    Service Menus 
    8.13 Service Log menu 
    The service log contains information about the monitor errors and events that occurred since the last 
    factory reset or service log reset. The service log is saved in the main CPU board’s SRAM memory. 
    NOTE: 
    • Cold start is a start-up with a power-off time of more than 15 minutes. The trend memory is 
    cleared and monitoring starts with the user default settings. 
    • Warm start is a start-up with a power-off time of less than 15 minutes. The trend information and 
    possible temporary settings are still available. 
    Scroll Last Er Scrolls the error (event) information. 
    Scroll Counters Scrolls the error or event counters. 
    Record Log  Prints the service log information on paper (with the recorder option, N-XREC). 
    Reset Log  Clears the service log content (recommended after a performed maintenance). 
    Reset Run Time  Resets the monitor’s run-time counter. The counter is located in the main CPU 
    board’s SRAM and basically counts the amount of time the monitor has been on 
    since leaving the factory. 
    Error History 
     
    To enter this service menu from 
    the Service View menu, select 
    Service Log. 
    Last errors Shows the last monitor errors and the time of their occurrence. 
    Error counters Contains counters for each different (detected) monitor error. The time of occurrence 
    of the last error is shown beside each counter. 
    Fast cold start Gives the number of erroneous cold starts when power was off < 20 seconds. The 
    cause can be a failing lead-acid battery or a software problem that was solved by 
    the hardware watchdog circuitry. 
    Last log reset Shows the date and the time of last service log reset. 
    8-35  
    						
    							
    Cardiocap/5 Technical Reference Manual 
    Event History 
     
    To enter this service menu from 
    the Service View menu: 
    1. Select Service Log.  
    2. Select Event History. 
    Running hours Shows how many hours the monitor has been on since the date and the time of the 
    last run-time reset. 
    Last events Shows the last events and the time of their occurrence. 
    Event counters Contains counters for each different (detected) event. The time the last event 
    occurred is shown beside each counter. 
    Last log reset Shows the date and the time of the last service log reset. 
    Alarm History 
    Shows the monitor alarm log. The alarm log is reset if the monitor is in standby for more than 15 minutes. 
     
    To enter this service menu from the Service View menu: 
    1. Select Service Log.  
    2. Select Event History. 
    3. Select Alarm History. 
     
    8-36  
    						
    							
    Contents 
    Chapter 9. Spare parts  
    9.1 F-MX (hemodynamic) and F-MXG (hemodynamic with gas)......9-1 
    9.1.1 Mechanical spare parts........................................................................\
    .......9-1 
    9.1.2 Electronic spare parts........................................................................\
    .........9-3 
    9.1.3 Connector boards and PVX unit....................................................................9-6 
    9.1.4 Hardware accessories........................................................................\
    .........9-7 
    9.1.5 Interconnect cables........................................................................\
    ............9-8 
    9.2 Parameter unit (NESTPR) and invasive pressures option  
    (N-XP)........................................................................\
    .........9-9 
    9.3 Airway gas options (N-XC, N-XCO, N-XCAiO)..........................9-10 
    9.3.1 Planned maintenance kits........................................................................\
    .9-10 
    9.3.2 Airway gas spare parts........................................................................\
    ......9-10 
    9.4 Recorder option (N-XREC)...................................................9-12 
    9.5 NMT and pulse oximetry options..........................................9-13 
    9.5.1 Enclosure kits........................................................................\
    ...................9-13 
    9.5.2 NMT option (N-XNMT)........................................................................\
    ........9-13 
    9.5.3 Pulse oximetry options (N-XOSAT and N-XNSAT)..........................................9-14 
     
     
      
    						
    							
    Spare Parts 
    9-1 
    9. SPARE PARTS 
    Use the Order Number when ordering parts. Item numbers are used only to identify parts shown in the 
    illustrations. 
    NOTE:  Remanufactured or exchange parts may be available for some measurement units or boards.  
    Request information when you order the part. 
    9.1  F-MX (hemodynamic) and F- MXG (hemodynamic with gas) 
    9.1.1  Mechanical spare parts 
    Item Description Order Number 
    1 Recorder hole cover plate (plastic) M1022174 
    2 Recorder hole cover plate (metallic) 6050-0006-779 
    3 Front cover M1033263 
    4 Rubber pad 895174 
    5 D-fend housing unit, complete 894737 
    6 D-fend body connector 875052 
     O-ring for D-fend body connector 65312 
     Lock ring for D-fend body connector 63813 
    7 Shield for display cable 6050-0005-865 
    8 Back cover unit M1020890 
    9 Rear foot, base (included in Item 8) Order Item 8 
     10 Rubber pad 65141 
    11 Label set   
    Includes  equipotentiality sticker (shown), battery (not 
    shown), and patent sticker (not shown) 6050-0005-677 
    12 Dust filter M1022178 
    13 Back mounting 6050-0007-186 
    14 Rear panel label, F-MX M1022216 
     Rear panel label, F-MXG M1022180 
      
