Page 101
3 - 17 (4) When automatic saving is set, data will be saved after completion of measurement. Upon opening its file, a data processing window will appear as in Fig. 3-12. Although a peak table can be displayed, simultaneous display of rate calculation (kinetics) result and peak table cannot be selected. Fig. 3-12 shows a spectrum and the result of rate calculation. This is selectable from the View menu. Unless automatic saving is set, select the Save As command from the File menu...
Page 102
3 - 18 3.5 Data Printout Printout can be provided even when the function for automatic printout after measurement is not used. The following two procedures are available. (1) Select the Print command from the File menu. Or click the (print) button on the toolbar. The field active on the data processing window (profiled in blue when clicked, field (a) in Fig. 3-12) will be printed. The contents, that are displayed by selecting the Print Preview command from the...
Page 103
3 - 19 (2) Select the Report command in the Data menu. Or click the (report) button on the toolbar. The Print Preview window will appear as shown in Fig. 3-14. Then, click Print. Select the items to be printed at the Report tab under Properties on the Edit menu. Fig. 3-14 NOTE: If characters overflow from the print frame in repeat measurement or in overlaying, then change the printing orientation from Portrait to Landscape. 3.5
Page 104
3 - 20 3.6 Calculating the Phosphorescence Life This is the function for calculating phosphorescence life from the measurement result (phosphorescence attenuation curve). Replacing data value at T1 with I 0, I0/e time (T2) is found, and T1, T2 and τ (phosphorescence life) = T2 - T1 are displayed. Select the phosphorescence life command from the data (D) menu or click the button on the Processing toolbar. A window as in Fig. 3-15 will appear. Fig. 3-15 Move...
Page 105
3 - 21 For printout of the calculation result, click Print. For printout of the spectrum and calculation result, apply a check mark to “With spectrum record” and click Print. For scale expansion, specify the area to be enlarged with the mouse. Draw a rectangle with the mouse button held down as shown in Fig. 3-17. Fig. 3-17 When releasing the mouse button, the specified area will be enlarged as shown in Fig. 3-18. Fig. 3-18 Double-clicking on the spectrum selects...
Page 106
4 - 1 4. PHOTOMETRY 4.1 Flow of Operation This section explains the method of determining sample concentration with a calibration curve prepared. Following is the operational flow. Fig. 4-1 Operational Flow (photometry) End Start Method setting Click button. Photometric method Measuring wavelength Standard table Sample name/comment input Click button. Measurement Measurement of standard and...
Page 107
4 - 2 4.2 Creating an Analysis Method Set the analytical conditions (method) prior to measurement. From the Edit menu, select the Method command or click the button. The dialog box shown in Fig. 4-2 will appear. Fig. 4-2 Analysis Method The following tabs are provided on the Analysis Method window for the photometry mode. These tabs are explained on the subsequent pages. • General • Quantitation • Instrument • Standards • Monitor • Report 4.2 Creating an...
Page 108
4 - 3 (1) General The parameters shown in Fig. 4-3 are provided under the General tab. Fig. 4-3 General Tab (a) Measurement (mode) Select a measurement mode from among the four listed below. • Wavelength scan • Time scan • Photometry • 3-D Scan (displayed when the ‘3-D Scan program’ (option) is installed in combination with Model F-2500) Here, select “Photometry.” For Wavelength scan, Time scan and 3-D Scan, refer to Sections 2, 3 and 5. (b) Operator...
Page 109
4 - 4 (e) Sampling • Standard When sample is exchanged manually. (f) Comments Enter a description or notes on measuring conditions. (g) Load Click this button to read out saved analytical conditions. The file opening window appears upon clicking, so select an analytical conditions file. (h) Save By clicking this button after setting the parameters for all six tabs, the analytical conditions are saved. Clicking this button displays the “Save As” window in creation of a...
Page 110
4 - 5 (a) Quantitation type Select the method of creating a calibration curve. Wavelength, Peak area, Peak height or Derivative method is selectable. Spectra measured by the Peak area, Peak height and Derivative method can be checked. 1) Wavelength Upon selecting “None” for calibration curve type in this quantitation method, up to 6 wavelengths can be set. When any other than “None” is selected for quantitation type, up to 3 wavelengths can be set. (i) One wavelength...