Page 101

GE HEALTHCARE
DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
Section 4-2 - General Procedure 4-17
4-2-6-4 Formatting CD-R/DVD-R Disk (cont’d)
3.) Select the removable media from media list.
4.) Type a name for the removable media in Label field.
5.) Select Format button. Figure 4-20 Formatting Removable Media2 

Page 102

GE HEALTHCARE
DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
4-18Section 4-2 - General Procedure
4-2-7-2 Archiving Presets to an CD-R/MO/DVD-R Disk
1.) Insert an empty (blank) formatted CD-R/MO/DVD-R disk into the CD-R/MO/DVD-R device.
2.) Access to the Utility  Menu on the Touch Panel, and select  System>Backup/Restore . The Backup 
screen will be shown on the monitor.
3.) Select the item to back up either from Resource Files.
4.) Select the media to locate the items.
5.) Click on  Backup. The...

Page 103

GE HEALTHCARE
DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
Section 4-2 - General Procedure 4-19
4-2-7-3 Loading Presets from an CD-R/MO/DVD-R disk
1.) Insert the CD-R/MO/DVD-R disk with the archived Presets into the CD-R/MO/DVD-R.
2.) Access to the Utility  Menu on the Touch Panel, and select  System>Backup/Restore. The Restore 
sheet will be shown on the monitor.
3.) Select the item to restore either from resource Files.
4.) Click on  Restore. A message to make sure the restore process is...

Page 104

GE HEALTHCARE
DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
4-20Section 4-2 - General Procedure
Section 4-3
Functional Checks
4-3-1 Basic Controls
For a functional check of the system’s features, including the  Control Panel, Touch Panel , Monitor , 
Keyboard  and Trackball , refer to the LOGIQ™ 7 Quick Guide.
4-3-2 Performance Tests
4-3-2-1 Recommended Test Phantoms
GE Healthcare recommends the RMI 430GS phantom, but it is not required. It is the most current 
phantom recommended to our field...

Page 105

GE HEALTHCARE
DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
Section 4-2 - General Procedure 4-21
4-3-3 Mode Checks
For a functional check of the system’s different modes, refer to the LOGIQ™ 7 Quick Guide. The Quick 
Guide will familiarize you with image optimization for B-Mode, M-Mode , Color Flow, and  Doppler.
4-3-3-1 System Checks 
4-3-4 Basic Measurements
Basic Measurements for the LOGIQ™ 7 include  Distance and Tissue Depth, Circumference/Area 
(Ellipse and Trace) , Volume , Time Interval ,...

Page 106

GE HEALTHCARE
DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
4-22Section 4-2 - General Procedure
4-3-5 ECG Checks
Connect the ECG Harness and check:
4-3-6 Cineloop Check
For activating Cine, creating and storing Cine Loops, and information on the Cine Timeline, refer to the 
LOGIQ™ 7 Quick Guide.
4-3-7 Backend Processor Checks
• If all the previous tests have been passed successfully, the backend processor is most likely OK.
• See Service Platform.
• If the system seems to be operating...

Page 107

GE HEALTHCARE
DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
Section 4-2 - General Procedure 4-23
4-3-9 Peripheral Checks
Check that peripherals work as described below:
4-3-9-1 Peripheral/Option Checks If any peripherals or options are not part of the system configuration, the check can be omitted. Refer 
to the User Manual for a list of approved peripherals/options. Table 4-5   Peripheral checks
StepTask to do Expected Result(s)
1.Press Freeze Stop image acquisition.
2. Press (P1) or (P4) on the...

Page 108

GE HEALTHCARE
DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
4-24Section 4-2 - General Procedure
4-3-10 Mechanical Functions
4-3-10-1 Keyboard and Display Platform Console Check
Table 4-7   Display platform Maneuverability check
StepTask to do Expected Result(s)
1.Pull release lever under the OP Panel to 
release the lock. The height of the Monitor and OP Panel will 
be maneuverable.
Figure 4-24 Keyboard Lever
Keyboad Lever 

Page 109

GE HEALTHCARE
DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
Section 4-2 - General Procedure 4-25
4-3-10-2 Brakes and Direction Locks Checks
Check that: brakes and direction locks function as described below. Refer to Figure 4-25 for the 
locations of brake and swivel.
Table 4-8    Brakes and Direction Lock Check
StepTask to doExpected Result(s)
1.Press on pedal no.2
To engage the pedal in full lock
2. Press on pedal no.3
To release the brake
3. Press on pedal no.1
To engage swivel lock
4.
Press...

Page 110

GE HEALTHCARE
DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
4-26Section 4-2 - General Procedure
4-3-10-3 Caster and Caster Link Checks
Check if:
•For Caster:- No loose caster with it jacked up. 
- Smooth operation when moving the scanner.
• For Caster Link Assy: - Effective brake and direction lock function (See section  4-3-10-2). 
- The pedals can be smoothly returned.
- The lock pin must be removed. 
Start reading GE Logiq 7 Service Manual