1 - 42 1.11 Photometer Menu Commands The following commands are available in the Spectrophotometer menu. • Set Wavelength • Zero Adjust • Auto Pre-scan • Shutter Open/Close (1) Set Wavelength This command sets the excitation/emission wavelength to a specified wavelength. Click the Set Wavelength command or button, and the dialog box shown below will come up. (2) Auto Pre-scan This command automatically searches for an optimum excitation/emission wavelength...
Page 62
1 - 43 1.12 Utility Menu Commands The Utility menu contains the following commands: • Instrumental Response • Analog Output Gain • EX Power Monitor • Sens. Check • Lamp Usage • Lamp Off (only with Model F-2500) (1) Instrumental Response When you use this function. You need the Spectral correction accessories kit (P/N 251-0126). (2) Analog Output Gain Use this command for setting a data value corresponding to 1 V of analog output. For details, refer to...
Page 63
1 - 44 1.13 Termination of FL Solutions Program After completion of measurement, shut down the instrument in the following procedure. (1) From the File (F) menu, select the Exit (X) command. Fig. 1-42 Select Yes, and the FL Solutions program will then be terminated. (2) Turn off the power switch of the spectrophotometer main unit. (3) Click the Start button of the Windows 98 and select the Shut Down (U) item. In the Shut Down Windows dialog box, select...
Page 64
2 - 1 2. WAVELENGTH SCAN 2.1 Flow of Operation This section explains the condition setting and operating method for wavelength scan. Measurement is made in the following procedure. Fig. 2-1 Operational Flow (wavelength scan) End Start Method setting Click button. Scan range Data mode Scan speed Sample name/comment input Click button. Pre-scan Measurement Data processing ...
Page 65
2 - 2 2.2 Creating an Analysis Method Set the analytical conditions (method) prior to measurement. From the Edit menu, select the Method command or click the button. The dialog box shown in Fig. 2-2 will appear. Fig. 2-2 Analysis Method The following tabs are provided on the Analysis Method window for the wavelength scan mode. These tabs are explained on the subsequent pages. General Instrument Monitor Processing Report 2.2 Creating an...
Page 66
2 - 3 (1) General The parameters shown in Fig. 2-3 are provided under the General tab. Fig. 2-3 General Tab (a) Measurement (mode) Select a measurement mode from among the four listed below. Wavelength scan Time scan Photometry 3-D Scan (displayed when the ‘3-D Scan program’ (option) is installed in combination with Model F-2500) Here, select “Wavelength scan.” For Photometry, Time scan and 3-D Scan, refer to Sections 3, 4 and 5. (b) Operator...
Page 67
2 - 4 (e) Sampling Standard When sample is exchanged manually. (f) Comments Enter a description or notes on measuring conditions. (g) Load Click this button to read out saved analytical conditions. The file opening window appears upon clicking, so select an analytical conditions file. (h) Save By clicking this button after setting the parameters for all five tabs, the analytical conditions are saved. Clicking this button overwrites parameters on the existing ones....
Page 68
2 - 5 (a) Scan mode Select one of the following scan modes. In case of Synchronous, excitation wavelength appears on the abscissa of the graph. Excitation Emission Synchronous (b) Data mode Select one of the following data modes. The Luminescence mode is for measuring luminescence in the biological/chemical field and only the sample side signal is acquired without performing ratio photometry. The Phosphorescence mode is for acquiring only phosphorescence after...
Page 69
2 - 6 (e) EX End WL This is the end wavelength for wavelength scan on the excitation side. Input the longer wavelength side of measuring wavelength range. Model Input Range F-4500 210 to 900 nm F-2500 230 to 800 nm (f) EX WL An excitation wavelength can be input. It is the end wavelength for wavelength scan on the excitation side. Input the longer wavelength side of measuring wavelength range. Model Input Range F-4500 200 to 900 nm F-2500 220 to 800 nm...
Page 70
2 - 7 (i) Scan speed Set a wavelength scan speed. (Unit: nm/min) F-4500 F-2500 30000 3000 12000 1500 2400 300 1200 60 300 15 60 15 For the Phosphorescence data mode, select one of the following values. 240 nm/min 60 nm/min 15 nm/min (j) Delay After pressing the Measure button, measurement is started following the delay time set here. It is used for temperature stabilization, etc. In repeat measurement, it is the time until the start of the first...