3.7.2 Troubleshooting 
If a problem occurs during a functional examination, check the components of the monitor according 
to the following troubleshooting chart. If the problem persists, please refer to Part II of this Technical 
Reference Manual.
Trouble Treatment
Nothing functions. Unplug and re-plug Remote Controller Cable. Also confirm that the cable is intact.
Unplug and re-plug the Power Cord. Also confirm...

Page 42

Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
30
Document no. M1031517-02                                                                           
For your notes: 

Page 43

Interfacing
31
                                          Document no. M1031517-02
4 Interfacing
External devices can be interfaced with the S/5 Compact Anesthesia and S/5 Compact Critical Care 
Monitors via the Compact Monitor’s serial port, via the Interface Module, E-INT / M-INT, and via the 
Device Interfacing Solution, N-DISxxx.
 The serial port can be used for interface with:
•S/5 Anesthesia Delivery Unit
•Printers
•Computers
An E-INT / M-INT can be used for interface with:
•Datex-Ohmeda monitors...

Page 44

Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
32
Document no. M1031517-02                                                                           
Table 1 Transference of parameters, Datex-Ohmeda monitors
Table 2 Transference of parameters, external monitors
4.1.1 Connection to external Datex-Ohmeda monitors
Use the INT-External Device Cable.
1. Make sure that the power to both monitors is turned off.
2. Connect the 9 pin D-connector to one of the connectors on the Interface...

Page 45

Interfacing
33
                                          Document no. M1031517-02
4.1.2 Connection to Critikon Dinamap 1846SX, Abbott Oximetrix 3 and Baxter Explorer
Use the INT-External Device Cable.
1. Make sure that the power to both monitors is turned off.
2. Connect the 9 pin D-connector to one of the connectors on the Interface Module, E-INT / M-INT. 
Tighten the finger screws.
3. Connect the 25 pin D-connector to the connector on the external monitor. Tighten the finger 
screws.
4.1.3 Connection...

Page 46

Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
34
Document no. M1031517-02                                                                           
4.2 Interfacing external bedside devices via S/5 Device Interfacing 
Solutions, N-DISxxx
The Device Interfacing Solution, N-DISxxx provides means for transferring physiological, waveform 
and event data from various bedside patient care devices to the Datex-Ohmeda monitoring system. 
The real-time and trended data can be displayed on...

Page 47

Interfacing
35
                                          Document no. M1031517-02
For specific information on parameters transferred from the interfaced device to the Datex-Ohmeda 
monitor and the applicable software versions of the device refer to the Installation guide 
accompanying each DIS module. 
4.2.1 Device Interfacing Solution components
The Device Interfacing Solution consists of:
−a device specific interfacing module
−a device specific cable
−a bus cable
−a connector for another bus cable...

Page 48

Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
36
Document no. M1031517-02                                                                           
Figure 13 Connection cables and LED indicators
WARNINGThe Compact Monitor, interfacing modules and interfaced devices must be 
situated in the same patient environment (as defined in IEC 60601-1-1).
WARNINGConnecting electrical equipment together or using the same extension cord for 
more than one device may cause their leakage...

Page 49

Interfacing
37
                                          Document no. M1031517-02
For mounting accessories, please refer to the “Supplies and Accessories” catalog. See the figure 
below for an example of a device interfacing.           
Figure 14 An example of interfacing external devices with Device Interfacing Solution
(1)  Datex-Ohmeda S/5 Compact Anesthesia Monitor (with software L-CANE02(A) or later)
(2) Aestiva/5 anesthesia machine
(3) RGM monitor
(4) Interfacing module
NOTE:  You can connect up to...

Page 50

Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
38
Document no. M1031517-02                                                                           
4.2.5 Functional check
There are two ways to check the function of the S/5 Device Interfacing Solution:
1. Press the 
Monitor Setup key.
2. Select Interfacing and open the Status Page menu. The status page shows you the current 
communication status of the interfacing module (1 - 10). 
NOTE: The status message ‘Connected’ appears on...