Home > GE > Health/Medical Alert > GE Cardiocap 5 Service Manual

GE Cardiocap 5 Service Manual

    Download as PDF Print this page Share this page

    Have a look at the manual GE Cardiocap 5 Service Manual online for free. It’s possible to download the document as PDF or print. UserManuals.tech offer 45 GE manuals and user’s guides for free. Share the user manual or guide on Facebook, Twitter or Google+.

    							
    Measurement Parameters 
    6.2 Measurement principles for hemodynamic parameters 
    6.2.1 NIBP 
    NIBP (Non-Invasive Blood Pressure) is an indirect method for measuring blood pressure. 
    The NIBP measurement is performed according to the oscillometric measuring principle. The cuff is 
    inflated with a pressure slightly higher than the presumed systolic pressure and deflated at a speed 
    based on the patients pulse. Data are collected from the oscillations caused by the pulsating artery. 
    Based on these oscillations, values for systolic, mean, and diastolic pressures are calculated. 
    6.2.2 ECG 
    Electrocardiography analyzes the electrical activity of the heart by measuring the electrical potential 
    produced with electrodes placed on the surface of the body. 
    ECG reflects 
    • Electrical activity of the heart. 
    • Normal/abnormal function of the heart. 
    • Effects of anesthesia on heart function. 
    • Effects of surgery on heart function. 
    See the Users Reference Manual for electrode positions and other information. 
    6.2.3 Impedance respiration 
    Impedance respiration is measured across the thorax between three ECG electrodes. The respiration 
    signal is made by supplying current between two electrodes and by measuring the differential current 
    from the third electrode. The input current is 200 µA (31 kHz).  The impedance measured is the 
    impedance change caused by breathing. When the patient is breathing or is ventilated, the volume of 
    air in the lungs changes, resulting in impedance between the electrodes. From these impedance 
    changes, the respiration rate is calculated and the respiration waveform is displayed on the screen.  
    6.2.4 Temperature 
    Temperature is measured by a probe whose resistance varies when the temperature changes, called 
    NTC (Negative Temperature Coefficient) resistor. 
    The resistance can be measured by two complementary methods: 
    • Applying a constant voltage across the resistor and measuring the current that flows through it. 
    • Applying a constant current through the resistor and measuring the voltage that is generated 
    across it. 
    The two methods are combined in the form of a voltage divider. The NTC resistor is connected in series 
    with a normal resistor and a constant voltage is applied across them. The temperature dependent 
    voltage can be detected at the junction of the resistors, thus producing the temperature signal from 
    the patient. The signal is amplified by analog amplifiers and further processed by digital electronics. 
    6-3  
    						
    							
    Cardiocap/5 Technical Reference Manual 
    6.2.5 Invasive blood pressure (N-XP option) 
    A catheter is inserted into an artery or vein to measure invasive blood pressure. The invasive pressure 
    setup—consisting of tubing, a pressure transducer, and an intravenous bag of normal saline all 
    connected together by stopcocks—is attached to the catheter. The transducer is placed level with the 
    heart and electrically zeroed. 
    The transducer is a piezo-resistive device that converts the pressure signal to a voltage. The monitor 
    interprets the voltage signal so that pressure data and pressure waveforms can be displayed. 
    6.2.6 Pulse oximetry, standard 
    NOTE: Only one pulse oximetry source at a time is allowed by the Cardiocap/5. When the N-XOSAT or  
    N-XNSAT option is installed in the monitor, standard Cardiocap/5 pulse oximetry is not available. 
    A pulse oximeter measures the light absorption of blood at two wavelengths, one in the near infrared 
    (about 900 nm) region and the other in the red region (about 660 nm) of the light spectrum. These 
    wavelengths are emitted by LEDs in the SpO2 sensor. The light is transmitted through peripheral tissue 
    and is detected by a PIN diode opposite to the LEDs in the sensor. The pulse oximeter derives the 
    oxygen saturation (SpO2) using an empirically-determined relationship between the relative absorption 
    at the two wavelengths and the arterial oxygen saturation, SaO2. 
    To measure the arterial saturation accurately, pulse oximeters use the component of light absorption 
    giving variations synchronous with heart beat as primary information on the arterial saturation.  
     
