Home > GE > Health/Medical Alert > GE Cardiocap 5 Service Manual

GE Cardiocap 5 Service Manual

    Download as PDF Print this page Share this page

    Have a look at the manual GE Cardiocap 5 Service Manual online for free. It’s possible to download the document as PDF or print. UserManuals.tech offer 45 GE manuals and user’s guides for free. Share the user manual or guide on Facebook, Twitter or Google+.

    							
    Installation and Functional Check 
    ECG board test 
    16. Enter the ESTP: ECG service menu: 
    To return to the Service View menu, select Previous Menu as needed until you can select 
    Modules and then select ESTP : ECG. 
    or 
    Press the ComWheel and select: 
    Monitor Setup 
    Install/Service (Password 16-4-34) -  
    Service View (Password 26-23-8) 
    Modules 
    ESTP: ECG 
    Check the Service Data: 
    • The Timeouts, Bad checksums, and Bad c-s by mod values are not increasing faster than 50 
    per second. 
    • The ECG/RESP board memories passed the internal memory test (RAM, ROM, and EEPROM 
    all show OK). 
     
    17. Check that the Power Freq value (the module mains power frequency) has been set according to 
    the supply frequency. If necessary, change the setting by selecting Power Freq. 
     
    18. Check that Resp Available and Resp measurement both show ON. 
    NOTE: The Resp measurement shows OFF if Resp is not selected to the screen setup.  
    To set up the screen, press the ComWheel and select:  
    Monitor Setup 
    Screen Setup 
    Waveform Fields or Digit Fields 
     
    STP board test 
    19. To enter the ESTP: STP service menu, press the ComWheel and select: 
    Monitor Setup 
    Install/Service (Password 16-4-34) -  
    Service View (Password 26-23-8) 
    Modules 
    ESTP: STP 
    Check that the Timeouts, Bad checksums, and Bad c-s by mod values are not increasing faster 
    than 50 per second. 
    Check that the STP board memories passed the internal memory test (RAM, ROM, and EEPROM 
    all show OK). 
     
    2-19  
    						
    							
    Cardiocap/5 Technical Reference Manual 
    20. Check that the protection for temperature calibration is ON: 
    Protect key in the menu should state OFF 
    Protect mode should state ON. 
    If necessary, change the Protection mode in the Calibrations menu. 
     
    NIBP board test 
    21. Press Previous Menu as needed to return to the Modules menu, then select NIBP to enter the 
    NIBP service menu. 
    Check that the Timeouts, Bad checksums, and Bad c-s by mod values are not increasing faster 
    than 50 per second. 
    Check that the NIBP board memories passed the internal memory test (RAM, ROM, and EEPROM 
    all show OK). 
     
    Pulse oximetry test (if N-XOSAT or N-XNSAT option is included) 
    22. Configure the monitor screen to display the Pleth waveform and other pulse oximetry parameters 
    by pressing the ComWheel and selecting: 
    Monitor Setup 
    Screen Setup 
    Waveform Fields 
    Select a field (Field1, for example) and choose PLETH as the parameter. 
    Connect the appropriate SpO2 finger sensor to the monitor (use a Nellcor sensor for the N-XNSAT 
    option; a Datex-Ohmeda sensor for the N-XOSAT option). Attach the sensor to your finger.  
    Check that an acceptable SpO2 reading and a proper SpO2 waveform appear on the screen. 
     
    23. Enter the SpO2 options service menu by pressing the ComWheel and selecting:  
    Monitor Setup 
    Install/Service (Password 16-4-34) -  
    Service View (Password 26-23-8) 
    Modules 
    M-NSAT 
    Check that the Timeouts, Bad checksums, and Bad c-s by mod values are equal to 0 (zero). 
    Check that ROM shows OK, indicating it passed the internal memory test. 
     
    24. Check that the three error indicators (MP-203 Error, QUART Error, and I/O Error) show NO. 
    NOTE: MP-203 Error represents the NSAT or OSAT pulse oximetry board. 
     
