Page 91

GE HEALTHCARE
DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
Section 4-2 - General Procedure 4-7
4-2-3 Power Shutdown
Purpose: This is a description on how to Shutdown the system.
4-2-3-1 Complete Power Down 1.) Press the ON/OFF
 key at the front of the System for about two (2) seconds. Refer to Figure 4-2.
2.) Switch OFF the Main Circuit Breaker at the rear of the system Refer to Figure 4-1.
3.) Disconnect the Main Power Cable if needed. Refer to Figure 4-1.
4-2-4 System Stand-by
LOGIQ 7 is...

Page 92

GE HEALTHCARE
DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
4-8Section 4-2 - General Procedure
4-2-4 System Stand-by (cont’d)
7.) To power up the scanner, tap POWER button momentary, and wait for power up.  It will  approximately 45seconds for the scanner to power up.
Stand-by Sequence
NOTICEDo not press the Power button long time, as it will cause the system to power down.
Figure 4-11   Stand-by Sequence
Press POWER button
Select "Standby"
Ch e c k Mo n ito rOFFWait
Check TouchPanel OFF...

Page 93

GE HEALTHCARE
DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
Section 4-2 - General Procedure 4-9
4-2-5 Using CD-R/MOD/DVD Drive
4-2-5-1 Using CD-R or DVD Drive
1.) Push the EJECT
 button, the disk tray will appear.
2.) Put the disk onto the disk tray.
3.) Press the EJECT
 button to insert the disk into the CD-R or DVD device.
4.) There are a number of methods to eject a disk from the CD-R or DVD. Ejection is automatic in some  cases. Manual ejection methods, listed in preferred order of use, are:...

Page 94

GE HEALTHCARE
DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
4-10Section 4-2 - General Procedure
4-2-5-2 Using MOD Drive
1.) Before installing an MO disk in the MOD, check the MO disk for loose hardware or damaged labels which could jam inside the MO Drive. Also ensure that the slide switch in one corner of the disk is 
set so that the disk is write enabled (disk hole closed).
2.) Insert the disk into the MOD with the label facing up.
3.) The are a number of methods to eject a disk from the MOD....

Page 95

GE HEALTHCARE
DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
Section 4-2 - General Procedure 4-11
4-2-6 Archiving and Loading Presets for BT07 (Including R7.5.x)
NOTE: Always save presets before any software reload. This ensures the presets loaded after the software 
reload are as up–to–date as possible.
All user presets except changes to Summary, Anatomy, and Biometry pages, can be saved on an CD-
R/MO/DVD-R disk for reloading on the system.
4-2-6-1 Regional Preset - General (Supported from BT07)...

Page 96

GE HEALTHCARE
DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
4-12Section 4-2 - General Procedure
4-2-6-2 Using Regional Preset
Regional Preset is selected during Application Load (Refer to 8-8-13 - Installing R7.x.x Application 
Software for details).
Current Regional Preset is displayed under  Utility > System > About .
Factory Default Regional Preset can be re-loaded from  Utility > Systme > About > Additional About.
Figure 4-14   Current Regional Preset
NOTICEFor Upgraded BT07, this field...

Page 97

GE HEALTHCARE
DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
Section 4-2 - General Procedure 4-13
4-2-6-3 Cautions Using Regional Preset
• Do not attempt to modify Regional Preset on Upgraded BT07.
• Backup/Restore function should be used between the same region systems.  Unexpected setting may result if you restore the preset files to another region setting system.
4-2-6-4 Formatting CD-R/DVD-R Disk 1.) Insert an empty (blank) CD-R/DVD-R disk into the CD-R/DVD-R device.
2.) Access to the Utility...

Page 98

GE HEALTHCARE
DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
4-14Section 4-2 - General Procedure
4-2-6-5 Archiving Presets to an CD-R/DVD-R Disk
1.) Insert an empty (blank) formatted CD-R/DVD-R disk into the CD-R/DVD-R device.
2.) Access to the Utility  Menu on the Touch Panel, and select  System>Backup/Restore . The Backup 
screen will be shown on the monitor.
3.) Select the item to back up either from Resource Files.
4.) Select the media to locate the items.
5.) Click on  Backup. The backup...

Page 99

GE HEALTHCARE
DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
Section 4-2 - General Procedure 4-15
4-2-6-6 Loading Presets from an CD-R/DVD-R disk
1.) Insert the CD-R/DVD-R disk with the archived Presets into the CD-R/DVD-R.
2.) Access to the Utility  Menu on the Touch Panel, and select  System>Backup/Restore. The Restore 
sheet will be shown on the monitor.
3.) Select the item to restore either from resource Files.
4.) Click on  Restore. A message to make sure the restore process is displayed on...

Page 100

GE HEALTHCARE
DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
4-16Section 4-2 - General Procedure
4-2-7 Archiving and Loading Presets for BT06/V65x or lower
NOTE: Always save presets before any software reload. This ensures the presets loaded after the software 
reload are as up–to–date as possible.
All user presets except changes to Summary, Anatomy, and Biometry pages, can be saved on an CD-
R/MO/DVD-R disk for reloading on the system.
4-2-7-1 Formatting CD-R/MO/DVD-R Disk 1.) Insert an empty...