Home > GE > Medical Equipment > GE Logiq 7 Service Manual

GE Logiq 7 Service Manual

Here you can view all the pages of manual GE Logiq 7 Service Manual. The GE manuals for Medical Equipment are available online for free. You can easily download all the documents as PDF.

Page 241

GE HEALTHCARE
DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
Section 8-2 - Software Loading Procedures for BT09 (R8.x.x or later) 8-35
8-2-10-2 Confirmation of the software version (cont’d)
2.) Touch the  System button on the Touch Panel.
3.) Click on  About tab. The software version will be shown on the monitor as shown. Confirm the 
following:
- Software Version ( R
8.x.x)
- Preset Resion (Americas, Europe, Asia, Japan, or None)
- Image Part Number (described on the Base System Software Image...

Page 242

GE HEALTHCARE
DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
8-36Section 8-2 - Software Loading Procedures for BT09 (R8.x.x or later)
8-2-10-3 Computer Name and Connection Settings
1.) Touch Utility > Connectivity.
2.) Click on TCP/IP.
3.) Type  Computer Name.
4.) Verify that Enable DHCP  has no check mark. If checked, remove the mark.
5.) Check if Computer name, IP Address, Subnet Mask, and Default Gateway  are proper ones 
which you wrote down in section 8-2-6-4, Saving Connectivity.
6.) Click...

Page 243

GE HEALTHCARE
DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
Section 8-2 - Software Loading Procedures for BT09 (R8.x.x or later) 8-37
8-2-10-3 Computer Name and Connection Settings (cont’d)
7.) Touch  Scan on the Touch Panel to return to the scan screen.
8.) Shut down the scanner.
9.) Power ON the scanner, then press  Patient key and verify that the following screen appears.
Figure 8-96   Patient  

Page 244

GE HEALTHCARE
DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
8-38Section 8-2 - Software Loading Procedures for BT09 (R8.x.x or later)
8-2-10-4 Check LCD Monitor Hardware Setting
For 19 inch Monitor
To adjust the brightness:
1.) Press the Toggle button (2) or (3)  Once for brightness adjustment.
a.) Verify that Brightness is  80 (Dark Room) or 100 (Bright Room) .
If it is not, increase brightness to press the adjustment button (3).
If it is not, decrease brightness to press the adjustment button...

Page 245

GE HEALTHCARE
DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
Section 8-2 - Software Loading Procedures for BT09 (R8.x.x or later) 8-39
8-2-10-6 VCR  Parameters Setting (for the VCR-equipped system ONLY)
1.) Touch Utility > System.
2.) Click on Peripherals  tab.
3.) Set the proper parameters: -PAL or NTSC
- VCR type
4.) Shut down the scanner, then restart the scanner.
5.) Verify that the VCR can be properly operated with no error (remote, play...) .
8-2-10-7 Regional Setup (For the system with...

Page 246

GE HEALTHCARE
DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
8-40Section 8-2 - Software Loading Procedures for BT09 (R8.x.x or later)
8-2-10-7 Regional Setup (For the system with language other than English ONLY)  (cont’d)
1.) Select Utility> System> General>Regional Option Menu ,
2.) Click on  Languages tab.
3.) Check  Install files for East Asian languages.
4.) Click on  Apply.
5.) The system requires the re-boot automatically, shut down the system.
Figure 8-100   Regional Option
NOTICEYou do...

Page 247

GE HEALTHCARE
DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
Section 8-2 - Software Loading Procedures for BT09 (R8.x.x or later) 8-41
8-2-10-7 Regional Setup (For the system with language other than English ONLY)  (cont’d)
6.) In  Utility> System> General , set the Language as JPN. Save this setting, but do not reboot the 
system yet.
7.) Press  Regional Options , under Standards and Formats select  Japanese, under Location select 
Japan . Press Apply .Figure 8-101   General
Figure 8-102...

Page 248

GE HEALTHCARE
DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
8-42Section 8-2 - Software Loading Procedures for BT09 (R8.x.x or later)
8-2-10-7 Regional Setup (For the system with language other than English ONLY)  (cont’d)
8.) Select the Language tab , press Details, under Installed Services select the Japanese keyboard , 
under Default input language select  Japanese, press Apply , Press OK.
9.) Select the  Advanced tab , then select Japanese in the Language for non-Unicode programs pull-
down...

Page 249

GE HEALTHCARE
DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
Section 8-2 - Software Loading Procedures for BT09 (R8.x.x or later) 8-43
8-2-10-7 Regional Setup (For the system with language other than English ONLY)  (cont’d)
13.) Select  Automatic Setup for Japanese Language  button. Then IME configuration shall be 
automatically done.
This configuration shall setup the keyboard assignment (remove  
R

Page 250

GE HEALTHCARE
DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
8-44Section 8-2 - Software Loading Procedures for BT09 (R8.x.x or later)
8-2-10-7 Regional Setup (For the system with language other than English ONLY)  (cont’d)
For Russian or Greek Languages
1.) In  Utility--> System--> General , set the Language as Russian or Greek. Save this setting, but do 
not reboot the system yet.
2.) Press  Regional Options , under Standards and Formats select  Russian or Greek, under Location 
select  Russia or...