Home > GE > Health/Medical Alert > GE Cardiocap 5 Service Manual

GE Cardiocap 5 Service Manual

    Download as PDF Print this page Share this page

    Have a look at the manual GE Cardiocap 5 Service Manual online for free. It’s possible to download the document as PDF or print. UserManuals.tech offer 45 GE manuals and user’s guides for free. Share the user manual or guide on Facebook, Twitter or Google+.

    							
    Functional Check Form 
    NeuroMuscular Transmission (NMT) test 
     OK N.A. Fail 
    29. Recognition    
    30. Communication and memory    
    31. ElectroSensor recognition    
    32. Stimulus current test    
    33. TOF measurement with NMT simulator 
    T1%  950-1059 Supramax. current  < 70 mA 
    T2%  950-1059 TOF%  95-105 
    T3%  950-1059 Count  4 
    T4%  950-1059 T1%  95-105 
    Ratio%  950-1059 
    Noise  < 50 
    34. Stimulus pulse width 
     100 µs 300 µs Allowed range 
    TOF%   95-105 
    Count   4 
    T1%   95-105 
     OK N.A. Fail 
    35. No response to stimulus    
    36. DBS measurement with NMT simulator 
    T1%  950-1059 DBS%  95-105 
    T2%  950-1059 Count  2 
    Ratio%  950-1059 T1%  95-105 
    37. ST measurement with NMT simulator 
    T1%  950-1059 Count  1 
    T1%  95-105 
     OK N.A. Fail 
    38. Automatic measurement off    
    39. MechanoSensor recognition    
    Appendix A 3  
    						
    							
    Functional Check Form 
    Performance Checks 
    General OK N.A. Fail 
    1. Screen setup    
    2. Waveforms    
    InvBP measurement    
    3. Test with patient simulator    
    SpO2 measurement    
    4. SpO2 sensor detection    
    5. Test measurement     
    6. “Probe off” detection    
    NIBP measurement    
    7. Test measurement and cuff identification    
    Gas measurement    
    8. D-fend check    
    9. CO2 waveform    
     
    Notes  
     
     
     
     
     
     
     
     
     
     
     
     
    Signature  Date  
     
    Appendix A 4    
    						
    							
    EMC Guidance 
    EMC GUIDANCE  
    Table 1. Guidance and manufacturer’s declaration – electromagnetic emissions 
     
    Guidance and manufacturer’s declaration – electromagnetic emissions 
    The Cardiocap/5 is intended for use in the electromagnetic environment specified below. The customer 
    or the user of the Cardiocap/5 should assure that it is used in such an environment. 
    Emissions test Compliance Electromagnetic environment - guidance 
    RF emissions 
    CISPR 11 
    Group 1 The Cardiocap/5 uses RF energy only for its internal 
    function. Therefore, its RF emissions are very low and are 
    not likely to cause any interference in nearby electronic 
    equipment. 
    RF emissions 
    CISPR 11 
    Class A 
    Harmonic emissions 
    IEC 61000-3-2 
    Class A 
    Voltage fluctuations/ 
    flicker emissions 
    IEC 61000-3-3 
    Complies 
    The Cardiocap/5 is suitable for use in all establishments 
    other than domestic and those directly connected to the 
    public low-voltage power supply network that supplies 
    buildings used for domestic purposes. 
     
    Appendix B 1  
    						
    							
    Functional Check Form 
    Table 2. Guidance and manufacturer’s declaration – electromagnetic immunity 
    Guidance and manufacturer’s declaration – electromagnetic immunity 
    The Cardiocap/5 is intended for use in the electromagnetic environment specified below. The 
    customer or the user of the Cardiocap/5 should assure that it is used in such an environment. 
    Immunity test IEC 60601 test 
    level 
    Compliance level Electromagnetic environment - 
    guidance 
    Electrostatic 
    discharge (ESD) 
    IEC 61000-4-2 
    ±6 kV contact 
     
    ±8 kV air 
    ±6 kV contact 
     
    ±8 kV air 
    Floors should be wood, concrete or 
    ceramic tile. If floors are covered 
    with synthetic material, the relative 
    humidity should be at least 30 %. 
    Electrical fast 
    transients/bursts 
    IEC 61000-4-4 
    ±2 kV for power 
    supply lines 
    ±1 kV for 
    input/output lines 
    ±2 kV for power 
    supply lines(1 
    ±1 kV for 
    input/output lines 
    Mains power quality should be that of a 
    typical commercial or hospital 
    environment. 
    Surge 
    IEC 61000-4-5 
    ±1 kV differential 
    mode 
    ±2 kV common 
    mode 
    ±1 kV differential 
    mode 
    ±2 kV common 
    mode 
    Mains power quality should be that 
    of a typical commercial or hospital 
    environment. 
    Voltage dips, 
    short 
    interruptions and 
    voltage variations 
    on power supply 
    lines 
    IEC 61000-4-11 
    95 % dip in UT) 
    for 0.5 cycle 
     
