Page 141
Service Menu 21 Document no. M1031519-01 Service Data Sensor ADULT/PEDIATRIC according to the selected measurement mode (sensor). Insp/Exp Flow Gains shown apply to the selected sensor. Aw Press Zero and Flow Zero are the result of zeroing in the user service menu. They can be adjusted in the factory calibration menu, but not permanently stored. Aw Press Gain is directly the value used in sw. It can be adjusted, but not permanently stored. Exp Flow and Insp...
Page 142
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors 22 Document no. M1031519-01 4.2 ECG Module Power freq: Set power frequency; 50 Hz/60 Hz. Filter low : Set filter low frequency; 0.05 Hz/0.5 Hz. Filter high: Set filter high frequency; 30 Hz (40 Hz if power freq is 60 Hz) /100 Hz or 150 Hz @ NE12STPR. Service Data Power freq, and Cable type show the values chosen or detected, Filter low and high defines the...
Page 143
Service Menu 23 Document no. M1031519-01 4.2.1 ECG Setup Filter filters the ECG signal high frequency noise and slow respiratory artefacts. Monit (monitor) filter is used in routine monitoring. It effectively filters the artefacts caused by the electrosurgery unit and respiration. Diagn (diagnostic) filter is used if more accurate information of the waveform is needed (e.g. of P-wave or AV block). The diagnostic filter is more susceptible both to high...
Page 144
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors 24 Document no. M1031519-01 4.3 ESTP Module Service menu for the multiparameter hemodynamic module’s oxygen saturation, temperature and invasive pressure measurements. NOTE: Pressure Module, M-P or E-P, and Pressure Temperature Module, M-PT or E-PT, have their own service menu; ”4.4 P/PT.” and so does Dual Pressure Module, M-PP or E-PP; ”4.5 PP.”...
Page 145
Service Menu 25 Document no. M1031519-01 Balance between leds is adjusted by changing the intensity of red/infrared. Intensity of infrared (Ired int.) is in the range of 40 to 255 and red intensity (red int.) is in the range of 40 to 255. DC gain shows the gain of DC signal adjusted by the module. IDC is the value of infrared signal. RDC is the dc value of red signal. AC gain is the gain of infrared and red ac signals. AC gain values can be 1 or 0. Value 1 means...
Page 146
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors 26 Document no. M1031519-01 4.3.1 Calibrations How to calibrate T1/ T2 The calibrations are possible only when the protection is set OFF. The temperature calibration requires accurate test plugs of value 25 °C and 45 °C. 1. Select Calibrate T1/Calibrate T2. 2. Insert the test plug 25 °C into the T1/T2 connector. 3. Press the ComWheel. 4. Insert the test plug...
Page 147
Service Menu 27 Document no. M1031519-01 4.4 P/PT Service menu for Pressure Module, M-P or E-P, and Pressure Temperature Module, M-PT or E-PT. Calibrations: see section “P/PT Calibrations”. Temp Test activates the automatic temperature test for the temperature channels T3 and T4. The result from the test is shown in the service data field. NOTE: The Temp Test needs to be selected twice before the test starts. Record Data: Record Data prints out the shown...
Page 148
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors 28 Document no. M1031519-01 RAM indicates the state of the RAM memory. ROM indicates whether the checksum at the EPROM is in accordance with the one the software has calculated. EEPROM indicates if the values stored in the permanent memory are valid. The state is either OK, Fail or ? (module not in place or a communication error). 4.4.1 P/PT Calibrations How...
Page 149
Service Menu 29 Document no. M1031519-01 4.5 PP Service menu for Dual Invasive Pressure Module, M-PP or E-PP. Calibrations: see section ”PP Calibrations.” Record Data prints out the service data and circuit board information (id., serial number, and software id.) on the recorder. Service Data Gain is a coefficient to compensate for gain error. Usually the values for P5 and P6 are between 17000 and 25000. Zero indicates the offset compensation value for...
Page 150
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors 30 Document no. M1031519-01 4.5.1 PP Calibrations How to calibrate The calibrations require a pressure transducer (with an appropriate cable) and a pressure manometer. 1. Connect the pressure transducer with the pressure manometer to the P5 / P6 connector. Select Calibrate P5 / Calibrate P6. Leave the transducer at room air pressure. 2. Push the ComWheel to...