Home > GE > Health/Medical Alert > GE Cardiocap 5 Service Manual

GE Cardiocap 5 Service Manual

    Download as PDF Print this page Share this page

    Have a look at the manual GE Cardiocap 5 Service Manual online for free. It’s possible to download the document as PDF or print. UserManuals.tech offer 45 GE manuals and user’s guides for free. Share the user manual or guide on Facebook, Twitter or Google+.

    							
    Overview 
    1.3 Monitor structure 
    The Cardiocap/5 can be equipped with several factory-configured options. The block diagram and 
    descriptions that follow represent the maximum functionality of the monitor with all options installed. 
     
    Figure 1-1. Cardiocap/5 monitor structure 
    The main software and measurement technologies are based on AS/3 hardware and software. Some 
    parameter-measuring unit boards are interchangeable with AS/3 module boards, however, the units 
    cannot be replaced with the corresponding modules as the hardware of the assemblies is different. 
    1.3.1 Measurement parameter units 
    The maximum Cardiocap/5 parameter measurement configuration consists of the NESTPR, PVX, and 
    CAiO units plus ONE of the following units:  NMT or NSAT or OSAT.  
    The NESTPR, CAiO, and NMT units are connected to the CPU through the Mother board and 
    communicate with the CPU over a standard AS/3 module bus. The NSAT or OSAT pulse oximetry unit is 
    also connected to the CPU through the Mother board and communicates with the CPU over a standard 
    AS/3 module bus, which is located on the SpO2 interface board.  
    Each parameter measurement board contains a CPU that processes measurement data for the 
    parameter(s) associated with that board before sending the data to the main CPU. 
    NESTPR unit 
    The NESTPR unit contains three boards for measuring hemodynamic parameters: 
    • The ECG board measures ECG (3-lead and 5-lead) and impedance respiration. 
    • The STP board measures oxygen saturation, temperature, and invasive blood pressure. 
    • The NIBP board measures noninvasive blood pressure.  
    1-3  
    						
    							
    Cardiocap/5 Technical Reference Manual 
    PVX unit for Patient Spirometry (N-XV option) 
    The PVX unit connects to the CAiO unit and measures the patient’s airway flow and pressure (Patient 
    Spirometry).  
    CAiO unit (N-XC, N-XCO, and N-XCAiO options) 
    The CAiO unit measures airway gases. It is capable of measuring CO2, N2O, O2, anesthetic agents (AA) 
    and also identifying the present anesthetic agent. 
    Datex-Ohmeda enhanced pulse oximetry (N-XOSAT option) 
    The OSAT unit measures oxygen saturation and pulse rate using Datex-Ohmeda enhanced pulse 
    oximetry technology. 
    Nellcor® compatible pulse oximetry (N-XNSAT option) 
    The NSAT unit measures oxygen saturation and pulse rate using signal processing electronics and 
    software that are based on Nellcor stand-alone oximeters.  
    NeuroMuscular Transmission (N-XNMT option) 
    The NMT unit measures the relaxation status (TOF, DBS, and ST) of patients. When used with a regional 
    block cable, the unit acts as a nerve locator.  
    1.3.2 Communication 
    The CPU communicates with the hemodynamic parameters measuring unit (NESTPR) and airway gas 
    measuring unit (CAiO) over a standard AS/3 module bus. It is based on the widely used industry 
    standard RS485, which uses a differential serial method to transfer data and is quite robust.  
    RS485 serial communication supports multidrop or party line connections. This means all units 
    connected to the module bus use the same two physical wires for communication purposes. The 
    module bus uses a 500 kbps data transfer rate. 
    Communication with the I/O board and the DC/DC board takes place over an internal synchronous 
    serial bus. The same bus also controls display brightness and audio signals by means of D/A 
    converters located on the CPU board.  
    Communication with the Net takes place over a separate synchronous serial channel. Communication 
    with the Net is possible, if the CPU board is equipped with an Ethernet controller. 
    Communication with the Recorder takes place over an asynchronous serial channel. 
    1.3.3 CPU board 
    The control functions of the monitor are centralized on the CPU board. The CPU: 
    • Controls the power on/off sequencing. 
    • Controls the brightness of the LCD screen by means of the Backlight board. 
    • Controls the Inverter board that provides the high voltage for the display backlights. 
    • Reads input from the keyboard and ComWheel. 
    • Controls the serial channels and I/O functions of the monitor. 
    Two PCMCIA-compatible data card slots on the board are for loading software and transferring data. 
    1-4  
    						
