Page 91
Planned maintenance instructions 5 Document no. M1031518-02 1.4 Planned maintenance parts 1.4.1 PM parts for Compact Airway Modules, E-CXXXXX/ M-CXXXXX without CO2 absorber - Anesthesia NOTE: Corresponding PM kit is available with Order No. 8001758. 1.4.2 PM parts for Compact Airway Modules, E-CXXXXX/ M-CXXXXX without CO2 absorber - Critical Care NOTE: Corresponding PM kit is available with Order No. 8001759. Part Order No. For product(s) Special tube (Nafion)...
Page 92
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors 6 Document no. M1031518-02 1.4.3 PM parts for Compact Airway Modules, E-CXXXXX/ M-CXXXXX with CO2 absorber - Anesthesia NOTE: Corresponding PM kit is available with Order No. 8001760. 1.4.4 PM parts for Compact Airway Modules, E-CXXXXX/ M-CXXXXX with CO2 absorber - Critical Care NOTE: Corresponding PM kit is available with Order No. 8001761. Part Order No....
Page 93
Planned maintenance instructions 7 Document no. M1031518-02 1.4.5 PM parts for Airway Module, E-miniC/ M-miniC For details on recommended accessories see the “Supplies and Accessories” catalog. Part Order No. For product(s) Special tube (Nafion 2 pcs) 733382 All Airway modules Mini D-fend O-ring (2 pcs) 656565 E-miniC/ M-miniC Mini D-fend 8002174 (pkg of 10 pcs) E-miniC/ M-miniC Zero valve air filter M1011471 E-miniC, every 3 years Zero valve air filter 571421...
Page 94
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors 8 Document no. M1031518-02 2 Planned maintenance check list 2.1 Visual inspection/preparation 2.1.1 General NOTE: Wear a static control wrist strap when handling PC boards. Electrostatic discharge may damage components on the board. Make sure that the monitor is switched to standby. Disconnect the mains power cord. If the monitor is connected to the...
Page 95
Planned maintenance check list 9 Document no. M1031518-02 −Check the D-fend O-rings and replace them, if necessary. −Replace the D-fend and sampling line. −Clean or replace the fan filter. NOTE: Use only approved sampling lines to ensure proper functioning. Use a 2 m/7 ft. sampling line with Compact Airway Modules E-/ M-COVX and E-/ M-CAiOVX. 3. The manufacturer recommendation is to replace the CO 2 absorber every 4 years. Replace the absorber, if necessary....
Page 96
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors 10 Document no. M1031518-02 −Clean the thermal printhead and the small glass window in front of the static brush with a cotton swab dipped in isopropyl alcohol, if necessary. NOTE: Avoid contact with the rubber paper roller. Be careful to limit the application of alcohol to the thermal printhead and the window. −Reinstall the paper roll. 2.1.6 Modules with...
Page 97
Planned maintenance check list 11 Document no. M1031518-02 Recorder - Record Waveforms - Waveform 1 - ECG1 - Waveform 2 - P1 - Waveform 3 - P2 Invasive Pressures - P1 ‘ART’ Setup - Label - ART - P2 ‘CVP’ Setup - Label - CVP - P3 Setup - Label - PA - P4 Setup - Label - P4 - P5 Setup - Label - P5 - P6 Setup - Label - P6 Pulse Oximetry - Pleth Scale - AUTO or Others - SPO2 Setup - Pleth Scale - AUTO Airway Gas - Spirometry View - Scaling - Indep. - Paw Scale - 20 -...
Page 98
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors 12 Document no. M1031518-02 2.2.2 Display 1. Check that the picture on the screen is adjusted correctly. 2.2.3 Keyboard(s) 1. Tests with the Command Board: −Press the Help key. Turn the ComWheel in both directions and check that the cursor in the menu moves correspondingly. Select Normal Screen and check that the menu disappears from the screen. Check the...
Page 99
Planned maintenance check list 13 Document no. M1031518-02 3. Wait until the message ‘Calibrating gas sensor’ disappears from the screen, then enter the Compact Airway Module General service menu. Check that the module configuration displayed corresponds with the Compact Airway Module type being used. 4. Check that the Time-outs, Bad checksums and Bad c-s by mod values are not increasing faster than by 5 per second. 5. Enter the Gases service menu: Check that...
Page 100
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors 14 Document no. M1031518-02 NOTE: When reassembling the module, make sure that the tubes are not pinched between the module box and internal parts. Insert the Compact Airway Module into the frame. Switch the monitor back on and wait until the message ‘Calibrating gas sensor’ disappears from the screen. 13. Block the tip of the sampling line with your finger...