GE Logiq 7 Service Manual
Here you can view all the pages of manual GE Logiq 7 Service Manual. The GE manuals for Medical Equipment are available online for free. You can easily download all the documents as PDF.
Page 211
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL Section 8-2 - Software Loading Procedures for BT09 (R8.x.x or later) 8-5 8-2-10 - System Set- tingsSoftware Versionexpected version R8.x.xver. Check Local Data FlowLocal Archive – Int.HDD OK NG Computer Name and connectivitySet parameters OK NG Software OptionsCheck OK NG Set VCR parameters Set parameters OK N/A Set LanguageSet Language options OK N/A Printer RegistrationSet printer OK N/A Report...
Page 212
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL 8-6Section 8-2 - Software Loading Procedures for BT09 (R8.x.x or later) 8-2-5 Software Loading Steering Guide This section and the following pages should provide detailed guide of the system and application installation procedures. Figure 8-55 BT07 Software Loading Procedures 8-2-6 - PRE-INSTALLATION PROCEDURES 8-2-7 - INSTALL BASE SYSTEM OS + ApplicationInstalling 8-2-10 - SYSTEM SETTINGS 8-2-11 - FUNCTIONAL CHECK 8-2-9 - INSTALLING...
Page 213
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL Section 8-2 - Software Loading Procedures for BT09 (R8.x.x or later) 8-7 8-2-5-1 Processes prior to R8.x.x Installation 8-2-5-2 Processes to install R8.x.xFigure 8-56 Pre-installation Procedures NOTICE*1: Loading R8.x.x application software procedures requires, at minimum, C: Drives to be ghosted (as opposed to full ghost). In such case, patient image data, preset parameters, and customer data will be maintained. Nonetheless, for...
Page 214
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL 8-8Section 8-2 - Software Loading Procedures for BT09 (R8.x.x or later) 8-2-5-3 After installation of R8.x.x Figure 8-58 Post Installation Procedures NOTICE*1: Verify that Probes are not connected to the scanner. Other peripheral devices such as printers should be turned ON at this time. *2: If Operator logon window appears, select adm(administrator) and enter the password that is configured already (default value is no...
Page 215
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL Section 8-2 - Software Loading Procedures for BT09 (R8.x.x or later) 8-9 8-2-6 Pre-installation Procedures The following section describes activities necessary before system and application installation. 8-2-6-1 Disabling USB Devices 1.) Turn OFF all USB peripherals connected to the scanner. 8-2-6-2 Save Customer Data / Setting Save the following user data. - Patient image data (Refer to operation manual.) Please confirm the user if the...
Page 216
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL 8-10Section 8-2 - Software Loading Procedures for BT09 (R8.x.x or later) 8-2-6-4 Saving Connectivity Before starting upgrading, write down the Connectivity Setting for back-up. 1.) Touch Utility > Connectivity on the touch panel and click the TCP/IP tab. 2.) Write down the following parameters: Figure 8-60 TCP/IP ParametersDescriptions (Entry for Stand-Alone System) Computer NameEnable DHCP This parameter must NOT be selected....
Page 217
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL Section 8-2 - Software Loading Procedures for BT09 (R8.x.x or later) 8-11 8-2-6-5 Software Option 1.) Touch Utility > Admin. 2.) Click on System Admin tab and write down HW Number and Installed Option Keys.. Figure 8-61 System Admin ItemsDescriptions HW Number Installed Option Key Table 8-4 System Admin
Page 218
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL 8-12Section 8-2 - Software Loading Procedures for BT09 (R8.x.x or later) 8-2-6-6 Regional Preset 1.) Go to Utility > System > About . 2.) Write down the current Preset Region. Figure 8-62 Preset Region
Page 219
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL Section 8-2 - Software Loading Procedures for BT09 (R8.x.x or later) 8-13 8-2-6-7 Printer Registration 1.) Select Utility > Connectivity > Button . 2.) Write down the printer names which are registered to Print1 through Print4. 3.) Click on >> to add Printflow View. Figure 8-63 Printer Registration ButtonActive Image Printer Print1 Print2 Print3 Print4 Table 8-5 Printer Registration Select Print x button to display the printer...
Page 220
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL 8-14Section 8-2 - Software Loading Procedures for BT09 (R8.x.x or later) 8-2-7 Install Base System 1.) Power OFF the scanner, wait more than 5 seconds, then power ON the scanner. 2.) Open the DVD drive and set the Ghost DVD. DVD drive reads Ghost DVD automatically and go to next step. 3.) After loading software, when the following screen appears, press the Enter key. NOTICEBefore starting to install base system, all printer must be...