    							
    Overview 
    Distributed processing 
    The parameter and airway measuring units contain their own microprocessor systems for performing 
    low level functions, such as waveform filtering and pneumatics control. At the same time, the main 
    CPU performs higher level tasks (trending and alarm control, for example). 
    1.3.4 Display 
    The main CPU directly controls the monitor display, a 10.4 inch color LCD. Supply voltages for the 
    display are connected via the CPU board and the Backlight board. The Inverter board provides high 
    voltage for the display backlights. The CPU controls the display brightness by adjusting the backlight 
    voltage. 
    1.3.5 I/O board 
    The I/O board contains D/A converters for analog outputs and the audio output amplifier. It also 
    contains connectors for the network identification plug, serial I/O, analog output, and external 
    keyboard. 
    Analog outputs are created by transferring digital data from the CPU to the D/A converters on the I/O 
    board through the internal synchronous serial bus. The network identification plug is connected to the 
    CPU over a separate synchronous serial channel. 
    1.3.6 DC/DC board 
    The DC/DC board converts 15 VDC coming from the AC/DC unit to different supply voltages for the 
    monitor. All outputs are short-circuit and over-voltage protected. The CPU controls the output voltages. 
    If the mains voltage drops, the 12 VDC back-up battery automatically supplies power to the monitor. 
    The battery will run the monitor for at least 15 minutes. The battery is always charged when mains 
    voltage is connected. The temperature sensor that measures the monitor’s internal temperature is 
    located on the board. The DC/DC board communicates with the CPU over the internal synchronous 
    serial bus.  
    1.3.7 AC/DC unit 
    The AC/DC unit converts the mains voltage to 15 VDC that is fed to the DC/DC board. The input 
    voltage range of the unit is 100 to 240 VAC. The DC/DC board shuts down the AC/DC unit when there 
    is over voltage detected on the 15 VDC output. The shutdown mode is reset by detaching the mains 
    power cord for 30 seconds. 
    1.3.8 Recorder (N-XREC option) 
    The Recorder prints trend data and record parameter waveforms. It connects to the CPU through an 
    asynchronous serial channel. Recorder supply voltages connect through a Recorder board that is 
    permanently attached to the recorder mounting box. The board contains a voltage filter and a delay 
    circuit for 12 V. 
    1-5  
    						
    							
    Cardiocap/5 Technical Reference Manual 
    1.4 Symbol definitions 
    Symbols on equipment 
     
    Attention! Read accompanying instructions, including all warnings and cautions, 
    before using this device.  
    This symbol has the following meanings when it appears on the screen: 
    • On the front panel indicates that protection against cardiac defibrillator discharge is 
    due in part to the accessories for pulse oximetry (SpO2), temperature (T) and 
    invasive pressure (P) measurement. 
    • When displayed beside the O2 value, indicates that the FiO2 low-alarm limit is set 
    below 21%. 
    • When displayed next to the HR value, indicates that there is a risk that the monitor 
    counts pacemaker spikes (pacer is set ON R) or the monitor counts T-waves (a wide 
    QRS is selected). 
     
    Type BF applied part (IEC 60601-1). Defibrillator-proof protection against electric shock. 
     
    Type CF applied part (IEC 60601-1). Defibrillator-proof protection against electric shock. 
     
    Main Menu. Located beside the ComWheel to indicate you can open the Main Menu by 
    pressing the ComWheel when no other menu is displayed. 
     
    Power On/Standby. 
    Pb 
    This battery contains lead. Separate from other waste for disposal according to local 
    regulations. 
    Pb 
    This battery contains lead and can be recycled.  
     
    Dangerous voltage. 
     Gas outlet (in airway gas models only). 
     Ethernet connectors. 
     
    Equipotentiality. Monitor can be connected to potential equalization conductor. 
     Alternating current. 
     Fuse. 
    1-6  
    						
    							
    Overview 
     
    ESD warning symbol for electrostatic sensitive devices. Pins of connectors identified 
    with the ESD warning symbol should not be touched. Connections should not be 
    made to these connectors unless ESD precautionary procedures are used. See 
    Safety precautions: ESD precautionary procedures in the “User’s Reference 
    Manual” for details. 
     
