Page 241
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL Section 8-2 - Software Loading Procedures for BT09 (R8.x.x or later) 8-35 8-2-10-2 Confirmation of the software version (cont’d) 2.) Touch the System button on the Touch Panel. 3.) Click on About tab. The software version will be shown on the monitor as shown. Confirm the following: - Software Version ( R 8.x.x) - Preset Resion (Americas, Europe, Asia, Japan, or None) - Image Part Number (described on the Base System Software Image...
Page 242
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL 8-36Section 8-2 - Software Loading Procedures for BT09 (R8.x.x or later) 8-2-10-3 Computer Name and Connection Settings 1.) Touch Utility > Connectivity. 2.) Click on TCP/IP. 3.) Type Computer Name. 4.) Verify that Enable DHCP has no check mark. If checked, remove the mark. 5.) Check if Computer name, IP Address, Subnet Mask, and Default Gateway are proper ones which you wrote down in section 8-2-6-4, Saving Connectivity. 6.) Click...
Page 243
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL Section 8-2 - Software Loading Procedures for BT09 (R8.x.x or later) 8-37 8-2-10-3 Computer Name and Connection Settings (cont’d) 7.) Touch Scan on the Touch Panel to return to the scan screen. 8.) Shut down the scanner. 9.) Power ON the scanner, then press Patient key and verify that the following screen appears. Figure 8-96 Patient
Page 244
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL 8-38Section 8-2 - Software Loading Procedures for BT09 (R8.x.x or later) 8-2-10-4 Check LCD Monitor Hardware Setting For 19 inch Monitor To adjust the brightness: 1.) Press the Toggle button (2) or (3) Once for brightness adjustment. a.) Verify that Brightness is 80 (Dark Room) or 100 (Bright Room) . If it is not, increase brightness to press the adjustment button (3). If it is not, decrease brightness to press the adjustment button...
Page 245
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL Section 8-2 - Software Loading Procedures for BT09 (R8.x.x or later) 8-39 8-2-10-6 VCR Parameters Setting (for the VCR-equipped system ONLY) 1.) Touch Utility > System. 2.) Click on Peripherals tab. 3.) Set the proper parameters: -PAL or NTSC - VCR type 4.) Shut down the scanner, then restart the scanner. 5.) Verify that the VCR can be properly operated with no error (remote, play...) . 8-2-10-7 Regional Setup (For the system with...
Page 246
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL 8-40Section 8-2 - Software Loading Procedures for BT09 (R8.x.x or later) 8-2-10-7 Regional Setup (For the system with language other than English ONLY) (cont’d) 1.) Select Utility> System> General>Regional Option Menu , 2.) Click on Languages tab. 3.) Check Install files for East Asian languages. 4.) Click on Apply. 5.) The system requires the re-boot automatically, shut down the system. Figure 8-100 Regional Option NOTICEYou do...
Page 247
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL Section 8-2 - Software Loading Procedures for BT09 (R8.x.x or later) 8-41 8-2-10-7 Regional Setup (For the system with language other than English ONLY) (cont’d) 6.) In Utility> System> General , set the Language as JPN. Save this setting, but do not reboot the system yet. 7.) Press Regional Options , under Standards and Formats select Japanese, under Location select Japan . Press Apply .Figure 8-101 General Figure 8-102...
Page 248
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL 8-42Section 8-2 - Software Loading Procedures for BT09 (R8.x.x or later) 8-2-10-7 Regional Setup (For the system with language other than English ONLY) (cont’d) 8.) Select the Language tab , press Details, under Installed Services select the Japanese keyboard , under Default input language select Japanese, press Apply , Press OK. 9.) Select the Advanced tab , then select Japanese in the Language for non-Unicode programs pull- down...
Page 249
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL Section 8-2 - Software Loading Procedures for BT09 (R8.x.x or later) 8-43 8-2-10-7 Regional Setup (For the system with language other than English ONLY) (cont’d) 13.) Select Automatic Setup for Japanese Language button. Then IME configuration shall be automatically done. This configuration shall setup the keyboard assignment (remove R
Page 250
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL 8-44Section 8-2 - Software Loading Procedures for BT09 (R8.x.x or later) 8-2-10-7 Regional Setup (For the system with language other than English ONLY) (cont’d) For Russian or Greek Languages 1.) In Utility--> System--> General , set the Language as Russian or Greek. Save this setting, but do not reboot the system yet. 2.) Press Regional Options , under Standards and Formats select Russian or Greek, under Location select Russia or...