Home > GE > Health/Medical Alert > GE Cardiocap 5 Service Manual

GE Cardiocap 5 Service Manual

    Download as PDF Print this page Share this page

    Have a look at the manual GE Cardiocap 5 Service Manual online for free. It’s possible to download the document as PDF or print. UserManuals.tech offer 45 GE manuals and user’s guides for free. Share the user manual or guide on Facebook, Twitter or Google+.

    							
    Service Menus 
    8.9.5 NIBP Buttons/Leds menu and data 
     
    To enter this service menu from 
    the Service View menu: 
    1. Select Modules. 
    2. Select NIBP.  
    3. Select Buttons/Leds. 
    The Auto ON/OFF, Manual ON/OFF, STAT ON/OFF, and Measur. ON/OFF menu selections have no 
    effect. 
    Buttons/Leds Data screen 
    See NIBP service menu, data, and checks earlier in this chapter for information on Pressure, Zero, 
    Protect handle, Calibr. prot., +15 V power, AD0 to AD7, Timeouts, etc. 
    The Auto On/Off and Start Cancel fields are not used with Cardiocap/5. 
    8-25  
    						
    							
    Cardiocap/5 Technical Reference Manual 
    8.9.6 NIBP Pneumatics menu and data 
     
    To enter this service menu from 
    the Service View menu: 
    1. Select Modules. 
    2. Select NIBP.  
    3. Select Pneumatics. 
    Start Pump Starts the pump; changes to Stop Pump when the pump is on. 
    Open Exh1 Opens exhaust valve 1; changes to Close Exh1 when the valve is open. 
    Open Exh2 Opens exhaust valve 2; changes to Close Exh2 when the valve is open. 
    Set Valve Adjusts the pulse-valve opening between 0 (fully closed) and 255 (fully open). To 
    adjust, push the ComWheel, turn it to adjust the value shown on the screen, and 
    push it to set the value. 
    Reset Clock Zeros the time on the display. 
    Pneumatics Data screen 
    See NIBP service menu, data, and checks earlier in this chapter for information on Pressure, Zero, 
    Protect handle, Calibr. prot., +15 V power, AD0 to AD7, Timeouts, etc. 
    Pump shows the state of the pump (ON/OFF). 
    Exh1 Valve and Exh2 Valve show the states of exhaust valve 1 and exhaust valve 2 (OPEN/CLOSED). 
    Pulse Valve shows how much the valve is opened (0 to 255) during Valve Setting. 
    Watchdog BEEP check 
    Connect a manometer to the front panel and pump in pressure. When the AD5 value changes from a 
    negative value to a positive value (at about 15 mmHg/2.0 kPa), a beep sounds. This is the watchdog 
    threshold pressure. Beyond this pressure the watchdog is active and cuts pressure at about 2 minutes 
    (adult). 
    8-26  
    						
    							
    Service Menus 
    Interval 20 mmHg -> 185 mmHg check 
    1. Wrap an adult cuff around a pipe and connect the cuff to the monitor.  
    2. Select Start Pump at different combinations of the valves open/closed and push the ComWheel.  
    The time it takes for the pressure to go from 20 mmHg to 185 mmHg (2.7 to 24.7 kPa) is counted 
    and displayed. When all valves are closed, the pump should pump the pressure in about 1 to 4 
    seconds.  
    3. To stop the pump, choose Stop Pump and push the ComWheel. 
    8.9.7 NIBP Watchdog menu and data 
     
    To enter this service menu from 
    the Service View menu: 
    1. Select Modules. 
    2. Select NIBP.  
    3. Select Watchdog. 
    Test ADULT Tests the watchdog timer in adult mode (120 to 140 seconds). 
    Test INFANT Tests the watchdog timer in infant mode (about 60 to 70 seconds). 
    Stop Test Stops the test. 
    Watchdog Data screen 
    See NIBP service menu, data, and checks earlier in this chapter for information on Pressure, Zero, 
    Protect handle, Calibr. prot., +15 V power, AD0 to AD7, Timeouts, etc. 
    Watchdog Interval shows the time the +15 Vdirty stays on during the test. 
    Watchdog timer tests (adult and infant) 
    Choose Test ADULT and push the ComWheel. Watchdog interval starts counting seconds and 
    continues counting as long as the +15 Vdirty is on. The time should be 120 to 140 seconds. 
    Choose Test INFANT and push the ComWheel. Watchdog interval starts counting seconds and 
    continues counting as long as the +15 Vdirty is on. The time should be 60 to 70 seconds. 
    8-27  
    						
    							
    Cardiocap/5 Technical Reference Manual 
    8.10 NMT menu and data 
     
    To access this menu from the Service 
    View menu: 
    1. Select Modules. 
    2. Select NMT. 
     