    Figure 6-3. Absorption of infrared light in the finger 
    6-4  
    						
    							
    Measurement Parameters 
    A general limitation of the above pulse oximetry principle is that due to the use of only two 
    wavelengths, only two hemoglobin species can be discriminated by the measurement.  
    Modern pulse oximeters are empirically calibrated against fractional saturation (SaO2frac) or against 
    functional saturation (SaO2func) as shown below: 
    SaO2frac = HbO2/(HbO2 + Hb + Dyshemoglobin) 
    SaO2func = HbO2/(HbO2 + Hb). 
    Functional saturation (SaO2func) is less sensitive to changes of carboxyhemoglobin and 
    methemoglobin concentrations in blood. 
    The oxygen saturation percentage SpO2 is calibrated against the functional saturation (SaO2func). The 
    advantage of this method is that the accuracy of SpO2 measurement relative to SaO2func can be 
    maintained even at rather high concentrations of carboxyhemoglobin in blood. Independent of the 
    calibration method, a pulse oximeter is not able to correctly measure oxygen content of the arterial 
    blood at elevated carboxyhemoglobin or methemoglobin levels. 
    Plethysmographic pulse wave 
    The plethysmographic waveform is derived from the IR signal and reflects the blood pulsation at the 
    measuring site. Thus the amplitude of the waveform represents the perfusion. 
    Sensor 
    SpO2 sensors contain the light source LEDs, which are located opposite the photodiode detector. 
    Different kinds of sensors are available from GE Healthcare, including clip-on and wrap styles. 
    Pulse rate 
    The pulse rate calculation is done by peak detection of the plethysmographic pulse wave. The signals 
    are filtered to reduce noise and checked to separate artifacts. 
    6.2.7 Pulse oximetry, Datex-Ohmeda enhanced (N-XOSAT option) 
    NOTE: Only one pulse oximetry source at a time is allowed by the Cardiocap/5. When the N-XOSAT 
    option is installed in the monitor, standard pulse oximetry and N-XNSAT are not available. 
    The Datex-Ohmeda enhanced pulse oximetry option uses a two-wavelength pulsatile system—red and 
    infrared light—to distinguish between oxygenated (O2Hb) and reduced (HHb) hemoglobin, each of 
    which absorbs different amounts of light emitted from the oximeter sensor. The SpO2 and pulse rate 
    are determined by the oximeter through sensor signal processing and microprocessor calculations. 
    The SpO2 sensor contains a light source and a photodetector: 
    • The light source consists of red and infrared light-emitting diodes (LEDs). 
    • The photodetector is an electronic device that produces an electrical current proportional to 
    incident light intensity. 
    The two light wavelengths generated by the sensor light source (the red and infrared LEDs) pass 
    through the tissue at the sensor site. The light is partially absorbed and modulated as it passes 
    through the tissue. 
    Arterial blood pulsation at the sensor site modulates transmission of the sensor’s light. Since other 
    fluids and tissues present generally don’t pulsate, they don’t modulate the light passing through that 
    location. The pulsatile portion of the incoming signal is used to detect and isolate the attenuation of 
    light energy due to arterial blood flow.   
    6-5  
    						
    							
    Cardiocap/5 Technical Reference Manual 
     
    Figure 6-4. Comparative light absorption 
    The sensor’s photodetector collects and converts the light into an electronic signal. Since O2Hb and 
    HHb allow different amounts of light to reach the photodetector at the selected wavelengths, the 
    electronic signal varies according to which light source is “on” (red or infrared) and the oxygenation of 
    the arterial hemoglobin. The oximeter uses this information to calculate the relative percentage of 
    O2Hb and HHb. 
     
    Figure 6-5. Extinction versus wavelength graph 
    The photodetector sends the electronic signal, which contains the light intensity information, to the 
    oximeter. The oximeter’s electronic circuitry processes the electronic signal, calculates the SpO2 and 
    pulse rate values, and displays them on the screen. 
    Calibration 
    Datex-Ohmeda enhanced pulse oximetry uses two wavelength ranges, 650 nm to 665 nm and 930 
    nm to 950 nm, both with an average power of less than 1 mW. These wavelengths are used to 
    calculate the presence of oxyhemoglobin (O2Hb) and reduced hemoglobin (HHb). 
    6-6  
    						
    							
    Measurement Parameters 
    A CO-oximeter typically uses four or more wavelengths of light and calculates reduced hemoglobin 
    (HHb), oxyhemoglobin (O2Hb), carboxyhemoglobin (COHb), and methemoglobin (MetHb). Therefore, 
    pulse oximetry readings and CO-oximetry readings will differ in situations where a patient’s COHb or 
    MetHb are increased. Increased patient COHb leads to falsely increased SpO2 in all pulse oximeters. 
    Datex-Ohmeda enhanced pulse oximetry uses functional calibration, which is represented 
    mathematically as the percentage of hemoglobin capable of carrying oxygen that is carrying oxygen. 
     
