GE Cardiocap 5 Service Manual
Have a look at the manual GE Cardiocap 5 Service Manual online for free. It’s possible to download the document as PDF or print. UserManuals.tech offer 45 GE manuals and user’s guides for free. Share the user manual or guide on Facebook, Twitter or Google+.
Measurement Parameters Before each stimulation, the sequence offset, noise, and threshold for response detection is measured. Offset is a baseline of the noise measurement. Noise is calculated by the same algorithm as the response signal itself. The response detection threshold is calculated based on the noise. If the response is not greater than the threshold, it is interpreted as no response. The EMG response is measured as integrated muscle activity. The EMG measurement starts 3 ms after the stimulation and lasts 15 ms. The 3-ms delay helps prevent the effect of stimulation artifact. When using a MechanoSensor, response is measured as movement of the thumb, which is the area of positive signal. Regional block A regional block cable can be used as a nerve locator in local anesthesia. A maximum current of 5.0 mA is given every second, every other second, or every third second. The response measurement is ocular. 6-13
Cardiocap/5 Technical Reference Manual 6.4 Components for NESTPR hemodynamic parameters 6.4.1 Serial communication NESTPR Figure 6-11. Serial communication and opto isolation Serial communication between the NESTPR unit and the CPU is done by a RS485-type bus whose buffers get their supply voltage (+5 VDC) from the DC/DC board. In the isolation section, their supply voltage (+5 V) is obtained from the isolated power supply. The data transmission rate is 500 kbps. The serial communication buffers are also controlled by the Reset signal so that when the Reset is active, the buffer does not transfer data. Reset is also an RS485. Additionally, an auxiliary logic power reset keeps the reset active for about 500 ms despite the state of reset in the module bus. Time constant determines the power-up reset time. Components prevent the unit from sending data during reset. ECG board communication with the module bus is made through RXD and TXD pins. 6-14
Measurement Parameters 6.4.2 NIBP board Address Bus Figure 6-12. NIBP board functional block diagram Pressure transducers The NIBP board contains two piezo-resistive pressure transducers (B1 and B2). Transducer B1 measures the pressure of the blood pressure cuff and the pressure fluctuations caused by arterial wall movement. Transducer B2 detects the cuff hose type as well as Cuff loose and Cuff occlusion conditions. The transducers are temperature-compensated internally. They are supplied by a constant voltage and their output voltage changes up to 40 mV maximum (50 kPa, 375 mmHg). Signal processing The NIBP board is controlled with 80C51FA microprocessor at 16 MHz oscillator frequency. The microprocessor system is equipped with its own power-up reset in addition to the external RS485 reset line. Two signals from the pressure transducers are amplified and sent to the A/D converter. After the converter, digitized signals are sent to the microprocessor for data processing. Before the converter, one signal adjusts the offset to the pressure safety level. 6-15
Cardiocap/5 Technical Reference Manual Software control Software controls the valves and pump. In addition to ON/OFF signals for each component, a common power switch for the valves and the pump is used at pump/valve failures. Memory NIBP program memory (EPROM) size is 128k × 8. The RAM (32k × 8 bit) stores variable NIBP measurement values. The EEPROM (64 × 16 bit) stores the calibration values for the pressure transducers, the pulse valve constants gained during measurements, the PC board identification, and the module serial number. Watchdog timer The NIBP board is equipped with a software-independent safety circuit to disconnect supply voltages from the pump and the valves if the cuff has been pressurized longer than preset time. The pressure limit is specified to a maximum of 15 mmHg (2.0 kPa). As soon as the cuff pressure rises over 15 mmHg (2.0 kPa), a timer starts counting. The timer is adjusted to stop the pump and open the valves in 2 minutes 10 seconds in adult/child mode and in 1 minute 5 seconds in infant mode. Valves Exhaust valves 1 and 2 empty the cuff and the joining chamber after the measurement. Exhaust valve 1 is used as a safety valve in infant mode; it opens at 165 mmHg (22.0 kPa). Exhaust valve 2 is used as a safety valve in adult mode; it opens at 320 mmHg (42.7 kPa). The bleed valve empties the cuff during measurement. The zero valve connects pressure transducer B1 to open air. Power supply section All connections are established via a 25-pin connector (D-type, female). The unit needs a +5 V, ± 15 V, and +15 VD (dirty) power supply to operate. The pump and the valves use a separate +15 VD power line. Supply voltages are generated in the DC/DC board. The reference voltages ± 5 Vref and +10 Vref are generated on the NIBP board. 6-16
Measurement Parameters 6.4.3 ECG board Patient signals are connected to overload protection circuits (resistors and gas-filled surge arresters) and analog switches to instrumentation amplifiers. The signals are amplified by 480 and limited by slew rate. Then they are A/D-converted, analyzed, and transferred to the module bus in digital form. Figure 6-13. ECG board block diagram 6-17
