Section 2-2 - General Console Requirements 2-5
2-2-3  EMI Limitations
Ultrasound machines are susceptible to Electromagnetic Interference (EMI) from radio frequencies, 
magnetic fields, and transient in the air wiring. They also generate EMI. The LOGIQ™ 7 complies with 
limits as stated on the EMC label. However there is no guarantee that interface will not occur in a 
particular installation.
Possible EMI sources should be identified...

Page 52

GE HEALTHCARE
DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
2-6Section 2-2 - General Console Requirements
2-2-4 Probes Environmental Requirements
Table 2-6   Operation and storage Temperatures for 2D Probes
OperationalStorageTransport
Temperature
10 - 40 oC
50 - 104 oF -10 - 60 
oC
14 - 140oF -40 - 60 
oC
-40 - 140 oF
Humidity30 - 85%
non-condensing 30 - 90%
non-condensing30 - 90%
non-condensing
Pressure700 - 1060hPa 700 - 1060hPa 700 - 1060hPa
Table 2-7   Operation and storage Temperatures for...

Page 53

GE HEALTHCARE
DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
Section 2-3 - Facility Needs 2-7
Section 2-3
Facility Needs
2-3-1 Purchaser Responsibilities
The work and materials needed to prepare the site is the responsibility of the purchaser. Delay, 
confusion, and waste of manpower can be avoided by completing pre installation work before delivery. 
User the Pre Installation checklist to verify that all needed steps have been taken,
Purchaser reasonability includes:
• Procuring the materials...

Page 54

GE HEALTHCARE
DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
2-8Section 2-3 - Facility Needs
2-3-2 Required Features
• Dedicated single branch power outlet of adequate amperage (see  Table 2-3) meeting all local and 
national codes which is located less than 2.5 m (8 ft.) from the unit’s proposed location
• Door opening is at least 76 cm (30 in) wide
• Proposed location for unit is at least 0.3 m (1 ft.) from the wall for cooling
• Power outlet and place for any external peripheral are within 2 m...

Page 55

GE HEALTHCARE
DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
Section 2-3 - Facility Needs 2-9
2-3-4 Recommended and Alternate Ultrasound Room Layout
Recommended standard floor plan and a minimal floor plan for ultrasound equipment:
Figure 2-1   RECOMMENDED ULTRASOUND ROOM  LAYOUT
LOGIQ
LOGIQ 

Page 56

GE HEALTHCARE
DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
2-10Section 2-3 - Facility Needs
2-3-5 Networking Pre-installation Requirements
2-3-5-1 Purpose of DICOM Network Function
DICOM services provide the operator with clinically useful features for moving images and patient 
information over a hospital network. Examples of DICOM services include the transfer of images to 
workstations for viewing or transferring images to remote printers. As an added benefit, transferring 
images in this...

Page 57

GE HEALTHCARE
DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
Section 2-3 - Facility Needs 2-11
2-3-5-2 DICOM Option Pre-installation Requirements (cont’d)
.
Figure 2-2   Worksheet for DICOM Network Information
LOGIQ™ 7
Host Name
AE TitleLocal PortIP Address
Net Mask. .
..
.
..
ROUTING INFORMATION
ROUTER2
ROUTER3 ROUTER1
.
..
.
..
.
..
.
..
.
..
.
..
.
.. Destination 
IP Addresses
GATEWAY IP Addresses
Default
DICOM APPLICATION INFORMATION NAME
Store 2 Store 3 Store 1
MAKE/REVISION
IP ADDRESSESPORT...

Page 58

GE HEALTHCARE
DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
2-12Section 2-3 - Facility Needs
This page was intentionally left blank. 

Page 59

GE HEALTHCARE
DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
Section 3-1 - Overview 3-1
Chapter 3
Installation
Section 3-1
Overview
3-1-1 Purpose of Chapter 3
This chapter contains information needed to install the unit. Included are references to a procedure that 
describes how to receive and unpack the equipment and how to file a damage or loss claim.
How to prepare the facility and unit of the actual installation, and how to check and test the unit, probes, 
and external peripherals for...

Page 60

GE HEALTHCARE
DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
3-2Section 3-1 - Overview
3-1-3 Installation Warnings
1.) Since the LOGIQ™ 7 weighs approximately 225 kg. (496 lb) without options, preferably two people 
should unpack it. Two people are also preferable for installing any additional bulky items.
2.) There are no operator serviceable components. To prevent shock, do not remove any covers or  panels. Should problems or malfunctions occur, unplug the power cord. Only qualified service...