GE S5 Service Manual
Here you can view all the pages of manual GE S5 Service Manual. The GE manuals for Medical Equipment are available online for free. You can easily download all the documents as PDF.
Page 51
Interfacing 39 Document no. M1031517-02 4.3 Interfacing Datex-Ohmeda Anesthesia Delivery Unit It is possible to interface the Datex-Ohmeda Anesthesia Delivery Unit to the S/5 Compact Anesthesia Monitor via the serial interface port. The data link is bi-directional. The parameters transferred to the Anesthesia Delivery Unit are summarized in Table 4 and the events transferred to the S/5 Compact Anesthesia Monitor are summarized in Table 5. The events are...
Page 52
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors 40 Document no. M1031517-02 Table 5 Events transferred from S/5 Anesthesia Delivery Unit to the S/5 Compact Anesthesia Monitor 4.4 Interfacing Dräger Cicero, Cato, Julian and Narkomed 2C (by NAD) It is possible to interface Dräger Cicero and Cato to the S/5 Compact Anesthesia Monitor or to the S/5 Compact Critical Care Monitor via the Interface Module, E-INT...
Page 53
Interfacing 41 Document no. M1031517-02 4.4.1 Interconnection Use the INT-External Device Cable, except for Julian and Cicero EM, in which case use the INT-Julian/Cicero EM cable. 1. Make sure that the power to both monitors is turned off. 2. Connect the 9 pin D-connector to one of the connectors on the Interface Module, E-INT / M-INT. Tighten the finger screws. 3. Connect the 25 pin D-connector to a corresponding connector on the anesthesia machine. The...
Page 54
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors 42 Document no. M1031517-02 Table 7 Parameters transferred from Dräger Cato, Julian and Narkomed 2C (NAD) monitor to the S/5 Compact Anesthesia Monitor or S/5 Compact Critical Care Monitor NOTE: The transference of parameters illustrated in Table 4 and Table 5 represents the maximum. Some of these parameters may not be available in all device versions. NOTE:...
Page 55
Interfacing 43 Document no. M1031517-02 4.5.3 Connection to parallel printers A parallel printer is connected to the serial interface port connector X11 on the frame unit via the Serial-to-Parallel Converter, order code 78030, model PI130-R2 or PI1115A. Contact your authorized GE Healthcare distributor for advice on suitable parallel printers. 4.5.4 Installing the Serial-to-Parallel Converter Order code 78030, model PI130-R2 A common serial modem cable, gender...
Page 56
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors 44 Document no. M1031517-02 To connect the S/5 Compact Monitor to printer: 1. Make sure that power is switched off on both devices. 2. Connect a standard PC-to-parallel-printer cable (order code 713701) to the parallel printer. 3. Check the DIP-switch settings on the converter (order code 78030, model PI1115A). 4. Connect a common RS-232 modem cable’s (order...
Page 57
Interfacing45 Document no. M1031517-02 4.7 Output signals Analog/ digital output signals on the connectors X4 and X8 can be used for interfacing with other devices. The pin assignments are il lustrated in tables/pictures below. Table 8 Coding element connector, X4 Table 9 Synchronization connector, X8 4.7.1 Digital outputs The digital output signals are as follows: Defibrillation Sync (X8 pin 2) The defibrillation sync signal is generated by the ECG. When acti...
Page 58
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors 46 Document no. M1031517-02 4.7.2 Analog outputs Direct ECG (X4 pin 7 and X8 pin 3) Delay (max.): 15 ms Gain ECG (out)/ECG (in): 1 V/1 mV Pacer: 5 V and 2 ms pulse The signal requires input impedance of 100 kΩ. NOTE: The ECG signal is based on the ECG measurement of the Hemodynamic Modules, M-ESTPR (all variations), M-NES(12)TPR (all variations), E-PRESTN/...
Page 59
Functional check 47 Document no. M1031517-02 5 Functional check These instructions include procedures for a functional check for Datex-Ohmeda S/5 Compact Anesthesia Monitor and S/5 Compact Critical Care Monitor. The functional check is recommended to be performed after monitor installation. These instructions include a “Functional check form, Datex-Ohmeda S/5 CAM, CCCM” to be filled in when performing the procedures. The symbol in the instructions means that...
Page 60
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors 48 Document no. M1031517-02 E-PSM(P)/ E-PRESTN/ M-PRESTNMulti-Link ECG accessories, IEC - Multi-link 3-leadwire set 412682-003 - Multi-link 5-leadwire set 412681-003 - Multi-link 5-leadwire set, C2-C6 416467-004 E-PRESTN/ M-PRESTN w/ (E12) - Multi-link 12-lead ECG trunk cable 416035-002 or Multi-Link ECG accessories, AHA E-PSM(P)/ E-PRESTN/ M-PRESTN -...