GE S5 Service Manual
Here you can view all the pages of manual GE S5 Service Manual. The GE manuals for Medical Equipment are available online for free. You can easily download all the documents as PDF.
Page 131
Service Menu 11 Document no. M1031519-01 1.3.2 Communication Interface status indicates the status of the data link between the CPU and MEM sections. If the MemCard option is installed, the status should always be on ACTIVE. If the status blinks between ACTIVE and CLOSED, a communications error has occurred. Message types indicates the type of data packets that have been sent (Tx) and received (Rx) since last monitor start. Data types are listed on the...
Page 132
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors 12 Document no. M1031519-01 1.4 Power supply The menu shows the voltages and the temperature measured by the power management board (PMB Voltages) and the main CPU board in the Display unit(DU Voltages). The measurement starts about 100 ms after a start-up. The values in the column under Mean are the mean values of the last one second, the Min column...
Page 133
Service Menu 13 Document no. M1031519-01 1.4.1 Battery 1.5 PCMCIA Board Select Battery: Selection of battery type. Capacity: Selection of battery capacity. Previous Menu: Return to the previous menu. Service Data Type shows the type of the battery used. Capacity refers to battery capacity in Ah. Left [%]: Evaluated capacity left in battery. Temperature [°C] measured inside the battery. Current [A] shows the charging state and charging current. Voltage [V]...
Page 134
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors 14 Document no. M1031519-01 2 Display Geometry views the geometry of the display. Colors views the color of the screen. Readability views the readability of the screen.
Page 135
Service Menu 15 Document no. M1031519-01 3 Keyboard Keyboard The service menu for testing the command board functions. Upper Led is for testing the upper alarm LED (red) on the command board. When the text is highlighted, the upper alarm LED can be turned on and off by pressing the ComWheel. Lower Led is for testing the lower alarm LED (yellow) on the command board. When the text is highlighted, the lower alarm LED can be turned on and off by pressing the...
Page 136
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors 16 Document no. M1031519-01 3.1 Keyboard Log 3.2 Keyboard Type Keyboard Scroll Log All the keyboard presses and the commands given by the ComWheel are recorded in the Keyboard Log. The keyboard log is saved in the permanent memory of the monitor. The length of the log is 1150 events. The log is FIFO type. Scroll Stat enables to scroll the keyboard events....
Page 137
Service Menu 17 Document no. M1031519-01 4 Parameters NOTE: Parameter values in Service Data fields are only for reference in this section. 4.1 Gas Unit Service menu for airway gas modules’ communication and airway gas and spirometry measurements.
Page 138
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors 18 Document no. M1031519-01 4.1.1 General Service Data Module configuration shows which measurement options are available, i.e. are detected by the module. Timeouts is a cumulative number that indicates how many times the module has not responded to the monitors inquiry. Bad checksums is a cumulative number that indicates how many times communication from...
Page 139
Service Menu 19 Document no. M1031519-01 4.1.2 Gases Noise Measactivates the noise measurement. Sample gain adj adjusts the sampling pump gain, i.e. for adjusting the sample flow measurement. Fall time Meas *activates the fall time measurement. Pump ctrlA manual control for the sampling pump. Zero valve ctrlA manual control for the zero valve. Occl valve ctrl*A manual control for the occlusion valve. Record Dataprints out the shown service data and board...
Page 140
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors 20 Document no. M1031519-01 4.1.3 Spirometry Insp and exp flow gains can be adjusted, if calibration is needed. A calibration pump or spirometry tester is used and readings are observed from the display. If a deviation exists, gains are adjusted accordingly. Gain scaling is 1.000, when the factory settings are in effect (1.050 in modules that contain the...