Home > GE > Health/Medical Alert > GE Cardiocap 5 Service Manual

GE Cardiocap 5 Service Manual

    Download as PDF Print this page Share this page

    Have a look at the manual GE Cardiocap 5 Service Manual online for free. It’s possible to download the document as PDF or print. UserManuals.tech offer 45 GE manuals and user’s guides for free. Share the user manual or guide on Facebook, Twitter or Google+.

    							
    Installation and Functional Check 
    Analog outputs 
    Each analog output signal is scaled linearly between –5 and +5 volts. The resolution is 4096 steps 
    over 10 volts, or approximately 0.00244 volts per step. All signal levels are updated every 10 ms. 
    OFF Default state. No signal is present at the analog output pin. 
    HR according to selected source (display value)  
    The original scale of 0 to 300 beats is scaled between 0 and 3 volts. 
    ECG1, ECG2, ECG3  
    The original scale of –5000 microvolts to +5000 microvolts is scaled between –5 and +5 volts. 
    P1 lre, P2 lre (Invasive pressure real-time values, low resolution)  
    The original scale of –20 mmHg to +320 mmHg (-2.7 to +42.7 kPa) is scaled between –0.2 
    and +3.2 volts. 
    P1 hre, P2 hre (Invasive pressure real-time values, high resolution)  
    The original scale of –20 mmHg to +50 mmHg (-2.7 to +6.7 kPa) is scaled between –2 and 
    +5 volts. 
    Pleth The original scale of –100% to 100% is scaled between –5 and +5 volts. 
    SpO2>40, SpO2>60, SpO2>80 (beat-to-beat, display value, 10 s average)  
    The original scale of 40 to 100% (SpO2>40), 60 to 100% (SpO2>60) or 80 to 100% 
    (SpO2>80) is scaled between –5 and +5 volts. 
    CO2 The original scale of 0% to 10% is scaled between 0 and +5 volts. Values greater than 10% 
    are set to 10%. (Airway gas special indications are applied. See Special indications for 
    analog outputs below.) 
    AA (anesthetic agent)  
    The original scale of 0% to 10% is scaled between 0 and +5 volts. Values greater than 10% 
    are set to 10%. (Airway gas special indications are applied. See Special indications for 
    analog outputs below.) 
    O2 The original scale of 0% to 100% is scaled between 0 and +5 volts. (Airway gas special 
    indications are applied. See Special indications for analog outputs below.) 
    N2O  The original scale of 0% to 100% is scaled between 0 and +5 volts. (Airway gas special 
    indications are applied. See Special indications for analog outputs below.) 
    Paw (airway pressure)  
    The original scale of –20 cmH2O to +80 cmH2O is scaled between –5 and +5 volts. (Airway 
    gas sensor failure is applied. See Special indications for analog outputs below.) 
    Flow The original scale of –100 l/minute to +100 l/minute is scaled between –5 and +5 volts. 
    (Airway gas sensor failure is applied. See Special indications for analog outputs below.) 
    Volume The original scale of –2.5 liters to +2.5 liters is scaled between –5 and +5 volts. (Airway gas 
    sensor failure is applied. See Special indications for analog outputs below.) 
    Resp The original scale of –5000 mohms is scaled between –5 and +5 volts. 
    RR The respiration rate display of 0 to 150 breaths per minute is scaled between 0 and +1.5 volts. 
    T1, T2 The original temperature scale of 0 °C to 50 °C is scaled between 0 and +5 volts. 
    TEST SIGNALS –5 V, 0 V, +5 V 
    Steady signals with one of the listed values. 
    TEST 1 Test signal of a triangle shape with a base width of 1 second (0 V minimum; +5 V maximum). 
    2-9  
    						
