears, it...
Page 228
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL 8-22Section 8-2 - Software Loading Procedures for BT09 (R8.x.x or later) 8-2-8-5 TCP/IP Filter Setup 1.) On the desktop, left-click on My Network Places, then right-click on Properties... 2.) Double-click Local Area Connection or Local Area Connection #2 (if you do NOT find Local Area Connection). 3.) Select Internet Protocol (TCP/IP) , then click Properties button. Figure 8-76 Local Area Connection Figure 8-77 Internet...
Page 229
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL Section 8-2 - Software Loading Procedures for BT09 (R8.x.x or later) 8-23 8-2-8-5 TCP/IP Filter Setup (cont’d) 4.) Click Advanced... button. 5.) Click Options tab.Figure 8-78 Advanced Figure 8-79 Options
Page 230
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL 8-24Section 8-2 - Software Loading Procedures for BT09 (R8.x.x or later) 8-2-8-5 TCP/IP Filter Setup (cont’d) 6.) Select TCP/IP Filtering , then click Properties button. Figure 8-80 TCP/IP Filtering