    Symbol for non-ionizing electromagnetic radiation. Interference may occur in the 
    vicinity of equipment marked with the symbol. 
     
    This symbol indicates that the waste of electrical and electronic equipment must not 
    be disposed as unsorted municipal waste and must be collected separately. Please 
    contact an authorized representative of the manufacturer for information concerning 
    the decommissioning of your equipment. 
    Symbols on screens 
     
    When displayed on the upper left corner of the screen, indicates alarms are silenced. 
    When in the menu or digit fields, indicates that the alarm source has been turned off. 
     Sub menu. Selecting an alternative with this symbol in a menu opens a new menu. 
     The monitor is connected to the Monitor Network. 
     
    Data card (green) and/or Menu card (white) is inserted. 
     Indicates the beats detected. 
     Respiration rate is measured using impedance respiration measurement. 
     Back-up battery operation and remaining capacity. 
     Back-up battery charging. 
    Symbols on transport packaging 
     The contents of the package are fragile and have to be handled with care. 
     Indicates the correct upright position of the transport package. 
     
    The package must be kept in a dry environment. 
     
    The package should be kept within the indicated temperature limitations. 
    1-7  
    						
    							
    Cardiocap/5 Technical Reference Manual 
    1.5 Safety precautions 
    1.5.1 Warnings 
    Refer to the User’s Reference Manual for additional warnings to be observed while monitoring a patient. 
    A WARNING indicates a situation in which the user or the patient may be in danger of injury or 
    death. 
    Installation 
    The monitor or its components should not be used adjacent to or stacked with other equipment. If 
    adjacent or stacked use is necessary, the monitor and its components should be observed to verify 
    normal operation in the configuration in which it will be used. 
    Pins of connectors identified with the ESD warning symbol should not be touched. Connections should 
    not be made to these connectors unless ESD precautionary procedures are used. For details, see 
    section “1.5.3 ESD precautionary procedures”. 
    After transferring or reinstalling the monitor, always check that it is properly connected and all parts 
    are securely attached. Pay special attention to this in case of stacked mounting. 
    Do not use the monitor in high electromagnetic fields (for example, during MRI). 
    A printer or computer must be supplied from an additional transformer providing at least basic 
    isolation (isolating or separating transformer). 
    If you accidentally drop the monitor, have it checked by authorized service personnel prior to clinical 
    use.  
    To avoid explosion hazard, do not use the monitor in presence of flammable anesthetics. The monitor 
    measures only non-flammable anesthetics. 
    Do not touch the patient, table, instruments, modules or the monitor during defibrillation. 
    Power connection 
    Before connecting the power cord to the mains outlet, check that the local voltage and frequency 
    rating corresponds with the rating stated on the device plate on the rear panel of the monitor. 
    Use only hospital-grade grounded power outlets and power cord. Do not remove the grounding pin 
    from the power plug. 
    Use only an intact power cord. Replace the cord if it is cracked, frayed, broken, or otherwise damaged. 
    Do not apply tension to the power cord otherwise the cord may get damaged. 
    Do not use an additional multiple socket outlet, extension cord or adapters of any kind. 
    Before starting to use the system, ensure that the whole combination complies with the international 
    standard IEC 60601-1-1 and with the requirements of the local authorities. Do not connect any 
    external devices to the system other than those specified. 
    To avoid the risk of electric shock, this equipment must only be connected to a supply mains with 
    protective earth. 
    External connection 
    Do not connect any external devices to the monitor other than those specified. 
    1-8  
    						
