Page 161

Service Menu
41
                                          Document no. M1031519-01
4.8 NMT
Service menu for Neuromuscular Transmission Module, 
E-NMT or M-NMT.
Start Curr. test is a test where the module itself checks 
the difference between a given current and the measured 
current. A 3 k
Ω resistance should be connected between 
the stimulus electrodes before starting the test. All the 
currents checked will be displayed on the service data 
screen. After the current value, the status of the test is...

Page 162

Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
42
Document no. M1031519-01                                                                           
Timeouts is a cumulative number that indicates how many times the module has not responded to the monitors 
inquiry.
Bad checksums is a cumulative number that indicates how many times communication from the module to the 
monitor has failed. 
Bad c-s by mod is a cumulative number that indicates how many communication errors the...

Page 163

Service Menu
43
                                          Document no. M1031519-01
4.8.1 NMT Setup
Start-up is for starting the neuromuscular transmission 
measurement.
Stop/Continue is for stopping/continuing the 
measurement.
Stimulus Mode is for changing the stimulus mode. The 
choices are: TOF - Train of Four, DBS -Double Burst 
Stimulation or ST - Single Twitch mode.
Cycle Time is for changing the time interval for NMT 
measurement. The choices are: 
for TOF and DBS: Manual and from 10 sec to 15...

Page 164

Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
44
Document no. M1031519-01                                                                           
4.9 Oxygen Saturation Modules, E-NSAT, M-NSAT, M-OSAT 
N/OSAT Data
NOTE: Preamp Error is indicated on the service menu, 
even though module contains MP-204 board. 
NOTE: The status and error indicators are not valid for the 
M-OSAT, E-NSAT and the M-NSAT..04 revision.
PR shows the pulse rate value [bpm] calculated from the 
pleth....

Page 165

Service Menu
45
                                          Document no. M1031519-01
4.10 More 

Page 166

Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
46
Document no. M1031519-01                                                                           
4.11 Interface module, E-INT, M-INT 
Gases, SpO2, NIBP, Spirometry, SvO2/C.O. indicate the 
parameters for which service data is available. The data 
which can be seen on those pages is raw data from the 
interfaced monitors, which will be processed for the 
normal screen.
Service Data
I-INT:  Indicates the status of the interface...

Page 167

Service Menu
47
                                          Document no. M1031519-01
4.11.1 Gases
4.11.2 Sp02
The data on this page is the raw data received from 
the interfaced monitor. This data will be processed 
further for the normal screen.
The data on this page is the raw data received 
from the interfaced monitor. This data will be 
processed further for the normal screen. 

Page 168

Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
48
Document no. M1031519-01                                                                           
4.11.3 NIBP
4.11.4 Spirometry
The data on this page is the raw data received from 
the interfaced monitor. This data will be processed 
further for the normal screen.
The data on this page is the raw data received from 
the interfaced monitor. This data will be processed 
further for the normal screen. 

Page 169

Service Menu
49
                                          Document no. M1031519-01
4.11.5 Sv02/C.O.
The data on this page is the raw data received from 
the interfaced monitor. This data will be processed 
further for the normal screen. 

Page 170

Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
50
Document no. M1031519-01                                                                           
4.12 Tonometry
Table 2 Module general status
HWModule hardware status. See Table 3 Module hardware status for a detailed description of the 
message.
bit 0-5 Not used
bit 6 State error
bit 7 Communication failure
bit 8 Power failure
bit 9 Clock failure
bit 10 EEPROM checksum failure
bit 11 EEPROM writing failure
bit 12 ROM failure...