Page 91

3 - 7 
(f) Delay  
After pressing the Measure button, measurement is 
started following the delay time set here.    It is used for 
temperature stabilization, etc.    In repeat measurement, 
it is the time until the start of the first measurement.   
The delay time is invalid for the second and 
subsequent measurements.   
Input range:    0 to 9999 s   
 
(g) EX Slit  
Select a slit width for the excitation side.    Unit:    nm   
 
F-4500 F-2500 
1.0 2.5  
2.5 5.0 
5.0 10.0 
10.0 20.0 
20.0  
 
 
(h) EM...

Page 92

3 - 8 
(j) Response  
Select a response time from the following.   
 
F-4500 F-2500 
0.04 0.04 
0.01 0.08 
0.05 0.4 
0.1 2 
0.5 8 
2.0 Auto 
8.0  
Auto  
 
 
By selecting Auto, a response time is set automatically 
according to scan speed.    Select Auto usually.   
 
(k)  Shutter control (only with F-4500)   
The shutter can be automatically closed in other than 
measurement for minimizing sample deterioration due 
to the energy of excitation beam and opened when 
measurement starts.    When you put a...

Page 93

3 - 9 
(3) Monitor  
 
When selecting the Monitor tab, the window in Fig. 3-5 
appears.   
 
 
 
Fig. 3-5  Monitor Tab 
 
 
Assign the following parameters.   
 
(a)  Y-Axis Max (maximum)   
Input a maximum value for Y-axis on the monitor 
window.  
 
(b)  Y-Axis Min (minimum)   
Input a minimum value for Y-axis on the monitor 
window.  
 
(c)  X-Axis Max (maximum)   
Input a maximum value for X-axis on the monitor 
window.  
 
(d)  X-Axis Min (minimum)   
Input a minimum value for X-axis on the monitor...

Page 94

3 - 10 
(f)  Open data processing window after data acquisition   
Select whether or not to conduct data processing after 
sample measurement.    When selected (check mark is 
applied), data is automatically saved after 
measurement.  
 
NOTE:  When the remaining memory capacity of PC 
in operation is very small (system resource is 
less than 20%), the data processing window 
will not open even if this item is turned ON.   
Either close the presently opened data 
processing window or restart the FL...

Page 95

3 - 11 
(a) Processing choices  
A list of data processing (Savitsky-Golay smooth, 
Mean smooth, Median smooth, Derivative) is shown.     
Select a data processing item, and click the rightward-
pointing arrow key between the Processing choices 
and Processing steps display fields.    Then, the 
selected method appears in the Processing steps field.   
 
(b) Processing steps  
The processing sequence is displayed.    To delete a 
processing method, first select the method, and then 
click the...

Page 96

3 - 12 
(5) Report  
 
Select the Report tab, and the window in Fig. 3-7 will 
appear.  
 
 
 
Fig. 3-7  Report Tab 
 
 
This window is used to select an output method and items 
of measurement result.     
 
(a) Output  
1) Print Report  
A report will be printed.    For printing method, 
refer to 3.6.   
2)  Use Microsoft® Excel   
Data will be transferred to Microsoft® Excel.     
For transfer method, refer to 6.13.   
3)  Use print generator sheet 
Select this method for using the report generator...

Page 97

3 - 13 
(c)  Printable (transferable) items   
1) Include data 
The date and time of analysis can be printed.   
Printing time will always be printed.       
2) Include method  
3) Include graph  
4)  Include data listing   
• Constant  
Data will be printed at fixed intervals between 2 
wavelength points.    Input Data interval, Start 
wavelength and End wavelength.     
• Select data  
Data at the specified wavelength will be printed.   
Up to 12 wavelengths can be specified.   
5)  Include peak table...

Page 98

3 - 14 
3.3  Defining Your Samples  
 
When “Use Sample Table” has been set for the Analysis Method, 
the samples must be defined before measurement.   
 
(1)  Open the monitor window and click the          button.   
 
(2)  When the Sample Table window in Fig. 3-8 appears, fill in 
the blank section.    For instance, if there are five samples, 
click the Insert button five times.   
 
 
 
Fig. 3-8    Sample Table Window 
 
 
(3)  Input the Sample, Comments and File Name for each 
sample.    The sample...

Page 99

3 - 15 
(4)  When input is finished, check the entries and then click the 
OK button.   
If a sample table is not used, then click the          button.   
The window in Fig. 3-9 will appear.    Make entries for 
Sample and Comment.    For automatic saving, put a check 
mark in the “Auto file saving after measurement” box and 
specify a file name for saving.   
 
 
 
Fig. 3-9 
3.3  

Page 100

3 - 16 
3.4    Conducting a Time Scan   
 
(1)  There are two methods of starting the scan.     
 
(a)  Select the Measure command from the 
Spectrophotometer menu.   
 
(b)  Click the     (measurement) button on the 
measurement toolbar.   
 
For halting the measurement midway, click the          button 
on the toolbar.   
 
(2)  Upon clicking the          button, the window shown in   
Fig. 3-10 appears when using a sample table.    Now click 
Yes.  
 
 
 
Fig. 3-10 
 
 
(3)  The dialog box as in Fig....