GE Cardiocap 5 Service Manual
Have a look at the manual GE Cardiocap 5 Service Manual online for free. It’s possible to download the document as PDF or print. UserManuals.tech offer 45 GE manuals and user’s guides for free. Share the user manual or guide on Facebook, Twitter or Google+.
Installation and Functional Check ECG board test 16. Enter the ESTP: ECG service menu: To return to the Service View menu, select Previous Menu as needed until you can select Modules and then select ESTP : ECG. or Press the ComWheel and select: Monitor Setup Install/Service (Password 16-4-34) - Service View (Password 26-23-8) Modules ESTP: ECG Check the Service Data: • The Timeouts, Bad checksums, and Bad c-s by mod values are not increasing faster than 50 per second. • The ECG/RESP board memories passed the internal memory test (RAM, ROM, and EEPROM all show OK). 17. Check that the Power Freq value (the module mains power frequency) has been set according to the supply frequency. If necessary, change the setting by selecting Power Freq. 18. Check that Resp Available and Resp measurement both show ON. NOTE: The Resp measurement shows OFF if Resp is not selected to the screen setup. To set up the screen, press the ComWheel and select: Monitor Setup Screen Setup Waveform Fields or Digit Fields STP board test 19. To enter the ESTP: STP service menu, press the ComWheel and select: Monitor Setup Install/Service (Password 16-4-34) - Service View (Password 26-23-8) Modules ESTP: STP Check that the Timeouts, Bad checksums, and Bad c-s by mod values are not increasing faster than 50 per second. Check that the STP board memories passed the internal memory test (RAM, ROM, and EEPROM all show OK). 2-19
Cardiocap/5 Technical Reference Manual 20. Check that the protection for temperature calibration is ON: Protect key in the menu should state OFF Protect mode should state ON. If necessary, change the Protection mode in the Calibrations menu. NIBP board test 21. Press Previous Menu as needed to return to the Modules menu, then select NIBP to enter the NIBP service menu. Check that the Timeouts, Bad checksums, and Bad c-s by mod values are not increasing faster than 50 per second. Check that the NIBP board memories passed the internal memory test (RAM, ROM, and EEPROM all show OK). Pulse oximetry test (if N-XOSAT or N-XNSAT option is included) 22. Configure the monitor screen to display the Pleth waveform and other pulse oximetry parameters by pressing the ComWheel and selecting: Monitor Setup Screen Setup Waveform Fields Select a field (Field1, for example) and choose PLETH as the parameter. Connect the appropriate SpO2 finger sensor to the monitor (use a Nellcor sensor for the N-XNSAT option; a Datex-Ohmeda sensor for the N-XOSAT option). Attach the sensor to your finger. Check that an acceptable SpO2 reading and a proper SpO2 waveform appear on the screen. 23. Enter the SpO2 options service menu by pressing the ComWheel and selecting: Monitor Setup Install/Service (Password 16-4-34) - Service View (Password 26-23-8) Modules M-NSAT Check that the Timeouts, Bad checksums, and Bad c-s by mod values are equal to 0 (zero). Check that ROM shows OK, indicating it passed the internal memory test. 24. Check that the three error indicators (MP-203 Error, QUART Error, and I/O Error) show NO. NOTE: MP-203 Error represents the NSAT or OSAT pulse oximetry board. 2-20
Installation and Functional Check Gas measurement and spirometry test 25. Connect a sample line to the D-fend. If the monitor is configured with Patient Spirometry (N-XV option), press the ComWheel to enter the Main Menu and select: Monitor Setup Screen 1 Setup Waveform Field Field 2 (choose Paw) Field 3 (choose Flow) NOTE: If you just turned on the monitor, wait until the Calibrating gas sensor message disappears from the screen. Then, enter the Gas Unit – General service menu by pressing the ComWheel and selecting: Monitor Setup Install/Service (Password 16-4-34) Service View (Password 26-23-8 Modules Gas Unit General Check that the configuration shown corresponds with the configuration of your monitor. 26. Check that the Timeouts, Bad checksums, and Bad c-s by mod values are not increasing faster than 50 per second. A value increasing faster than this indicates a failure in Module Bus communication. Refer to the Troubleshooting chapter. 27. Select Previous Menu to return to the Gas Unit menu, then select Gases. Check that the displayed Ambient value corresponds with the current ambient pressure (± 20 mmHg). 28. Check that the Amb-Work value in the service menu is within the range of 40-75 mmHg. 2-21
