Section 3-4 - Completing the Installation 3-13
3-4-2 Optional Peripherals/Peripheral Connection (cont’d)
Check if the fixing belts are loosen. If it is, tighten the belts again to secure the peripheral(s). If 
necessary, remove the peripheral from the device room to tighten the belts.
3-4-2-1 Approved on-board peripherals Refer to Section 5-3 Peripheral Compatibility .
Figure 3-6   How to Hold Peripherals
Belts
Belts BeltsConnect the...

Page 72

GE HEALTHCARE
DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
3-14Section 3-4 - Completing the Installation
3-4-2-1 Approved on-board peripherals  (cont’d)
Connecting Cables
3-4-2-2 Reference off-board peripherals and options None.
CAUTIONEquipment damage possibility. Be sure to use the following recommended connecting cables to connect 
recording devices and a network with LOGIQ™ 7 console.
Table 3-4   List of Connecting Cables
NamePart No.FigureNOTE
Power Supply Cable P9509EE Connected to power...

Page 73

GE HEALTHCARE
DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
Section 3-4 - Completing the Installation 3-15
3-4-3 Available Probes
See in specification in the LOGIQ™ 7 Reference Manual for Probes and intended use.
See Chapter 9 - Renewal Parts for Part Numbers to be used when ordering new or replacement probes.
Table 3-5   List of Transducers supported (All models)
Probe Name
Material of 
Headshell
Area of UsingTYPECatalog NumberPart Number
3C PESABDOMINAL CONVEX H79802P
H40412LB 2286353
2286354...

Page 74

GE HEALTHCARE
DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
3-16Section 3-4 - Completing the Installation
NOTE: PES: Polyethersulfone NORYL: Modified Polyphenylene Oxide PU: Polyurethane
PBT: Polybutylene Terephthalate ABS: Acrylonitrile Butadiene Styrene
NOTE: Some probes indicated on the table above have two different part numbers. The upper row  shows the part numbers of probes for Japan. The lower row shows the part numbers of probes 
for regions other than Japan. Probes which have only one...

Page 75

GE HEALTHCARE
DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
Section 3-4 - Completing the Installation 3-17
3-4-4 Video Specification
3-4-5 Software Option Configuration
3-4-5-1 Onsite check and configuration
Select Utility > Admin > System Admin  and check the option software to be installed.
Table 3-9   Video Specifications
CRTLCD
Timing Parameter
800x600
75Hz
Timing Parameter
1280 x 1024 60Hz
Horizontal Rate [kHz] 46.88Horizontal Rate [kHz] 64
Horizontal Period [ µs] 21.33 Horizontal Period...

Page 76

GE HEALTHCARE
DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
3-18Section 3-5 - Installation Paperwork
Section 3-5
Installation Paperwork
NOTE: During and after installation, the documentation (i.e. Users Manual, Installation Manuals...) for 
the peripheral units must be kept as part of the original system documentation. This will ensure 
that all relevant safety and user informations are available during the operation and service of 
the complete system.
3-5-1 Peripherals/Accessories Connector...

Page 77

GE HEALTHCARE
DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
Section 3-5 - Installation Paperwork 3-19
3-5-1-1 Rear Panel Connector
Located on the rear panel are video input and output connectors, audio input and output, camera 
expose connectors, footswitch connector power connector and control connections for VCR, printer, 
and service tools. 
This section indicates the pin assignment for each connector.
Figure 3-7   Rear Pannel Connector (BT04 or later)
Video In
Rec/Expose
VGA
S-Video
Out...

Page 78

GE HEALTHCARE
DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
3-20Section 3-5 - Installation Paperwork
3-5-1-1 Rear Panel Connector (cont’d)
NOTE: Each outer (case) ground line of peripheral/accessory connectors are protectively grounded.  Signal ground lines are not isolated, except the Service port (3). All of signal lines (include signal 
GND) of the Service port are isolated.
The specified peripherals/accessories only can be connected to the USB ports.Rear Pannel Connector for BT03 or lower...

Page 79

GE HEALTHCARE
DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
Section 3-5 - Installation Paperwork 3-21
3-5-1-1 Rear Panel Connector (cont’d)
Pin Assignment of S-Video Connector Connector: S-Terminal, 4-pin
Pin Assignment of VGA Connector Connector: Shrank D-Sub, 15-pin
Pin Assignment of Service/VCR 1/VCR 2 Connector, USB1.1 Connector: 4 pin Table 3-10   Pin Assignment of S-Video Connector
Pin No Output/Input Signal Description
1SVIDEO OUT/IN YG Y (Luma) GND
2 SVIDEO OUT/IN CG C (Chroma) GND
3...

Page 80

GE HEALTHCARE
DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
3-22Section 3-5 - Installation Paperwork
3-5-1-1 Rear Panel Connector (cont’d)
Pin Assignment of Digital Int. Connector Connector: 6-pin
Pin Assignment of Ethernet Connector: RJ-45 Modular, 8-pin
Pin Assignment of Insite Connector: RJ-11 Modular, 6-pin Table 3-13   Pin Assignmen t of IEEE-1394 Connector)
Pin NoOutput Signal Description
1VP 1394 Power Supply
2 VG 1394 Power Ground
3 TPB 1394 Data B (-)
4 TPB 1394 Data B (+)
5 TPA 1394...