GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL Section 6-4 - Reloading the Probe Data 6-9 Section 6-4 Reloading the Probe Data This procedure will be used for troubleshooting when a image sensitivity is NOT recovered. For details, refer to Section 7, Troubleshooting. This rewrites the probe data (probe delay data for beam forming) saved on the flash memories of DDBF and TRAP boards. 1.) Insert the Service Dongle into the USB port, located at the rear panel of the scanner. 2.) Power...
Page 172
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL 6-10Section 6-5 - Monitor and LCD Adjustments Section 6-5 Monitor and LCD Adjustments 6-5-1 CRT Monitor Contrast and Brightness Adjustment Before calibrate the Monitor, check your PCVIC or DGVIC Part Number on the Utility Screen. The typical setting value changes with its Part Number. (To enter this screen, select Utility > System > About > Additional About Information.) To adjust the contrast and brightness: 1.) Select Calibration from...
Page 173
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL Section 6-5 - Monitor and LCD Adjustments 6-11 6-5-1 CRT Monitor Contrast and Brightness Adjustment (cont’d) 2.) Press the Toggle button (1) for contrast and brightness. Confirm that the contrast (or brightness) indicator is displayed on the monitor. If the brightness is displayed, press the toggle button again. 3.) Set the Contrast according to the table below. 4.) Set the Brightness at 100. Then slowly decrease the Brightness until...
Page 174
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL 6-12Section 6-5 - Monitor and LCD Adjustments 6-5-219 inch LCD Monitor Brightness Adjustment To adjust the brightness: 1.) Press the Toggle button (2) or (3) Once for brightness adjustment. a.) Verify that Brightness is 80. If it is not, increase brightness to press the adjustment button (3). If it is not, decrease brightness to press the adjustment button (2). b.) Press the button (1) Once. The brightness display disappears. NOTE: 19...
Page 175
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL Section 6-5 - Monitor and LCD Adjustments 6-13 6-5-317 inch LCD Monitor Contrast and Brightness Adjustment To adjust the contrast and brightness: 1.) Press the Toggle button (1) Once for brightness adjustment. a.) Verify that Brightness is 55. If it is not, increase brightness to press the adjustment button (3). b.) Press the button (1) again for confirmation. 2.) Press the Toggle button (1) Twice for contrast adjustment. a.) Verify...
Page 176
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL 6-14Section 6-5 - Monitor and LCD Adjustments 6-5-4 LCD Monitor Resolution Should one encounter “small screen” on the LCD monitor, screen resolution must be adjusted in order to view full size. 1.) From the touch panel, select Utility > System > About. 2.) Click Additional About Info. 3.) Select High Resolution. 6-5-5 LCD Touch Panel Adjustment There are two types of the touch panel LCD in LOGIQ 7 (manufactured by Gaudi or Technart)....
Page 177
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL Section 6-5 - Monitor and LCD Adjustments 6-15 6-5-5-1 For Gaudi LCD 1.) Refer to 4-2-2-3 Entering Maintenance Mode to enter the windows desktop. 2.) The following screen should appear when exiting to Windows. 3.) Check the followings: - Proper balance of the LCD display as shown in Figure 6-17 - No Jitter (Phenomenon that a character does not move sideways) NOTE: If the location of the LCD screen is appropriately set in center, do not...
Page 178
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL 6-16Section 6-5 - Monitor and LCD Adjustments 6-5-5-1 For Gaudi LCD (cont’d) 5.) Find out the notch to access OSD Switch located at backside of the LCD as shown in Figure 6-18. Press the switch inside of the notch. NOTE: Be careful not to cut fingers with notch. 6.) OSD menu is shown on the LCD display. 7.) Remove the maintenance rod, and re-install the OP Panel temporarily to use the LCD touch panel. Refer to Replacement Procedure in...
Page 179
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL Section 6-5 - Monitor and LCD Adjustments 6-17 6-5-5-1 For Gaudi LCD (cont’d) 11.) After finishing the Auto Adjustment procedures, the following menu is displayed on the LCD. 12.) Make sure that the LCD position is fixed and shown at the center location. 13.) If it is fixed appropriately, select OK then touch ENTER to go on to the next step. If the display is not adjusted yet, select CANCEL then touch ENTER to repeat the adjustment...
Page 180
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL 6-18Section 6-5 - Monitor and LCD Adjustments 6-5-5-2 For Technart 6-5-5-2-1 Auto Adjustment 1.) Refer to 4-2-2-3 Entering Maintenance Mode to enter the windows desktop. 2.) The following screen should appear when exiting to Windows. 3.) Press the AUTO button once on the LCD touch panel board. Figure 6-21 Start-up Screen Figure 6-22 Auto Button Auto Button