Page 111
GE Logiq E9/Vivid E9 Training Manual © 2016 Conquest Imaging 104 Front End Parts Replacement The term front end as whole refers to the input and signal processing side of the system including probes, signal relays, ADC/DAC Vivid E9 Front End Processor (FEP) To prepare: Power down the Vivid E9 Disconnect all probes and external I/O cabling. Remove the Right Side Cover First remove the FEP cover. An EMI cover is used to ground the PCIe (GFI) cable to the FEP cover: Unscrew the fixing...
Page 112
GE Logiq E9/Vivid E9 Training Manual © 2016 Conquest Imaging 105 Use top and bottom ejectors on GRLY board to slightly pull out front planes. Use last GTX board and repeat same process as GRLY. The card connectors have color-keys to prevent installation in an incorrect rack position. If the system is powered on with a card placed in the wrong position, the VIVID E9 will be destroyed.
Page 113
GE Logiq E9/Vivid E9 Training Manual © 2016 Conquest Imaging 106 Repeat those steps until the two XD_Bus boards release from remaining FEP boards. Gently release each of XD_Bus boards from connectors on GRLY and GTX. Be sure to place boards on ESD safe surface.
Page 114
GE Logiq E9/Vivid E9 Training Manual © 2016 Conquest Imaging 107 Module 9 System Adjustments This module covers adjustments that can be made to the GE E9 systems. Touch Screen Calibration The Touch Screen Calibration is found on the Operator Panel Test Dialog. Follow these steps to open the Operator Panel Test Dialog: Select Config (F2). Log on as ADM and select System > Test. Select Operator Panel Test. This will open the Operator Panel Test Dialog: The Calibrate button is used to...
Page 115
GE Logiq E9/Vivid E9 Training Manual © 2016 Conquest Imaging 108 Select Calibrate. The Touch screen goes blank and display a cross in the screen’s upper left corner. Put your finger on the centre of the cross and press slightly. This calibrates this point and the cross moves to the next point on the screen. Continue until all five points on the screen have been calibrated. Select Done.to complete the calibration. DC Offset Calibration This procedure calibrates each ADC channel of the GRX...
Page 116
GE Logiq E9/Vivid E9 Training Manual © 2016 Conquest Imaging 109 Diagnostics. Select the “+” before Service Diagnostics folder to open the list of choices. Select the “+” before Analog Receive to open the test choices for the analog receiver. Select DC Offset Calibration and then follow the on- screen instructions. folder. Select Execute to run the calibration. After completing the calibration, you must restart the system. Operator Panel Adjustments This section covers...
Page 117
GE Logiq E9/Vivid E9 Training Manual © 2016 Conquest Imaging 110 Normal Operator Panel XY Movement When the Operator Panel is in the locked position if you press the left button (lock) of the Frogleg Controls, it will cause the motorized park lock to release and momentarily releases the brakes in the froglegs permitting the OP to move in the XY direction. When the console is not locked (floating) the lock button will only release the brakes in the froglegs to allow easy movement. The lock...
Page 118
GE Logiq E9/Vivid E9 Training Manual © 2016 Conquest Imaging 111 XY Brake Adjustment Locate the hex adjustment screw behind the XY Assembly. Use a 3 mm hex tool to adjust the XY Lock release for optimal lock and minimal slack in transport mode. Clockwise will tighten the lock and will reduce slack, but will also require more force for engaging the lock. Rotating counterclockwise will increase the slack. Too tight will reduce the slack and make it harder to engage the lock. Make the...
Page 119
GE Logiq E9/Vivid E9 Training Manual © 2016 Conquest Imaging 112 Glossary Apodization – A weighting function used as means of reducing side lobes in the beam. Application Entity – A node in a DICOM network. Back End – System block on the user interface side that contains master controller, signal processing, image memory and video layout, peripheral and user interface. Beamforming – A common signal processing technique used to enable directionally or spatially selected signals to be sent or...
Page 120
GE Logiq E9/Vivid E9 Training Manual © 2016 Conquest Imaging 113 twice the frequency of the transmitted sound. I2C – Inter-IC bus, a two wire serial bus for communication between integrated circuits. Developed by Philips in the 1980’s it is now an industry standard. In Plane Switching – A type of thin film transistor LCD screen that has particularly good wide viewing angle and accurate color reproduction. Loops – Multiframe objects (e.g. video) M Mode – is also known as T-M mode or time-motion...