Page 81
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL Section 3-5 - Installation Paperwork 3-23 3-5-1-1 Rear Panel Connector (cont’d) Pin Assignment for Camera B/W NOTE: Output level of control signals indicated in the above tables are TTL level. Pin Assignment of Insite 3-5-1-2 Front Connector Panel Located on the front panel are Microphone, LED, and Reset. Table 3-16 Pin Assignment of Mini-Jack for Controlling B/W Camera Pin No Output Signal 1P RIN T 2 Signal GND Table 3-17 Pin...
Page 82
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL 3-24Section 3-5 - Installation Paperwork 3-5-1-3 B/W Printer Connector Panel 3-5-1-4 Footswitch Connector Panel Round 5 pin connector. NOTE: Output level of control signals indicated in the above tables are TTL level. Table 3-18 Pin Assignment of Mini-Jack for Controlling B/W Printer Pin No Output Signal 1P RIN T 2 Signal GND Table 3-19 Pin Assignment of Mini-Jack for Footswitch Pin NoOutput Signal 1S W1-W H 2S W2-R D 3S W3-G N 4...
Page 83
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL Section 3-5 - Installation Paperwork 3-25 This page was intentionally left blank.
Page 84
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL 3-26Section 3-5 - Installation Paperwork
Page 85
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL Section 4-1 - Overview 4-1 Chapter 4 Functional Checks Section 4-1 Overview 4-1-1 Purpose for Chapter 4 This chapter provides procedures for quickly checking major functions of the LOGIQ™ 7 console, diagnostics by using the built-in service software, and power supply adjustments. 4-1-2 Special Equipment Required • An empty (blank) MO Disk. • At least one transducer. (normally you should check all the transducers used on the system.)...
Page 86
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL 4-2Section 4-2 - General Procedure Section 4-2 General Procedure 4-2-1 Lockout/Tagout Requirements Follow OSHA Lockout/Tagout requirements by ensuring you are in total control of the plug. 4-2-2 Power On/Boot Up NOTE: After turning off the system, wait at least ten seconds before turning it on again. The system may not be able to boot if power is recycled too quickly. 4-2-2-1 Power Up 1.) Connect the Main Power Cable at the rear of the...
Page 87
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL Section 4-2 - General Procedure 4-3 4-2-2-1 Power Up (cont’d) 4.) Press the ON/OFF key at the front of the System once. 4-2-2-2 Power Up Sequence NOTE: For consoles with BEP4 (BT09), power switch lamp is darker compared to existing BT04-BT07 consoles with BEP3. This is NOT abnormality and system will function normally regardless of the brightness of the power button. 1.) The Start Up Screen will be shown on the Monitor display when...
Page 88
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL 4-4Section 4-2 - General Procedure 2.) After initialization is complete, all lighted buttons on the Control Panel light and the default B-Mode screen or Patient screen (no probes are connected) is displayed on the monitor display. 4-2-2-3 Entering Maintenance Mode 4-2-2-3-1 R6.2.x or later 1.) Insert the service dongle into any USB port. 2.) Wait for scanner boot-up. 3.) Tap ON/OFF the power switch. 4.) Click on Exit. 5.) Enter the...
Page 89
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL Section 4-2 - General Procedure 4-5 4-2-2-3 Entering Maintenance Mode (cont’d) 7.) Click on maintenance . 8.) Click on Exit to window . Figure 4-6 Clicking on Maintetance Figure 4-7 Clicking on Exit to Window 7 7
Page 90
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL 4-6Section 4-2 - General Procedure 4-2-2-3 Entering Maintenance Mode (cont’d) 4-2-2-3-2 R 4.x.x to R6.0.x 1.) The Start Up Screen will be shown on the Monitor display when the system is turned ON. 2.) The “Probing for maintenance access” dialog appears. Enter Password, then click on OK. 3.) Then Start Loader display will be shown on the Monitor display.Click on Maintenance. 4.) Click on Exit to Windows in maintenace diaglog...