Page 91

Planned maintenance instructions
5
                                          Document no. M1031518-02
1.4 Planned maintenance parts
1.4.1 PM parts for Compact Airway Modules, E-CXXXXX/ M-CXXXXX without CO2 absorber
 - Anesthesia
NOTE: Corresponding PM kit is available with Order No. 8001758.
1.4.2 PM parts for Compact Airway Modules, E-CXXXXX/ M-CXXXXX without CO2 absorber
 - Critical Care
NOTE: Corresponding PM kit is available with Order No. 8001759. Part Order No. For product(s)
Special tube (Nafion)...

Page 92

Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
6
Document no. M1031518-02                                                                           
1.4.3 PM parts for Compact Airway Modules, E-CXXXXX/ M-CXXXXX with CO2 absorber
 - Anesthesia
NOTE: Corresponding PM kit is available with Order No. 8001760.
1.4.4 PM parts for Compact Airway Modules, E-CXXXXX/ M-CXXXXX with CO2 absorber
 - Critical Care
NOTE: Corresponding PM kit is available with Order No. 8001761. Part Order No....

Page 93

Planned maintenance instructions
7
                                          Document no. M1031518-02
1.4.5 PM parts for Airway Module, E-miniC/ M-miniC
For details on recommended accessories see the “Supplies and Accessories” catalog. Part Order No. For product(s)
Special tube (Nafion 2 pcs) 733382 All Airway modules
Mini D-fend O-ring (2 pcs) 656565 E-miniC/ M-miniC 
Mini D-fend 8002174 (pkg of 10 pcs) E-miniC/ M-miniC 
Zero valve air filter M1011471 E-miniC, every 3 years
Zero valve air filter 571421...

Page 94

Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
8
Document no. M1031518-02                                                                           
2 Planned maintenance check list
2.1 Visual inspection/preparation
2.1.1 General
NOTE:  Wear a static control wrist strap when handling PC boards. Electrostatic discharge may 
damage components on the board.
Make sure that the monitor is switched to standby.
Disconnect the mains power cord. If the monitor is connected to the...

Page 95

Planned maintenance check list
9
                                          Document no. M1031518-02
−Check the D-fend O-rings and replace them, if necessary.
−Replace the D-fend and sampling line.
−Clean or replace the fan filter.
NOTE: Use only approved sampling lines to ensure proper functioning.
Use a 2 m/7 ft. sampling line with Compact Airway Modules E-/ M-COVX and E-/ M-CAiOVX.
3. The manufacturer recommendation is to replace the CO
2 absorber every 4 years. Replace the 
absorber, if necessary....

Page 96

Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
10
Document no. M1031518-02                                                                           
−Clean the thermal printhead and the small glass window in front of the static brush with a 
cotton swab dipped in isopropyl alcohol, if necessary.
NOTE: Avoid contact with the rubber paper roller. Be careful to limit the application of alcohol to 
the thermal printhead and the window.
−Reinstall the paper roll.

2.1.6 Modules with...

Page 97

Planned maintenance check list
11
                                          Document no. M1031518-02
Recorder - Record Waveforms - Waveform 1 - ECG1
- Waveform 2 - P1
- Waveform 3 - P2
Invasive Pressures - P1 ‘ART’ Setup - Label - ART
- P2 ‘CVP’ Setup - Label - CVP
- P3 Setup - Label - PA
- P4 Setup - Label - P4
- P5 Setup - Label - P5
- P6 Setup - Label - P6
Pulse Oximetry - Pleth Scale - AUTO
or
Others - SPO2 Setup - Pleth Scale - AUTO
Airway Gas - Spirometry View - Scaling - Indep.
- Paw Scale - 20
-...

Page 98

Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
12
Document no. M1031518-02                                                                           
2.2.2 Display
1. Check that the picture on the screen is adjusted correctly.

2.2.3 Keyboard(s)
1. Tests with the Command Board:
−Press the 
Help key. Turn the ComWheel in both directions and check that the cursor in the 
menu moves correspondingly. Select Normal Screen and check that the menu 
disappears from the screen.
Check the...

Page 99

Planned maintenance check list
13
                                          Document no. M1031518-02
3. Wait until the message ‘Calibrating gas sensor’ disappears from the screen, then enter the 
Compact Airway Module General service menu.
Check that the module configuration displayed corresponds with the Compact Airway Module 
type being used.
4. Check that the Time-outs, Bad checksums and Bad c-s by mod values are not increasing faster 
than by 5 per second.
5. Enter the Gases service menu: Check that...

Page 100

Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
14
Document no. M1031518-02                                                                           
NOTE: When reassembling the module, make sure that the tubes are not pinched between the 
module box and internal parts.
Insert the Compact Airway Module into the frame. Switch the monitor back on and wait until the 
message ‘Calibrating gas sensor’ disappears from the screen.
13. Block the tip of the sampling line with your finger...