    							 GE Logiq E9/Vivid E9 Training Manual 
    © 2016 Conquest Imaging 24 
    Raw Compression Enables compression of raw data images upon 
    storage and export.  
    Max Frame Rate Select 25, 30 or Full (original acquisition) from the 
    pop-up menu. 
    Compression Select compression type or no compression. 
    Quality Set picture quality from 1 to 100%. A low picture quality level 
    allows high data compression, while a high picture quality restrains 
    the compression. 
    Allow Multiframe : Allow cineloop storage. 
    Connection Settings Definition: 
    Retry Set maximum number of connection retries, time interval 
    between tentative and time-out. 
    DICOM Settings Definitions: 
    AE Title Application Entity Title  
    Port The Port no. is allocated during DICOM configuration. Refer to 
    your network specifications. 
    Verification Verify the connection to another DICOM application 
    Storage commitment Send a request to a PACS, asking it to 
    permanently archive image(s) 
    MPPS Modality Perform Procedure Step: send information (typically 
    to a HIS) that a scheduled exam has been started, performed or 
    interrupted. 
    DICOM SR settings:x  
     Allow SR: enable DICOM SR. 
     Allow SR private data: send the current exam data in a private 
    format. This option is by default unchecked and should only be 
    used with DICOM storage devices that can handle private data 
    format. 
     Signed Doppler velocities: send signed Doppler velocities. 
     Use older SR version: when checked a Use older SR version pull-
    down menu is displayed. The current exam data will be sent in the 
    same format as the selected SR version. Details about format and 
    content of the SR version can be found in the corresponding user 
    manual of the selected version 
    Raw compression is active 
    only if the setting Allow 
    raw data is checked.  
    						
    							 GE Logiq E9/Vivid E9 Training Manual 
    © 2016 Conquest Imaging 25 
    Additional Outputs 
    Use Additional outputs to configuration of the P1, P2 and Record 
    buttons on the Control panel. Several outputs such as Video Print, 
    Laser print, DICOM storage...etc. can be associated to these buttons 
    (i.e. pressing P1 can result in printing to a printer and storage to a 
    DICOM media). 
     
    P1 / P2 / Record button configuration 
    In Button field select P1, P2 or Record. 
    Select an output device in the Available output field and press the 
    Right arrow button to assign the device to the selected button. The 
    Properties window for the selected device is displayed, if the 
    selection is configurable. 
    Adjust the device specific parameters and select OK. 
    Adjust the image specific parameters (see table below). 
    Configuration parameters: 
    Format Select between: 
     Raw DICOM 
    1. Select between P1, P2 
    and Record buttons. 
    2. Available output 
    devices that can be 
    assigned to the current 
    button. 
    3. Output devices assigned 
    to the current button. 
    4. Add or remove selected 
    device to/from the current 
    button. 
    5. Adjust the device 
    settings of the selected 
    assigned device. 
    6. Select the type of 
    images to produce and 
    adjust image settings. 
    7. Printer configuration  
    						
    							 GE Logiq E9/Vivid E9 Training Manual 
    © 2016 Conquest Imaging 26 
     DICOM 
    Image compression Select compression mode from the pop-up menu. 
    Quality When JPEG compression is selected, adjust the picture quality 
    between 1 and 100%. A low picture quality level allows high data 
    compression, while a high picture quality restrains the compression. 
    Image frames Select between: 
     Single: stores single frame only 
     Multiple: stores cineloop 
     Secondary Capture: screen shot 
    Capture Area Select between: 
     Video Area 
     Entire Screen 
    Unlock Patient Record 
    If for any reason an examination is not properly finished, the patient 
    record will be locked and cannot be opened again until it is unlocked. 
    To unlock patient records: 
     Press Utility/Config on the Touch panel and select the category 
    Admin. Select the tab Unlock Patient. 
     Select the patient record(s) you want to unlock. 
     Select Unlock to unlock the selected patient record(s) or select 
    Unlock all to unlock all patient records and click OK in the 
    confirmation window. 
    Logiq E9 Connectivity 
    TCPIP: allows you to configure the Internet Protocol. 
    Device: allows you to set up devices. 
    Service: allows you to configure a service for example; DICOM 
    services such as printers, worklist, and other services such as video 
    print and standard print from the list of supported services.  
    You can search for a 
    specific patient record or a 
    group of patient record 
    using the search filters. 
    To set up connectivity for 
    your institution you must 
    login with administrator 
    privileges.  
    						
