GE Logiq 7 Service Manual
Here you can view all the pages of manual GE Logiq 7 Service Manual. The GE manuals for Medical Equipment are available online for free. You can easily download all the documents as PDF.
Page 141
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL Section 5-4 - Block Diagrams and Theory 5-25 5-4-9 Interconnect Cabling (BT09 or later) (cont’d) Figure 5-44 USB Connection (BT09 or later) AA BB USB type A (Ver. 2.0) USB type B (Ver. 2.0) AA AA AA AA AAAAAAAA AAAA 1 2 3 4 56 7 8910 11 12 OP Panel DVD Drive Color Printer Front Panel (Voice Command Mic,USB Memory or Service) Rear Panel Out VCR1 VCR2 Service BW Printer 4D BoxBB USB Option Dongle BB (reserve) BB BBAA (reserve)AAA BB BB...
Page 142
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL 5-26Section 5-4 - Block Diagrams and Theory 5-4-10 Difference points between BT07 (or la ter) and BT06 (or earlier) Systems for R7.5.x 17 to 19 LCD Monitor Upgrade For BT06 or earlier system, the DVI cable is connected directly to the monitor from the BECOMP3 via the DVI-HDMI adapter. Figure 5-45 R7.5.x 17” to 19” LCD Monitor Upgrade (For BT07 or later) Figure 5-46 R7.5.x 17” to 19” LCD Monitor Upgrade (For BT06 or earlier)
Page 143
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL Section 5-5 - Common Service Platform 5-27 Section 5-5 Common Service Platform 5-5-1 Introduction The Service Platform contains a set of software modules that are common to all PC backend ultrasound and cardiology systems. This web-enabled technology provides linkage to e-Services, e-Commerce, and the iCenter, making GE’s scanners more e-enabled than ever. The Common Service Platform will increase service productivity and reduce...
Page 144
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL 5-28Section 5-5 - Common Service Platform 5-5-2-2 Service Login (cont’d) 4.) Select GE Service at the “Select User Level” field. 5.) Enter the password for the Service Platform. 6.) Click on Okay. 5-5-2-2-2 For BT03 or lower: Click on the button with the wrench icon in the status bar at the bottom of the scanner console. This button links the user or the Field Engineer (FE) to the service login screen. 5-5-2-3 Access / Security The...
Page 145
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL Section 5-5 - Common Service Platform 5-29 5-5-3 Service Home Page When the Common Service Desktop is started, the Service Home Page appears. The Home Page contains the software revision along with the hardware inventory and the results of the latest System Health Information. The navigation bar at the top of the screen allows the user to select from several tools and applications. NOTE: When using the Common Service Desktop do NOT...
Page 146
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL 5-30Section 5-5 - Common Service Platform 5-5-4 Error Logs Page From the Error Logs page the Log Viewer displays five categories with pull-down sub-menus. The Service Interface has a log viewing tool driven by the following high-level requirements. • Simple filtering of the scanner log(s) with filtering capabilities being a function of login access permissions. • Logs are viewable by all service modes. • Allow for multiple instances of...
Page 147
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL Section 5-5 - Common Service Platform 5-31 5-5-4-1 Logs The six sub-menus of the Logs category are System, Power, Temperature, Probe, Board, and DICOM. Figure 5-51 Log Sub-Menus (System) LogsDescriptions System Software bug information for software engineering to debug the software.Power Not used for LOGIQ 7. Infomatics Exam informations. Whenever the New patient or En d Current Patient is pressed, they are reported as a log.
Page 148
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL 5-32Section 5-5 - Common Service Platform TemperatureError (!!) The error occurred in the temperature with “!!” mark. In the time when the error occurred, the message “Temp is above safe limit. Check air-filter! Shutdown in #seconds.” appears in the status bar and the system is automatically shut down. The temperature which the error occurs: • 75 degrees or higher for Rack Temperature (1) • 65 degrees or higher for Rack Temperature...
Page 149
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL Section 5-5 - Common Service Platform 5-33 5-5-4-1-1 Log Viewing The Service Platform has a log viewing tool driven by the following high-level requirements. • Simple filtering of the scanner log(s) with filtering capabilities being a function of login access permissions. • Logs are viewable by all service modes. • Allow for multiple instances of the log viewer. • Color-coded log entries for severity levels: * Severity 1 - Color coded...
Page 150
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL 5-34Section 5-5 - Common Service Platform 5-5-4-4 Filter Click on he sub-menu, System Logs, to enter Filter menu. Filtering is used to extract select data from the Error Logs. There are four different error levels and 11 packages that can be filtered. 5-5-4-5 Exit The sub-menu, Exit Log Viewer , returns the user to the Common Service Desktop home page. Figure 5-54 Filter Category Figure 5-55 Exit Category BT04/BT06/BT06-2 or later