GE Cardiocap 5 Service Manual
Have a look at the manual GE Cardiocap 5 Service Manual online for free. It’s possible to download the document as PDF or print. UserManuals.tech offer 45 GE manuals and user’s guides for free. Share the user manual or guide on Facebook, Twitter or Google+.
Service Menus 8.13 Service Log menu The service log contains information about the monitor errors and events that occurred since the last factory reset or service log reset. The service log is saved in the main CPU board’s SRAM memory. NOTE: • Cold start is a start-up with a power-off time of more than 15 minutes. The trend memory is cleared and monitoring starts with the user default settings. • Warm start is a start-up with a power-off time of less than 15 minutes. The trend information and possible temporary settings are still available. Scroll Last Er Scrolls the error (event) information. Scroll Counters Scrolls the error or event counters. Record Log Prints the service log information on paper (with the recorder option, N-XREC). Reset Log Clears the service log content (recommended after a performed maintenance). Reset Run Time Resets the monitor’s run-time counter. The counter is located in the main CPU board’s SRAM and basically counts the amount of time the monitor has been on since leaving the factory. Error History To enter this service menu from the Service View menu, select Service Log. Last errors Shows the last monitor errors and the time of their occurrence. Error counters Contains counters for each different (detected) monitor error. The time of occurrence of the last error is shown beside each counter. Fast cold start Gives the number of erroneous cold starts when power was off < 20 seconds. The cause can be a failing lead-acid battery or a software problem that was solved by the hardware watchdog circuitry. Last log reset Shows the date and the time of last service log reset. 8-35
Cardiocap/5 Technical Reference Manual Event History To enter this service menu from the Service View menu: 1. Select Service Log. 2. Select Event History. Running hours Shows how many hours the monitor has been on since the date and the time of the last run-time reset. Last events Shows the last events and the time of their occurrence. Event counters Contains counters for each different (detected) event. The time the last event occurred is shown beside each counter. Last log reset Shows the date and the time of the last service log reset. Alarm History Shows the monitor alarm log. The alarm log is reset if the monitor is in standby for more than 15 minutes. To enter this service menu from the Service View menu: 1. Select Service Log. 2. Select Event History. 3. Select Alarm History. 8-36
Contents Chapter 9. Spare parts 9.1 F-MX (hemodynamic) and F-MXG (hemodynamic with gas)......9-1 9.1.1 Mechanical spare parts........................................................................\ .......9-1 9.1.2 Electronic spare parts........................................................................\ .........9-3 9.1.3 Connector boards and PVX unit....................................................................9-6 9.1.4 Hardware accessories........................................................................\ .........9-7 9.1.5 Interconnect cables........................................................................\ ............9-8 9.2 Parameter unit (NESTPR) and invasive pressures option (N-XP)........................................................................\ .........9-9 9.3 Airway gas options (N-XC, N-XCO, N-XCAiO)..........................9-10 9.3.1 Planned maintenance kits........................................................................\ .9-10 9.3.2 Airway gas spare parts........................................................................\ ......9-10 9.4 Recorder option (N-XREC)...................................................9-12 9.5 NMT and pulse oximetry options..........................................9-13 9.5.1 Enclosure kits........................................................................\ ...................9-13 9.5.2 NMT option (N-XNMT)........................................................................\ ........9-13 9.5.3 Pulse oximetry options (N-XOSAT and N-XNSAT)..........................................9-14