    Start Curr. test A test that checks the difference between a given current and the measured 
    current.  
     Connect a 3 kΩ resistance between the stimulus electrodes before starting the 
    test. All currents that are checked will be displayed on the service data screen. 
    After the current value, the test status is shown.  
     If the test is not passed, factory calibration is needed.  Contact GE Healthcare. 
    T1%/ref  NMT MechanoSensor setting. Always use a setting of 3 for normal monitoring. 
    Settings 1 and 2 are for research purposes only. 
    Record Data Prints the displayed service data and circuit board information (id., serial number, 
    and software id.) to the optional Recorder. 
    NMT Service Data screen 
    Cable The type of cable being used. 
    Cable check value The bit amount. You can check the value for each cable using this chart: 
    EMG –100 to +100 
    Cable OFF > 1950 
    Piezo 900 to 1100 
    Regional block 600 to 800 
      
    Module Keys  Not applicable to the Cardiocap/5. 
    Stimulus mode The stimulus mode selected in the NMT Setup menu. 
    Measurement Indicates status (ON/OFF). 
    8-28  
    						
    							
    Service Menus 
    T1%, etc. The measured response. A value of 1000 corresponds to 100%. Each is a ratio of 
    the Absolute T1, etc. to the Reference (Ref). As drugs are administered, the 
    typical response for T1 lessens, then drops more rapidly for each successive 
    measured response (T2%, T3%, then T4%).  
    Ratio% The common label for TOF/DBS%.  
    Absolute T1, etc. The voltage measured from the A/D converter. The area-under-the-curve of the 
    response to stimulation as the muscle-relaxant anesthetic is being administered. 
    Ref (Reference) The area-under-the-curve calculation from the response of the NMT 
    stimulation before the administration of inhibiting anesthetics.  
    Noise The interference just before the measurement. A typical value is < 10. Represents 
    the measurement of the area-under-the-curve of the response when no stimulus 
    is given, but the circuitry noise is picking up some response. The noise factor 
    varies from sensor to sensor. 
    Offset An average of the noise measurement; the voltage baseline. An offset of 2520 is 
    2520 mV, or 2.5 V.  This is the measureable baseline voltage of the sensor and 
    varies from sensor to sensor.  
    Curr set The selected current, a value of 700 corresponds to 70 mA. 
    Curr meas The measured current, a value of 700 corresponds to 70 mA. 
    Pulses Pulses the module has produced. 
    Piezo probe T1% Information related to MechanoSensor settings. 
    Ref. search Information related to MechanoSensor settings. 
    Timeouts, Bad checksums, and Bad c-s by mod are cumulative numbers that indicate the following: 
    Timeouts—how many times the NMT unit did not respond to an inquiry from the 
    monitor. 
    Bad checksums—how many times communication from the NMT unit to the 
    monitor failed. 
    Bad c-s by mod—how many communication errors the NMT unit detected. 
    Counting starts at power on and is reset to zero at power off. Nonzero values do 
    not indicate a failure, but continuous counting (more than 50 per second) 
    indicates a serial communication failure or that the NMT unit is not installed. Also, 
    communicaton errors from other modules can cause these numbers to rise. 
    RAM The state of RAM memory.  
    ROM Indicates whether the checksum in the EPROM is in accordance with the value 
    calculated by the software.  
    EEPROM Indicates if the values stored in permanent memory are valid. The state is either 
    OK, Fail or ? (module not installed or a communication error). 
    8-29  
    						
    							
    Cardiocap/5 Technical Reference Manual 
    8.10.1 NMT Setup Menu 
     
    To access this menu from the Service View menu: 
    1. Select Modules. 
    2. Select NMT. 
    3. Select NMT Setup. 
     
    Start-up Starts an NMT measurement. 
    Continue Continues an NMT measurement. 
    Stimulus Mode Selects the mode of stimulation:  TOF (train of four), DBS (double burst 
    stimulation), or ST (single twitch). 
    Cycle Time Sets the time intervals for the selected Stimulus Mode. For TOF and DBS, the 
    selections are Manual, 10, 12, 15, or 20 seconds, and 1, 5, or 15 minutes. The 
    ST cycle times are 1, 10, and 20 seconds. 
    When the cycle time is less than one minute, the Count must be below the Count 
    limit in two successive measurements before the Recovery Note is enabled. To 
    give the note, the number of Counts must exceed or match the limit twice in one 
    minute. The note disappears when two measurements are below the limit again. 
    If the cycle time is one minute or more, or if the measurement is done manually, 
    at least one Count must be below the chosen limit before the Recovery Note is 
    enabled. To give the note, one Count must exceed or match the Count limit. The 
    note disappears when one measurement is below the limit again. 
    Current Sets the stimulus current, which is either supramaximal (automatic start-up 
    search) or manually selected from 10 mA through 70 mA at 5 mA intervals. 
    Pulse Width Increases or decreases the effect of the stimulation. Selections are 100 µs,  
    200 µs, and 300 µs. 
     
