Home > GE > Ultrasound System > GE Vivid E9 User Manual

GE Vivid E9 User Manual

    Download as PDF Print this page Share this page

    Have a look at the manual GE Vivid E9 User Manual online for free. It’s possible to download the document as PDF or print. UserManuals.tech offer 45 GE manuals and user’s guides for free. Share the user manual or guide on Facebook, Twitter or Google+.

    							 GE Logiq E9/Vivid E9 Training Manual 
    © 2016 Conquest Imaging 64 
    Module 8 Parts Replacement 
    This section provides procedures and information for replacing spare 
    parts on the E9 systems. Always follow the standard ESD and safety 
    procedures when replacing parts or modules. Pay close attention to 
    the part number when replacing parts: 
     Most parts are labeled with the part number. 
     Always replace like for like or check compatibility. 
     Visually inspect all parts before installing.  
     Some parts are different depending on BT level and/or SW. 
     Some parts are only compatible with Logiq E9 or Vivid E9. 
    System Power Down 
    The system should be completely powered down before opening the 
    chassis and beginning any parts replacement: 
     Press the Power button on the Control Panel to power off the 
    system. 
     Select Shutdown from the touchscreen to shut down the system. 
    The system is not fully shut down until the light on the operator 
    panel turns off. 
     After the system has fully shut down switch the MAINS circuit 
    breaker on the back of the system to OFF. 
     Unplug the AC power cord from wall socket.  
    Chassis Cover Removal 
    This section describes how to open up the chassis for repairs and 
    maintanance. These should be removed in the following order: 
    1. Side Cover 
    2. Top Cover 
    Check the model number 
    and other vital 
    information before 
    ordering parts for the 
    system. 
    If the system is not fully 
    shut-down when you 
    switch off the mains 
    breaker loss of data on 
    the HD can occur. 
    Wait 30 seconds or so for 
    capacitors to discharge 
    before opening system. 
    If you need more detail 
    refer to the company 
    service manual.  
    						
    							 GE Logiq E9/Vivid E9 Training Manual 
    © 2016 Conquest Imaging 65 
    3. Foot Rest Bumper 
    4. Front Cover  
    5. Filter Cover  
    6. Rear Cover 
     
    Replace the covers in reverse order. 
    Side Covers 
    The side covers are fastened with two “click” locks, located at the 
    lower end of the cover. To remove the side covers: 
     
     Insert a #1 phillips screwdriver into the rectangular hole in the 
    side bumper until it reaches the lock mechanism. Push the 
    handle on the screwdriver downwards to release the lock. 
     Lift the cover backwards and up to remove it from the system. 
    Top Cover 
    To remove the top cover: 
     First remove the side covers.  
     Remove the two screws located near the handle and gently pull 
    back and up. 
     
    Foot Rest Bumper 
    The foot rest bumper is fixed with snap locks. To remove the foot rest 
    bumper: 
     Pull the foot rest bumper upwards and over the pedals to release 
    Side cover removal 
    procedure is easier if you 
    release the rear lock first. 
     
    You may need to push 
    down the pedals in order 
    to remove the foot rest 
    bumper.  
    						
    							 GE Logiq E9/Vivid E9 Training Manual 
    © 2016 Conquest Imaging 66 
    the snap locks, while freeing the side latches. 
     
    Front Cover 
    Follow these steps to remove the front cover: 
     Remove the other panels and the foot rest bumper. 
     Unscrew the two (2 pc) screws that fix the front cover assembly 
    to the chassis. 
     
     Pull the upper end of the front cover out and upwards to free it 
    from the pedals and the frame. 
    Back Filter Cover 
    To remove the filter cover: 
      Gently pull the filter cover out and away from the system. 
    To replace the Filter Cover:  
     Place into position and press the filter cover’s top corners until 
    the filter cover locks engage. 
    In the illustration the side 
    cover is removed to show 
    the Side Latch. It is not 
    necessary to remove the 
    side cover to remove the 
    foot rest bumper. 
    Inspect the Filter, and 
    clean if necessary.  
    						
    							 GE Logiq E9/Vivid E9 Training Manual 
    © 2016 Conquest Imaging 67 
     
    Rear Cover 
    There are four screws; two on each side of the cover. Remove the 
    screws from both sides of the rear cover and lift it away from the 
    chassis. 
     
    When replacing the rear 
    cover make sure the 
    manual release handle of 
    the Z Mechanism extends 
    through the TOP air vent 
    on the rear cover.  
    						
    							 GE Logiq E9/Vivid E9 Training Manual 
    © 2016 Conquest Imaging 68 
    Remove Bulkhead Cover 
    At the rear of the system, release the console’s frogleg mechanism by 
    inserting a screwdriver into the release point and pressing until 
    release. Pull the console out to its extended position to gain access to 
    the screws in the next step. 
     
    Use thumbs to press upper lock tabs toward the front of the system 
    to release the top of the Bulkhead Cover. 
     
    Pull the Bulkhead Cover away from the system. 
     
     
    Slots for Bulkhead 
    Cover 
    XY Release point  
    						
    							 GE Logiq E9/Vivid E9 Training Manual 
    © 2016 Conquest Imaging 69 
    Remove Handle, Left/Right Top Handle  
    Preparations: 
    1.) Power down the Vivid E9 and disconnect the Mains Power Cable. 
    2.) Remove the buttons below the Touch Panel. 
    3.) Remove the Upper Operator Panel. 
    4.) Remove the Lower Operator Panel. 
    Hatches are used to attach the top handles to the lower User 
    Interface. 
    Be careful when doing this procedure. It is easy to break the plastic 
    hatches  
    The screwdriver positions shown below Release Hatches indicate 
    where the hatches are located. 
    Release Hatches 
     
     Use a thin screwdriver to release the hatches, one by one as 
    illustrated, until you can remove the handle. 
    Handle Left Top / Handle Right Top reinstallation 
     Remove the Gel Cups (if present). 
     Install the Handle, Left Top / Handle, Right Top by pressing them 
    into the lower UI Frame and replace the Gel Cups. 
     
