Page 51

Interfacing
39
                                          Document no. M1031517-02
4.3 Interfacing Datex-Ohmeda Anesthesia Delivery Unit
It is possible to interface the Datex-Ohmeda Anesthesia Delivery Unit to the S/5 Compact Anesthesia 
Monitor via the serial interface port. The data link is bi-directional. The parameters transferred to the 
Anesthesia Delivery Unit are summarized in Table 4 and the events transferred to the S/5 Compact 
Anesthesia Monitor are summarized in Table 5. The events are...

Page 52

Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
40
Document no. M1031517-02                                                                           
Table 5 Events transferred from S/5 Anesthesia Delivery Unit to the S/5 Compact 
Anesthesia Monitor
4.4  Interfacing Dräger Cicero, Cato, Julian and Narkomed 2C (by NAD)
It is possible to interface Dräger Cicero and Cato to the S/5 Compact Anesthesia Monitor or to the 
S/5 Compact Critical Care Monitor via the Interface Module, E-INT...

Page 53

Interfacing
41
                                          Document no. M1031517-02
4.4.1 Interconnection
Use the INT-External Device Cable, except for Julian and Cicero EM, in which case use the 
INT-Julian/Cicero EM cable.
1. Make sure that the power to both monitors is turned off.
2. Connect the 9 pin D-connector to one of the connectors on the Interface Module, E-INT / M-INT. 
Tighten the finger screws.
3. Connect the 25 pin D-connector to a corresponding connector on the anesthesia machine. The...

Page 54

Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
42
Document no. M1031517-02                                                                           
Table 7 Parameters transferred from Dräger Cato, Julian and Narkomed 2C (NAD) monitor 
to the S/5 Compact  Anesthesia Monitor or S/5 Compact Critical Care Monitor
NOTE: The transference of parameters illustrated in Table 4 and Table 5 represents the maximum. 
Some of these parameters may not be available in all device versions.
NOTE:...

Page 55

Interfacing
43
                                          Document no. M1031517-02
4.5.3 Connection to parallel printers
A parallel printer is connected to the serial interface port connector X11 on the frame unit via the 
Serial-to-Parallel Converter, order code 78030, model PI130-R2 or PI1115A.
Contact your authorized GE Healthcare distributor for advice on suitable parallel printers.
4.5.4 Installing the Serial-to-Parallel Converter
Order code 78030, model PI130-R2
A common serial modem cable, gender...

Page 56

Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
44
Document no. M1031517-02                                                                           
To connect the S/5 Compact Monitor to printer:
1. Make sure that power is switched off on both devices.
2. Connect a standard PC-to-parallel-printer cable (order code 713701) to the parallel printer.
3. Check the DIP-switch settings on the converter (order code 78030, model PI1115A).
4. Connect a common RS-232 modem cable’s (order...

Page 57

Interfacing45
                                          Document no. M1031517-02
4.7 Output signals
Analog/ digital output signals on the connectors X4  and X8 can be used for interfacing with other 
devices. The pin assignments are il lustrated in tables/pictures below.
Table 8 Coding element connector, X4
Table 9 Synchronization connector, X8
4.7.1 Digital outputs
The digital output signals are as follows:
Defibrillation Sync (X8 pin 2)
The defibrillation sync  signal is generated by the ECG. When acti...

Page 58

Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
46
Document no. M1031517-02                                                                           
4.7.2 Analog outputs
Direct ECG (X4 pin 7 and X8 pin 3)
Delay (max.):  15 ms
Gain ECG (out)/ECG (in):  1 V/1 mV 
Pacer: 5 V and 2 ms pulse
The signal requires input impedance of 100 kΩ.
NOTE: The ECG signal is based on the ECG measurement of the Hemodynamic Modules, M-ESTPR (all 
variations), M-NES(12)TPR (all variations), E-PRESTN/...

Page 59

Functional check
47
                                          Document no. M1031517-02
5 Functional check
These instructions include procedures for a functional check for Datex-Ohmeda S/5 Compact 
Anesthesia Monitor and S/5 Compact Critical Care Monitor. The functional check is recommended to 
be performed after monitor installation.
These instructions include a “Functional check form, Datex-Ohmeda S/5 CAM, CCCM” to be filled in 
when performing the procedures. The symbol 
 in the instructions means that...

Page 60

Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
48
Document no. M1031517-02                                                                           
E-PSM(P)/ E-PRESTN/ M-PRESTNMulti-Link ECG accessories, IEC
- Multi-link 3-leadwire set 412682-003
- Multi-link 5-leadwire set 412681-003
- Multi-link 5-leadwire set, C2-C6 416467-004
E-PRESTN/ M-PRESTN w/ (E12) - Multi-link 12-lead ECG trunk cable 416035-002
or Multi-Link ECG accessories, AHA
E-PSM(P)/ E-PRESTN/ M-PRESTN -...