    The calculation of SpO2 assumes 1.6% carboxyhemoglobin (COHb), 0.4% methemoglobin (MetHb), 
    and no other pigments. These values are based on the Datex-Ohmeda Pulse Oximeter Empirical 
    Calibration Study. Appreciable variation from these values will influence SpO2 accuracy. 
    6.2.8 Pulse oximetry, Nellcor compatible (N-XNSAT option) 
    NOTE: Only one pulse oximetry source at a time is allowed by the Cardiocap/5. When the N-XNSAT 
    option is installed in the monitor, standard pulse oximetry and N-XOSAT are not available. 
    The N-XNSAT pulse oximetry option uses the Nellcor pulse oximetry algorithm and should be used with 
    Nellcor pulse oximetry sensors only. Refer to the Cardiocap/5 User’s Reference Manual for a list of 
    Nellcor sensors approved for use with this option. 
    6-7  
    						
    							
    Cardiocap/5 Technical Reference Manual 
    6.3 Measurement principles for airway gases, spirometry, and NMT 
    6.3.1 CO2, N2O, and agent measurement 
    TPX is a side-stream gas analyzer that measures real-time concentrations of CO2, N2O and anesthetic 
    agents (Halothane, Enflurane, Isoflurane, Desflurane, and Sevoflurane). The TPX analyzer identifies 
    anesthetic agents or mixtures of two anesthetic agents and measures their concentrations. It also 
    detects mixtures of more than two agents and issues an alarm. 
     
    Figure 6-6. TPX sensor principle 
    TPX is a nondispersive infrared analyzer that measures absorption of the gas sample at seven infrared 
    wavelengths that are selected using optical narrow band filters. 
    The infrared radiation detectors are thermopiles. 
    Concentrations of CO2 and N2O are calculated from absorption measured at 3 to 5 µm. 
     
    Figure 6-7. Absorbance of N2O and CO2 
    Identification of anesthetic agents and calculation of their concentrations is performed by measuring 
    absorptions at five wavelengths in the 8 to 9 µm band, then using a set of five equations to solve the 
    concentrations. 
    6-8  
    						
    							
    Measurement Parameters 
     
    Figure 6-8. Infrared absorbance of AAs 
    To achieve measuring accuracy, numerous software compensation parameters are individually 
    determined for each TPX during factory calibration. 
    6.3.2 O2 measurement 
    The differential oxygen measuring unit uses the paramagnetic principle in a pneumatic bridge 
    configuration. The signal picked up with a differential pressure transducer is generated in a measuring 
    cell with a strong magnetic field that is switched on and off at a frequency of 165 Hz. The output signal 
    is a DC voltage proportional to the difference in O2 concentration between the two gases to be 
    measured. 
     
    Figure 6-9. O2 measurement principle 
    6-9  
    						
    							
    Cardiocap/5 Technical Reference Manual 
    6.3.3 Patient spirometry measurement 
    In mechanical ventilation, breaths are delivered to the patient by a ventilator with a proper tidal volume, 
    respiration rate (RR), and inspiration/expiration ratio in a time determined by the ventilator settings.  
    Patient Spirometry monitors patient ventilation and displays these parameters: 
    • Expiratory and inspiratory tidal volume (TV) in ml 
    • Expiratory and inspiratory minute volume (MV) in l/minute 
    • Expiratory spontaneous minute volume in l/minute 
    • Expiratory volume in first second (V1.0) in percent (%) 
    • Inspiration/expiration ratio (I:E) 
    Airway pressure 
    • Peak pressure (Ppeak) 
    • Mean airway pressure (Pmean)—Critical care software only 
    • End inspiratory pressure (Pplat) 
    • PEEPi, PEEPe—Critical care software only 
    NOTE: PEEPi = intrinsic PEEP, PEEP-PEEPe 
    • Positive end expiratory pressure (PEEP)—Anesthesia software only 
    • Real time airway pressure waveform (Paw) 
    PEEP, Ppeak, Pmean, and Pplat are measured by a pressure transducer on the PVX board. Atmospheric 
    pressure is used as a reference. The pressure measurement is made from the airway part that is 
    closest to the patient between the patient circuit and the intubation tube. 
    Airway flow 
    • Real time flow waveform (V) 
    • Compliance (Compl) 
    • Airway resistance (Raw) 
    • Pressure volume loop 
    • Flow volume loop 
    The kinetic gas pressure is measured using the Pitot effect. A pressure transducer measures the Pitot 
    pressure. The obtained pressure signal is linearized and corrected according to the density of the gas. 
    Flow speed is calculated from these pressure values, then the tidal volume (TV) value is integrated. The 
    minute volume (MV) value is calculated and averaged using the TV and respiratory rate values. 
    6-10  
    						
