Page 71

 GE Logiq E9/Vivid E9 Training Manual 
© 2016 Conquest Imaging 64 
Module 8 Parts Replacement 
This section provides procedures and information for replacing spare 
parts on the E9 systems. Always follow the standard ESD and safety 
procedures when replacing parts or modules. Pay close attention to 
the part number when replacing parts: 
 Most parts are labeled with the part number. 
 Always replace like for like or check compatibility. 
 Visually inspect all parts before installing.  
 Some parts...

Page 72

 GE Logiq E9/Vivid E9 Training Manual 
© 2016 Conquest Imaging 65 
3. Foot Rest Bumper 
4. Front Cover  
5. Filter Cover  
6. Rear Cover 
 
Replace the covers in reverse order. 
Side Covers 
The side covers are fastened with two “click” locks, located at the 
lower end of the cover. To remove the side covers: 
 
 Insert a #1 phillips screwdriver into the rectangular hole in the 
side bumper until it reaches the lock mechanism. Push the 
handle on the screwdriver downwards to release the lock. 
 Lift...

Page 73

 GE Logiq E9/Vivid E9 Training Manual 
© 2016 Conquest Imaging 66 
the snap locks, while freeing the side latches. 
 
Front Cover 
Follow these steps to remove the front cover: 
 Remove the other panels and the foot rest bumper. 
 Unscrew the two (2 pc) screws that fix the front cover assembly 
to the chassis. 
 
 Pull the upper end of the front cover out and upwards to free it 
from the pedals and the frame. 
Back Filter Cover 
To remove the filter cover: 
  Gently pull the filter cover out and away...

Page 74

 GE Logiq E9/Vivid E9 Training Manual 
© 2016 Conquest Imaging 67 
 
Rear Cover 
There are four screws; two on each side of the cover. Remove the 
screws from both sides of the rear cover and lift it away from the 
chassis. 
 
When replacing the rear 
cover make sure the 
manual release handle of 
the Z Mechanism extends 
through the TOP air vent 
on the rear cover.  

Page 75

 GE Logiq E9/Vivid E9 Training Manual 
© 2016 Conquest Imaging 68 
Remove Bulkhead Cover 
At the rear of the system, release the console’s frogleg mechanism by 
inserting a screwdriver into the release point and pressing until 
release. Pull the console out to its extended position to gain access to 
the screws in the next step. 
 
Use thumbs to press upper lock tabs toward the front of the system 
to release the top of the Bulkhead Cover. 
 
Pull the Bulkhead Cover away from the system. 
 
 
Slots for...

Page 76

 GE Logiq E9/Vivid E9 Training Manual 
© 2016 Conquest Imaging 69 
Remove Handle, Left/Right Top Handle  
Preparations: 
1.) Power down the Vivid E9 and disconnect the Mains Power Cable. 
2.) Remove the buttons below the Touch Panel. 
3.) Remove the Upper Operator Panel. 
4.) Remove the Lower Operator Panel. 
Hatches are used to attach the top handles to the lower User 
Interface. 
Be careful when doing this procedure. It is easy to break the plastic 
hatches  
The screwdriver positions shown below...

Page 77

 GE Logiq E9/Vivid E9 Training Manual 
© 2016 Conquest Imaging 70 
Palm Rest Assembly 
Prepare: power down the Vivid E9 
2.) Disconnect the Mains Power Cable from the wall outlet. 
3.) Disconnect all probes and external I/O cabling. 
4.) Remove the Upper Operator Panel. 
5.) Remove the Lower Operator Panel. 
6.) Remove Handle Left Top and Handle Right Top. 
Palm Rest Assembly Removal 
To remove the palm rest: 
 Release the hatch lock mechanisms securing the palm rest 
assembly using the same method and...

Page 78

 GE Logiq E9/Vivid E9 Training Manual 
© 2016 Conquest Imaging 71 
LCD Monitor Replacement 
Prepare for replacement: 
 Move the top console to its lower locked position. 
 Power down the system. 
 Disconnect all probes and external I/O cabling. 
Logiq E9 LCD Monitor Replacement 
Logiq E9 systems come with two types of monitors: 
 Original LCD Monitor (P/N 5167953) 
 LCD Monitors used in R4 original consoles, LCD Monitor V2 
(5392293-21, used in R4.x and later production)  
LCD Monitor (P/N 5167953)...

Page 79

 GE Logiq E9/Vivid E9 Training Manual 
© 2016 Conquest Imaging 72 
 The LCD monitor arm lock is LOCKED. 
 Install the LCD monitor assembly on the monitor bracket. Verify 
that all four fixing screws have engaged in their slots. 
 Use a level if available on the monitor while you tighten the four 
screws. 
 Tilt the monitor to horizontal position and connect the cables. If 
the cables were replaced replace the tiewraps and cable clamps, 
if present. 
 Reinstall WLAN dongle, if the system is equipped...

Page 80

 GE Logiq E9/Vivid E9 Training Manual 
© 2016 Conquest Imaging 73 
 Remove the two fixing screws from the back side of the LCD 
Monitor assembly and lift the cover.  
 Disconnect the two cables shown below: 
 
 Tilt the monitor back to its vertical position before removing the 
screws. 
 Lift the LCD monitor assembly upwards away from the LCD arm 
bracket. 
 Replace with the new LCD monitor. 
Replacing the LCD Arm assembly 
Vivid Preparations 
To prepare for replacement: 
1.) Power down the Vivid E9...
Start reading GE Vivid E9 User Manual