    Measurement Parameters 
     
    Figure 6-25. Signal processing 
     
    Figure 6-26. Control logic 
     
    Figure 6-27. Calibration data stored in EEPROM 
    6-33  
    						
    							
    Cardiocap/5 Technical Reference Manual 
    6.7 NMT components 
    6.7.1 NMT board 
    The NMT circuit board consists of the following functional sections: 
    • Constant current stimulator. 
    • Measuring electronics for the EMG signals. 
    • Microprocessor for stimulation and measurement control, and for counting measurement results. 
    • Serial communication. 
    The serial bus speed is 500 kbps and the bus itself is half duplex (data can be transferred in both 
    directions but only one way at a time). 
    Serial communication 
     
    Figure 6-28. NMT serial communication and opto isolation 
    Stimulator 
    The constant current stimulator generates pulses whose amplitude is independent of the load. The 
    main components of the stimulator are a transformer, a capacitor and a transistor. The transformer 
    produces a high voltage which charges the capacitor and the transistor adjusts the pulse width and 
    amplitude of the current. 
     
    Figure 6-29. NMT stimulator block diagram 
    6-34  
    						
    							
    Measurement Parameters 
    6.7.2 Front panel connector (NMT) 
       
    NMT connector Pin No. Signal 
     
    1 
    2 
    3 
    4 
    5 
    6 
    7 
    8 
    9 
    10 
    EMG signal + 
    EMG signal – 
    Not used 
    Stimulus + 
    Stimulus – 
    Ground 
    Not connected 
    Sensor identification 
    +5 V 
    Mechanical signal 
    6.7.3 Module bus connector 
     
       
    Pin No. I/O Signal 
    1 I RESET_RS485 
    2 I –15 VDC 
    3 I +15 VDIRTY 
    4 I +15 VDC 
    5 I/O –DATA_RS485 
    6 I/O DATA_RS485 
    7 – Ground & Shield 
    8 I –RESET_RS485 
    9 I CTSB 
    10 O RTSB 
    11 I RXDB 
    12 O TXDB 
    13 – Ground & Shield 
    14 I +32 VDIRTY 
    15 I GroundDIRTY 
    16 I CTSC 
    17 O RTSC 
    18 I RXDC 
    19 O TXDC 
    20 – ON/STANDBY 
    21 – PWM_ECG 
    22 – RXDD_RS232 
    23 – TXDD_RS232 
    24 I +5 VDC 
    25 I +5 VDC 
       
     
    6-35  
    						
    							
    Cardiocap/5 Technical Reference Manual 
     
     
    6-36  
    						
    							
    Contents 
    Chapter 7. Service Procedures  
    7.1 General service information..................................................7-1 
    7.2 Monitor software installation................................................7-1 
    “Cold start” instructions........................................................................\
    .....7-1 
    “Warm start” instructions........................................................................\
    ...7-1 
    7.3 Replacement procedures......................................................7-2 
    7.3.1 Disassembling the frame assembly.............................................................7-2 
    7.3.2 Mother board removal........................................................................\
    .......7-2 
    7.3.3 ECG, NIBP, and STP board replacement......................................................7-3 
    NESTPR unit removal........................................................................\
    .........7-3 
    ECG board replacement........................................................................\
    .....7-3 
    NIBP board replacement........................................................................\
    ....7-3 
    STP board replacement........................................................................\
    ......7-3 
    7.3.4 Backplate unit removal........................................................................\
    ......7-3 
    7.3.5 Gas unit CPU board, pump, and fan replacement.........................................7-4 
    Airway-gas unit removal........................................................................\
    .....7-4 
    Gas unit CPU board replacement................................................................7-4 
    Pump unit replacement........................................................................\
    ......7-4 
    Fan replacement........................................................................\
    ...............7-4 
    7.3.6 AC/DC unit replacement........................................................................\
    ....7-5 
    7.3.7 DC/DC board replacement........................................................................\
    .7-5 
    7.3.8 Recorder unit replacement........................................................................\
    .7-5 
    7.3.9 Inverter board and Backlight board replacement..........................................7-5 
    Inverter board replacement........................................................................\
    7-5 
    7.3.10 CPU battery replacement........................................................................\
    ...7-6 
    7.3.11 CPU board replacement........................................................................\
    .....7-7 
    7.3.12 LCD display, LCD display backlight, and keyboard replacement....................7-7 
    LCD display removal........................................................................\
    ..........7-7 
    Replacing LCD display backlights...............................................................7-7 
    Replacing the keyboard........................................................................\
    .....7-7 
    7.3.13 Enhanced pulse oximetry (N-XOSAT) or SpO2 interface board replacement.....7-8 
    7.3.14 Nellcor pulse oximetry (N-XNSAT) or SpO2 interface board replacement..........7-9 
    7.3.15 NMT board or NMT cable assembly replacement..........................................7-9 
    7.4 Adjustments and calibration...............................................7-10 
    7.4.1 Pressure safety-level detection OFFSET.................................................7-10 
    7.4.2 NIBP calibration check........................................................................\
    .....7-11 
    Checking the primary pressure channel calibration.....................................7-11 
    Using a manometer to check NIBP calibration............................................7-11 
    7.4.3 NIBP calibration........................................................................\
    ..............7-12 
    7.4.4 Temperature calibration........................................................................\
    ...7-13 
    7.4.5 Invasive pressure calibration....................................................................7-13 
    7.4.6 General checks for gas system components..............................................7-14 
      