      
    						
    							 GE Logiq E9/Vivid E9 Training Manual 
    © 2016 Conquest Imaging 70 
    Palm Rest Assembly 
    Prepare: power down the Vivid E9 
    2.) Disconnect the Mains Power Cable from the wall outlet. 
    3.) Disconnect all probes and external I/O cabling. 
    4.) Remove the Upper Operator Panel. 
    5.) Remove the Lower Operator Panel. 
    6.) Remove Handle Left Top and Handle Right Top. 
    Palm Rest Assembly Removal 
    To remove the palm rest: 
     Release the hatch lock mechanisms securing the palm rest 
    assembly using the same method and cautions as with the 
    handles and remove the palm rest assembly. 
    To reinstall the palm rest assembly: 
     Ensure the XYZ Buttons cable runs along the lower tray cable 
    channel so that the cable is not pinched when the palm rest is 
    replaced. Reinstall the other components in the opposite order 
    from above. 
      
    						
    							 GE Logiq E9/Vivid E9 Training Manual 
    © 2016 Conquest Imaging 71 
    LCD Monitor Replacement 
    Prepare for replacement: 
     Move the top console to its lower locked position. 
     Power down the system. 
     Disconnect all probes and external I/O cabling. 
    Logiq E9 LCD Monitor Replacement 
    Logiq E9 systems come with two types of monitors: 
     Original LCD Monitor (P/N 5167953) 
     LCD Monitors used in R4 original consoles, LCD Monitor V2 
    (5392293-21, used in R4.x and later production)  
    LCD Monitor (P/N 5167953) Replacement 
     Tilt the LCD monitor forward to horizontal position. Unscrew the 
    fixing screw on the rear side of the LCD monitor assembly ad 
    remove the monitor cables cover. 
     Disconnect the cables to the monitor.  
     Remove the two clamps. 
     If the device is equipped with the WLAN option, remove the 
    WLAN dongle and transfer it to the new monitor. Lift the dongle, 
    slightly to disengage the Velcro, and pull the dongle out of the 
    USB port. Tilt the monitor back to vertical position. 
     Lock the monitor back into it’s vertical position. 
     Loosen the four screws by turning each screw between one half 
    and one turn counter-clockwise.  
     Lift the LCD Monitor assembly upwards until you can lift it away 
    from the Monitor Bracket. 
    Installing the new monitor: 
    To be sure the system is level check that: 
     The system is on a level surface. 
     The wheels are in-line. 
    Note: For Logiq monitor 
    replacement check if the 
    system uses WLAN. You 
    may need to remove the 
    dongle from beneath the 
    cover of the monitor and 
    replace it in the new 
    monitor.  
    If you are replacing any of 
    the cables, you will also 
    need to remove two 
    additional tie-wraps at the 
    bracket. 
    It is not necessary to 
    completely remove the 
    screws. 
    Ensure the LCD Monitor is 
    installed level with the 
    system before tightening 
    the four screws:  
    						
    							 GE Logiq E9/Vivid E9 Training Manual 
    © 2016 Conquest Imaging 72 
     The LCD monitor arm lock is LOCKED. 
     Install the LCD monitor assembly on the monitor bracket. Verify 
    that all four fixing screws have engaged in their slots. 
     Use a level if available on the monitor while you tighten the four 
    screws. 
     Tilt the monitor to horizontal position and connect the cables. If 
    the cables were replaced replace the tiewraps and cable clamps, 
    if present. 
     Reinstall WLAN dongle, if the system is equipped with the WLAN 
    option and replace the back cover. 
    LCD Monitor V2 (5392293-21) Replacement 
     Remove the two LCD cable cover fixing screws. 
     The LCD Cable is secured to the LCD with a retaining screw, an 
    EMC shielded retainer and two phillips screws. 
     Remove the screws and retainer. 
     Disconnect the cable from the LCD. 
     Keep the LCD monitor supported and remove the two M5x20 
    screws and spring “lock” washers which mount the LCD to the 
    arm assembly. 
    When installing the new monitor simply reverse the steps above and 
    keep the system level by ensuring that: 
     The system is on a level surface. 
     The wheels are in-line. 
     The LCD monitor arm lock is LOCKED. 
    Vivid E9 LCD Monitor Replacement 
    The backside cover of the LCD monitor assembly covers the two 
    cables to the monitor. To gain access to the cable connectors, first 
    remove the LCD rear cover: 
     Tilt the LCD monitor forward to it is in the horizontal position. 
    WARNING: When 
    removing the LCD from 
    the arm make sure the 
    arm assembly is in the 
    locked position! The 
    assembly can spring open 
    and cause damage and 
    potential injury.  
    						
    							 GE Logiq E9/Vivid E9 Training Manual 
    © 2016 Conquest Imaging 73 
     Remove the two fixing screws from the back side of the LCD 
    Monitor assembly and lift the cover.  
     Disconnect the two cables shown below: 
     
     Tilt the monitor back to its vertical position before removing the 
    screws. 
     Lift the LCD monitor assembly upwards away from the LCD arm 
    bracket. 
     Replace with the new LCD monitor. 
    Replacing the LCD Arm assembly 
    Vivid Preparations 
    To prepare for replacement: 
    1.) Power down the Vivid E9 
    Screws  
    						
    All GE manuals Comments (0)

    Related Manuals for GE Vivid E9 User Manual