Cardiocap/5 Technical Reference Manual NeuroMuscular Transmission (NMT) test (if N-XNMT option is included) 29. Cycle power to the monitor and wait until the normal monitoring screen appears. Then, configure the monitor for NMT measurement: • Press the ComWheel and select Monitor Setup. Select Screen Setup. Select Digit Fields, select a field, and choose NMT. • To preset the NMT measurement settings, press NMT. Select Stimulus Mode and choose TOF. Select Cycle Time and choose 10 sec. • Select NMT Setup. Select Current and choose S(70 mA). Select Pulse width and choose 200 µS. Select Stim. Beep Volume and choose 2. Check that the NMT header with related information is displayed in the chosen digit field. 30. To enter the NMT service menu, press the ComWheel and select Monitor Setup. Then select: Install/Service (password 16–4–34) Service (password 26-23-8) Modules NMT Check that the Timeouts, Bad checksums, and Bad c-s by mod values are not increasing faster than by 50 per second. Check that the NMT memory passed the internal memory test (RAM, ROM and EEPROM all state OK. 31. Check that: • In the digit fields, the NMT header, NMT-related information, and “Cable off” are shown. • In the service menu, Cable states OFF. Connect an ElectroSensor to the NMT sensor cable and plug the cable into the NMT connector on the front panel. Check that the digit field message changes to “Measurement OFF” and Cable in the service menu states EMG and ELECTR. OFF. 32. Perform the stimulus current test: Connect a 3 kΩ resistor between the ElectroSensor stimulus electrode leads (brown and white). In the NMT service menu, select Start Curr. test. Check that the test was successful with all three test currents, that is, Current test (mA): states 30 OK, 50 OK, and 70 OK. Connect the NMT ElectroSensor leads to the NMT simulator. Set the switch on the simulator to “Fade off” and turn the knob to “max.” In the service menu, check that Cable states EMG. 2-22
Installation and Functional Check 33. Start NMT measurement (TOF) by selecting NMT Setup and selecting Start-up. Press Previous Menu to return to the NMT service menu. In the digit field, when the “Supramax search” message changes to “Setting reference,” check that the detected supramaximal current is less than 70 mA (the Current set value on the service screen is less than 700). Check that four successive stimulus pulses are generated in 10-second intervals. During each stimulus pulse, a tone sounds and a small asterisk () appears in the digit field. In the digit field, check that TOF% is within 95-105, Count is 4, and T1% is within 95-105. In the NMT service menu, check that T1%, T2%, T3%, T4% and Ratio% are all within 950-1059. Check that the Noise value stays under 50. 34. In the NMT service menu, select NMT Setup and change the stimulus Pulse Width to 100 µs. Then press Previous Menu to return to the NMT service menu. In the digit field, check that TOF% is still within 95-105, Count is 4, and T1% is within 95-105. Change the stimulus pulse width to 300 µs and check the same parameters. 35. Turn the knob on the NMT simulator to 0 (zero). In the NMT service menu, check that the values for T1%, T2%, T3%, T4% turn to 0 and the Ratio% shows – – –. In the digit field, TOF%= should also state – – –; Count and T1% should show 0. 36. Turn the NMT simulator knob back to “max.” Change the stimulus mode to Double Burst Stimulation (select NMT Setup – Stimulus Mode – DBS). Check that only two stimulus pulses are given in a 10-second interval. In the NMT service menu, check that T1%, T2%, and Ratio% are still within 950-1059. In the digit field, check that DBS% is within 95-105, Count is 2, and T1% is within 95-105. 37. Change the stimulus mode to Single Twitch stimulation (ST). Check that the monitor starts giving only one stimulus pulse with a 1-second interval. Note the time when the ST stimulation started. In the service menu, check that the T1% value is within 950-1059. In the digit field, the Count value should be 1 and T1% within 95-105. 38. Continue single twitch stimulation. Five minutes after the start of ST stimulation, check that the NMT measurement stops and the “Measurement OFF” message appears in the NMT digit field. 39. Replace the NMT ElectroSensor with the NMT MechanoSensor. In the NMT service menu, check that Cable states PIEZO. 2-23
