Page 131

Service Menu
11
                                          Document no. M1031519-01
1.3.2 Communication
Interface status indicates the status of the data link 
between the CPU and MEM sections. If the MemCard 
option is installed, the status should always be on 
ACTIVE. If the status blinks between ACTIVE and CLOSED, 
a communications error has occurred.
Message types indicates the type of data packets that 
have been sent (Tx) and received (Rx) since last monitor 
start. Data types are listed on the...

Page 132

Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
12
Document no. M1031519-01                                                                           
1.4 Power supply
The menu shows the voltages and the temperature 
measured by the power management board (PMB 
Voltages) and the main CPU board in the Display 
unit(DU Voltages). The measurement starts about 
100 ms after a start-up. The values in the column 
under Mean are the mean values of the last one 
second, the Min column...

Page 133

Service Menu
13
                                          Document no. M1031519-01
1.4.1 Battery
1.5 PCMCIA Board
Select Battery: Selection of battery type.
Capacity: Selection of battery capacity. 
Previous Menu: Return to the previous menu.
Service Data
Type shows the type of the battery used. 
Capacity refers to battery capacity in Ah.
Left [%]: Evaluated capacity left in battery.
Temperature [°C] measured inside the battery.
Current [A] shows the charging state and charging 
current.
Voltage [V]...

Page 134

Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
14
Document no. M1031519-01                                                                           
2 Display
Geometry views the geometry of the display.
Colors views the color of the screen.
Readability views the readability of the screen. 

Page 135

Service Menu
15
                                          Document no. M1031519-01
3 Keyboard
Keyboard
The service menu for testing the command board 
functions.
Upper Led is for testing the upper alarm LED (red) on the 
command board. When the text is highlighted, the upper 
alarm LED can be turned on and off by pressing the 
ComWheel.
Lower Led is for testing the lower alarm LED (yellow) on 
the command board. When the text is highlighted, the 
lower alarm LED can be turned on and off by pressing the...

Page 136

Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
16
Document no. M1031519-01                                                                           
3.1 Keyboard Log
3.2 Keyboard Type
Keyboard Scroll Log
All the keyboard presses and the commands given 
by the ComWheel are recorded in the Keyboard 
Log. The keyboard log is saved in the permanent 
memory of the monitor. The length of the log is 
1150 events. The log is FIFO type.
Scroll Stat enables to scroll the keyboard events....

Page 137

Service Menu
17
                                          Document no. M1031519-01
4 Parameters
NOTE: Parameter values in Service Data fields are only for reference in this section.
4.1 Gas Unit
Service menu for airway gas modules’ communication 
and airway gas and spirometry measurements. 

Page 138

Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
18
Document no. M1031519-01                                                                           
4.1.1 General
Service Data
Module configuration shows which measurement 
options are available, i.e. are detected by the module.
Timeouts is a cumulative number that indicates how 
many times the module has not responded to the 
monitors inquiry. 
Bad checksums is a cumulative number that indicates 
how many times communication from...

Page 139

Service Menu
19
                                          Document no. M1031519-01
4.1.2 Gases
Noise Measactivates the noise measurement.
Sample gain adj adjusts the sampling pump gain, i.e. 
for adjusting the sample flow 
measurement.
Fall time Meas *activates the fall time measurement. 
Pump ctrlA manual control for the sampling 
pump.
Zero valve ctrlA manual control for the zero valve.
Occl valve ctrl*A manual control for the occlusion 
valve.
Record Dataprints out the shown service data and 
board...

Page 140

Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
20
Document no. M1031519-01                                                                           
4.1.3 Spirometry
Insp and exp flow gains can be adjusted, if calibration is 
needed. A calibration pump or spirometry tester is used 
and readings are observed from the display. If a deviation 
exists, gains are adjusted accordingly. Gain scaling is 
1.000, when the factory settings are in effect (1.050 in 
modules that contain the...