    40 % UT 
    (60 % dip in UT) 
    for 5 cycles 
     
    70 % UT 
    (30 % dip in UT) 
    for 25 cycles 
     
    95 % dip in UT) 
    for 5 sec 
    95 % dip in UT) 
    for 0.5 cycle 
     
    40 % UT 
    (60 % dip in UT) 
    for 5 cycles 
     
    70 % UT 
    (30 % dip in UT) 
    for 25 cycles 
     
    95 % dip in UT) 
    for 5 sec 
    Mains power quality should be that 
    of a typical commercial or hospital 
    environment. If user of the 
    Cardiocap/5 requires continued 
    operation during power mains 
    interruptions, it is recommended that 
    the Cardiocap/5 be powered from an 
    uninterruptible power supply or a 
    battery. 
    Power frequency 
    (50/60 Hz) 
    magnetic field 
    IEC 61000-4-8 
    3 A/m 3 A/m Power frequency magnetic field 
    should be at levels characteristic of 
    a typical location in a typical 
    commercial or hospital environment. 
    NOTE UT is the a.c. mains voltage prior to application of the test level. 
    (1 ±1 kV for ECG and RESP measurement and K-CREMCO. 
     
    Appendix B 2    
    						
    							
    EMC Guidance 
    Table 3. Guidance and manufacturer’s declaration – electromagnetic immunity 
    Guidance and manufacturer’s declaration – electromagnetic immunity 
    The Cardiocap/5 is intended for use in the electromagnetic environment specified below. The customer or the user 
    of the Cardiocap/5 should assure that it is used in such an environment. 
    Immunity test IEC 60601 test 
    level 
    Compliance 
    level 
    Electromagnetic environment - guidance 
       Portable and mobile RF communications equipment 
    should be used no closer to any part of the 
    Cardiocap/5, including cables, than the 
    recommended separation distance calculated from 
    the equation applicable to the frequency of the 
    transmitter. 
    Recommended separation distance 
    Conducted RF 
    IEC 61000-4-6 
    3 Vrms 
    150 kHz to 80 MHz 
    3 Vrms d = 1.2 P   
     3 Vrms 150 kHz to 80 MHz 1 Vrms (2 d = 3.5 P  
    Radiated RF 
    IEC 61000-4-3 
    3 V/m 
    80 MHz to 2.5 GHz 
    3 V/m d = 1.2 P     80 MHz to 800 MHz 
    d = 2.3 P     800 MHz to 2.5 GHz 
       where P is the maximum output power rating of the 
    transmitter in watts (W) according to the transmitter 
    manufacturer and d is the recommended separation 
    distance in metres (m). 
    Field strengths from fixed RF transmitters, as 
    determined by an electromagnetic site survey, a 
    should be less than the compliance level in each 
    frequency range. b 
     
    Interference may occur in the vicinity of equipment 
    marked with the following symbol: 
     
    NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.  
    NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption 
    and reflection from structures, objects and people. 
    a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land 
    mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicated theoretically 
    with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site 
    survey should be considered. If the measured field strength in the location in which the Cardiocap/5 is used 
    exceeds the applicable RF compliance level above, the Cardiocap/5 should be observed to verify normal 
    operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or 
    relocating the Cardiocap/5. 
    b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m. 
    (2 For N-XOSAT SpO2 measurement. 
     
    Appendix B 3  
    						
    							
    Functional Check Form 
    Table 4. Recommended separation distances between portable and mobile RF communications 
    equipment and the Cardiocap/5. 
    Recommended separation distances between portable and mobile RF communications 
    equipment and the Cardiocap/5. 
    The Cardiocap/5 is intended for use in an electromagnetic environment in which radiated RF 
    disturbances are controlled. The customer or the user of the Cardiocap/5 can help prevent 
    electromagnetic interference by maintaining a minimum distance between portable and mobile RF 
    communications equipment (transmitters) and the Cardiocap/5 as recommended below, according to the 
    maximum output power of the communications equipment. 
    Separation distance according to frequency of transmitter 
    m 
    Rated maximum 
    output power of 
    transmitter 
    W 
    150 kHz to 80 MHz
    d = 1.2 P  
    d = 3.5 P (2 
    80 MHz to 800 MHz
    d = 1.2 P  
    800 MHz to 2.5 GHz 
    d = 2.3 P  
    0.01 0.12 
    0.35 (2 0.12 0.23 
    0.1 0.38 
    1.1 (2 0.38 0.73 
    1 1.2 
    3.5 (2 1.2 2.3 
    10 3.8 
    11 (2 3.8 7.3 
    100 12 
    35 (2 12 23 
    For transmitters rated at a maximum output power not listed above, the recommended separation 
    distance d in meters (m) can be estimated using the equation applicable to the frequency of the 
    transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the 
    transmitter manufacturer. 
    NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies. 
    NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by 
    absorption and reflection from structures, objects and people. 
    (2 For N-XOSAT SpO2 measurement. 
    Appendix B 4    
    						