    							
    Overview 
    Electrical shock hazard 
    When you connect equipment to the Cardiocap/5 input and output connectors, you are configuring a 
    medical system and are responsible for ensuring that the system complies with IEC/EN 60601-1-1 
    and with local requirements. 
    Do not touch any exposed wire or conductive surface while covers are off and the monitor is energized. 
    The voltages present can cause injury or death. 
    Always perform an electrical safety check and leakage current test of the monitor after service. 
    Fuse replacement 
    Replace a fuse only with one of the same type and rating. 
    Explosion hazard 
    To avoid explosion hazard do not use the monitor in the presence of flammable anesthetics. 
    Patient safety 
    Do not perform any testing or maintenance on the monitor while it is being used on a patient. 
    Never install the monitor so that it is above the patient.  
    The monitor must not be used without manufacturer approved mounting attached. 
    Operation of the monitor outside the specified values may cause inaccurate results. 
    To prevent erroneous readings, do not use physically damaged sensors or sensor cables. Discard a 
    damaged sensor or sensor cable immediately. Never repair a damaged sensor or cable; never use a 
    sensor or cable repaired by others. A damaged sensor or a sensor soaked in liquid may cause burns 
    during electrosurgery. 
    PATIENTS WITH PACEMAKERS OR ARRHYTHMIAS: Monitor may count the pacemaker pulses as heart 
    beats during cardiac arrest, some arrhythmias, and with certain types of pacemakers particularly in 
    ON R mode. Do not rely entirely upon rate meter alarms. Keep patients with pacemakers and 
    arrhythmias under close surveillance. 
    PACEMAKER PATIENTS: The impedance respiration measurement may cause rate changes in Minute 
    Ventilation Rate Responsive Pacemakers. In this case, set the pacemaker rate responsive mode off or 
    turn the monitor impedance respiration measurement off. 
    Temperature probes 
    To prevent injury, use Datex-Ohmeda temperature probes only. 
    Cleaning and service 
    Only trained personnel with proper tools and test equipment should perform the tests and repairs 
    described in this manual. Unauthorized service may void the monitor warranty. 
    Always unplug the monitor before cleaning or service. After cleaning or service ensure that every part 
    of the monitor is dry before reconnecting it to the power supply. 
    Do not touch any exposed wire or conductive surface while any cover is removed and the monitor is 
    energized. The voltages present can cause injury or death. 
    Pins of connectors identified with the ESD warning symbol should not be touched. Connections should 
    not be made to these connectors unless ESD precautionary procedures are used. For details, see 
    section “1.5.3 ESD precautionary procedures”. 
    1-9  
    						
    							
    Cardiocap/5 Technical Reference Manual 
    Always perform an electrical safety check and a leakage current test on the monitor after service. 
    Handle the water trap and its contents as you would any body fluid. Infectous hazard may be present. 
    Accessories 
    Use only accessories, mounts and defibrillator-proof cables and invasive pressure transducers 
    approved by GE Healthcare. For a list of approved supplies and accessories, see the Supplies and 
    Accessories catalog. Other cables, transducers and accessories may cause a safety hazard, damage 
    the equipment or system, result in increased emissions or decreased immunity of the equipment or 
    system or interfere with the measurement. Protection against cardiac defibrillator discharge is due in 
    part to the accessories for pulse oximetry (SpO2), temperature (T) and invasive pressure (P) 
    measurement. 
    Single-use accessories are not designed to be reused. Reuse may cause a risk of contamination 
    and/or affect the measurement accuracy. 
    1.5.2 Cautions 
    Refer to the User’s Reference Manual for additional cautions to be observed while monitoring a patient. 
    A CAUTION indicates a condition that may lead to equipment damage or malfunction. 
    General 
    US Federal law restricts this device to sale by or on the order of a licensed medical practitioner. 
    Do not apply pressurized air to any outlet or tubing connected to the monitor. Pressure may destroy 
    sensitive elements. 
    Turn off the power before making any rear panel connections. 
    Use only cables and accessories approved by GE Healthcare. Other cables and accessories may 
    damage the system or interfere with measurement.  
    Vibrations during transport may disturb SpO2, ECG, impedance respiration, and NIBP measurements. 
    Installation 
    Leave space for air circulation to prevent the monitor from overheating. 
    Before use 
    Allow two minutes for warm-up and note any error messages or deviations from normal operation. 
    Airway gas measurement 
    Strong scavenging suction may change the operating pressure of the monitor and cause inaccurate 
    readings or internal damage. 
    Autoclaving and sterilizing 
    Do not steam autoclave or gas sterilize the monitor. 
    Cleaning and service 
    Do not use hypochlorite, ammonia-based, phenol-based, or acetone-based cleaners. These cleaners 
    may damage the surface of the monitor. 
    Do not immerse any part of the monitor in liquid or allow liquid to enter the interior. 
    1-10  
    						
    							
    Overview 
    Do not apply pressurized air to any outlet or tubing connected to the monitor. 
    Clean the fan dust filter on the rear panel once a month or whenever necessary. 
    Electrostatic discharge through the PC boards may damage the components. Before handling printed 
    circuit boards, wear a static control wrist strap. Handle all boards by their nonconductive edges and 
    use antistatic containers when transporting them. 
    Do not break or bypass the patient isolation barrier when testing PC boards. 
    If liquid has accidentally entered the system or its parts, disconnect the power cord from the power 
    supply and have the equipment serviced by authorized service personnel. 
    Batteries 
     
