GE S5 Service Manual
Here you can view all the pages of manual GE S5 Service Manual. The GE manuals for Medical Equipment are available online for free. You can easily download all the documents as PDF.
Page 31
System installation 19 Document no. M1031517-02 Figure 6 External connections of Compact Monitor frame F-CM(C)1 rev.00...03 (1) Dust filters (2) Recorder unit (optional) (3) Connector for factory use only (4) Synchronization connector (5) Connector for external keyboard (anesthesia record keeping solution keyboard), K-CREMCO or Barcode Reader (6) Potential equalization connector (7) Serial port connector (8) Receptacle for power cord (9) NET ID connector...
Page 32
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors 20 Document no. M1031517-02 3.3.2 Connecting to mains Connect the power cord to the mains power inlet on the side of the power supply unit and to the wall socket. NOTE: Before taking the monitor into use for the first time, the batteries should be fully charged. Keep the monitor connected to the mains until the Battery charging symbol disappears, or in STBY...
Page 33
System installation 21 Document no. M1031517-02 3.3.5 Inserting the parameter modules Insert each plug-in parameter module into a module slot. Firmly press the module in position. NOTE: Ensure that the module is properly oriented (i.e. the module release latch is facing downward). Do not use two or more modules with identical functions in the monitor. Modules with identical functions are: •Hemodynamic multiparameter modules, E-PRESTN, E-PRETN, E-RESTN, M-PRESTN,...
Page 34
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors 22 Document no. M1031517-02 10. Set the monitors network communication according to the network software in use, if necessary. Enter the service menu: Monitor Setup -Install/Service (16-4-34) - Service (26-23-8)- Frame - Network- Config and set •Network software S-CNET99 -> DRI Level = 1999 •Network software S-CNET01 -> DRI Level = 2001 •Network software...
Page 35
System installation 23 Document no. M1031517-02 5. Wait until the normal screen appears and check that a message Field upgrade OK is displayed as a note of successful upgrade. 6. Remove the option software card. 7. Enter the service menu and make sure that both options were loaded (Monitor Setup -Install/Service - Service - Frame - Options - Net+DC). 8. Install the device plate for option software on inside the cover for card drive slots. Close the cover. 9....
Page 36
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors 24 Document no. M1031517-02 4. Wait until the normal screen appears and check that a message Field Upgrade OK is displayed as a note of successful upgrade. 5. Remove the option software card. 6. Switch the monitor to standby. 7. Detach the cover for card drive slots and install the antenna cover. 8. Insert the wireless LAN antenna card into the foremost card...
Page 37
System installation 25 Document no. M1031517-02 3.4 S/5 Remote Controller, K-CREMCO To connect S/5 Remote Controller, K-CREMCO, to an S/5 Compact Monitor: 1. Make sure that the monitor is switched to standby. 2. Connect the K-CREMCO to the 5-pin DIN connector X9 that is located on the right side of the Compact Monitor display unit. 3.5 Anesthesia Record Keeping Keyboard, K-ARKB To connect Anesthesia Record Keeping Keyboard, K-ARKB, to an S/5 Compact Monitor: 1....
Page 38
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors 26 Document no. M1031517-02 3.7 S/5 Compact Airway Modules, E-xxxx / M-xxxxx This chapter provides information for installing Compact Airway Modules E-xxxxx. Figure 9 Compact Airway Module, E-XXXX Connection to frame 1. Ensure that the module is properly orientated (i.e. module release latch facing downward). 2. Align the module insertion guide slot with the...
Page 39
System installation 27 Document no. M1031517-02 Scavenging through the ventilator reservoir Figure 10 Scavenging through ventilator reservoir 1. Connect an exhaust line to the sample gas outlet on the module’s front panel. 2. Attach the other end of the line to the ventilator reservoir. Make sure that the reservoir tube diameter is at least 2 - 3 times larger than the exhaust line. Scavenging through the anesthesia gas scavenging system Anesthesia machines are...
Page 40
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors 28 Document no. M1031517-02 Connecting directly to the vacuum scavenging system 1. Connect the exhaust line to the monitor’s sample gas outlet. 2. Connect the exhaust line only to an open scavenging system where gas is removed at room pressure. Do not connect the monitor directly to a vacuum scavenging system. Returning sampling gas to the patient circuit The...