Page 181

GE HEALTHCARE
DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
Section 6-5 - Monitor and LCD Adjustments 6-19
6-5-5-2 For Technart (cont’d)
4.) “AutoAdjust OK” message will appear. If it does not appear. press  AUTO button again.
Figure 6-23   AutoAdjust OK message
Figure 6-24   Auto Button
Auto Button 

Page 182

GE HEALTHCARE
DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
6-20Section 6-5 - Monitor and LCD Adjustments
6-5-5-2 For Technart (cont’d)
5.) Display the LCD monitor menu and visually check if auto adjustment succeeds.
- If no character blurring exits, auto adjustment succeeds. Go to  6-5-5-2-3 Completing 
adjustment .
- If character blurring exits, auto adjustment fails. Perform  6-5-5-2-2 Manual Deblurring (For 
Technart ONLY) .
Figure 6-25   Blurring/No Blurring
Blurring at the right side of the...

Page 183

GE HEALTHCARE
DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
Section 6-5 - Monitor and LCD Adjustments 6-21
6-5-5-2-2 Manual Deblurring (For Technart ONLY)
Perform manual deblurring only when auto adjustment fails.
1.) Refer to 4-2-2-3  Entering Maintenance Mode  to enter the windows desktop.
2.) Select Program > Accessories > Paint  to run the paint software.
3.) Phase Check: a.) Drag and drop the paint screen on the touch panel screen.
b.) Click on square symbol at the top right corner of the...

Page 184

GE HEALTHCARE
DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
6-22Section 6-5 - Monitor and LCD Adjustments
6-5-5-2 For Technart (cont’d)
c.) Click on the line icon and select the thinest line.
d.) Draw a vertical line at the right side of the screen below and check if it has no deblurring.
- If deblurring exists, go to next step.
- If no deblurring exists, go to step 4- HTotal Check Figure 6-28   Line
Figure 6-29   Drawing Line 

Page 185

GE HEALTHCARE
DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
Section 6-5 - Monitor and LCD Adjustments 6-23
6-5-5-2 For Technart (cont’d)
e.) Display the LCD monitor menu and press the MENU button once. The following menu appears.
f.) Repeat to press the MENU button until the “Phase” blinks.
g.) Repeat to press the PLUS button to find the value with no character blurring. For example, this value is X.
h.) Set the value ( X + 14) to Phase.
i.) Press the AUTO button once to save the setting.
j.)...

Page 186

GE HEALTHCARE
DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
6-24Section 6-5 - Monitor and LCD Adjustments
6-5-5-2 For Technart (cont’d)
4.) Htotal Check:
a.) Display the LCD monitor menu and press the MENU button once. The following menu appears.
b.) Check HTotal value:
- If it is 16, go to step5 - Hposition/VPosition check
- If it is NOT 16, go to next step.
c.) Repeat to press the MENU button until the “HTotal” blinks.
d.) Repeat to press the PLUS or MINUS button to set  16.
e.) Press the AUTO...

Page 187

GE HEALTHCARE
DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
Section 6-5 - Monitor and LCD Adjustments 6-25
6-5-5-2 For Technart (cont’d)
b.) Check the upper, lower, left, and right edge of the screen. They should be centered evenly in the monitor.
- If they are vertically out of alignment, adjust VPosition.
- If they are horizontally out of alignment, adjust HPosition.
c.) Display the LCD monitor menu and press the MENU button once. The following menu appears.
d.) Repeat to press the MENU button...

Page 188

GE HEALTHCARE
DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
6-26Section 6-5 - Monitor and LCD Adjustments
6-5-5-2-3 Completing adjustment
1.) Display the LCD monitor menu and press the left button (MENU). The following menu appears.
2.) Verify that:- Contrast: 210
- ContrastRed: 145
- ContrastGrn: 130
- ContrastBlu: 160
3.) Verify that: - The display does not move up, down, or sideways.
- All menu are shown in the LCD display.
- Turning the brightness knob clockwise fades in and turning it...

Page 189

GE HEALTHCARE
DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
Section 6-6 - BW Printer Setting / Adjustment 6-27
Section 6-6
BW Printer Setting / Adjustment
6-6-1 Parameters for UP-D897
1.) When system started, Go to Utility > System >Peripherals, then hit Printers button in the Setup 
group.
2.) Click  SONY UP-D897  in the Printer List, then right click and select Properties .
3.) Select  General tab, then hit  Printing Preferences ... 
4.) Select  Layout tab, then ...
-Set  1920x1280  into Paper...

Page 190

GE HEALTHCARE
DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
6-28Section 6-6 - BW Printer Setting / Adjustment
6-6-1 Parameters for UP-D897 (cont’d)
5.) Select  Density Adjust  tab, then ...
-Set  TONE1 into GAMMA
-Set  7 into Sharpness
Figure 6-36   Printing Preferences 