    							
    Cardiocap/5 Technical Reference Manual 
    TEST 2 Test signal of a triangle shape with a base width of 4 seconds (–5 V minimum; +5 V maximum). 
    Special indications for analog outputs 
    Start-up indication occurs when the monitor is started by a power on or by an internal restart (caused 
    by a fatal failure). The indication consists of three triangle-shaped signals with a base width of 1 
    second, base of 0 volts, and a height of 5 volts. 
    Airway gas calibration:  During calibration of any gases, a square wave is generated. The minimum 
    value is 0 volts, the maximum value is +2 volts, the minimum phase length equals the maximum 
    phase length, and the frequency is 0.25 Hz. 
    Airway gas zeroing:  During zeroing of any gases, a square wave is generated. The minumum value is  
    0 volts, the maximum value is +5 volts, the minimum phase length equals the maximum phase length, 
    and the frequency is 0.25 Hz. 
    Airway gas occlusion:  During occlusion of any gases, a triangle-shaped signal is generated with a 
    base width of 4 seconds, a minimum value of 0 volts, and a maximum value of +5 volts. 
    Airway gas air leak:  During an air leak of any gases, a triangle-shaped signal is generated with a base 
    width of 3 seconds, a minimum value of 0 volts, and a maximum value of +5 volts. 
    Airway gas sensor failure:  During a sensor failure, a triangle-shaped signal is generated with a base 
    width of 2 seconds, a minimum value of 0 volts, and a maximum value of +5 volts. 
    2.4.4 Setting the analog output signals 
    To set the analog output signals: 
    1. Press the ComWheel and select Monitor Setup from the Main Menu. 
    2. Select Install/Service and enter the password (16-4-34). 
    3. Select Installation. 
    4. Select Analog Outputs and set the channels. 
    5. Press the Normal Screen key. 
    2-10  
    						
    							
    Installation and Functional Check 
    2.5 Connector pin assignments 
    2.5.1 Analog/digital output connector (X1) 
    CON./ I/O BOARD SIGNAL DIRECTION LEVEL FUNCTION 
    X1/1 GND – 0V GND 
    X1/2 N/C    
    X1/3 DSYNCOUT O CMOS Defibrillation sync. 
    X1/4 N/C    
    X1/5 NCALLOUT O CMOS Nurse Call 
    X1/6 N/C    
    X1/7 GND – 0V GND 
    X1/8 to X1/10 N/C    
    X1/11 NCALLB1 – – Nurse call relay 
    X1/12 NCALLB2 – – Nurse call relay 
    X1/13 N/C    
    X1/14 A1OUT O  Analog output 
    X1/15 +5VOUT – +5V +5 V output voltage 
    X1/16 N/C    
    X1/17 A3OUT O  Analog output 
    X1/18 N/C    
    X1/19 A2OUT O  Analog output 
    X1/20 A0OUT O  Analog output 
    X1/21 to X1/37 N/C    
    X1/38 ECGOUT O  Analog ECG-signal output 
    X1/39 to X1/43 N/C    
    X1/44 GND – 0V GND 
     
    2-11  
    						
    							
    Cardiocap/5 Technical Reference Manual 
    2.5.2 RS-232 serial communication interface/local printer connector (X2) 
    Connector X2 provides an RS-232 serial communication link with handshaking. 
     
     
    D9 male connector pinout (front view) 
      
    CON./ I/O BOARD SIGNAL DIRECTION LEVEL FUNCTION 
    X2/1 GND – 0V Ground 
    X2/2 RXD I RS232 Serial bus, Receive 
    X2/3 TXD O RS232 Serial bus, Transmit 
    X2/4 +5VS – +5V +5 V output voltage, maximum 200 mA 
    X2/5 GND – 0V Ground 
    X2/6 N/C    
    X2/7 RTS O RS232 Serial bus, Request To Send 
    X2/8 CTS I RS232 Serial bus, Clear To Send  
    X2/9 N/C    
         
    2.5.3 Network identification plug connector (X3) 
    The coding element interface is used for bedside address-coding for the Ethernet. The element is 
    connected to the 9-pin female connector. 
     
     
    D9 female connector pinout (front view) 
      
    CON./ I/O BOARD SIGNAL DIRECTION LEVEL FUNCTION 
    X3/1 IDCS O CMOS ID chip select 
    X3/2 IDCLK O CMOS ID clock 
    X3/3 IDDI I CMOS ID data in 
    X3/4 IDDO O CMOS ID data out 
    X3/5 GND - 0V Ground 
    X3/6 +5VB - +5V +5V output voltage 
    X3/7 N/C    
    X3/8 N/C    
    X3/9 GND - 0V Ground 
    2-12  
    						
    							
    Installation and Functional Check 
    2.5.4 Ethernet connector (X4) 
    The Ethernet interface meets IEEE 802.3 specifications (10BASE-T) with hospital-grade approved 
    power and data transformers. 
     