    • A lithium battery on the CPU Board. Dispose of the faulty IC containing the battery 
    according to local regulations. 
    The battery package of the power supply unit in this device contains lead, which is hazardous to the 
    environment. Dispose of the battery according to local regulations. 
    To replace the batteries safely, please refer to the instructions in this manual. 
    • Do not short-circuit the battery terminals. Short-circuiting the battery may produce a very high 
    current, which damages the battery and may cause injury to personnel. 
    • Do not dispose of the battery into open flame, nor put the battery near fire, as it may explode. 
    • Do not disassemble the battery. It contains electrolyte, which may damage clothing or cause 
    injury to skin or eyes. If exposed to electrolyte, wash the injured area with plenty of water and 
    contact a doctor. 
    See Symbols on equipment earlier in this chapter. 
    Special components and modifications 
    Special components used in this monitor are vital to assure reliability and safety. GE Healthcare 
    assumes no responsibility for damage if replacement components not approved by GE Healthcare are 
    used. 
    The manufacturer accepts no responsibility for modifications made to the monitor outside the factory. 
    Storage and transport 
    Do not store or transport the monitor outside the specified temperature, pressure and humidity 
    ranges: 
    Temperature -10 to +50 °C (14 to 122°F) 
    Ambient pressure 660 to 1060 hPa (500 to 800 mmHg) 
     660 to 1060 mbar 
    Relative humidity 0 to 85 % non-condensing 
    1.5.3 ESD precautionary procedures 
    • To avoid electrostatic charges building up, it is recommended to store, maintain and use the 
    equipment at a relative humidity of 30% or greater. Floors should be covered by ESD 
    dissipative carpets or similar. Non-synthetic clothing should be used when working with the 
    component. 
    • To prevent applying a possible electrostatic discharge to the ESD sensitive parts of the 
    equipment, one should touch the metallic frame of the component or a large metal object 
    1-11  
    						
    							
    Cardiocap/5 Technical Reference Manual 
    located close to the equipment. When working with the equipment and specifically when the 
    ESD sensitive parts of the equipment may be touched, a grounded wrist strap intended for 
    use with ESD sensitive equipment should be worn. Refer to the documentation provided with 
    the wrist straps for details of proper use. 
    ESD precautionary procedure training 
    • It is recommended that all potential users receive an explanation of the ESD warning symbol 
    and training in ESD precautionary procedures. 
    • The minimum contents of an ESD precautionary procedure training should include an 
    introduction to the physics of electrostatic charge, the voltage levels that can occur in normal 
    practice and the damage that can be done to electronic components if they are touched by 
    an operator who is electrostatically charged. Further, an explanation should be given of 
    methods to prevent build-up of electrostatic charge and how and why to discharge one’s 
    body to earth or to the frame of the equipment or bond oneself by means of a wrist strap to 
    the equipment or the earth prior to making a connection. 
    1.5.4 Disposal 
    Dispose of the whole device, parts of it, its packing material and this manual in accordance with local 
    environmental and waste disposal regulations. 
    1.5.5 Points to note 
    Medical electrical equipment needs special precautions regarding electromagnetic compatibility and 
    needs to be installed and put into service by qualified personnel according to the electromagnetic 
    compatibility information provided in Chapter 2. 
    Portable and mobile RF communications equipment can affect the medical electrical equipment.  
    Service and reparations are allowed for authorized service personnel only. 
    1.6 Specifications 
    All product specifications are subject to change without prior notice. 
    1.6.1 F-MX and F-MXG frames 
    Power supply 
    Rated voltages and frequencies:  100 to 240 VAC 60/50 Hz 
    Allowed voltage fluctuations:  ± 10%   
    Maximum power consumption:  80 VA 
    Fuses (2):  T2AH/250V 
    Environmental conditions 
    Operating temperature:  +10 to +40 °C  (50 to 104 °F) 
    Storage and transport temp:  –10 to +50 °C  (14 to 122 °F) 
    Relative humidity:  10 to 85 % noncondensing, in airway 0 to 100 % condensing 
    Atmospheric pressure:  660 to 1060 hPa (500 to 800 mmHg) 
    1-12  
    						
    All GE manuals Comments (0)