    						
    							
    Cardiocap/5 Technical Reference Manual 
     
     
    9-2  
    						
    							
    Spare Parts 
    9.1.2 Electronic spare parts 
    Item Description Order Number 
    15 Potential equalization connector 6027-0000-018 
    16 Fuse T2AH 511200 
    17 AC/DC unit M1035974 
    18 Mother board M1020911 
    19 DC/DC board M1020913 
    20 NIBP cuff hose connector 64654 
    21 NIBP adapter,  F-MX 895173 
     NIBP adapter, F-MXG 895172 
    22 Parameter unit, NESTPR  M1033884 
    23 Battery 3Ah 12V  17013 
    24 Airway gas unit, N-XC (ID 896227) 898954 
     Airway gas unit, N-XCO (ID 896228) 898955 
     Airway gas unit, N-XCAiO (ID 896226) 898956 
    25 Fan unit M1020905 
    26 I/O board  M1020901 
    27 Net board M1021956 
    28 Loudspeaker unit  M1021089 
     
     
    9-3  
    						
    							
    Cardiocap/5 Technical Reference Manual 
    9-4 
    Electronic spare parts (continued) 
    Item Description Order Number 
    29a EMI Seal for display and recorder box, D/d = 1.8/0.7 72511 
    29b EMI Seal for housing, D/d = 1.60/0.64 72510 
    30 LCD display shield 572757 
    31 LCD display unit 6017-0000-052 
    32 Backlight for LCD display (includes 2 bulbs) 572791 
    33 Display adapter board M1020893 
    34 Inverter board 572754 
    35 Backlight board M1021912 
    36 Lithium battery for SRAM on CPU board 197230 
    37 CPU board  (including HW for Network functionality) 
    NOTE: The appropriate service soft ware kit(s) are required when 
    replacing the CPU board (and may be required when replacing the 
    battery, depending on the monitor software level): 
    M1020891 
     1.  Cardiocap/5 Monitor Softwa re for Anesthesia S/XANE01S 
     2.  Cardiocap/5 Monitor Software  for Critical Care S/XCCA01S 
     3. Cardiocap/5 N-XNET Software 
    (order if N-XNET is installed) S/XNETS 
     4. Cardiocap/5 N- XDNET Software  
    (order if N-XDNET is installed) S/XDNETS 
    37 CPU board  (not including HW for Network functionality) 
    NOTE: The appropriate service soft ware kit(s) are required when 
    replacing the CPU board (and may be required when replacing the 
    battery, depending on the monitor software level): 
    M1115092 
     1.  Cardiocap/5 Monitor Softwa re for Anesthesia S/XANE01S 
     2.  Cardiocap/5 Monitor Software  for Critical Care S/XCCA01S 
    38 ComWheel switch, rotary 113291 
    39 Keyboard M1021955 
    40 ComWheel knob 896291 
    115 Flex board, in the version ..01 only M1029296  
    						
    							
    Spare Parts 
     
    9-5  
    						
    							
    Cardiocap/5 Technical Reference Manual 
    9.1.3 Connector boards and PVX unit 
    Item Description Order Number 
    41 Temperature connector board M1021914 
    42 PVX unit for spirometry (ID 895771) 898958 
    43 Invasive blood pressure connector board F-MX M1021913 
     Invasive blood pressure connector board F-MXG M1021917 
    44 SpO2 connector board, standard SpO2 M1032534 
     SpO2 connector board /cable assembly, N-XNSAT option 6050-0005-830 
     SpO2 connector board/cable assembly, N-XOSAT option 6050-0005-920 
    45 ECG connector board M1021518 
    46 NMT connector board/cable assembly 6050-0005-890 
    114 Front mask for models with standard SpO2 895171 
    114 Front mask for models with N-XOSAT/NSAT  6050-0005-944 
     
     
     
    9-6  
    						