    							 GE Logiq E9/Vivid E9 Training Manual 
    © 2016 Conquest Imaging 27 
    Dataflow: allows you to adjust the settings of the selected dataflow 
    and associated services. Selecting a dataflow customizes the 
    ultrasound system to work according to the services associated with 
    the selected dataflow. 
    Button: allows you to assign a pre-configured output service (or set of 
    output services) to the Print keys on the control panel. 
    Removable Media: enables formatting (DICOM, database, or blank 
    formatting) and DICOM verification of removable media. 
    Miscellaneous: allows you to set up the patient exam menu options, 
    print and store options, and the order of the columns in the 
    examination list on the Patient menu. 
    Configure these screens from left to right, starting with the TCPIP tab 
    first. 
    TCPIP 
    This configuration category enables users with administrativerights to 
    set the TCPIP for the system and connected remote archive. 
     
     Type the name of the Ultrasound system in the Computer Name 
    field. 
     Type the IP-Address (acquire unique static IP address from 
    hospital network administrator), Subnet Mask, and Default 
    Gateway (if applicable). 
     Select Save settings and re-boot the ultrasound system. 
    TCPIP settings are not restored when restoring backups. This by 
    design to ensure that the IP address is unique. Reboot is required for 
    the changes to take effect. 
    Note: TCPIP settings do 
    not get restored when 
    restoring backups. This is 
    per system design. The 
    LOGIQ E9 IP address 
    MUST BE unique. 
    Note: Do not set up the 
    system with DHCP. The IP 
    address MUST BE static 
    for the diagnostic and 
    DICOM to function 
    correctly.  
    						
    							 GE Logiq E9/Vivid E9 Training Manual 
    © 2016 Conquest Imaging 28 
    Device 
    To add a new device: Press Add and type the device name in the 
    Name field. Type the device’s =P address in the IP Address field. 
      
    Preset Parameter Descriptions 
    Add - add a device. Remove - delete a device. 
    Ping: Use Ping to confirm that the device is connected. 
    Properties: Name Type the name of the device. 
    Properties: IP Address Type the device’s =P address. 
    Properties: AE Title AE Title of the LOGIQ E9. NOTE: Only available 
    forMyComputer. 
    Properties: Port Number IP Port Number Used for DICOM, set by 
    default to 104. NOTE: Only available for MyComputer. 
    Properties: MAC Address Unique network card address. NOTE: Only 
    available for MyComputer. 
    Service 
    For each Device that you have added to the system, you need to set 
    up the service(s) that the device supports. 
    Connectivity Services Preset Menu, My Computer 
     
    You must be an 
    administrator to update 
    these screens.  
    						
    							 GE Logiq E9/Vivid E9 Training Manual 
    © 2016 Conquest Imaging 29 
    Connectivity Services Preset Menu, New Device 
     
    The Services screen has the following sections of information: 
    Destination Device - lists information about destination devices. You 
    can select from a list of currently existing devices. 
    Service Type to Add - lists information about services for the 
    destination device. You can add services, select from a list of currently 
    existing services, and remove services. 
    Service Properties - lists properties for the service currently selected 
    in the Services section. The name and properties in this section 
    change, depending on what service is currently selected. 
     
    To add a service to a destination device:  
     Select the service from the pull-down menu and Add. Specify the 
    properties for this service and Save. Verify (ping) the service. 
    To remove a service:  
     Select the service.and press Remove and Save. 
    Basic DICOM Preset Parameter Descriptions 
    Name - Free text: give a descriptive name to the device. 
    AE Title - The Application Entity Title for the service. 
    Port Number - The port number of the service. 
    Maximum Retries Max # – the maximum number of times to try 
    establishing a connection to the service. 
    Retry Interval (sec) - Specify how often (in seconds) the system 
    should try to establish a connection to the service. 
    Timeout - The amount of time after which the system will stop trying 
    to establish a connection to the service. 
    DICOM Image Storage 
    DICOM Image Storage allows the system to send or receive 
    ultrasound images in a format that can be interpreted by PACS.  
    						
    							 GE Logiq E9/Vivid E9 Training Manual 
    © 2016 Conquest Imaging 30 
      
     
    DICOM Image Storage Preset Parameter Descriptions: 
    Allow Multiframe Select to allow cine loop storage. 
    Allow raw data Select to save data in both TruAccess (raw data) and 
    DICOM format. Clear to save in DICOM format only. 
    Compression Select the compression type: None, Rle, Jpeg, or 
    Jpeg2000. 
    Max Framerate Select the maximum frame rate: Full, 25, or 30. 
    Color Support Select: Mixed, Gray or Color 
    Reopen per image When selected, the storage protocol uses a new 
    association for each image sent. 
    Enable Structured Reporting Select for Structured Reporting. 
    Key Image Notes Image deletion notification. ONLY available for the 
    Direct Store Workflow and ONLY generated when there are images 
    deleted during the exam. Selecting this lets the reader at the PACS 
    system know which images have been deleted. An indicator is placed 
    on deleted images with a reason, “Rejected for Quality Reasons,” for 
    example. 
    DICOM Performed Procedure 
    DICOM Performed Procedure provides an acknowledgement that a 
    study has been performed. 
    DICOM Print 
    DICOM Print provides the ability to send or receive ultrasound image 
    data to DICOM printers.  
    						