    8-30  
    						
    							
    Service Menus 
    8.11 M-NSAT menu and data (N-XNSAT or N-XOSAT options) 
     
    To enter this service menu from the Service 
    View menu: 
    1. Select Modules. 
    2. Select M-NSAT. 
    PR Pulse rate value [bpm] calculated from pleth. 
    SpO2% Oxygen saturation value multiplied by 100. 
    NoProbe 
    PulseSearch 
    NoPulse 
    CheckProbe 
    Not applicable to the Cardiocap/5. 
    MP-203 Error Yes if an error message is sent by the SpO2 board. 
    NOTE:  MP-203 Error is used to represent any version of the Datex-Ohmeda or 
    Nellcor SpO2 board. 
    QUART Error Yes if an error is detected in the QUART data communications device located in the 
    SpO2 interface board. 
    I/O Error Yes when a communication error occurs between the SpO2 board and the SpO2 
    interface board. 
    Timeouts, Bad checksums, and Bad c-s by mod are cumulative numbers that indicate the following: 
    Timeouts—how many times the unit did not respond to an inquiry from the monitor. 
    Bad checksums—how many times communication from the SpO2 unit to the 
    monitor failed. 
    Bad c-s by mod—number of communication errors detected by the SpO2 unit. 
    Counting starts at power on and is reset to zero at power off. Nonzero values do not 
    indicate a failure, but continuous counting (> 50 per second) indicates a serial 
    communication failure or that the SpO2 unit is not installed. Also, communicaton 
    errors from other modules can cause these numbers to rise. 
    ROM Indicates whether the checksum in the EPROM is in accordance with the checksum 
    calculated by the software. The state is OK, Fail or ? (unit not installed or 
    communication error). 
    8-31  
    						
    							
    Cardiocap/5 Technical Reference Manual 
    8.12 More Modules menu 
     
    To enter this service menu from the Service View menu: 
    1. Select Modules. 
    2. Select More Modules.  
    NOTE: The Interface and Battery selections in the More Modules menu are not used in Cardiocap/5. 
    8-32  
    						
    							
    Service Menus 
    8.12.1 Memory Module menu and Module Status data 
     
    To enter this service menu from 
    the Service View menu: 
    1. Select Modules. 
    2. Select More Modules.  
    3. Select Memory Module. 
    Module Status screen 
    NOTE: All values can be - - - to indicate “No data is available.” 
    Module present Indicates whether the module is firmly attached to the monitor (YES or NO). 
    Module active Indicates whether module services are available (YES or NO). 
    ROM Status of the memory module’s ROM memory (OK or ERR). 
    RAM Status of the memory module’s RAM memory (OK or ERR). 
    PCMCIA Status of the memory module’s PCMCIA controller (OK or ERR). 
    EEPROM Status of the memory module’s EEPROM memory (OK or ERR). 
    SLOT1, SLOT2 The foremost slot and the rearmost slot, respectively. 
    Card type The type of card (MENU or DATA). If duplicated card is inserted, type DUPL. 
    File system The memory card file system (ATA is the only supported file system). If a non-ATA 
    card is used, UNKNOWN is shown. If the card is poorly attached, LOOSE is shown. 
    Card size The total amount of disk space in the card in kilobytes. 
    Card used The total amount of used disk space in the card in kilobytes. 
    Card full Indicates whether all disk space in the card is used (YES or NO). 
    Card empty Indicates the lack of menu files in the MENU card or no files in the DATA card. 
    Possible values are YES and NO. 
    Read error Indicates whether reading from the card failed (YES or NO). 
    Write error  Indicates whether writing to the card failed (YES or NO). 
    8-33  
    						
    							
    Cardiocap/5 Technical Reference Manual 
    8.12.2 Memory Module menu and Communication data 
     
    To enter this service menu from the 
    Service View menu: 
    1. Select Modules. 
    2. Select More Modules.  
    3. Select Memory Module.  
    4 Select Communication. 
    Communication screen 
    Interface status 
    The status of the data link between the monitor and the memory module. If memory module is 
    properly attached, the status should always be ACTIVE. If status blinks between ACTIVE and 
    CLOSED, a communications error has occurred: remove module briefly, and insert it back into the 
    monitor frame to check if the error disappears. 
    Message types  The type of data packets sent (Tx) and received (Rx) since the last monitor start: 
    Record K—communication between the Monitor and Record Keeper. 
    File Op.—operations of Patient data. 
    Service—Memory Module operations. 
    Modes—User Mode operations. 
    Module status—number of sent/received data packets that relate to the memory module status. 
    Packets total—total number of data packets sent/received since last monitor start. 
    Bytes total—total number of data bytes sent/received since last monitor start. 
    Transmission errors since the last monitor start: 
    Timeouts—The number of time-outs that occurred in memory-module data transmission. 
    Chksum err—The number of checksum errors in data packets from memory module. 
    Length err—The number of data packets with erroneous length from the memory module. 
    Duplicated—The number of duplicate data packets from the memory module. 
    8-34  
    						
    All GE manuals Comments (0)