    						
    							
    Cardiocap/5 Technical Reference Manual 
    Oxygen measuring (OM) unit ..................................................................... 7-14 
    TPX measurin g unit ........................................................................\
    .......... 7-14 
    PVX measurin g unit........................................................................\
    .......... 7-14 
    Gas unit CPU  board........................................................................\
    .......... 7-14 
    7.4.7 Gas sampling system  leak test .................................................................. 7-15 
    7.4.8 Flow rate checks and adjustments ............................................................ 7-15 
    Flow rate checks ........................................................................\
    .............. 7-15 
    Flow rate ad justment.......................................................................\
    ......... 7-15 
    7.4.9 Spirometry leakage test  and calibration..................................................... 7-16 
    Spirometry leakage test........................................................................\
    .... 7-16 
    Flow calibration ........................................................................\
    ............... 7-16 
    7.4.10 Gas calibration ........................................................................\
    ................ 7-16 
    Calibration gas regula tor flow check .......................................................... 7-16 
    Gas calibration ........................................................................\
    ................ 7-17 
    Calibration adju stments ........................................................................\
    ... 7-18 
    Autozeroing intervals ........................................................................\
    ........ 7-18 
    Table of Figures  
    Figure 7-1. CPU batte ry replacement ........................................................................\
    ........ 7-6 
    Figure 7-2. SpO2 interface board ........................................................................\
    .............. 7-8 
    Figure 7-3. NI BP board ........................................................................\
    .......................... 7-10 
    Figure 7-4. Calibr ation switch ........................................................................\
    ................. 7-12 
    Figure 7-5. Attachin g the regulator........................................................................\
    .......... 7-17 
    Figure 7-6. Connecting a sampli ng line and feeding gas................................................... 7-17 
     
      
    						
    							
    Service Procedures 
    7. SERVICE PROCEDURES 
    7.1 General service information 
    Field service of the Cardiocap/5 monitor is limited to replacing faulty circuit boards, mechanical parts, 
    or a real-time clock battery. Some boards may be returned to GE Healthcare for exchange. 
    For service advice from GE Healthcare, please provide the unit serial number, full type designation, 
    and a detailed fault description.  
    CAUTION: Only trained personnel with the appropriate tools and equipment shall perform the tests 
    and repairs outlined in this chapter. Unauthorized service may void warranty of the unit. 
    CAUTION: Electrostatic discharge through the PC boards may damage the components. Before 
    replacing PC boards, wear a static-control wrist strap. Handle all PC boards by their non-conductive 
    edges and use anti-static containers when transporting them. 
    CAUTION: When reassembling the monitor, make sure all cables and tubing are reconnected properly 
    and are not pinched between other components. 
    7.2 Monitor software installation 
    “Cold start” instructions 
    Follow these installation instructions if the monitor has been off for more than 15 minutes: 
    1. Insert the service software PC card into the data card slot closest to the display. 
    2. Power on the monitor. Both LEDs illuminate and the screen remains blank for approximately two 
    minutes while loading. 
    3. When the normal monitoring screen is displayed, remove the card. 
    4. Perform the factory reset. 
    “Warm start” instructions 
    Follow these instructions if the monitor has been off for less than 15 minutes: 
    1. Insert the service software PC card into the data card slot closest to the display. 
    2. To start the software loading you need to perform the factory reset. Push the ComWheel and 
    select: 
    Install/Service (16-4-34) 
    Service (26-23-8) 
    Monitor 
    Factory Reset 
    Both LEDs illuminate and the screen remains blank for approximately two minutes while loading. 
    3. When the normal monitoring screen is displayed, remove the PC card. 
    4. Perform the factory reset. 
    7-1  
    						
