Section 5-4 - Block Diagrams and Theory 5-25
5-4-9 Interconnect Cabling (BT09 or later) (cont’d)
Figure 5-44   USB Connection (BT09 or later)
AA
BB USB type A (Ver. 2.0)
USB type B (Ver. 2.0)
AA
AA
AA
AA
AAAAAAAA
AAAA
1
2
3 4
56
7 8910
11 12
OP Panel
DVD Drive
Color Printer Front Panel
(Voice Command Mic,USB Memory or Service)
Rear Panel
Out
VCR1
VCR2
Service
BW Printer 4D BoxBB
USB Option Dongle
BB
(reserve)
BB
BBAA
(reserve)AAA
BB
BB...

Page 142

GE HEALTHCARE
DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
5-26Section 5-4 - Block Diagrams and Theory
5-4-10 Difference points between BT07 (or la ter) and BT06 (or earlier) Systems for 
R7.5.x 17 to 19 LCD Monitor Upgrade
For BT06 or earlier system, the DVI cable is connected directly to the monitor from the BECOMP3 via 
the DVI-HDMI adapter.
Figure 5-45   R7.5.x 17” to 19” LCD Monitor Upgrade (For  BT07 or later)
Figure 5-46   R7.5.x 17” to 19” LCD Monitor Upgrade (For  BT06 or earlier) 

Page 143

GE HEALTHCARE
DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
Section 5-5 - Common Service Platform 5-27
Section 5-5
Common Service Platform
5-5-1 Introduction
The Service Platform contains a set of software modules that are common to all PC backend ultrasound 
and cardiology systems. This web-enabled technology provides linkage to e-Services, e-Commerce, 
and the iCenter, making GE’s scanners more e-enabled than ever. The Common Service Platform will 
increase service productivity and reduce...

Page 144

GE HEALTHCARE
DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
5-28Section 5-5 - Common Service Platform
5-5-2-2 Service Login (cont’d)
4.) Select GE Service  at the “Select User Level” field. 
5.) Enter the password for the Service Platform.
6.) Click on  Okay.
5-5-2-2-2 For BT03 or lower:
Click on the button with the wrench icon  in the status bar at the bottom of the scanner console. This button links 
the user or the Field Engineer (FE) to the service login screen.
5-5-2-3 Access / Security The...

Page 145

GE HEALTHCARE
DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
Section 5-5 - Common Service Platform 5-29
5-5-3 Service Home Page
When the Common Service Desktop is started, the Service Home Page appears. The Home Page 
contains the software revision along with the hardware inventory and the results of the latest System 
Health Information. The navigation bar at the top of the screen allows the user to select from several 
tools and applications.
NOTE: When using the Common Service Desktop do 
NOT...

Page 146

GE HEALTHCARE
DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
5-30Section 5-5 - Common Service Platform
5-5-4 Error Logs Page
From the Error Logs page the Log Viewer displays five categories with pull-down sub-menus. The 
Service Interface has a log viewing tool driven by the following high-level requirements.
• Simple filtering of the scanner log(s) with filtering capabilities being a function of login access 
permissions.
• Logs are viewable by all service modes.
• Allow for multiple instances of...

Page 147

GE HEALTHCARE
DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
Section 5-5 - Common Service Platform 5-31
5-5-4-1 Logs
The six sub-menus of the Logs category are System, Power, Temperature, Probe, Board, and DICOM.
Figure 5-51   Log Sub-Menus (System)
LogsDescriptions
System Software bug information for software engineering to debug the software.Power Not used for LOGIQ 7.
Infomatics Exam informations. Whenever the New patient or En d Current Patient is pressed, they are reported as a log. 

Page 148

GE HEALTHCARE
DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
5-32Section 5-5 - Common Service Platform
TemperatureError (!!)
The error occurred in the temperature with “!!” mark. In the time when the error occurred, the 
message “Temp is above safe limit. Check air-filter! Shutdown in #seconds.” appears in the status 
bar and the system is automatically shut down.
The temperature which the error occurs:
• 75 degrees or higher for Rack Temperature (1)
• 65 degrees or higher for Rack Temperature...

Page 149

GE HEALTHCARE
DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
Section 5-5 - Common Service Platform 5-33
5-5-4-1-1 Log Viewing
The Service Platform has a log viewing tool driven by the following high-level requirements.
• Simple filtering of the scanner log(s) with filtering capabilities being a function of login access permissions.
• Logs are viewable by all service modes.
• Allow for multiple instances of the log viewer.
• Color-coded log entries for severity levels: * Severity 1 - Color coded...

Page 150

GE HEALTHCARE
DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
5-34Section 5-5 - Common Service Platform
5-5-4-4 Filter
Click on he sub-menu,  System Logs, to enter Filter menu. Filtering is used to extract select data from 
the Error Logs. There are four different error levels and 11 packages that can be filtered.
5-5-4-5 Exit The sub-menu,  Exit Log Viewer , returns the user to the Common Service Desktop home page.
Figure 5-54   Filter Category
Figure 5-55   Exit Category
BT04/BT06/BT06-2
or later 