GE Logiq E9/Vivid E9 Training Manual © 2016 Conquest Imaging 54 Electromagnetic Interference Electromagnetic Interference (EMI) between wireless electronic transmitting devices and medical equipment can cause degradation of the ultrasound image. The system is in compliance with existing EMI/EMC requirements. However, the use of this system in proximity of an electromagnetic field can cause degradation of the ultrasound image at times. Review the environment in which the system is being used, to...
Page 62
GE Logiq E9/Vivid E9 Training Manual © 2016 Conquest Imaging 55 Low humidity (hot and dry conditions) Improper grounding Unshielded cables Poor connections Use the following precautions in order to reduce ESD: Wrist straps Antistatic mats Antistatic spray on carpets or linoleum Before handling a board or you can touch a grounded surface such as the power supply housing to discharge any built-up charge. Fire Safety The following general fire safety practices are relevant for...
Page 63
GE Logiq E9/Vivid E9 Training Manual © 2016 Conquest Imaging 56 Module 7 Troubleshooting This module covers common troubleshooting tools and procedures for the E9 systems. Normal Power On/Boot Sequence The following describes the normal power on/boot sequence for the E9 systems The system has two power switches: The Circuit Breaker Switch is located on the back of the system next to the power cord plug-in. This allows to the system power supply to power up but does not turn on the system...
Page 64
GE Logiq E9/Vivid E9 Training Manual © 2016 Conquest Imaging 57 Turn System ON Switch ON the Mains power circuit breaker at the rear of the unit. The control panel ON/OFF button should become amber. You should also hear a “click” sound from the relays in the AC Power supply and the unit is ready to boot. The ON/OFFbutton will remain amber. Press once on the ON/OFF button on the Operator Panel to boot the unit. The ON/OFF button turns green when it is pressed. During a normal boot sequence, you...
Page 65
GE Logiq E9/Vivid E9 Training Manual © 2016 Conquest Imaging 58 Troubleshooting Log/Image Capture If the system should malfunction you can capture logs by pressing the Alt+D keys simultaneously. This Alt+D function should respond even in the event of a system lock-up. This will collect a screen capture of the monitor, system presets and several log files in a date and time stamped “.zip” file. Depending on the system the Alt+D function collects a screen capture of the image monitor,...
Page 66
GE Logiq E9/Vivid E9 Training Manual © 2016 Conquest Imaging 59 Trouble Shooting Network Issues The network tools include Wireshark which is a “network sniffer” that displays network traffic between the ultrasound system and other devices. It is useful for debugging connectivity issues. Its output can be saved to a file. To display the Wireshark interface press Alt-N: In the Capture/Interfaces window press the Options button that corresponds to the IP address. Uncheck Capture packets...
Page 67
GE Logiq E9/Vivid E9 Training Manual © 2016 Conquest Imaging 60 Press Apply. Once Wireshark displays DICOM traffic, you can select an individual line and expand the DICOM tree in the lower frame to view DICOM-specific data. Capture Restart clears the current traffic display, in preparation for the next DICOM exchange. Entries with “Reject” or “Abort” indicate issues. To save the network traffic currently displayed, stop the capture and press File / Save As. In the pop-up, select a...
Page 68
GE Logiq E9/Vivid E9 Training Manual © 2016 Conquest Imaging 61 This saves the D=COM traffic in a file with suffix “.pcap”, which can be analyzed by Wireshark or a similar network sniffer. Trouble Shooting Operator Panel Issues This section covers common operator panel issues. No Audio Check volume settings in the application and also in Windows. Use headphones, (the type that you use on a personal CD player or a laptop) to test the audio output directly from the back of the BEP. If no...
Page 69
GE Logiq E9/Vivid E9 Training Manual © 2016 Conquest Imaging 62 board. Always measure the speakers’ impedance; it should measure approximately 7 ohms. If speakers are bad, it is possible that the amplifier on the Upper OP Assembly might be defective, too. No Video on LCD Display If the video is too dim, has dimmed areas or there is no video at all, replace the HV LCD inverter. Check the cabling within the system. If the HV LCD inverter is burnt, there is a possibility that the Digital...
Page 70
GE Logiq E9/Vivid E9 Training Manual © 2016 Conquest Imaging 63 board. If only one probe fails to be recognized, replace the probe.