Page 21
Introduction 9 Document no. M1031517-02 IPX class: IPX0 IPX1 IPX2 IPX3 IPX4 IPX7 IPX8Degree of protection against harmful ingress of water as detailed in the IEC 60529: - Ordinary equipment - Protection against vertically falling water drops. - Protection against vertically falling water drops when enclosure tilted up to 15 °. - Protected against spraying water. - Protected against splashing water. - Protected against the effects of temporary immersion in...
Page 22
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors 10 Document no. M1031517-02 1.2 Safety The following list contains general warnings and cautions you should know before installing, maintaining or servicing the system. Warnings and cautions specific to the use of the system can be found in the User’s Guide and User’s Reference Manual. 1.2.1 Safety precautions Warnings WARNINGA WARNING indicates a situation...
Page 23
Introduction 11 Document no. M1031517-02 Laser radiation • When using the ARK Barcode Reader, N-SCAN, do not stare into the beam. The N-SCAN is a Class 2 laser product. External connection • Do not connect any external devices to the monitor other than those specified. Fuse replacement • Replace a fuse only with one of the same type and rating. Explosion hazard • Do not use the monitor in the presence of flammable anesthetics. Patient safety • Do not perform any...
Page 24
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors 12 Document no. M1031517-02 •Single use accessories are not designed to be reused. Reuse may cause a risk of contamination and/or affect the measurement accuracy. Cautions CAUTION A CAUTION indicates a condition that may lead to equipment damage or malfunction. Installation • Leave space for air circulation to prevent the monitor from overheating. •Before...
Page 25
Introduction 13 Document no. M1031517-02 •Do not short-circuit the battery terminals, this may produce a very high current, which will damage the battery. •Do not dispose of the battery into open flame, nor put the battery near fire, as it may explode. •Do not dismantle the battery. See also section “Symbols”. Storage and transport Do not store or transport the monitor outside the specified temperature, pressure and humidity ranges: Temperature -10...+50...
Page 26
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors 14 Document no. M1031517-02 2 System description 2.1 Introduction Datex-Ohmeda monitors build up a freely configurable modular system. The architecture is designed to enable different module combinations so that the user is able to get the desirable parameter and feature set. This modular approach makes it possible to add new features when they are needed....
Page 27
System description 15 Document no. M1031517-02 2.3 Distributed processing A system assembled from S/5 products is a multiprocessor system. All parameter modules have their own microprocessor, which performs functions such as module key control, waveform filtering, parameter related computing and pneumatic control, etc. At the same time the main CPU performs higher level tasks such as trending and alarm control. While the parameter modules and CPU are...
Page 28
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors 16 Document no. M1031517-02 2.5 Software loading The program memory on the CPU board is loaded with monitor software at the factory. The software is used for running all the functions that are integrated into the PC board. For service and upgrade procedures, the CPU board is fitted with a PCMCIA card drive through which new software can be loaded. 2.6...
Page 29
System installation 17 Document no. M1031517-02 3 System installation 3.1 Unpacking instructions 1. Confirm that the packing box is undamaged. If the box is damaged, contact the shipper. 2. Open the top of the box and carefully unpack all components. 3. Confirm that all components are undamaged. If any of the components are damaged, contact the shipper. 4. Confirm that all components are included. If any of the components are missing, contact your GE Healthcare...
Page 30
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors 18 Document no. M1031517-02 3.3.1 Compact Monitor connections Figure 5 External connections of Compact Monitor frame F-CM(C)1 rev.04 (1) Air filter (2) NET connector (3) NET ID connector (not used with N-CMW) (4) Recorder (optional) (5) SYNC connector (6) Connector for external keyboard (anesthesia record keeping solution keyboard), K-CREMCO or Barcode...