Service Menu 51 Document no. M1031519-01 Table 3 Module hardware status Module keyModule keyboard status. See Table 4 Module keyboard status for a detailed description of the message. Table 4 Module keyboard status Error Statuses (HEX): General errorModule general error status. See Table 5 Module general error status for a detailed description of the message. Table 5 Module general error status Pneuma errorModule pneumatics error status. See Table 6 Module...
Page 172
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors 52 Document no. M1031519-01 Table 6 Module pneumatics error status HW errorModule hardware error status. See Table 7 Module hardware error status for a detailed description of the message. Bit Error label Description What to do 0 pumping error Stepper motor is not able to generate negative pressure at the pump pressure sensor during zeroing.Check the...
Page 173
Service Menu 53 Document no. M1031519-01 Table 7 Module hardware error status TestbitTestbit is a module production phase test status and is always 0000 (HEX). If the testbit status is ≠ 0000, then the module should be returned to the factory. Voltages Vdd; 5 V digital units driving voltage received from the module frame. 5vdc; 5 Volts of the module derived internally from 15vdc. 15vdc; 15 Volts of the module received from the module frame. Pneuma voltage; 15 V...
Page 174
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors 54 Document no. M1031519-01 4.12.1 Tonometry This is another route to reach Tonometry setup menu, Others - Tonometry. For a detailed description on how to use this menu see “User’s Reference Manual”. How to calibrate PgCO2 1. Connect the calibration gas sampling line to the regulator and to the module’s catheter connector. 2. Select PgCO2 Calibration. 3....
Page 175
Service Menu 55 Document no. M1031519-01 4.12.2 Press Sensor Start Calib.: A selection to start a module pressure sensor calibration sequence. The procedure is guided online. Record Data: A selection to print out the shown service data to the device defined in the ...Service - Record Data menu. TONO service data Amb Press; ambient pressure measured by the pressure sensor. All the pressure values are given in both mmHg and mbar on this page. Pump Press;...
Page 176
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors 56 Document no. M1031519-01 4.12.3 System Test Table 8 Description of system test Test label Description If test fails Pump unit function Stepper motor aspirates/pushes room air through the zeroing valve. If the stepper is working properly, it is able to generate small negative/positive pressure at the pump unit pressure sensor. Check the function of...
Page 177
Service Menu 57 Document no. M1031519-01 Cath conn. leak test The air tightness of the zeroing unit and catheter connector is tested.Make sure that the catheter connector is properly closed during the test. Check the zeroing unit and the catheter connector for leaks. Block test Air is aspirated through the zeroing valve. If pressure at the pump pressure sensor drops too much, the tubing/valves from the zeroing valve to the pump pressure sensor is...
Page 178
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors 58 Document no. M1031519-01 4.12.4 CO2 Sensor Lamp ON/OFF: A selection to toggle the sensor lamp on and off. Meas Signal: A selection to start an automatic signal measurement sequence. This sequence measures sensor lamp signal levels. Meas Zero: A selection to start an automatic zeroing measurement sequence. Meas Gas: A selection to start an automatic gas...
Page 179
Service Menu 59 Document no. M1031519-01 4.12.5 Pneumatics Zero valve: A selection to toggle the zero valve on/off. Press valve: A selection to toggle the pump valve on/off. Record Data: A selection to print out the shown service data to the device defined in the ...Service - Record Data menu. TONO Cathpres; pressure measured from the catheter line [mbar]. Ambpres; ambient pressure in the room [mbar]. Pumppres; pressure measured from the pumping unit [mbar]....
Page 180
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors 60 Document no. M1031519-01 4.12.6 AD Record Data: A selection to print out the shown service data to the device defined in the ...Service - Record Data menu. TONO All the numbers in this section are AD counts, i.e. computer internal data. EXTERNAL ADC CO2; CO 2 sensor CO2 signal. temp; CO 2 sensor temperature signal. cath press; pressure measured from the...