Spare Parts 9-1 9. SPARE PARTS Use the Order Number when ordering parts. Item numbers are used only to identify parts shown in the illustrations. NOTE: Remanufactured or exchange parts may be available for some measurement units or boards. Request information when you order the part. 9.1 F-MX (hemodynamic) and F- MXG (hemodynamic with gas) 9.1.1 Mechanical spare parts Item Description Order Number 1 Recorder hole cover plate (plastic) M1022174 2 Recorder hole cover plate (metallic) 6050-0006-779 3 Front cover M1033263 4 Rubber pad 895174 5 D-fend housing unit, complete 894737 6 D-fend body connector 875052 O-ring for D-fend body connector 65312 Lock ring for D-fend body connector 63813 7 Shield for display cable 6050-0005-865 8 Back cover unit M1020890 9 Rear foot, base (included in Item 8) Order Item 8 10 Rubber pad 65141 11 Label set Includes equipotentiality sticker (shown), battery (not shown), and patent sticker (not shown) 6050-0005-677 12 Dust filter M1022178 13 Back mounting 6050-0007-186 14 Rear panel label, F-MX M1022216 Rear panel label, F-MXG M1022180
Spare Parts 9.1.2 Electronic spare parts Item Description Order Number 15 Potential equalization connector 6027-0000-018 16 Fuse T2AH 511200 17 AC/DC unit M1035974 18 Mother board M1020911 19 DC/DC board M1020913 20 NIBP cuff hose connector 64654 21 NIBP adapter, F-MX 895173 NIBP adapter, F-MXG 895172 22 Parameter unit, NESTPR M1033884 23 Battery 3Ah 12V 17013 24 Airway gas unit, N-XC (ID 896227) 898954 Airway gas unit, N-XCO (ID 896228) 898955 Airway gas unit, N-XCAiO (ID 896226) 898956 25 Fan unit M1020905 26 I/O board M1020901 27 Net board M1021956 28 Loudspeaker unit M1021089 9-3
Cardiocap/5 Technical Reference Manual 9-4 Electronic spare parts (continued) Item Description Order Number 29a EMI Seal for display and recorder box, D/d = 1.8/0.7 72511 29b EMI Seal for housing, D/d = 1.60/0.64 72510 30 LCD display shield 572757 31 LCD display unit 6017-0000-052 32 Backlight for LCD display (includes 2 bulbs) 572791 33 Display adapter board M1020893 34 Inverter board 572754 35 Backlight board M1021912 36 Lithium battery for SRAM on CPU board 197230 37 CPU board (including HW for Network functionality) NOTE: The appropriate service soft ware kit(s) are required when replacing the CPU board (and may be required when replacing the battery, depending on the monitor software level): M1020891 1. Cardiocap/5 Monitor Softwa re for Anesthesia S/XANE01S 2. Cardiocap/5 Monitor Software for Critical Care S/XCCA01S 3. Cardiocap/5 N-XNET Software (order if N-XNET is installed) S/XNETS 4. Cardiocap/5 N- XDNET Software (order if N-XDNET is installed) S/XDNETS 37 CPU board (not including HW for Network functionality) NOTE: The appropriate service soft ware kit(s) are required when replacing the CPU board (and may be required when replacing the battery, depending on the monitor software level): M1115092 1. Cardiocap/5 Monitor Softwa re for Anesthesia S/XANE01S 2. Cardiocap/5 Monitor Software for Critical Care S/XCCA01S 38 ComWheel switch, rotary 113291 39 Keyboard M1021955 40 ComWheel knob 896291 115 Flex board, in the version ..01 only M1029296
Cardiocap/5 Technical Reference Manual 9.1.3 Connector boards and PVX unit Item Description Order Number 41 Temperature connector board M1021914 42 PVX unit for spirometry (ID 895771) 898958 43 Invasive blood pressure connector board F-MX M1021913 Invasive blood pressure connector board F-MXG M1021917 44 SpO2 connector board, standard SpO2 M1032534 SpO2 connector board /cable assembly, N-XNSAT option 6050-0005-830 SpO2 connector board/cable assembly, N-XOSAT option 6050-0005-920 45 ECG connector board M1021518 46 NMT connector board/cable assembly 6050-0005-890 114 Front mask for models with standard SpO2 895171 114 Front mask for models with N-XOSAT/NSAT 6050-0005-944 9-6