Page 61
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL Section 3-1 - Overview 3-3 Section 3-2 Receiving and Unpacking the Equipment When a new system arrives, check that any components are not damaged and are not in short supply. If shipping damage or shortage occurs, contact the address shown in Chapter 1. Unpacking Procedures: 1.) Cut the two Metal Bands. 2.) Lift the Cap up and off. 3.) Remove the six (6) Plastic Joints from the Outer Sleeves. 4.) Remove the Outer Sleeves. 5.) Remove the...
Page 62
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL 3-4Section 3-1 - Overview Section 3-2 Receiving and Unpacking the Equipment (cont’d) NOTE: Check the shipping container for special instructions. Verify that the container is intact. In some cases a secondary container may be used. If so, ask the carrier for unpacking instructions. NOTICEFor the protection of LCD monitor, re-use the packing sleeve whenever possible, in case of shipment after unpacking. Figure 3-2 Labels on...
Page 63
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL Section 3-1 - Overview 3-5 3-2-1 Safety Reminders DANGERWHEN USING ANY TEST INSTRUMENT THAT IS CAPABLE OF OPENING THE AC GROUND LINE (I.E., METER’S GROUND SWITCH IS OPEN), DO NOT TOUCH THE UNIT! CAUTIONTwo people should unpack the unit because of its weight. Two people are required whenever a part weighing 19kg (35 lb.) or more must be lifted. CAUTIONIf the unit is very cold or hot, do not turn on it s power until it has had a chance...
Page 64
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL 3-6Section 3-1 - Overview 3-2-2 Moving into Position In general, a single adult can move the LOGIQ™ 7 along an even surface with no steep grades. At least two people should move the machine when large humps, grooves, or grades will be encountered. (It is better to pull from the rear rather than push from the front of the unit). Before moving, store all loose parts in the unit. Wrap transducers in soft cloth or foam to prevent damage....
Page 65
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL Section 3-1 - Overview 3-7 3-2-3 LCD Monitor Tilt Caution CAUTIONDo NOT put your hand here! Your fingers might be pinched when adjusting angle of the LCD monitor. Figure 3-3 LCD Monitor Tilt Caution For R7.5.x 19inch LCD system
Page 66
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL 3-8Section 3-1 - Overview 3-2-4 LCD Monitor Swing CautionLCD Monitor Tilt Caution CAUTIONArm can be swung 180 degree left and right, and stick out from the side frame. Pay attention to the location from the wall in the room for room layout, and notice the attention to customer for Examination Table side. CAUTIONLCD will hit the KBD in case of moving over the right/left side.
Page 67
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL Section 3-1 - Overview 3-9 3-2-5 LCD Monitor Removal Caution 3-2-6 Shipping Delivery Requirements Do NOT hold the scanner at the LCD monitor or rotaing arm using the belt as shown. Always hold it at the body part of the console. CAUTIONAlways lock the LCD monitor arm before removing the LCD monitor. This prevents the monitor arm from being suddenly and fastly extended when removing the LCD monitor. Failure to heed this warning may...
Page 68
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL 3-10Section 3-3 - Preparing for Installation Section 3-3 Preparing for Installation 3-3-1 Verify Customer Order Compare items received by the customer to that which is listed on the delivery order. Report any items that are missing, back ordered or damaged. 3-3-2 Physical Inspection 3-3-2-1 System Voltage Settings • Verify that the scanner is set to the correct voltage. The Voltage settings for the LOGIQ™ 7 Scanner is found on the label...
Page 69
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL Section 3-4 - Completing the Installation 3-11 Section 3-4 Completing the Installation 3-4-1 Probe (Transducer) Connection 1.) Connect a transducer to the upper transducer receptacle as follows: a.) Ensure that the transducer twist lock lever to the horizontal position. b.) Insert the transducer connector on the receptacle guide pin until it touches the receptacle mating surface. c.) Twist the transducer twist lock lever to vertical...
Page 70
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL 3-12Section 3-4 - Completing the Installation 3-4-2 Optional Peripherals/Peripheral Connection Depending on the customer order, the peripheral(s) already has been installed onto the scanner before shipment. If it does not, select the proper location to install the peripherals, following the table below. Figure 3-5 Optional Peripherals/Peripheral Connection Location for peripheralOne peripheral (Color printer ONLY)One peripheral (VCR...