    							
    Cardiocap/5 Technical Reference Manual 
    7.3 Replacement procedures 
    NOTE:  Reassembly is essentially the reverse of disassembly. When reassembling the unit, make sure 
    that the cables and tubes, if any, are not pinched between the boards. 
    Tools 
    • Screwdriver (pozidrive) 
    • Hexagon driver (3.0 mm) 
    • Pliers 
    • Cable tie 
    7.3.1 Disassembling the frame assembly 
    1. Remove the 4 hexagon and 13 pozidrive screws that hold the back cover. 
    2. To pull the front-cover unit apart from the inner-parts unit, remove the 4 screws that hold the front-
    cover unit and the inner-parts unit together.  
    
    						
    							
    Service Procedures 
    7.3.3 ECG, NIBP, and STP board replacement 
    The NESTPR unit must be removed before you can replace the ECG, NIBP, or STP boards. 
    NESTPR unit removal 
    You may separate the display from the frame to gain better access to the NESTPR unit, but it is not 
    necessary. To remove the unit without separating the display: 
    1. Disconnect the 5 patient-connector cables from the front panel. 
    Important: For reassembly, note the connection of temperature cable T1 (black wires) and, for 
    monitors with two temperature measurements, temperature cable T2 (yellow wires). T2 is not 
    connected for monitors that have only one temperature measurement. 
    2. Disconnect the NIBP tubes from the module. 
    NOTE: The foremost NIBP tube is connected to the upper NIBP connector. 
    3. Carefully lift the unit up. 
    ECG board replacement 
    1. Remove 4 screws from the ECG board and disconnect the flat cable from the STP board. 
    2. Detach the ECG board. 
    NIBP board replacement 
    1. To access the NIBP board, remove the ECG board. 
    2. Remove 4 screws from the corners of the NIBP unit. 
    3. Disconnect the ribbon cable from the EST board. 
    4. Disconnect the NIBP tubes from the adapter. 
    NOTE: Notice the correct tube connections; lower one to the inner one.  
    5. Pull the NIBP board carefully out. 
    STP board replacement 
    1. Detach the ECG board. 
    2. Disconnect NIBP tubes from the both sides of the plastic strut. Remove the plastic strut carefully 
    from the NESTPR unit’s frame. 
    3. Disconnect the ribbon cable from the STP board 
    4. Remove 4 screws from the corners of the STP board and detach the STP board. 
    7.3.4 Backplate unit removal 
    The backplate unit contains the fan, speaker, and I/O board  
    1. Remove the 3 screws that hold the backplate unit to the AC/DC unit. 
    2. Push the backplate unit forward and bend the retainers in both sides to loosen the unit. 
    3. Carefully pull out the backplate unit. 
    4. Disconnect the CPU–NET board cable from the I/O board. 
    5. Disconnect the fan power cable from the DC/DC board. 
    6. To remove the sample gas outlet tube, unfasten the lock washer from the gas outlet connector 
    (X6) and pry the connector out of the cover. 
    7-3  
    						
    							
    Cardiocap/5 Technical Reference Manual 
    7.3.5 Gas unit CPU board, pump, and fan replacement 
    The airway gas unit must be removed before you can replace the gas unit CPU board, the pump, or the 
    fan. 
    Airway-gas unit removal 
    1. Move the backplate unit out of the way to access the airway gas unit (see Backplate unit removal). 
    2. Unfasten the 3 screws and washers that hold the locking chock and remove the chock. 
    3. If installed, disconnect the hook-and-loop strap. 
    4. Disconnect the 2 tubes from the rear of the airway gas unit. 
    5. Grip the airway gas unit and pull it out sideways. 
    Gas unit CPU board replacement 
    1. Remove 4 screws from the corners of the OM board. 
    2. Disconnect the following cables from the gas unit CPU board: 
    
    						