    2-20  
    						
    							
    Installation and Functional Check 
    Gas measurement and spirometry test 
    25. Connect a sample line to the D-fend.  
    If the monitor is configured with Patient Spirometry (N-XV option), press the ComWheel to enter 
    the Main Menu and select: 
    Monitor Setup 
    Screen 1 Setup 
    Waveform Field 
    Field 2 (choose Paw) 
    Field 3 (choose Flow) 
    NOTE: If you just turned on the monitor, wait until the Calibrating gas sensor message 
    disappears from the screen. Then, enter the Gas Unit – General service menu by pressing the 
    ComWheel and selecting: 
    Monitor Setup 
    Install/Service (Password 16-4-34) 
    Service View (Password 26-23-8 
    Modules 
    Gas Unit 
    General 
    Check that the configuration shown corresponds with the configuration of your monitor. 
     
    26. Check that the Timeouts, Bad checksums, and Bad c-s by mod values are not increasing faster 
    than 50 per second. A value increasing faster than this indicates a failure in Module Bus 
    communication. Refer to the Troubleshooting chapter. 
     
    27. Select Previous Menu to return to the Gas Unit menu, then select Gases. Check that the 
    displayed Ambient value corresponds with the current ambient pressure (± 20 mmHg). 
     
    28. Check that the Amb-Work value in the service menu is within the range of 40-75 mmHg. 
     
    2-21  
    						
    							
    Cardiocap/5 Technical Reference Manual 
    NeuroMuscular Transmission (NMT) test (if N-XNMT option is included) 
    29. Cycle power to the monitor and wait until the normal monitoring screen appears. Then, configure 
    the monitor for NMT measurement: 
    • Press the ComWheel and select Monitor Setup. 
    Select Screen Setup. 
    Select Digit Fields, select a field, and choose NMT. 
    • To preset the NMT measurement settings, press NMT.  
    Select Stimulus Mode and choose TOF. 
    Select Cycle Time and choose 10 sec. 
    • Select NMT Setup. 
    Select Current and choose S(70 mA). 
    Select Pulse width and choose 200 µS. 
    Select Stim. Beep Volume and choose 2. 
    Check that the NMT header with related information is displayed in the chosen digit field.  
     
    30. To enter the NMT service menu, press the ComWheel and select Monitor Setup. Then select: 
     Install/Service (password 16–4–34) 
    Service (password 26-23-8) 
    Modules 
    NMT 
    Check that the Timeouts, Bad checksums, and Bad c-s by mod values are not increasing faster 
    than by 50 per second. Check that the NMT memory passed the internal memory test (RAM, ROM 
    and EEPROM all state OK. 
     
    31. Check that: 
    • In the digit fields, the NMT header, NMT-related information, and “Cable off” are shown. 
    • In the service menu, Cable states OFF. 
    Connect an ElectroSensor to the NMT sensor cable and plug the cable into the NMT connector on 
    the front panel. Check that the digit field message changes to “Measurement OFF” and Cable in 
    the service menu states EMG and ELECTR. OFF. 
     
    32. Perform the stimulus current test: 
    Connect a 3 kΩ resistor between the ElectroSensor stimulus electrode leads (brown and white). 
    In the NMT service menu, select Start Curr. test. Check that the test was successful with all three 
    test currents, that is, Current test (mA): states 30 OK, 50 OK, and 70 OK. 
    Connect the NMT ElectroSensor leads to the NMT simulator. Set the switch on the simulator to 
    “Fade off” and turn the knob to “max.” In the service menu, check that Cable states EMG. 
     
    2-22  
    						
    							
    Installation and Functional Check 
    33. Start NMT measurement (TOF) by selecting NMT Setup and selecting Start-up. Press Previous 
    Menu to return to the NMT service menu. 
    In the digit field, when the “Supramax search” message changes to “Setting reference,” check 
    that the detected supramaximal current is less than 70 mA (the Current set value on the service 
    screen is less than 700).   
    Check that four successive stimulus pulses are generated in 10-second intervals. During each 
    stimulus pulse, a tone sounds and a small asterisk () appears in the digit field. In the digit field, 
    check that TOF% is within 95-105, Count is 4, and T1% is within 95-105. 
    In the NMT service menu, check that T1%, T2%, T3%, T4% and Ratio% are all within 950-1059.  
    Check that the Noise value stays under 50. 
     
