Page 71
GE Logiq E9/Vivid E9 Training Manual © 2016 Conquest Imaging 64 Module 8 Parts Replacement This section provides procedures and information for replacing spare parts on the E9 systems. Always follow the standard ESD and safety procedures when replacing parts or modules. Pay close attention to the part number when replacing parts: Most parts are labeled with the part number. Always replace like for like or check compatibility. Visually inspect all parts before installing. Some parts...
Page 72
GE Logiq E9/Vivid E9 Training Manual © 2016 Conquest Imaging 65 3. Foot Rest Bumper 4. Front Cover 5. Filter Cover 6. Rear Cover Replace the covers in reverse order. Side Covers The side covers are fastened with two “click” locks, located at the lower end of the cover. To remove the side covers: Insert a #1 phillips screwdriver into the rectangular hole in the side bumper until it reaches the lock mechanism. Push the handle on the screwdriver downwards to release the lock. Lift...
Page 73
GE Logiq E9/Vivid E9 Training Manual © 2016 Conquest Imaging 66 the snap locks, while freeing the side latches. Front Cover Follow these steps to remove the front cover: Remove the other panels and the foot rest bumper. Unscrew the two (2 pc) screws that fix the front cover assembly to the chassis. Pull the upper end of the front cover out and upwards to free it from the pedals and the frame. Back Filter Cover To remove the filter cover: Gently pull the filter cover out and away...
Page 74
GE Logiq E9/Vivid E9 Training Manual © 2016 Conquest Imaging 67 Rear Cover There are four screws; two on each side of the cover. Remove the screws from both sides of the rear cover and lift it away from the chassis. When replacing the rear cover make sure the manual release handle of the Z Mechanism extends through the TOP air vent on the rear cover.
Page 75
GE Logiq E9/Vivid E9 Training Manual © 2016 Conquest Imaging 68 Remove Bulkhead Cover At the rear of the system, release the console’s frogleg mechanism by inserting a screwdriver into the release point and pressing until release. Pull the console out to its extended position to gain access to the screws in the next step. Use thumbs to press upper lock tabs toward the front of the system to release the top of the Bulkhead Cover. Pull the Bulkhead Cover away from the system. Slots for...
Page 76
GE Logiq E9/Vivid E9 Training Manual © 2016 Conquest Imaging 69 Remove Handle, Left/Right Top Handle Preparations: 1.) Power down the Vivid E9 and disconnect the Mains Power Cable. 2.) Remove the buttons below the Touch Panel. 3.) Remove the Upper Operator Panel. 4.) Remove the Lower Operator Panel. Hatches are used to attach the top handles to the lower User Interface. Be careful when doing this procedure. It is easy to break the plastic hatches The screwdriver positions shown below...
Page 77
GE Logiq E9/Vivid E9 Training Manual © 2016 Conquest Imaging 70 Palm Rest Assembly Prepare: power down the Vivid E9 2.) Disconnect the Mains Power Cable from the wall outlet. 3.) Disconnect all probes and external I/O cabling. 4.) Remove the Upper Operator Panel. 5.) Remove the Lower Operator Panel. 6.) Remove Handle Left Top and Handle Right Top. Palm Rest Assembly Removal To remove the palm rest: Release the hatch lock mechanisms securing the palm rest assembly using the same method and...
Page 78
GE Logiq E9/Vivid E9 Training Manual © 2016 Conquest Imaging 71 LCD Monitor Replacement Prepare for replacement: Move the top console to its lower locked position. Power down the system. Disconnect all probes and external I/O cabling. Logiq E9 LCD Monitor Replacement Logiq E9 systems come with two types of monitors: Original LCD Monitor (P/N 5167953) LCD Monitors used in R4 original consoles, LCD Monitor V2 (5392293-21, used in R4.x and later production) LCD Monitor (P/N 5167953)...
Page 79
GE Logiq E9/Vivid E9 Training Manual © 2016 Conquest Imaging 72 The LCD monitor arm lock is LOCKED. Install the LCD monitor assembly on the monitor bracket. Verify that all four fixing screws have engaged in their slots. Use a level if available on the monitor while you tighten the four screws. Tilt the monitor to horizontal position and connect the cables. If the cables were replaced replace the tiewraps and cable clamps, if present. Reinstall WLAN dongle, if the system is equipped...
Page 80
GE Logiq E9/Vivid E9 Training Manual © 2016 Conquest Imaging 73 Remove the two fixing screws from the back side of the LCD Monitor assembly and lift the cover. Disconnect the two cables shown below: Tilt the monitor back to its vertical position before removing the screws. Lift the LCD monitor assembly upwards away from the LCD arm bracket. Replace with the new LCD monitor. Replacing the LCD Arm assembly Vivid Preparations To prepare for replacement: 1.) Power down the Vivid E9...