Page 221

GE HEALTHCARE
DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
Section 8-2 - Software Loading Procedures for BT09 (R8.x.x or later) 8-15
8-2-7 Install Base System (cont’d)
4.) The following screen appears. Press 
B then Enter to initialize the C drive
5.) Check if the message appears as reloading C: partition only, then press Enter  key.
If the message appears as reloading the entire disk, press  CTRL-ALT-DEL to reboot the sysytem 
and retry from procedures described in previous page. Figure 8-65...

Page 222

GE HEALTHCARE
DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
8-16Section 8-2 - Software Loading Procedures for BT09 (R8.x.x or later)
8-2-7 Install Base System (cont’d)
6.) If the following screen appears, verify that  OK buttton is selected, then press  Enter key. (If it is not 
selected, select OK  button using the arrow keys then press Enter  key.)
7.) Approx. fifteen minutes later, the following screen appears. Eject the  Base System Software Load 
Image DVD from the drive.
8.) To restart the...

Page 223

GE HEALTHCARE
DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
Section 8-2 - Software Loading Procedures for BT09 (R8.x.x or later) 8-17
8-2-8 Modify System/Windows Settings
After Base System is successfully installed, various parameters must be properly set before proceeding to Application 
load.  The below outlines necessary system/windows setting.
8-2-8-1 UPS Setting UPS automatically setup by Ghost. You do not need to setup it manually. You can see the message 
below on the desktop.
Please...

Page 224

GE HEALTHCARE
DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
8-18Section 8-2 - Software Loading Procedures for BT09 (R8.x.x or later)
8-2-8-2 Drive Letter Setting
If you CANNOT see this dialog at the first boot up, this dialog appears at 2nd boot up.
Please choose  OK.
You should check the Drive Letter manually by Windows Explorer for double check. - DVD Drive should be assigned to G:
If Drive letter is wrong, please double-click Remap Drive Letter  icon on the desktop.
8-2-8-3 Hardware Device...

Page 225

GE HEALTHCARE
DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
Section 8-2 - Software Loading Procedures for BT09 (R8.x.x or later) 8-19
2.) The System Properties window appears. Select Hardware > Divice Manager...
3.) Verify that X or ! is not displayed.
Figure 8-72   Divice Manager
NOTICEThe Eusc  (Mlultifunction adapters > Eusc) might contain “ !” mark, but it is no problem. Skip any of 
device driver installation procedures described below.
“!” mark on Eusc is no problem. 

Page 226

GE HEALTHCARE
DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
8-20Section 8-2 - Software Loading Procedures for BT09 (R8.x.x or later)
8-2-8-3 Hardware Device Check:  (cont’d)
4.) If  X or  ! is displayed:
- Left-click on the device with  x mark, then rigtht-click on Enable.
- Left-click on the device with  ! mark, then rigtht-click on Properties  to perform the device driver 
reconfiguration. When some files are required, designate the file in the directories of 
C:\WINDOWS\SYSTEM32  or...

Page 227

GE HEALTHCARE
DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
Section 8-2 - Software Loading Procedures for BT09 (R8.x.x or later) 8-21
8-2-8-4 LAN Parameter Setting
1.) On the desktop, left-click on  My Network Places, then right-click on  Properties...
2.) Verify that two network icons appears. (If “Local Area Connection” icon does not appear, the  hardware failure might occur.)
3.) Using the right trackball button, double-click on  Local Area Connection.
4.) When the following window appears, it...

Page 228

GE HEALTHCARE
DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
8-22Section 8-2 - Software Loading Procedures for BT09 (R8.x.x or later)
8-2-8-5 TCP/IP Filter Setup
1.) On the desktop, left-click on  My Network Places, then right-click on  Properties...
2.) Double-click  Local Area Connection  or Local Area Connection #2 (if you do NOT find Local 
Area Connection).
3.)  Select  Internet Protocol (TCP/IP)  , then click Properties button.
Figure 8-76   Local Area Connection
Figure 8-77   Internet...

Page 229

GE HEALTHCARE
DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
Section 8-2 - Software Loading Procedures for BT09 (R8.x.x or later) 8-23
8-2-8-5 TCP/IP Filter Setup (cont’d)
4.) Click Advanced...  button.
5.)  Click  Options tab.Figure 8-78   Advanced
Figure 8-79   Options
 
  

Page 230

GE HEALTHCARE
DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
8-24Section 8-2 - Software Loading Procedures for BT09 (R8.x.x or later)
8-2-8-5 TCP/IP Filter Setup (cont’d)
6.) Select TCP/IP Filtering , then click Properties button.
Figure 8-80   TCP/IP Filtering 