    Ethernet connector (front view) 
      
    PIN SIGNAL DIRECTION FUNCTION 
    1 Tx + O Transmit data 
    2 Tx – O Transmit data 
    3 Rx + I Receive data 
    4 N/C   
    5 N/C   
    6 Rx – I Receive data 
    7 N/C   
    8 N/C   
        
    2.5.5 Remote Control connector (X5) 
    CON./ I/O BOARD SIGNAL DIRECTION LEVEL FUNCTION 
    X5/1 CLK I/O TTL Keyboard clock 
    X5/2 DATA I/O TTL Keyboard data 
    X5/3 –    
    X5/4 GND – 0V Ground 
    X5/5 +5 VB – +5V +5V supply voltage 
     
    2-13  
    						
    							
    Cardiocap/5 Technical Reference Manual 
    2.6 Functional check 
    The functional check consists of instructions to verify the correct operation of the Cardiocap/5. The 
    functional check is recommended after installing the monitor and after long storage.  
    The first part of the instructions contains procedures that are performed through the service menu. 
    These checks are especially recommended if the monitor has been stored and not used for a long 
    time. 
    When installing a new monitor, performance checks with a simulator and accessories are typically 
    enough to ensure the correct function of the monitor. We recommend that you always perform the 
    complete functional check to confirm that no hardware failures occurred during transport. 
    The instructions are for the maximum functional configuration of the Cardiocap/5. Perform the 
    procedures in ascending order and skip items that do not correspond to your monitor configuration.  
    Using the Functional Check Form  
    A Functional Check Form is included at the end of this chapter. Copy this form and complete the 
    “checklist” as you perform the functional check procedures. 
    NOTE:  means to sign the form after performing the procedure. 
    Recommended tools 
    Tool Order Number 
    SpO2 finger sensor for the pulse oximetry option. 
    NOTE:  With the N-XNSAT option, use only Nellcor sensors (DS-100A, for 
    example). 
    With the N-XOSAT option, use only Datex-Ohmeda sensors (OXY-F1-H, for example). 
    For standard pulse oximetry, use only Datex-Ohmeda sensors (OXY-F4-N, for example). 
     
    Simulator capable of simulating ECG, RESP, and invBP Obtain locally 
    Adult NIBP cuff 572435 
    Adult NIBP hose 877235 
    Temperature test plug set 884515 
    MemCard – Menu, English (only for monitors with Data card option) 893860 
    MemCard – Data, English (only for monitors with Data card option) 887045 
    Sampling line 3.0 m (only for monitors with gas option) 73319 
      
    If the NMT option (N-XNMT) is installed, you will need these additional items: 
      
    Tool Order Number 
    NMT simulator 871251 
    NMT ElectroSensor 888416 
    NMT MechanoSensor 888418 
    NMT Sensor Cable 888414 or 888415 
    3 kΩ 0.25W 1% resistor Obtain locally 
    2-14  
    						
    							
    Installation and Functional Check 
    2.6.1 Functional inspection 
    1. Connect the power cord. 
    Check that the green external power indicator/battery charge status LED turns on or starts 
    flashing. 
     
    2. Switch the monitor on. Check that the monitor starts up as follows: 
    • The start-up sound is heard from the loudspeaker 
    • The normal monitoring screen appears 
    • The alarm LEDs turn on and off after about 20 sec 
    • No error messages appear onto the screen 
    • If the monitor contains a recorder, start-up information is printed. Verify that the time and 
    date on the printout are correct. 
     
    3. Check the loudspeaker by adjusting the alarm volume in the Alarms Setup menu: 
    Press the ComWheel to enter the Main Menu and select: 
    Alarms Setup 
    Alarm Volume 
    Test the whole volume scale from 1 to 10 by turning the ComWheel and check that the alarm 
    volume changes correspondingly. The alarm sound should be clear and audible at all settings. 
    Select Main Menu. 
     
    4. Check that the time and date are correct and adjust, if necessary: 
    Press the ComWheel to enter the Main Menu and select: 
    Monitor Setup 
    Time and Date 
    NOTE: To prevent the loss of trend data, you cannot change the time and date after starting a new 
    case or admitting a new patient. 
    NOTE: If the clock shows a time of 0:00 continuously after start-up, the SRAM/Timekeeper battery 
    on the CPU board needs to be replaced.  
    To return to the Main Menu, select Previous Menu. 
     
