GE Logiq 7 Service Manual
Here you can view all the pages of manual GE Logiq 7 Service Manual. The GE manuals for Medical Equipment are available online for free. You can easily download all the documents as PDF.
Page 101
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL Section 4-2 - General Procedure 4-17 4-2-6-4 Formatting CD-R/DVD-R Disk (cont’d) 3.) Select the removable media from media list. 4.) Type a name for the removable media in Label field. 5.) Select Format button. Figure 4-20 Formatting Removable Media2
Page 102
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL 4-18Section 4-2 - General Procedure 4-2-7-2 Archiving Presets to an CD-R/MO/DVD-R Disk 1.) Insert an empty (blank) formatted CD-R/MO/DVD-R disk into the CD-R/MO/DVD-R device. 2.) Access to the Utility Menu on the Touch Panel, and select System>Backup/Restore . The Backup screen will be shown on the monitor. 3.) Select the item to back up either from Resource Files. 4.) Select the media to locate the items. 5.) Click on Backup. The...
Page 103
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL Section 4-2 - General Procedure 4-19 4-2-7-3 Loading Presets from an CD-R/MO/DVD-R disk 1.) Insert the CD-R/MO/DVD-R disk with the archived Presets into the CD-R/MO/DVD-R. 2.) Access to the Utility Menu on the Touch Panel, and select System>Backup/Restore. The Restore sheet will be shown on the monitor. 3.) Select the item to restore either from resource Files. 4.) Click on Restore. A message to make sure the restore process is...
Page 104
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL 4-20Section 4-2 - General Procedure Section 4-3 Functional Checks 4-3-1 Basic Controls For a functional check of the system’s features, including the Control Panel, Touch Panel , Monitor , Keyboard and Trackball , refer to the LOGIQ™ 7 Quick Guide. 4-3-2 Performance Tests 4-3-2-1 Recommended Test Phantoms GE Healthcare recommends the RMI 430GS phantom, but it is not required. It is the most current phantom recommended to our field...
Page 105
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL Section 4-2 - General Procedure 4-21 4-3-3 Mode Checks For a functional check of the system’s different modes, refer to the LOGIQ™ 7 Quick Guide. The Quick Guide will familiarize you with image optimization for B-Mode, M-Mode , Color Flow, and Doppler. 4-3-3-1 System Checks 4-3-4 Basic Measurements Basic Measurements for the LOGIQ™ 7 include Distance and Tissue Depth, Circumference/Area (Ellipse and Trace) , Volume , Time Interval ,...
Page 106
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL 4-22Section 4-2 - General Procedure 4-3-5 ECG Checks Connect the ECG Harness and check: 4-3-6 Cineloop Check For activating Cine, creating and storing Cine Loops, and information on the Cine Timeline, refer to the LOGIQ™ 7 Quick Guide. 4-3-7 Backend Processor Checks • If all the previous tests have been passed successfully, the backend processor is most likely OK. • See Service Platform. • If the system seems to be operating...
Page 107
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL Section 4-2 - General Procedure 4-23 4-3-9 Peripheral Checks Check that peripherals work as described below: 4-3-9-1 Peripheral/Option Checks If any peripherals or options are not part of the system configuration, the check can be omitted. Refer to the User Manual for a list of approved peripherals/options. Table 4-5 Peripheral checks StepTask to do Expected Result(s) 1.Press Freeze Stop image acquisition. 2. Press (P1) or (P4) on the...
Page 108
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL 4-24Section 4-2 - General Procedure 4-3-10 Mechanical Functions 4-3-10-1 Keyboard and Display Platform Console Check Table 4-7 Display platform Maneuverability check StepTask to do Expected Result(s) 1.Pull release lever under the OP Panel to release the lock. The height of the Monitor and OP Panel will be maneuverable. Figure 4-24 Keyboard Lever Keyboad Lever
Page 109
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL Section 4-2 - General Procedure 4-25 4-3-10-2 Brakes and Direction Locks Checks Check that: brakes and direction locks function as described below. Refer to Figure 4-25 for the locations of brake and swivel. Table 4-8 Brakes and Direction Lock Check StepTask to doExpected Result(s) 1.Press on pedal no.2 To engage the pedal in full lock 2. Press on pedal no.3 To release the brake 3. Press on pedal no.1 To engage swivel lock 4. Press...
Page 110
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL 4-26Section 4-2 - General Procedure 4-3-10-3 Caster and Caster Link Checks Check if: •For Caster:- No loose caster with it jacked up. - Smooth operation when moving the scanner. • For Caster Link Assy: - Effective brake and direction lock function (See section 4-3-10-2). - The pedals can be smoothly returned. - The lock pin must be removed.