    34. In the NMT service menu, select NMT Setup and change the stimulus Pulse Width to 100 µs. 
    Then press Previous Menu to return to the NMT service menu. 
    In the digit field, check that TOF% is still within 95-105, Count is 4, and T1% is within 95-105. 
    Change the stimulus pulse width to 300 µs and check the same parameters. 
     
    35. Turn the knob on the NMT simulator to 0 (zero). In the NMT service menu, check that the values for 
    T1%, T2%, T3%, T4% turn to 0 and the Ratio% shows – – –. In the digit field, TOF%= should also 
    state – – –; Count and T1% should show 0. 
     
    36. Turn the NMT simulator knob back to “max.” Change the stimulus mode to Double Burst 
    Stimulation (select NMT Setup – Stimulus Mode – DBS).  
    Check that only two stimulus pulses are given in a 10-second interval. In the NMT service menu, 
    check that T1%, T2%, and Ratio% are still within 950-1059. In the digit field, check that DBS% is 
    within 95-105, Count is 2, and T1% is within 95-105. 
     
    37. Change the stimulus mode to Single Twitch stimulation (ST). Check that the monitor starts giving 
    only one stimulus pulse with a 1-second interval. Note the time when the ST stimulation started. 
    In the service menu, check that the T1% value is within 950-1059. In the digit field, the Count 
    value should be 1 and T1% within 95-105. 
     
    38. Continue single twitch stimulation. Five minutes after the start of ST stimulation, check that the 
    NMT measurement stops and the “Measurement OFF” message appears in the NMT digit field. 
     
    39. Replace the NMT ElectroSensor with the NMT MechanoSensor. In the NMT service menu, check 
    that Cable states PIEZO. 
     
    2-23  
    						
    							
    Cardiocap/5 Technical Reference Manual 
    2.6.2 Performance checks 
    1. Configure the monitor screen to display all parameter information. 
    Press the ComWheel to enter the Main Menu and select: 
    Monitor Setup – Screen Setup – Waveform Fields/Digit Fields 
    An example is shown below: 
    Waveform Fields Digit Fields 
    Field 1 – ECG1 Field 1 – T1 (or Temp) 
    Field 2 – Paw Field 2 – NIBP 
    Field 3 – P1 Field 3 – Resp 
    Field 4 – P2 Field 4 – Gases 
    Field 5 – Pleth  
    Field 6 – CO2  
    Select Split Screen – None 
    To return to the Normal Screen, press the Normal Screen key. 
    Connect the patient simulator and check that all parameter information is displayed as configured. 
    NOTE: InvBP waveforms are not shown without a patient simulator. 
    Preset the measurement settings for parameters as follows, for example: 
    ECG - ECG Setup - HR Source - AUTO 
    Invasive Pressures 
    P1 ‘ART’ Setup – Label – ART 
    P2 ‘CVP’ Setup – Label – CVP 
    Press the ComWheel to enter the Parameters menu and select:  
    Pulse Oximetry – Pleth Scale – AUTO 
    Airway Gas – Spirometry Setup – Scaling – Indep. 
     Paw Scale – 20 
     Flow Scale – 15 
    Resp Setup  – Size – 1.0 
     Resp Rate Source – AUTO 
     Detection Limit – AUTO 
    NOTE: The RESP waveform or Digit Field on the screen must be selected before you can turn on 
    the respiration measurement. 
     
    2. Connect the patient simulator to the ECG and invBP connectors, connect an SpO2 finger sensor to 
    the SpO2 connector, and connect the temperature test plug to a Temp connector. Attach the SpO2 
    sensor to your finger. Check that: 
    • The simulated waveforms are good. 
    • The SpO2 value is in the expected range 
    • The Pleth waveform is normal. 
     
    2-24  
    						
    							
    Installation and Functional Check 
    3. Set the invBP simulator to +100 mmHg (+13.3 kPa) static pressure. Push the ZERO button and 
    check that: 
    • The invBP waveforms set on the baseline. 
    • The digital values go to zero. 
    Set the simulator to 0 mmHg (0 kPa) and press ZERO. After the zeroing is completed, turn the 
    simulator to dynamic mode and check that the waveform is normal. 
     