    2-15  
    						
    							
    Cardiocap/5 Technical Reference Manual 
    5. Check that the supply voltages are within the given limit values: 
    Press the ComWheel to enter the Main Menu and select: 
    Monitor Setup 
    Install/Service (Password 16-4-34) 
    Service View (Password 26-23-8) 
    Monitor  
    Voltages 
    Voltages Mains power ON Mains power OFF 
    VDD/BAT 14.3 to 16.2 10 to 13 
    VIN 15VB 10 to 14.5 10 to 13.5 
    VDD 15.0 to 16.5  
    +12V 11.4 to 12.6  
    +15VD 14.25 to 15.85  
    +15V 14.4 to 15.6  
    -15V 15.6 to 14.4  
    +2.5VREF 2.40 to 2.54  
    TEMP 10 to 60  
    BATVOLT 10 to 14.5 10 to 13.5 
       
    VDD/TEMP OK 0 1 
    CHG ON 1 0 
       
    D Sync test count 0  
     
    6. Check the back-up battery: 
    Disconnect the power cord (without switching the monitor to standby). 
    Check that the monitor continues to run normally on battery power. The battery indicator should 
    appear in the upper right corner of the screen: 
     
    Reconnect the power cord and check that during charging, the charging symbol is displayed and 
    the battery charge status LED starts flashing: 
     
     
    2-16  
    						
    							
    Installation and Functional Check 
    Recorder test (if N-XREC option is included) 
    7. Open the paper compartment cover and check that the “Recorder: Cover open” message is 
    displayed. Close the cover. 
     
    8. To record waveforms, press the ComWheel to enter the Main Menu and select: 
    Record/Print 
    Record Waveforms 
    Record Wave 
    Check that the recording quality is acceptable.  
    To stop recording, select Stop Wave. 
    Press the Record Waveform/Stop key and verify that the selected waveforms print.  
    Press the key again to stop recording. 
    Press the Record Trend/Stop key and verify that the selected trends print.  
    Press the key again to stop recording. 
     
    Memory card (PCMCIA) test (if N-XDNET option is included) 
    9. Press the ComWheel to enter the Main Menu and select:  
    Monitor Setup 
    Install/Service (Password 16-4-34) 
    Service View (Password 26-23-8) 
    Modules 
    Memory Module 
    Check that the module is recognized (YES) and that the memory on the memory board and the 
    PCMCIA controller passed the tests (OK): 
    Module present YES  ROM  OK  PCMCIA  OK 
    Module active YES  RAM OK  EEPROM OK 
     
    10. Select Communication and check that Interface status states Active continuously and the error 
    counter values on the bottom part of the menu are stable. 
     
    11. Select Module Status and insert a memory card labeled ”Menu” in the memory card slot.  
    Check that within one minute: 
    • The “Menu card inserted” message appears in the message field. 
    • The white Menu card symbol appears on the upper right corner of the screen if the battery is 
    not charging (the battery charging symbol is not present).  
    Wait until the information regarding the memory card slot is fully updated in the service menu, 
    then check that the screen shows the Card type is MENU and the File system is ATA. Check that 
    the rest of the information for the memory card slot is reliable and no errors have been detected. 
    2-17  
    						
    							
    Cardiocap/5 Technical Reference Manual 
    Repeat the test using a Data card in the other memory card slot. Verify that the green Data card 
    symbol is displayed if the battery charging symbol is not present, the “Data card inserted” 
    message appears, and the Card type is DATA.). 
     
    Network test (if N-XNET or N-XDNET option is included) 
    12. Connect the Monitor-Network cable and the Identification plug to the monitor. To check the 
    connection to the network, check the Network connection LEDs between the connectors: 
    • Yellow LED should flash intermittently. 
    • Green LED should be lit continuously. 
     
    13. Check that the Datex-Ohmeda Network symbol  is displayed on the upper right 
    corner of the screen.  
    NOTE: If the battery is being charged, the battery charging symbol is displayed instead of the 
    network symbol. 
    A message regarding the connection to the Datex-Ohmeda Information Center should appear in 
    the message field on the screen. 
     
    14. To enter the Communication service menu and check Network information, press the ComWheel 
    to enter the Main Menu and select: 
    Monitor Setup 
    Install/Service (Password 16-4-34) 
    Service View (Password 26-23-8) 
    Monitor 
    Communication 
    Network 
    Check that: 
    • The Location ID number matches the ID plug connected to X3. 
    • The packets and bytes IN is increasing slowly. 
    • The packets and bytes OUT is increasing quickly. 
    • Connections shows the names of the connected networks. 
    • The data error counters (CRC, Frame, and Transm.) are stable. 
    NOTE: The counters may show values greater than 0. However, any values increasing continuously 
    indicate a problem. 
     
    15. Check that the hardware error counters (Intern., Missed, FIFO, and Overrun) all show 0 (zero). A 
    value greater than zero for any counter indicates a problem on the CPU board. 
     
    2-18  
    						
    All GE manuals Comments (0)