Cardiocap/5 Technical Reference Manual Analog ECG section The ECG cable is connected to connector pins E1 to E6 on the input board, which contains an overload protection circuit. Analog switches connect leads to amplifiers. The state of the switches depends on the cable type. Lead-off, noise, and pacemaker are detected by a slew rate detector. Lower frequency is determined by a high-pass (HP) filter 0.5 Hz (monitor bandwidth) or 0.05 Hz (diagnostic or ST bandwidth). ECG filtering The Cardiocap/5 monitor has three ECG filtering modes: MONITORING 0.5 to 30 Hz (with 50 Hz reject filter) 0.5 to 40 Hz (with 60 Hz reject filter) DIAGNOSTIC 0.05 to 100 Hz ST FILTER 0.05 to 30 Hz (with 50 Hz reject filter) 0.05 to 40 Hz (with 60 Hz reject filter) Filtering reduces high-frequency noise and low-frequency (for example, respiratory) movement artifacts. The monitoring filter is used in normal monitoring. The diagnostic filter is used if more accurate diagnostic information is needed. The ST filter gives more accurate information of ST segment, but reduces high-frequency noise. The high-pass filters 0.5 Hz and 0.05 Hz are done with hardware. The monitor sends a command to the NESTPR unit determining which corner frequency (0.5 Hz or 0.05 Hz) is to be used. The 50 Hz and 60 Hz reject filters are low-pass filters with zero at 50 Hz or 60 Hz correspondingly and they are done with software. They are for the mains supply filtering. When these filters are used, 3 dB value for low-pass filter is 30 Hz or 40 Hz. Software filters are not used in diagnostic mode. Then, the upper frequency is limited by hardware and the –3 dB frequency is 100 Hz. Respiration section Analog switches control the current supply source of the impedance respiration measurement. The lead selection for the 3-lead cable can be seen from the following table: Selected lead Current source between Signal measured from I R - L F II R - F L III L - F R Position on body surface IEC Standard coding AAMI standard coding right arm R = red RA = white left arm L = yellow LA = black left leg F = green LL = red 6-18
Measurement Parameters When the 5-lead cable is used, the current source is between L-F and the signal is measured from the N, independently on the lead selection. The respiration amplifier consists of the operational amplifiers and the components around them. An analog switch controls the gain of the first stage of the preamplifier. The synchronous rectifier consists of the analog switches, which detect the respiration signal from 31 kHz amplitude modulated raw signal. The amplifier stage consists of the differential amplifier and the last amplifier. The differential amplifier consists of the operational amplifiers and the components around them. This stage is AC-coupled on both sides for minimizing the offset voltages. The last amplifier amplifies the signal derived from the differential amplifier stage. The respiration signal is zeroed at the beginning of the measurement. Zeroing is also used for fast recovery of the measurement after the motion artifact. This is done in the amplifier section. NOTE: The respiration measurement is switched OFF for 20 seconds when defibrillation is detected at the defibrillation detector. Microprocessor section The microprocessor contains RAM and EPROM. The processor uses external EEPROM memory. The microprocessors internal 8-channel A/D converter converts the ECG signals to digital form. Isolated section See STP board later in this chapter. The patient isolation of ECG is 5 kV. WARNING: Do not touch battery-operated monitor during defibrillation procedure. Power supply section See STP board later in this chapter. There is a test connector (X20) on the board for voltages +5 VREF, +5V, +12V, GND and –12V. 6-19
Cardiocap/5 Technical Reference Manual 6.4.4 STP board To NIBP Board Figure 6-14. STP board block diagram Microprocessor unit The STP board uses an Intel 80C196KC-16 processor with three A/D converters, external memories, an 8-bit data bus, a 16 MHz oscillator, an open collector reset, and a watchdog timer. The processors internal UART communicates with the CPU board. High speed I/O obtains the pulse control sequence for pulse oximetry measurement. The oscillator provides its timing clock. 6-20
Measurement Parameters Isolation section The outputs of the two opto isolators are analog signals in the isolated section, however, the signal is processed on logical high-low level. Reset line is an open collector type with a pull-up resistor, allowing the processor to use its internal watchdog function. Power supply section Isolated supply voltages of the boards are developed from +15 Vdirty voltage from the DC/DC board. The power supply is a switched mode circuit, where the FET transistor switch is controlled by an oscillator using a bipolar timer. The frequency of the oscillator is about 30 kHz and the pulse ratio is 50%. Control of the FET switch is slowed to suppress spurious interference. A special pulse transformer is used in the circuit. In the secondary circuit, normal linear regulators are used except for +5 V (low drop type linear regulator). 6.4.5 STP board—Pulse oximetry measurement section Figure 6-15. Pulse oximetry measurement block diagram LED control signals The processor sends pulse-width modulated signals, IRED intensity and RED intensity, that are converted to DC voltage and filtered. Switches send either RED or IRED intensity forward to the amplifier in the LED driving circuit. 6-21
Cardiocap/5 Technical Reference Manual LED driving circuit The differential amplifier circuit measures the voltage difference that corresponds to the LED current. Its output is sent back to the processor in 0 to 5 V level. There are feedback circuits from LED current measurement and LED intensity control. Background light is measured by picking up a sample from the signal. The sample is modified to 0 to 5 V level and sent to the processor. Measured signal preamplification The preamplifier is a current-to-voltage converter with gain selection. The higher gain is used for measuring thin tissue. Digitally controlled amplifier The D/A converter is a digitally-controlled amplifier after which there is another constant amplifier. Red and infrared channel separation Switches separate red and infrared channels. The operational amplifier functions as a buffer and after this, an infrared DC signal is sent to the processor. A capacitor separates out the AC signal, which is sent to the processor after amplification. There is a switch to choose the amplification constant. 6.4.6 STP board—Temperature measurement section The value of NTC resistor in the probe depends on the patients temperature, which is measured with the following principle. The temperature signal is produced by voltage dividers, part of which is the temperature probe used on the patient. The output is amplified by the calibrated amplifier whose offset voltage makes its output spread on both sides of zero. A wider output range (measurement range) means better resolution. Figure 6-16. Temperature measurement principle 6-22