GE HEALTHCARE
DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
Section 3-5 - Installation Paperwork 3-23
3-5-1-1 Rear Panel Connector (cont’d)
Pin Assignment for Camera B/W
NOTE: Output level of control signals indicated in the above tables are TTL level. Pin Assignment of Insite
3-5-1-2 Front Connector Panel Located on the front panel are Microphone, LED, and Reset.  Table 3-16   Pin Assignment of Mini-Jack for Controlling B/W Camera
Pin No
Output Signal
1P RIN T
2 Signal GND
Table 3-17   Pin...

Page 82

GE HEALTHCARE
DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
3-24Section 3-5 - Installation Paperwork
3-5-1-3 B/W Printer Connector Panel
3-5-1-4 Footswitch Connector Panel
Round 5 pin connector.
NOTE: Output level of control signals indicated in the above tables are TTL level. Table 3-18   Pin Assignment of Mini-Jack for Controlling B/W Printer
Pin No
Output Signal
1P RIN T
2 Signal GND
Table 3-19   Pin Assignment of Mini-Jack for Footswitch
Pin NoOutput Signal
1S W1-W H
2S W2-R D
3S W3-G N
4...

Page 83

GE HEALTHCARE
DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
Section 3-5 - Installation Paperwork 3-25
This page was intentionally left blank. 

Page 84

GE HEALTHCARE
DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
3-26Section 3-5 - Installation Paperwork 

Page 85

GE HEALTHCARE
DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
Section 4-1 - Overview 4-1
Chapter 4
Functional Checks
Section 4-1
Overview
4-1-1 Purpose for Chapter 4
This chapter provides procedures for quickly checking major functions of the LOGIQ™ 7 console, 
diagnostics by using the built-in service software, and power supply adjustments. 
4-1-2 Special Equipment Required
• An empty (blank) MO Disk.
• At least one transducer.
(normally you should check all the transducers used on the system.)...

Page 86

GE HEALTHCARE
DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
4-2Section 4-2 - General Procedure
Section 4-2
General Procedure
4-2-1 Lockout/Tagout Requirements
Follow OSHA Lockout/Tagout requirements by ensuring you are in total control of the plug.
4-2-2 Power On/Boot Up
NOTE: After turning off the system, wait at least ten seconds before turning it on again. The system 
may not be able to boot if power is recycled too quickly.
4-2-2-1 Power Up 1.) Connect the Main Power Cable at the rear of the...

Page 87

GE HEALTHCARE
DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
Section 4-2 - General Procedure 4-3
4-2-2-1 Power Up (cont’d)
4.) Press the ON/OFF 
key at the front of the System once.
4-2-2-2 Power Up Sequence NOTE: For consoles with BEP4 (BT09), power switch lamp is darker compared to existing BT04-BT07  consoles with BEP3.  This is NOT abnormality and system will function normally regardless of 
the brightness of the power button.
1.) The Start Up Screen will be shown on the Monitor display when...

Page 88

GE HEALTHCARE
DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
4-4Section 4-2 - General Procedure
2.) After initialization is complete, all lighted buttons on the Control Panel light and the default B-Mode 
screen or Patient screen (no probes are connected) is displayed on the monitor display.
4-2-2-3 Entering Maintenance Mode 4-2-2-3-1 R6.2.x or later 1.) Insert the service dongle into any USB port.
2.) Wait for scanner boot-up.
3.) Tap ON/OFF the power switch.
4.) Click on  Exit.
5.) Enter the...

Page 89

GE HEALTHCARE
DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
Section 4-2 - General Procedure 4-5
4-2-2-3 Entering Maintenance Mode (cont’d)
7.) Click on  maintenance .
8.) Click on  Exit to window .
Figure 4-6   Clicking on Maintetance
Figure 4-7   Clicking on Exit to Window
7 7 

Page 90

GE HEALTHCARE
DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
4-6Section 4-2 - General Procedure
4-2-2-3 Entering Maintenance Mode (cont’d)
4-2-2-3-2 R 4.x.x to R6.0.x 1.) The Start Up Screen will be shown on the Monitor display when the system is turned ON.
2.) The “Probing for maintenance access” dialog appears. Enter  Password, then click on  OK.
3.) Then Start Loader display will be shown on the Monitor display.Click on  Maintenance.
4.) Click on  Exit to Windows  in maintenace diaglog...