GE Logiq E9/Vivid E9 Training Manual © 2016 Conquest Imaging 14 Power from the wall outlet (100 to 240 VAC, 50/60 Hz) is connected to the Main Power Supply. The MPS delivers the needed voltages to the rest of the Logiq E9/Vivid E9: Internal Peripherals (110-115 VAC) Front End Rack (DC power with several voltages) Front End Rack (TXPSV1 and TXPSV2 for the transmitters) (TSV1 and TSV2 for the transmitters) Front End Rack (PMXVOUT for the probe channel multiplexers) Back End Processor...
Page 22
GE Logiq E9/Vivid E9 Training Manual © 2016 Conquest Imaging 15 8. ACFAIL_N output signal goes high. 9. Controller turns OFF +48V. 1. BEP detects short-term contact of Power Switch. 2. BEP OS dialog box prompts user for power down. 3. BEP, via PCI express, commands the GFI or MRX to stop scanning, disconnect TX circuits from TS, stop strobe TS_ON_STRB. 4. BEP, via USB, commands Power Supply to shutdown TSV1 and TSV2 (SetTxPS 0x68). 5. Controller turns OFF PMX voltages. 6. BEP, via USB,...
Page 23
GE Logiq E9/Vivid E9 Training Manual © 2016 Conquest Imaging 16 Module 3 Operating Modes The following general operating modes are available on E9 systems: B-Mode B-Mode is a two-dimensional image of the amplitude of the echo signal. It is used for location and measurement of anatomical structures and for spatial orientation during operation of other modes. In B mode, a two-dimensional cross-section of a three- dimensional soft tissue structure such as the heart is displayed in real time....
Page 24
GE Logiq E9/Vivid E9 Training Manual © 2016 Conquest Imaging 17 M-Mode In M-mode, soft tissue structure is shown as a scrolling display, with depth on the Y-axis and time on the X-axis. It is mostly used for cardiac measurements. M-mode is also known as T-M mode or Time- Motion mode. Ultrasound echoes of different intensities are mapped to different gray scale values in the display. M-mode displays time motion information derived from a stationary beam. M-mode is normally used in conjunction...
Page 25
GE Logiq E9/Vivid E9 Training Manual © 2016 Conquest Imaging 18 Power Doppler Power Doppler is the same as Color Doppler except that it uses the amplitude of the signal to detect movement. The power in the remaining signal after wall filtering is then averaged over time to present a steady state image of blood flow distribution. It is independent of velocity and direction of flow, so there is no signal aliasing. It is independent of angle allowing the detection of smaller velocities than Color...
Page 26
GE Logiq E9/Vivid E9 Training Manual © 2016 Conquest Imaging 19 The benefits of CW Doppler include high sensitivity to low velocities and detection of high velocities without aliasing. CW Doppler cannot distinguish between the sending and receiving signals or extraneous echoes, nor does CW Doppler produce a precise image like Pulsed Wave Doppler. Other Modes 4D: The E9 Ultrasound System can be used to acquire multiple, sequential 2D (B-Mode) (BMode) images which can be combined to reconstruct...
Page 27
GE Logiq E9/Vivid E9 Training Manual © 2016 Conquest Imaging 20 Module 4 Network Configuration The following sections provide some basic biomedical networking background information along with information and procedures specific to the GE LogiqE9/VividE9 ultrasound systems. DICOM The Digital Imaging and Communications in Medicine (DICOM) is a standard that specifies a consistent file structure for biomedical images and important associated information that must remain associated with the...
Page 28
GE Logiq E9/Vivid E9 Training Manual © 2016 Conquest Imaging 21 select a pre-configured dataflow to automatically customize the unit to work according to the settings associated with this dataflow. For the Logiq systems the dataflows must be configured during system set up. Vivid E9 Connectivity Vivid E9 Dataflow Descriptions This section describes the pre-configured dataflows for th Vivid E9 systems. LocalArchive-Int.HD No DICOM multi-frame is stored in this configuration. The local...
Page 29
GE Logiq E9/Vivid E9 Training Manual © 2016 Conquest Imaging 22 Worklist/Local Archive - LocalHD Search in the DICOM Modality Worklist, the patient found is copied into the local database. The patient information and examination results are stored to the local database. Images are stored to the local hard disk. Worklist/Remote Archive - Remote Storage This dataflow is used in a network environment that includes Vivid HL7 Gateway. The patient list in the Search/Create Patient window is coming...
Page 30
GE Logiq E9/Vivid E9 Training Manual © 2016 Conquest Imaging 23 Adjusting the Assigned Devices Select the device in the Selected devices field. Press Properties to display the Properties window. Adjust the device specific parameters listed below: General Settings Definitions: Name: give a descriptive name for the device. IP address: Select from drop-down menu Database Name: Automatically selected according to the IP address File destination: Automatically selected according to the IP...