Cardiocap/5 Technical Reference Manual 2.6.2 Performance checks 1. Configure the monitor screen to display all parameter information. Press the ComWheel to enter the Main Menu and select: Monitor Setup – Screen Setup – Waveform Fields/Digit Fields An example is shown below: Waveform Fields Digit Fields Field 1 – ECG1 Field 1 – T1 (or Temp) Field 2 – Paw Field 2 – NIBP Field 3 – P1 Field 3 – Resp Field 4 – P2 Field 4 – Gases Field 5 – Pleth Field 6 – CO2 Select Split Screen – None To return to the Normal Screen, press the Normal Screen key. Connect the patient simulator and check that all parameter information is displayed as configured. NOTE: InvBP waveforms are not shown without a patient simulator. Preset the measurement settings for parameters as follows, for example: ECG - ECG Setup - HR Source - AUTO Invasive Pressures P1 ‘ART’ Setup – Label – ART P2 ‘CVP’ Setup – Label – CVP Press the ComWheel to enter the Parameters menu and select: Pulse Oximetry – Pleth Scale – AUTO Airway Gas – Spirometry Setup – Scaling – Indep. Paw Scale – 20 Flow Scale – 15 Resp Setup – Size – 1.0 Resp Rate Source – AUTO Detection Limit – AUTO NOTE: The RESP waveform or Digit Field on the screen must be selected before you can turn on the respiration measurement. 2. Connect the patient simulator to the ECG and invBP connectors, connect an SpO2 finger sensor to the SpO2 connector, and connect the temperature test plug to a Temp connector. Attach the SpO2 sensor to your finger. Check that: • The simulated waveforms are good. • The SpO2 value is in the expected range • The Pleth waveform is normal. 2-24
Installation and Functional Check 3. Set the invBP simulator to +100 mmHg (+13.3 kPa) static pressure. Push the ZERO button and check that: • The invBP waveforms set on the baseline. • The digital values go to zero. Set the simulator to 0 mmHg (0 kPa) and press ZERO. After the zeroing is completed, turn the simulator to dynamic mode and check that the waveform is normal. 4. Check that the “No probe” message is displayed when an SpO2 sensor is not connected. Connect an SpO2 finger sensor to the monitor (but not to a finger). Check that “Probe off” or “Check probe” is displayed. The Pulse search message may be displayed first. 5. Attach the SpO2 sensor to your finger and check that the pleth waveform is displayed and the SpO2 value is in the expected range (95-99%). Check that the HR value is calculated from SpO2 when the ECG and InvBP (P1/P2) cables are not connected. 6. Remove the SpO2 sensor from your finger and check that “Probe off” or “Check probe” is displayed. Disconnect the sensor from the monitor and check that “No probe” is displayed. 7. Attach an adult NIBP cuff onto your arm and perform one NIBP measurement. Check that the cuff is identified (Adult appears in the NIBP digit field for a short time). Check that the measurement result is reasonable. 8. Block the tip of the sampling line with your finger and check that the “Sample line blocked” message is displayed on the monitor screen within 30 seconds. Detach the D-fend and check that the ”Check D-fend” message is displayed on the monitor screen within 30 seconds. 9. Reattach the D-fend. Breathe shortly into the sampling line. Check that the CO2 waveform moves up on the screen. After completing the check, be sure to fill in all necessary documents. 2-25
Functional Check Form FUNCTIONAL CHECK FORM Datex-Ohmeda Cardiocap/5 Customer Service Service Engineer Date Monitor Configuration Monitor model: Hemodynamic model F-MX Hemodynamic model with gas measurement F-MXG Measurement options Measurement options (SpO2) N-XP Two invasive pressure channels and second temperature (T2) N-XOSAT Datex-Ohmeda enhanced pulse oximetry N-XC CO2 N-XNSAT Nellcor compatible pulse oximetry N-XCO CO2, N2O, Patient Oxygen Data collection and data management options N-XCAiO CO2, anesthetic agents, agent identification, N2O, Patient Oxygen N-XREC Recorder N-XV Patient Spirometry N-XNET Network N-XNMT NeuroMuscular Transmission (NMT) N-XDNET Data card and Network Functional Inspection OK = Test OK N.A. = Test not applicable Fail = Test Failed General OK N.A. Fail 1. Battery charge status LED 2. Start-up 3. Loudspeaker 4. Real-time clock 5. Voltage checks Acceptance limits TEMP 10 to 60 °C BATVOLT with mains power ON 10 to 14.5 V BATVOLT with mains power OFF 10 to 13.5 V 6. Back-up battery Appendix A 1
Functional Check Form Recorder test OK N.A. Fail 7. Recorder message 8. Recording Memory card (PCMCIA) test 9. Memory 10. Communication 11. Memory card recognition Network test 12. Connection to network 13. Communication 14. Data error counters 15. Hardware error counters ECG board test 16. Communication and memories 17. Power frequency 18. Resp measurement recognition STP board test 19. Communication and memory 20. Protection for temperature calibration NIBP board test 21. Communication and memory Pulse oximetry options test 22. Recognition and test measurement 23. Communication and memory 24. Error status Gas measurement and spirometry test 25. Configuration 26. Communication 27. Ambient pressure Current ambient pressure ± 20mmHg 28. Amb-Work 40 to 75 mmHg Appendix A 2