    							
    Contents 
    Chapter 3. Planned Maintenance  
    3.1 Introduction ........................................................................\
    . 3-1 
    3.1.1 Using the Planned Maintenance Form .......................................................... 3-1 
    3.1.2 Recommended parts........................................................................\
    ........... 3-1 
    3.1.3 Recommended tools and accessories .......................................................... 3-2 
    3.2 Planned maintenance checks................................................ 3-3 
    1. Visual inspection ........................................................................\
    .......... 3-3 
    2. Parts replacement ........................................................................\
    ........ 3-3 
    3. Functional inspection ........................................................................\
    ... 3-4 
    4. ECG measurement test ........................................................................\
    . 3-5 
    5. Temperature measurement test ............................................................. 3-5 
    6. Non-invasive blood pressure  (NIBP) measurement test............................ 3-6 
    7. SpO2 measurement test ........................................................................\
     3-6 
    8. Invasive blood pressure measurement test ............................................. 3-7 
    9. Gas measurement test........................................................................\
    .. 3-7 
    10. Anesthetic agent identification test ........................................................ 3-7 
    11. Spirometry test........................................................................\
    ............. 3-8 
    12. Gas sampling line and  D-fend check ...................................................... 3-8 
    13. Trend test ........................................................................\
    .................... 3-8 
    14. Watchdog test........................................................................\
    .............. 3-8 
    15. Recorder test ........................................................................\
    ............... 3-8 
    16. Network test........................................................................\
    ................. 3-9 
    17. Data card test ........................................................................\
    ............ 3-10 
    18. Service log check ........................................................................\
    ....... 3-10 
    19. Electrical safety check ........................................................................\
     3-10 
    Planned Maintenance Form ..........................................End of this chapter  
    Table of Figures  
    Figure 3-1.  Nafion tube and fi lters to replace in gas unit .................................................... 3-3 
     
      
    						
    							
    Planned Maintenance 
    3. PLANNED MAINTENANCE 
    3.1 Introduction 
    These instructions include procedures for planned maintenance (PM) of the Datex-Ohmeda 
    Cardiocap/5 monitor. Performance of planned maintenance procedures is recommended once each 
    year after installation of the monitor.  
    These instructions are for the maximum functional configuration of the monitor. Perform the 
    procedures in order and skip items that do not correspond with the configuration of your monitor. 
    Complete instructions on how to perform complex procedures are included in the Service Procedures 
    chapter later in this manual. 
    3.1.1 Using the Planned Maintenance Form 
    A Planned Maintenance Form is included at the end of this chapter. Complete this “checklist” as you 
    perform the maintenance procedures.  
    NOTE:  means to sign the form after performing the procedure. 
    3.1.2 Recommended parts 
    Part Order Number 
    Nafion™ tube 733382 
    Filter (2 pieces) 886136 
    OM reference filter 86901 
    Zero absorber 895933 
    D-fend O-ring (2 pieces) 65312 
    D-fend (black) 876446 
    D-fend+ (green) 881319 
    Sampling line 3.0 m 73319 
    Fan filter 896113 
    Recorder paper (if N-XREC option is installed) 74205 
    Planned Maintenance Kit, Anesthesia 8001760 
    Planned Maintenance Kit, Critical Care 8001761 
      
    NOTE: The contents of the Planned Maintenance Kits are listed in the Spare Parts chapter. 
    3-1  
    						
    							
    Cardiocap/5 Technical Reference Manual 
    3.1.3 Recommended tools and accessories 
    Tool Order Number 
    Patient simulator for ECG, Impedance Respiration, and BP Obtain locally 
    Pressure manometer Obtain locally 
    Temperature test set 884515 
    5-lead ECG cable Obtain locally 
    SpO2 finger sensor (compatible with the installed pulse oximetry) Obtain locally 
    InvBP transducer Obtain locally 
    Adult NIBP cuff 572435 
    Adult NIBP hose 877235 
    Infant NIBP cuff 877407 
    Infant NIBP hose 877514 
    Flowmeter Obtain locally 
    Flow cassette 50/1.1 873812 
    Extra silicon tubing Obtain locally 
    Calibration gas (CO2, N2O, O2, Des)  
    North America (must use with regulator 75553-01) 755571 
    Outside North America (must use with regulator 755533) 755583 
    Sampling line 3.0 m 73319 
    Spirometry tester 884202 
    Spirometry tube 884101 
    D-lite 733950 
    Screwdriver, pozidrive type Obtain locally 
    Hexagon wrench, M3 Obtain locally 
    Troubleshooting extension cable 884298 
     
    3-2  
    						
    All GE manuals Comments (0)