Page 71
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL Section 3-4 - Completing the Installation 3-13 3-4-2 Optional Peripherals/Peripheral Connection (cont’d) Check if the fixing belts are loosen. If it is, tighten the belts again to secure the peripheral(s). If necessary, remove the peripheral from the device room to tighten the belts. 3-4-2-1 Approved on-board peripherals Refer to Section 5-3 Peripheral Compatibility . Figure 3-6 How to Hold Peripherals Belts Belts BeltsConnect the...
Page 72
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL 3-14Section 3-4 - Completing the Installation 3-4-2-1 Approved on-board peripherals (cont’d) Connecting Cables 3-4-2-2 Reference off-board peripherals and options None. CAUTIONEquipment damage possibility. Be sure to use the following recommended connecting cables to connect recording devices and a network with LOGIQ™ 7 console. Table 3-4 List of Connecting Cables NamePart No.FigureNOTE Power Supply Cable P9509EE Connected to power...
Page 73
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL Section 3-4 - Completing the Installation 3-15 3-4-3 Available Probes See in specification in the LOGIQ™ 7 Reference Manual for Probes and intended use. See Chapter 9 - Renewal Parts for Part Numbers to be used when ordering new or replacement probes. Table 3-5 List of Transducers supported (All models) Probe Name Material of Headshell Area of UsingTYPECatalog NumberPart Number 3C PESABDOMINAL CONVEX H79802P H40412LB 2286353 2286354...
Page 74
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL 3-16Section 3-4 - Completing the Installation NOTE: PES: Polyethersulfone NORYL: Modified Polyphenylene Oxide PU: Polyurethane PBT: Polybutylene Terephthalate ABS: Acrylonitrile Butadiene Styrene NOTE: Some probes indicated on the table above have two different part numbers. The upper row shows the part numbers of probes for Japan. The lower row shows the part numbers of probes for regions other than Japan. Probes which have only one...
Page 75
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL Section 3-4 - Completing the Installation 3-17 3-4-4 Video Specification 3-4-5 Software Option Configuration 3-4-5-1 Onsite check and configuration Select Utility > Admin > System Admin and check the option software to be installed. Table 3-9 Video Specifications CRTLCD Timing Parameter 800x600 75Hz Timing Parameter 1280 x 1024 60Hz Horizontal Rate [kHz] 46.88Horizontal Rate [kHz] 64 Horizontal Period [ µs] 21.33 Horizontal Period...
Page 76
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL 3-18Section 3-5 - Installation Paperwork Section 3-5 Installation Paperwork NOTE: During and after installation, the documentation (i.e. Users Manual, Installation Manuals...) for the peripheral units must be kept as part of the original system documentation. This will ensure that all relevant safety and user informations are available during the operation and service of the complete system. 3-5-1 Peripherals/Accessories Connector...
Page 77
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL Section 3-5 - Installation Paperwork 3-19 3-5-1-1 Rear Panel Connector Located on the rear panel are video input and output connectors, audio input and output, camera expose connectors, footswitch connector power connector and control connections for VCR, printer, and service tools. This section indicates the pin assignment for each connector. Figure 3-7 Rear Pannel Connector (BT04 or later) Video In Rec/Expose VGA S-Video Out...
Page 78
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL 3-20Section 3-5 - Installation Paperwork 3-5-1-1 Rear Panel Connector (cont’d) NOTE: Each outer (case) ground line of peripheral/accessory connectors are protectively grounded. Signal ground lines are not isolated, except the Service port (3). All of signal lines (include signal GND) of the Service port are isolated. The specified peripherals/accessories only can be connected to the USB ports.Rear Pannel Connector for BT03 or lower...
Page 79
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL Section 3-5 - Installation Paperwork 3-21 3-5-1-1 Rear Panel Connector (cont’d) Pin Assignment of S-Video Connector Connector: S-Terminal, 4-pin Pin Assignment of VGA Connector Connector: Shrank D-Sub, 15-pin Pin Assignment of Service/VCR 1/VCR 2 Connector, USB1.1 Connector: 4 pin Table 3-10 Pin Assignment of S-Video Connector Pin No Output/Input Signal Description 1SVIDEO OUT/IN YG Y (Luma) GND 2 SVIDEO OUT/IN CG C (Chroma) GND 3...
Page 80
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL 3-22Section 3-5 - Installation Paperwork 3-5-1-1 Rear Panel Connector (cont’d) Pin Assignment of Digital Int. Connector Connector: 6-pin Pin Assignment of Ethernet Connector: RJ-45 Modular, 8-pin Pin Assignment of Insite Connector: RJ-11 Modular, 6-pin Table 3-13 Pin Assignmen t of IEEE-1394 Connector) Pin NoOutput Signal Description 1VP 1394 Power Supply 2 VG 1394 Power Ground 3 TPB 1394 Data B (-) 4 TPB 1394 Data B (+) 5 TPA 1394...