Page 41

GE HEALTHCARE
DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
Section 1-2 - Important Conventions 1-19
1-3-11 How to remove the Ghost CD-ROM
The Ghost CD-ROM (Base System Software Load Image CD-ROM) is mounted on the PC-BOX inside 
the scanner using velcro tapes. The upper side of the PC-BOX cover contains sharp edge causing a FE 
to have possibility of cutting his hand if he removes the CD-ROM with the PC-BOX cover closed.
Figure 1-6   How to Remove Ghost CD-ROM
Ghost CD-ROM
Sharp Edge
PC-BOX...

Page 42

GE HEALTHCARE
DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
1-20Section 1-2 - Important Conventions
1-3-11 How to remove the Ghost CD-ROM (cont’d)
The sharp edge of the PC Box Cover are chamfered before shipment. However, the Ghost CD-ROM 
must be removed with the PC-Box cover open according to the following steps.
1.) Disconnect the DVI-D connector, then remove the four mounting bolts (three hex bolts and one phillips screw).
2.) Open the PC box cover and remove the Ghost CD-ROM from the PC box....

Page 43

GE HEALTHCARE
DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
Section 1-2 - Important Conventions 1-21
Section 1-4
EMC, EMI, and ESD
1-4-1 Electromagnetic Compatibility
Electro Magnetic Compatibility describes a level of performance of a device within its electromagnetic 
environment. This environment consists of the device itself and its surroundings including other 
equipment, power sources and persons with which the device must interface. Inadequate compatibility 
results when a susceptible...

Page 44

GE HEALTHCARE
DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
1-22Section 1-5 - Customer Assistance
Section 1-5
Customer Assistance
This system is not repairable by the customer. If this equipment does not work as indicated in the 
Operator Manual, please contact your service support center. If the service engineer needs additional 
information to repair this equipment, please contact the following address (The necessary information 
will be provided to the Service Engineer as needed).
1-5-1 System...

Page 45

GE HEALTHCARE
DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
Section 1-5 - Customer Assistance 1-23
1-5-2 Contact Information (cont’d)
NOTE: If this equipment does not work as indicated in the Operator Manual(s), contact your support  center. Have the system ID number available when you call. 

Page 46

GE HEALTHCARE
DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
1-24Section 1-5 - Customer Assistance
This page was intentionally left blank. 

Page 47

GE HEALTHCARE
DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
Section 2-1 - Overview 2-1
Chapter 2
Pre Installation
Section 2-1
Overview
2-1-1 Purpose of this chapter 2
This chapter provides the information required to plan and prepare for the installation of a LOGIQ™ 7. 
Included are descriptions of the facility and electrical needs to be met by the purchaser of the unit. A 
checklist is also provided at the end of this section to help determine whether the proper planning and 
preparation is...

Page 48

GE HEALTHCARE
DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
2-2Section 2-2 - General Console Requirements
Section 2-2
General Console Requirements
2-2-1 Console Environmental Requirements 
2-2-1-1 Cooling
The cooling requirement for the LOGIQ™ 7 is 3500 BTU/hr. This figure does not include cooling needed 
for lights, people, or other equipment in the room. Each person in the room places an additional 300 
BTU/hr. demand on the cooling system.
2-2-1-2 Lighting Bright light is needed for system...

Page 49

GE HEALTHCARE
DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
Section 2-2 - General Console Requirements 2-3
2-2-2 Electrical Requirements
2-2-2-1 Electrical Requirements
NOTE: GE Medical Systems requires a dedicated power and ground for the proper operation of its  Ultrasound equipment. This dedicated power shall originate at the last distribution panel before 
the system. 
Sites with a mains power system 
with defined Neutral and Line:
The dedicated line shall consist of one phase, a neutral (not...

Page 50

GE HEALTHCARE
DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
2-4Section 2-2 - General Console Requirements
2-2-2-5 Site Power Outlets
A desiccated AC power outlet must be within reach of the unit without extension cords. Other outlets 
adequate for the external peripherals, medical and test equipment needed to support this unit must also 
be present within 1 m (3.2 ft.) of the unit. Electrical installation must meet all current local, state, and 
national electrical codes.
2-2-2-6 Unit Power Plug...