Page 51
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL Section 2-2 - General Console Requirements 2-5 2-2-3 EMI Limitations Ultrasound machines are susceptible to Electromagnetic Interference (EMI) from radio frequencies, magnetic fields, and transient in the air wiring. They also generate EMI. The LOGIQ™ 7 complies with limits as stated on the EMC label. However there is no guarantee that interface will not occur in a particular installation. Possible EMI sources should be identified...
Page 52
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL 2-6Section 2-2 - General Console Requirements 2-2-4 Probes Environmental Requirements Table 2-6 Operation and storage Temperatures for 2D Probes OperationalStorageTransport Temperature 10 - 40 oC 50 - 104 oF -10 - 60 oC 14 - 140oF -40 - 60 oC -40 - 140 oF Humidity30 - 85% non-condensing 30 - 90% non-condensing30 - 90% non-condensing Pressure700 - 1060hPa 700 - 1060hPa 700 - 1060hPa Table 2-7 Operation and storage Temperatures for...
Page 53
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL Section 2-3 - Facility Needs 2-7 Section 2-3 Facility Needs 2-3-1 Purchaser Responsibilities The work and materials needed to prepare the site is the responsibility of the purchaser. Delay, confusion, and waste of manpower can be avoided by completing pre installation work before delivery. User the Pre Installation checklist to verify that all needed steps have been taken, Purchaser reasonability includes: • Procuring the materials...
Page 54
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL 2-8Section 2-3 - Facility Needs 2-3-2 Required Features • Dedicated single branch power outlet of adequate amperage (see Table 2-3) meeting all local and national codes which is located less than 2.5 m (8 ft.) from the unit’s proposed location • Door opening is at least 76 cm (30 in) wide • Proposed location for unit is at least 0.3 m (1 ft.) from the wall for cooling • Power outlet and place for any external peripheral are within 2 m...
Page 55
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL Section 2-3 - Facility Needs 2-9 2-3-4 Recommended and Alternate Ultrasound Room Layout Recommended standard floor plan and a minimal floor plan for ultrasound equipment: Figure 2-1 RECOMMENDED ULTRASOUND ROOM LAYOUT LOGIQ LOGIQ
Page 56
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL 2-10Section 2-3 - Facility Needs 2-3-5 Networking Pre-installation Requirements 2-3-5-1 Purpose of DICOM Network Function DICOM services provide the operator with clinically useful features for moving images and patient information over a hospital network. Examples of DICOM services include the transfer of images to workstations for viewing or transferring images to remote printers. As an added benefit, transferring images in this...
Page 57
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL Section 2-3 - Facility Needs 2-11 2-3-5-2 DICOM Option Pre-installation Requirements (cont’d) . Figure 2-2 Worksheet for DICOM Network Information LOGIQ™ 7 Host Name AE TitleLocal PortIP Address Net Mask. . .. . .. ROUTING INFORMATION ROUTER2 ROUTER3 ROUTER1 . .. . .. . .. . .. . .. . .. . .. Destination IP Addresses GATEWAY IP Addresses Default DICOM APPLICATION INFORMATION NAME Store 2 Store 3 Store 1 MAKE/REVISION IP ADDRESSESPORT...
Page 58
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL 2-12Section 2-3 - Facility Needs This page was intentionally left blank.
Page 59
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL Section 3-1 - Overview 3-1 Chapter 3 Installation Section 3-1 Overview 3-1-1 Purpose of Chapter 3 This chapter contains information needed to install the unit. Included are references to a procedure that describes how to receive and unpack the equipment and how to file a damage or loss claim. How to prepare the facility and unit of the actual installation, and how to check and test the unit, probes, and external peripherals for...
Page 60
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL 3-2Section 3-1 - Overview 3-1-3 Installation Warnings 1.) Since the LOGIQ™ 7 weighs approximately 225 kg. (496 lb) without options, preferably two people should unpack it. Two people are also preferable for installing any additional bulky items. 2.) There are no operator serviceable components. To prevent shock, do not remove any covers or panels. Should problems or malfunctions occur, unplug the power cord. Only qualified service...