    							 GE Logiq E9/Vivid E9 Training Manual 
    © 2016 Conquest Imaging 31 
    DICOM Print Preset Parameter Descriptions: 
    Vendor Specify PACS Vendor. This enables annotation on patient 
    information since by definition patient information is missing on the 
    title bar. 
    Format Indicates how many prints to print per page, for example, 1,1, 
    1,2, 1,3, up to 7,5. Partial prints are displayed as one print job. 
    Priority Specify the print job priority: High, Medium, or Low. 
    Medium Select the print medium: Clear Film, Paper, or Blue Film. 
    Copies Enter the number of copies. 
    Orientation Specify whether to print the image Portrait (vertically) or 
    Landscape (horizontally). 
    Film Size Specify the dimensions of the film size. 
    Film Destination Specify the film destination for the exposed film 
     Magazine – Store in a film magazine 
     Processor – Develop in a film processor 
    Magnification Specify how the printer magnifies the image to fit it 
    onto the film. 
     Replicate – Interpolated pixel are copies of the adjacent 
    pixels 
     Bilinear – Interpolated pixels are created by bilinear 
    interpolations between the adjacent pixels 
     Cubic – Interpolated pixels are created by cubic 
    interpolations between the adjacent pixels 
     None – No interpolation 
    Smoothing Type Specify the printer’s magnification interpolation for 
    the output. 
    Trim Specify whether you want a trim box to be printed around each 
    image on the film: Yes or No. 
    Min Density Enter a number indicating the minimum density level of 
    the film. 
    Max Density Enter a number indicating the maximum density level of 
    the film. 
    Border Select to have the border area surrounding and between the 
    images of the film: Black or White. 
    Empty Image Select to have a Black or White empty image. 
    Color Select whether to have the image Color or Grey. 
    Configuration Information Enter vendor-specific image quality 
    settings. 
    Film Session Label Type a name for the group of film labels associated 
    with the print job. 
    Annotation Enable Lets you annotate the image.  
    						
    							 GE Logiq E9/Vivid E9 Training Manual 
    © 2016 Conquest Imaging 32 
    DICOM Query/Retrieve 
    DICOM Query/Retrieve provides a list of patients sorted by 
    query parameters. 
     
     
     
    If you have problems with slow responses from DICOM servers, you 
    can try increasing the time-out in the DICOM server properties dialog: 
    Utility -> Connectivity -> Service -> Properties-> Maximum Retries 
    and Timeout 
     Slow response time may result in images being re-sent automatically 
    and low transfer rates. The retry settings can be used to make jobs 
    retry on bad networks. 
     
    When portable (off-line), use the minimum time-out and no retries as 
    it will affect shutdown speed. 
     
    DICOM Query/Retrieve Preset Parameter Descriptions: 
    Maximum Results Specify the maximum number of patient records 
    the system will retrieve when searching the patient database, up to 
    300, with the default set at 100. 
    Search Criteria Displays the Search Criteria window, where you can 
    enter search parameters for the system to use when searching the 
    patient database. 
    DICOM Query/Retrieve Search Criteria 
    Select Search Criteria to Add Select the type of information that you 
    want to define for search parameters. The following searches are 
    Note: Some PACS vendors 
    only offer Query/Retrieve 
    as an option. You should 
    confirm that this service is 
    supported.  
    						
    							 GE Logiq E9/Vivid E9 Training Manual 
    © 2016 Conquest Imaging 33 
    allowed: Patient Name, Patient ID, Modality, Scheduled Procedure 
    Start Date, Start Time, and End Time. 
    Tags (at least one) The name of a tag selected to use for search 
    criteria. 
    Properties: Value Type the value of the Selected Tag item. For 
    example, if you select Referring Physician’s Name in the Select Tag 
    field, youcan enter the name of the physician in the Value field. 
    Properties: Don’t Use Select to turn off the selected search criteria. 
    To exclude a tag from the worklist query, select Don’t Use and then 
    select Add to List. 
    Add Select to add the tag and value to the list of search criteria. 
    Remove Select to remove the tag and value from the list of search 
    criteria. 
    Clear Clears all tags. 
    DICOM Storage Commitment 
    DICOM Storage Commitment provides acknowledgement from the 
    PACS server that the study has been accepted into archive. 
     
    DICOM Storage Commitment Preset Parameter Descriptions: 
    Associated Storage This selection is based on the services entered by 
    the user. 
     
    DICOM Worklist 
    DICOM Worklist provides a list of patients sorted by query 
    parameters.  
    						
    All GE manuals Comments (0)