Page 211

GE HEALTHCARE
DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
Section 8-2 - Software Loading Procedures for BT09 (R8.x.x or later) 8-5
8-2-10 - System Set-
tingsSoftware Versionexpected version R8.x.xver.
Check Local Data FlowLocal Archive – Int.HDD OK      NG
Computer Name and connectivitySet parameters OK      NG
Software OptionsCheck OK      NG
Set VCR parameters Set parameters OK      N/A
Set LanguageSet Language options OK      N/A
Printer RegistrationSet printer OK      N/A
Report...

Page 212

GE HEALTHCARE
DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
8-6Section 8-2 - Software Loading Procedures for BT09 (R8.x.x or later)
8-2-5 Software Loading Steering Guide
This section and the following pages should provide detailed guide of the system and application installation 
procedures.
Figure 8-55   BT07 Software Loading Procedures
8-2-6 - PRE-INSTALLATION 
PROCEDURES
8-2-7 - INSTALL BASE SYSTEM
OS + ApplicationInstalling
8-2-10 - SYSTEM SETTINGS
8-2-11 - FUNCTIONAL CHECK
8-2-9 - INSTALLING...

Page 213

GE HEALTHCARE
DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
Section 8-2 - Software Loading Procedures for BT09 (R8.x.x or later) 8-7
8-2-5-1 Processes prior to R8.x.x Installation
8-2-5-2 Processes to install R8.x.xFigure 8-56   Pre-installation Procedures
NOTICE*1:
 Loading R8.x.x application software procedures requires, at minimum, C: Drives to be ghosted (as 
opposed to full ghost).  In such case, patient image data, preset parameters, and customer data will be 
maintained.  Nonetheless, for...

Page 214

GE HEALTHCARE
DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
8-8Section 8-2 - Software Loading Procedures for BT09 (R8.x.x or later)
8-2-5-3 After installation of R8.x.x
Figure 8-58   Post Installation Procedures
NOTICE*1: Verify that Probes are not connected to the scanner. Other peripheral devices such as printers 
should be turned ON at this time.
*2:  If Operator logon window appears, select adm(administrator) and enter the password that is         
configured already (default value is no...

Page 215

GE HEALTHCARE
DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
Section 8-2 - Software Loading Procedures for BT09 (R8.x.x or later) 8-9
8-2-6 Pre-installation Procedures
The following section describes activities necessary before system and application installation.
8-2-6-1 Disabling USB Devices 1.) Turn OFF all USB peripherals connected to the scanner.
8-2-6-2 Save Customer Data / Setting Save the following user data.
- Patient image data (Refer to operation manual.)
Please confirm the user if the...

Page 216

GE HEALTHCARE
DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
8-10Section 8-2 - Software Loading Procedures for BT09 (R8.x.x or later)
8-2-6-4 Saving Connectivity
Before starting upgrading, write down
 the Connectivity Setting  for back-up.
1.) Touch  Utility > Connectivity  on the touch panel and click the  TCP/IP tab.
2.) Write down the following parameters:
Figure 8-60   TCP/IP
ParametersDescriptions (Entry for Stand-Alone System)
Computer NameEnable DHCP This parameter 
must NOT be selected....

Page 217

GE HEALTHCARE
DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
Section 8-2 - Software Loading Procedures for BT09 (R8.x.x or later) 8-11
8-2-6-5 Software Option
1.) Touch Utility > Admin.
2.) Click on System Admin  tab and write down HW Number  and Installed Option Keys..
Figure 8-61   System Admin
ItemsDescriptions
HW Number
Installed Option Key
Table 8-4   System Admin 

Page 218

GE HEALTHCARE
DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
8-12Section 8-2 - Software Loading Procedures for BT09 (R8.x.x or later)
8-2-6-6 Regional Preset
1.) Go to Utility > System > About .
2.) Write down the current Preset Region.
Figure 8-62   Preset Region 

Page 219

GE HEALTHCARE
DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
Section 8-2 - Software Loading Procedures for BT09 (R8.x.x or later) 8-13
8-2-6-7 Printer Registration
1.) Select Utility > Connectivity > Button .
2.) Write down the printer names which are registered to Print1 through Print4.
3.) Click on  >> to add Printflow View.
Figure 8-63   Printer Registration
ButtonActive Image Printer
Print1
Print2
Print3
Print4
Table 8-5   Printer Registration
Select Print x 
button to 
display the 
printer...

Page 220

GE HEALTHCARE
DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
8-14Section 8-2 - Software Loading Procedures for BT09 (R8.x.x or later)
8-2-7 Install Base System
1.) Power OFF the scanner, wait more than 5 seconds, then power ON the scanner.
2.) Open the DVD drive and set the Ghost DVD.
DVD drive reads Ghost DVD automatically and go to next step.
3.) After loading software, when the following screen appears, press the  Enter key.
NOTICEBefore starting to install base system, all printer must be...