    4. Check that the “No probe” message is displayed when an SpO2 sensor is not connected. 
    Connect an SpO2 finger sensor to the monitor (but not to a finger). Check that “Probe off” or 
    “Check probe” is displayed. The Pulse search message may be displayed first. 
     
    5. Attach the SpO2 sensor to your finger and check that the pleth waveform is displayed and the 
    SpO2 value is in the expected range (95-99%).  
    Check that the HR value is calculated from SpO2 when the ECG and InvBP (P1/P2) cables are not 
    connected. 
     
    6. Remove the SpO2 sensor from your finger and check that “Probe off” or “Check probe” is 
    displayed.  
    Disconnect the sensor from the monitor and check that “No probe” is displayed. 
     
    7. Attach an adult NIBP cuff onto your arm and perform one NIBP measurement. Check that the cuff 
    is identified (Adult appears in the NIBP digit field for a short time). 
    Check that the measurement result is reasonable. 
     
    8. Block the tip of the sampling line with your finger and check that the “Sample line blocked” 
    message is displayed on the monitor screen within 30 seconds. 
    Detach the D-fend and check that the ”Check D-fend” message is displayed on the monitor 
    screen within 30 seconds. 
     
    9. Reattach the D-fend. Breathe shortly into the sampling line. Check that the CO2 waveform moves 
    up on the screen. 
     
    After completing the check, be sure to fill in all necessary documents. 
     
    2-25  
    						
    							
    Functional Check Form 
    FUNCTIONAL CHECK FORM 
    Datex-Ohmeda Cardiocap/5 
    Customer  
    Service  
    Service Engineer  Date  
    Monitor Configuration 
    Monitor model: Hemodynamic model 
    F-MX  Hemodynamic model with gas measurement 
    F-MXG  
     
    Measurement options Measurement options (SpO2) 
    N-XP Two invasive pressure channels and 
    second temperature (T2)  N-XOSAT Datex-Ohmeda enhanced pulse 
    oximetry  
    N-XC CO2  N-XNSAT Nellcor compatible pulse 
    oximetry  
    N-XCO CO2, N2O, Patient Oxygen  Data collection and data management options 
    N-XCAiO CO2, anesthetic agents, agent 
    identification, N2O, Patient Oxygen  N-XREC Recorder  
    N-XV Patient Spirometry  N-XNET Network  
    N-XNMT NeuroMuscular Transmission (NMT)  N-XDNET Data card and Network  
    Functional Inspection 
    OK = Test OK N.A. = Test not applicable Fail = Test Failed 
    General OK N.A. Fail 
    1. Battery charge status LED    
    2. Start-up    
    3. Loudspeaker    
    4. Real-time clock    
    5. Voltage checks    
        Acceptance limits 
    TEMP  10 to 60 °C 
    BATVOLT with mains power ON  10 to 14.5 V 
    BATVOLT with mains power OFF   10 to 13.5 V 
    6. Back-up battery    
    Appendix A 1  
    						
    							
    Functional Check Form 
    Recorder test OK N.A. Fail 
    7. Recorder message    
    8. Recording    
    Memory card (PCMCIA) test    
    9. Memory    
    10. Communication    
    11. Memory card recognition    
    Network test    
    12. Connection to network    
    13. Communication    
    14. Data error counters    
    15. Hardware error counters    
    ECG board test    
    16. Communication and memories    
    17. Power frequency    
    18. Resp measurement recognition    
    STP board test    
    19. Communication and memory    
    20. Protection for temperature calibration    
    NIBP board test    
    21. Communication and memory    
    Pulse oximetry options test    
    22. Recognition and test measurement    
    23. Communication and memory    
    24. Error status    
    Gas measurement and spirometry test    
    25. Configuration    
    26. Communication    
    27. Ambient pressure  Current ambient pressure ± 20mmHg 
    28. Amb-Work  40 to 75 mmHg 
    Appendix A 2    
    						
    All GE manuals Comments (0)