    							
    Measurement Parameters 
    Compliance and airway resistance 
    Compliance tells how big a pressure difference is needed to deliver a certain amount of gas into the 
    patient. Compliance is calculated for each breath from the following equation: 
    ComplTVexp
    PPEEPiPEEPeplat
    =−−
     
    The airway resistance, Raw, is calculated using an equation that describes the kinetics of the gas flow 
    between the lungs and the D-lite flow sensor. The equation states that the pressure at the D-lite can at 
    any moment of the breath be approximated using the equation 
    p(t) = Raw • V’(t) + V(t) / Compl + PEEP 
    where p(t), V(t) and V(t) are the pressure, flow and volume measured at the D-lite at a time (t), 
    Raw is the airway resistance, Compl is the compliance, and PEEP is the total end-expiratory 
    pressure. 
    D-liteTM flow sensor 
    Patient Spirometry uses a D-lite flow sensor. The adult D-lite sensor measures adults; the pediatric 
    measures children. Single-use and reusable versions of these sensors are available. 
    The D-lite adapter measures kinetic pressure by a two-sided Pitot tube. Velocity is calculated from the 
    pressure difference according to Bernoullis equation:  
     
    where: v = velocity (m/second), dP = pressure difference (cmH2O), and ρ = density (kg/m3)  
    Flow is then determined using the calculated velocity: 
    F = v x A 
    where: F = flow (l/minute), v = velocity (m/second), and A = cross area (m2) 
    The volume information is obtained by integrating the flow signal. 
    6-11  
    						
    							
    Cardiocap/5 Technical Reference Manual 
    6.3.4 NeuroMuscular Transmission (NMT) measurement 
    The NMT feature provides peripheral nerve stimulation and response measurement that supports 
    electromyography EMG. You can also use the Cardiocap/5 NMT feature as a nerve locator for regional 
    nerve blocking with a regional block cable, however, in this case there is no response measurement. 
    Nerve stimulation 
    There are three NMT stimulus modes: Train of Four (TOF), Double Burst 3,3 (DBS) and Single Twitch (ST). 
    • Train of Four stimulus generates four stimulation pulses at 0.5 second intervals. The response is 
    measured after each stimulus and the ratio of the fourth and first response of the TOF sequence is 
    calculated (TOF%). If the first response does not exceed a certain signal level, TOF% is not 
    calculated due to poor accuracy. 
    • Double burst (3,3) stimulation includes two bursts with a 750 ms interval. Both bursts consist of 
    three pulses separated by 20 ms intervals. The responses of both bursts are measured, and the 
    ratio of the second and first response is calculated (DBS%). EMG responses are measured 
    immediately after the first stimulus pulse of both bursts. 
    • Single Twitch stimulation generates one stimulation pulse. The response is measured after the 
    stimulus. To prevent decurarization of the stimulated area, the measurement is automatically 
    stopped after 5 minutes stimulation in 1-second cycle time. 
    Tetanic/Post-Tetanic Count (PTC) 
    Tetanic/PTC can measure deeper relaxation than TOF. The tetanic stimulation is produced when Start 
    is chosen under Tetanic/PTC. The length of stimulation is 5 seconds. The stimulation generates 
    pulses with a frequency of 50 Hz and with a selected pulse width and current. After tetanic stimulation 
    and a three second delay, Single Twitch stimulation is produced to detect the post tetanic count (PTC). 
    PTC describes the number of responses detected after tetanic stimulation. If there is no response, the 
    measurement will be stopped. If responses do not fade away, a maximum of 20 responses will be 
    calculated. If more can be detected, the PTC value is displayed only as “> 20” and measurement 
    stops.  
    If the TOF, DBS, or ST measurement cycle was on when tetanic stimulation started, the cycle will 
    continue after the PTC. After completing the PTC measurement during 1 minute TOF, DBS, or another 
    PTC measurement is not possible. This is to avoid erroneous readings due to post-tetanic potentiation. 
    Response 
     
    Figure 6-10. Principle of response measurement 
    6-12  
    						
    All GE manuals Comments (0)